14 — Digital PCR

Counting without a curve.

+ The questionHow can splitting one sample into thousands of partitions count molecules without a standard curve?

Split, amplify, count

Real-time PCR infers the starting amount from how long the reaction takes to become visible, and that inference depends on efficiency and standards. Digital PCR asks a simpler question of many small reactions instead of a hard question of one large one. The reaction mix, with the sample in it, is divided into thousands of tiny partitions (Chapter 8). They are droplets of liquid separated by oil, or chambers etched in a chip, and each holds a fraction of a nanoliter. The whole set is cycled. A partition that received at least one target molecule amplifies it to a bright signal, while a partition that received none stays dark. The instrument then counts the bright partitions.

The count does not depend on efficiency in the way a Cq does (Chapter 13). A partition that amplifies a little slowly still ends up bright after forty cycles, and it is counted as positive just the same. And no standards are needed, because the answer is a count of molecules in a known volume, not a comparison with samples of known amount.

32 — Split, amplify, count
reaction mix split cycle 57 molecules: 150 empty, 43 hold one, 7 hold two 50 of 200 bright: 25 % two molecules, one bright partition 20,000 read, 5,000 bright 75 % dark → 0.288 molecules per partition 0.288 × 20,000 ≈ 5,750 molecules 0.85 nL droplets → about 340 per µL grid illustrates the proportions of the worked example
Template DNAReadoutFocal detail
Plate 32 — The mix is split into thousands of partitions, cycled, and the bright ones counted. Molecules land at random, so some bright partitions held two or more; the dark fraction gives the true average. With 5,000 of 20,000 partitions bright, 75 % dark means about 0.288 molecules per partition, about 5,750 molecules in all, 15 % more than the naive count.

Molecules arrive at random

The difficulty is that molecules do not distribute themselves one to a partition. They land at random, so at any useful concentration some partitions receive two or three molecules and still count as one bright partition. Counting bright partitions therefore undercounts the molecules, and the undercount grows with concentration. Digital PCR corrects for the undercount with the Poisson distribution, the statistics of random scattering that set the sampling floor in Chapter 6.

The correction works from the dark partitions, because those are unambiguous: a dark partition received exactly zero molecules. Chapter 6 showed that if the average is one molecule per partition, 37 % of partitions receive none, and each additional molecule in the average multiplies that fraction by 0.37 again. The relationship also works backward, so the fraction of dark partitions gives the average number of molecules per partition.

Solved example: correcting a count of bright partitions

A run chosen for illustration reads 20,000 partitions, of which 5,000 are bright, so 15,000, or 75 %, are dark. By the rule of Chapter 6, the share of dark partitions is 0.37, more exactly 0.368, raised to the power of the average number of molecules per partition. The average is therefore the power to which 0.368 must be raised to give 0.75, and trying values gives 0.2877, or about 0.288: 0.368 raised to the power 0.288 is 0.75. The partitions therefore held about 0.2877 × 20,000 ≈ 5,750 molecules, about 15 % more than the 5,000 bright partitions. With droplets of 0.85 nanoliters, 0.288 molecules per droplet is 0.288 ÷ 0.85 ≈ 0.34 molecules per nanoliter. A microliter is 1,000 nanoliters, so the reaction held about 340 molecules per microliter. Without the correction, the count would have missed about 5,750 − 5,000 = 750 molecules: the second and further molecules in partitions that held more than one.

The undercount grows quickly. At an average of 0.1 molecules per partition, about 5 % of the bright partitions hold more than one molecule. At an average of one, 42 % do, and at an average of 1.6, about 60 % do. In the popular picture of digital PCR, each partition holds one molecule or none, but that picture is true only at very low concentrations: at the concentrations that give the best precision, most bright partitions hold more than one molecule. The result therefore depends heavily on the correction.

33 — How much the correction does
0 20 40 60 80 100 0.05 0.1 0.2 0.5 1 2 5 10 PERCENT AVERAGE MOLECULES PER PARTITION (TENFOLD STEPS: 0.1, 1, 10) best precision 20,000 partitions: one dark partition left ≈ 9.9 partitions dark bright partitions holding more than one molecule 5 % 42 % about 20 % dark 60 % at best precision, most bright partitions hold more than one
Template DNAReadoutFocal detail
Plate 33 — As the average load rises, fewer partitions stay dark and more of the bright ones held more than one molecule: about 5 % at 0.1 molecules per partition, 42 % at 1 and about 60 % at 1.6, where precision is best and about 20 % of partitions are dark. With 20,000 partitions the measurement runs out near 9.9 molecules per partition.

Precision and range

Because the answer comes from counting, its precision follows the statistics of counting. Scientists at a digital PCR manufacturer modeled that precision and found that it is best at an average of about 1.6 molecules per partition, when about 20 % of partitions are dark, and that it improves with the square root of the number of partitions. Four times as many partitions therefore halve the relative uncertainty.

The range has firm limits at both ends. At the top, when nearly every partition is bright, the few dark ones carry all the information: with 20,000 partitions, a single remaining dark partition corresponds to an average of about 9.9 molecules per partition, and beyond that the measurement breaks down. A run with every partition bright, or every partition dark, carries no quantitative information at all. At the bottom, the limit is the volume actually analyzed, because no molecule can be counted that was not loaded. One manufacturer's plates analyze about 3 microliters per well in one format and about 24 microliters in another, and a test that analyzes the larger volume can detect rarer targets.

