In the first published application of PCR, in 1985, the tube was opened after every heating step and fresh enzyme was added by hand, twenty times in a twenty-cycle run. With a heat-stable enzyme, the tube now stays sealed, and the instrument has one control: temperature. This part shows how a repeating sequence of three temperatures runs three different processes and how two primers select one stretch of DNA. It then follows doubling to a billion, explains why the enzyme had to survive repeated heating to 95 °C before the process could be automated, and shows why every reaction eventually stops.
One knob, three steps.
+ The questionIf nothing can reach into the sealed tube, how can temperature alone run three different steps?
Three processes, three temperature dependences
Once a PCR tube is sealed, nothing more is added to it and nothing is taken out. The instrument cannot stir it, cannot add enzyme at the right moment, cannot remove a product. It can only change the temperature of the metal block the tube sits in. Yet a PCR needs three distinct things to happen, in order, in every cycle: the strands must separate, the primers must bind, and the polymerase must extend them.
What makes this possible is that each of the three processes responds to temperature differently. Strand separation happens only above the melting temperature of the target, so it needs the tube hot, near 95 °C. Primer binding needs the opposite: the short primers form stable duplexes with their targets only below their own melting temperature, typically somewhere around 55 to 65 °C. Extension needs the polymerase to be active, and a polymerase from a heat-loving bacterium works fastest at around 75 to 80 °C, slower at lower temperatures and hardly at all near room temperature. Because different temperature ranges switch the three processes on and off, changing a single variable in the right sequence can run them one after another. Temperature is a switch with three positions.
Denature
The cycle starts hot. At 94 to 98 °C the target duplexes open into single strands (Chapter 2). Once the liquid in the tube is at temperature, a short target separates within seconds, so the denaturing step of each cycle in a modern diagnostic test is brief: the CDC test specifies 3 seconds at 95 °C. The first denaturation of a run is longer, typically 2 minutes, for two reasons. First, long genomic DNA takes longer to open completely the first time, and so does DNA still packed with proteins from the sample. Second, many enzymes are supplied in a blocked form that only an initial period of heat releases, a feature Chapter 5 explains.
Every second spent hot costs enzyme activity, so the denaturing step is kept as short as the target allows. Short steps put a demand on the instrument: it must bring the liquid, not just the block, to temperature quickly, because a step of 3 seconds is meaningless if the sample needs 10 seconds to reach that temperature.
Anneal
Next the instrument cools the tube to the annealing temperature, where the primers bind. The temperature is chosen relative to the primers' own melting temperature. One manufacturer recommends, for its Taq polymerase, an annealing temperature about 5 °C below the lower of the two primers' melting temperatures. Primer binding is fast because the primers are present in enormous excess, so a few seconds are enough (Chapter 3).
The choice of temperature is a trade-off between yield and specificity, which here means how strictly the reaction copies only the target. Too low, and primers stay bound to partial matches elsewhere in the sample, giving products that are not the target. Too high, and even the correct sites are occupied only part of the time, so each cycle copies fewer of the targets. The fraction of targets copied in each cycle is the reaction's efficiency, and an annealing temperature that is too high lowers it. Chapter 5 shows what goes wrong at the low end.
Extend
Then the polymerase extends each bound primer along its template. The rate of extension depends strongly on temperature, and the numbers explain the shape of modern PCR protocols. In 1988 scientists at Cetus, the company that developed Taq polymerase for PCR, published measurements of its speed. They found that it added more than 60 nucleotides per second at 70 °C, about 24 per second at 55 °C, 1.5 per second at 37 °C and a quarter of a nucleotide per second at 22 °C. Scientists at Roche Molecular Systems, which made the enzyme, later characterized it in purified form. They measured about 150 nucleotides per second at 75 °C, close to its optimum.
At 70 °C, Taq adds more than 60 nucleotides per second. A product of 1,000 bases therefore needs about 1,000 ÷ 60, or about 17 seconds, and the common rule of thumb of one minute per thousand bases is deliberately generous. For a target of about 100 bases at 55 °C, where Taq adds about 24 nucleotides per second, the time is about 100 ÷ 24, or about 4 seconds. A short diagnostic target can therefore be extended at the annealing temperature, without a separate hotter step.
For long targets, a separate extension step near 72 °C is still used: a three-step cycle of denature, anneal and extend. For the short targets of diagnostic tests, the anneal and extend steps are usually merged into one. The CDC test specifies 45 cycles of just two steps: 95 °C for 3 seconds, then 55 °C for 30 seconds, during which the primers bind, the polymerase extends them and the instrument reads the fluorescence. The two-step cycle saves a third hold in every cycle, about 45 times per run, while the block travels about the same number of degrees.
Many accounts say 72 °C is the temperature at which Taq works fastest. The purified enzyme is most active at 75 to 80 °C. A temperature around 72 °C is a practical compromise: fast enough for extension, and cool enough that the primers do not fall off before the polymerase has extended them by a few bases, after which the lengthened strand is stable. One enzyme supplier specifies 68 °C in its own protocol. The accuracy of the copy depends mainly on the reaction's chemistry rather than on the extension temperature: on the reaction's pH, its dNTP concentration and the balance of magnesium to dNTPs.
The program is the only instruction
Put together, a PCR run is a temperature program: an initial hold, then a fixed sequence of two or three temperatures repeated 25 to 45 times, sometimes followed by a slow heating to check the product (Chapter 12). Endpoint PCR examines the product only after the last cycle. Its older protocols were written for slower instruments, held each step for 15 to 60 seconds and took one to two hours. A diagnostic real-time test on a fast instrument runs 40 to 45 short cycles in well under an hour, and research systems have run a complete PCR in less than a minute (Chapter 23).
Nothing in the tube tells the instrument how the reaction is going, and in an ordinary thermal cycler nothing in the instrument checks the reaction's progress. The instrument follows its program, and the program assumes that the liquid in every well actually reaches each set point and holds it for the stated time. Whether that assumption holds is the subject of Chapters 9 and 10.
A test's temperature profile consists of its set points, hold times and number of cycles. The profile was fixed when the test was developed, and the test's performance was established with that profile on a particular instrument. Changing the profile changes the test, and so does running the profile on an instrument that delivers it differently. The instructions for the CDC test even named the instrument's run mode: "Standard 7500", not the faster mode the same instrument offered (Chapter 17).
- The instrument controls only temperature, but strand separation, primer binding and extension each respond to temperature differently, so a sequence of temperatures runs them in order.
- Denaturation takes seconds near 95 °C once the liquid is at temperature, and annealing is set a few degrees below the primers' melting temperature, trading yield against specificity.
- Taq's extension rate rises steeply with temperature, so short diagnostic targets are extended during the annealing step, and the CDC test's cycle alternates 95 °C for 3 seconds with 55 °C for 30 seconds.
- The program is the only instruction, and it assumes that the liquid in every well actually reaches each set point.