The partition volume is the calibration

The phrase "absolute quantification without a standard curve" needs one qualification. Digital PCR counts molecules per partition, and converting that into molecules per microliter requires the volume of a partition. If the volume is wrong, every result is wrong by the same factor. In a 2015 study, scientists at the European Commission's Joint Research Centre measured the droplets of a widely used system under a microscope. They found a mean volume of 0.834 nanoliters, about 8 % smaller than the value then built into the instrument's software, and the difference was enough to explain a systematic disagreement with another digital platform. The software value was later changed to 0.85 nanoliters. Digital PCR needs no standard curve, but every partition must have the volume that the software assumes.

"Absolute" means "if the volume is right"

Digital PCR's count is absolute only to the accuracy of its partition volume and the uniformity of its partitions. Partitions also differ from one another slightly, and partitions that fall between bright and dark must be classified by a threshold the user or software sets. Each of these is a source of error that a standard curve would have absorbed and that digital PCR must control directly.

The instruments

The systems described here divide a reaction into between about 8,500 and 100,000 partitions, depending on the platform and the format, and they read them in one of two ways. The droplet systems that made the method widespread make about 20,000 droplets from a 20-microliter sample in a separate droplet generator. The droplets are cycled in an ordinary thermal cycler and then streamed single file past a detector in a reader, so the work passes from one instrument to the next, with plates moved between them by hand. QuantaLife developed this system and was sold to Bio-Rad in 2011. That year its scientists, by then at Bio-Rad, published the system's original description and reported droplets of about a nanoliter, generated and read at about a thousand per second. Later droplet readers detect six colors, according to their maker.

34 — Two ways to build a digital PCR system
DROPLETS, THREE INSTRUMENTS DROPLET GENERATOR sample oil 20 µL → about 20,000 droplets ORDINARY THERMAL CYCLER cycle DROPLET READER read one by one plate moved by hand plate moved by hand partition volume is the calibration: one study measured 0.834 nL, about 8 % smaller than the software assumed FIXED CHAMBERS, ONE INSTRUMENT image partition cycle partition, cycle and image without moving the plate one system: 20,480 microchambers, results in under 90 min systems described here: about 8,500 to 100,000 partitions volume analyzed sets the lower limit: about 3 µL or 24 µL per well in one maker's formats
HeatReadoutFocal detail
Plate 34 — Droplet systems make about 20,000 droplets from a 20 µL sample, cycle them in an ordinary cycler and stream them past a detector, with plates moved by hand between instruments. Integrated systems partition, cycle and image fixed chambers in one instrument. Either way, molecules per microliter depend on the partition volume the software assumes, and one study found the true volume about 8 % smaller than the software assumed.

Other systems fix the partitions in place and image them. Some integrate everything: the user loads a plate and the instrument partitions, cycles and images it. One offers plates with about 8,500 or 26,000 partitions per well. Another uses 20,480 microchambers per sample and gives results in under 90 minutes, according to its maker. A third partitions plates of 20,000, 28,000 or 100,000 nanowells in a separate step before its analyzer cycles and images them. Another approach packs droplets into a flat single layer in a chip and images it.

The choice between these systems is an engineering trade between throughput, hands-on time, the volume analyzed and the cost per sample. The market is still moving: in July 2025 one of the large suppliers announced it had acquired a smaller company whose droplet-imaging technology it sells as of October 2026, in a series that detects seven colors according to its maker.

Digital PCR justifies its cost where counting matters more than speed. One use is measuring a rare mutation among many normal copies of the same gene, for example a tumor mutation in DNA from a blood sample. Another is measuring copy numbers absolutely, without standards. A third is testing samples that carry substances that slow the reaction, which digital PCR is reported to tolerate better than real-time PCR. Digital PCR is slower and more expensive per sample than real-time PCR and covers a narrower range of concentrations in one run, so it complements real-time PCR rather than replacing it.

The central question, answered

The guide's central question can now be answered in full. Heating and cooling a sealed tube finds one stretch of DNA among billions because at the annealing temperature two primers bind stably only where their sequences match, and only the stretch between two primers facing each other is copied exponentially (Chapters 2 and 5). The same cycle copies that stretch a billion times because each round of separating, priming and extending lets every copy act as a template, so the count nearly doubles in each cycle, and a heat-stable enzyme survives the whole run (Chapters 4, 6 and 7).

The instrument counts how much of the target the sample held because a probe or dye makes light in proportion to the copies, and the cycle at which the light crosses a line, or the fraction of partitions that never light, reports the starting amount (Chapters 12 to 14). The rest of the guide deepens the second half of the answer: how the instrument is engineered so that its temperatures and its readings can be trusted, and how a test built on it earns that trust.

Choose digital for the right reasons

Digital PCR is worth its cost when the result must be an absolute count, when the target is a small fraction of a large background, or when the sample tends to inhibit the reaction. It requires attention to partition volume and uniformity, to the threshold that separates bright from dark, and to loading in the range where the statistics are precise.

+ What this chapter established
  • Digital PCR divides the sample into thousands of partitions, cycles them and counts the bright ones, and the count depends far less on efficiency than a Cq does and needs no standards.
  • Molecules land at random, so many bright partitions hold more than one molecule, and the fraction of dark partitions gives the true average, about 15 % above the naive count in the chapter's example.
  • Precision is best near 1.6 molecules per partition and improves with the square root of the number of partitions, and the volume analyzed sets the lower limit.
  • Results are absolute only if the partition volume is right: one study found an 8 % error in a widely used system.
Created with Claude (Anthropic).