+ Part III · The thermal cycler
The box that changes temperature.
In a study published in 2005, researchers tested six new cyclers and found two in which the liquid was still 13 to 20 °C short of the set point when the hold timer of the denaturing step started. On those two instruments a diagnostic test gave no positive results until the step was made hotter and longer. The chemistry assumes that every well reaches each temperature, holds it for the stated time and leaves it on schedule. This part opens the instrument that is supposed to make that true: the path heat takes from the liquid to the room and back, the thermoelectric modules that pump it, and the engineering of speed, evenness and accuracy, including how each is specified, measured and sometimes missed.
Inside the thermal cycler.
+ The questionWhat has to be inside a box whose job is to change temperature on a schedule?
The path heat takes
A PCR reaction is a drop of liquid, typically 10 to 50 microliters, in a thin-walled plastic tube. A thermal cycler moves heat into that liquid and then out again, about forty times per run, as quickly and evenly as possible. Everything inside the box serves that path, and the path is easiest to follow from the sample outward.
Instead of a loose tube, the reaction may sit in a plastic plate, a single tray of small tubes, each called a well. The tube, or the plate's well, sits in a shaped hole in a metal block, which holds 96 or 384 of them. Beneath the block are thermoelectric modules, usually called Peltier modules, which pump heat into the block or out of it depending on the direction of the current through them. Beneath the modules is a heat sink, a large finned piece of metal, and a fan that blows room air across the fins. Above the tubes is a heated lid, a heated cover that presses down on the tube caps. Somewhere in the block is a temperature sensor, and a controller reads it and sets the current in the modules.
Each layer resists the flow of heat and stores some of it, and together those resistances and stores of heat set how the instrument behaves. The plastic wall is a poor conductor, so the liquid always lags behind the block. The block must be massive enough to spread heat evenly across all its wells and light enough to change temperature quickly. The modules must pump enough heat to move the block several degrees per second. The heat sink must carry that heat away to the room. A design that improves one of these usually worsens another, and Chapter 16 shows how such trade-offs are budgeted.
The Peltier module
A Peltier module is a flat array of small blocks of two kinds of semiconductor, connected electrically in series and sandwiched between two thin ceramic plates. When a current flows, it carries heat from one plate to the other. When the current is reversed, the heat flows the other way. The same component therefore heats the block when the current runs one way and cools it when the current runs the other, and because it needs no compressor, no liquid and no moving parts, it suits a compact instrument that must switch direction every few seconds.
The physics comes with losses. The current that pumps heat also generates resistive heat in the semiconductor, and some heat leaks through the module from its hotter plate to its cooler one. A module maker's technical reference states the balance at the face being cooled: the heat removed from it is the heat the current carries, minus half the resistive heat, minus the heat that leaks back into it. The resistive heat works against the module when it cools the block and adds to the heat delivered when it warms it. A PCR block sits above the temperature of the heat sink, so while the block is cooled the leakage runs from the block toward the sink and slightly aids cooling, but the resistive loss dominates.
Cooling is therefore slower than heating in many instruments: one manufacturer specifies its 96-well block at 4.4 °C per second for heating and 2.2 °C per second for cooling. The heat sink must reject both the heat pumped out of the block and all the electrical power put in, so the fan and fins are sized for more heat than the block ever holds.
A Peltier module has no moving parts, but it does wear out. Each swing in temperature expands and contracts its layers by different amounts, because the semiconductor, the solder and the ceramic expand differently with heat. The stress, repeated in every cycle, cracks the joints. A module maker states in its reliability guide that thermal cycling is the operating mode with the highest failure rates, and that failures are often abrupt. The same maker tested cycling-rated modules swung between 30 and 100 °C: every module survived at least 25,000 cycles, and the group's mean time between failures was calculated at 68,000 cycles. In a university study, researchers cycled a commercial module to destruction: its performance had fallen by a fifth after 40,000 cycles, and it failed at 45,000.
An instrument chosen for illustration runs three 45-cycle tests a day. That is 3 × 45 = 135 thermal cycles a day, and 135 × 365 = 49,275, or about 49,000 cycles a year. In the tests above, the maker's modules had a calculated mean time between failures of 68,000 cycles, and the university's module failed at 45,000. One year of this use is therefore of the same order as the cycles to failure in those tests. Both tests swung their modules far wider than a PCR cycle's 40 degrees: 70 degrees in the module maker's test, and more in the university's. A narrower swing stresses the joints less, but the life is still finite. This is why PCR instruments use modules rated for cycling, mounted so that the layers can flex, and why the module, like the fan beside it, is treated as a part with a service life.
Peltier modules are often praised as having nothing to wear out. They have no moving parts, but thermal cycling is their harshest duty, and their life is finite and measured in cycles. The fan that cools the heat sink is a moving part. The modules and the fan together set the instrument's service interval.
The block, the lid and the sensor
The block's job is to make every well the same temperature at the same moment. Aluminum is common. Perkin-Elmer, an early maker of cyclers, filed a patent in 1992 that describes a block machined from a solid piece of a corrosion-resistant aluminum alloy, chosen for thermal uniformity. Silver conducts heat almost twice as well as pure aluminum: 429 watts per meter per kelvin against 237. Some instruments use silver or gold-plated silver blocks.
The speed of a change in temperature, in degrees per second, is called the ramp rate. One manufacturer lists the maximum ramp rate of one of its instruments at 2.6 °C per second with a silver block and 2.3 with an aluminum one, so nearly twice the conductivity gives about 13 % more speed. Conductivity is only one limit, beside the power of the modules and the heat the block must store. Some designs spread heat with a sealed chamber in which a fluid evaporates at the hot side and condenses at the cold, the principle of a heat pipe.
The heated lid sits on the tube caps or the plate seal at about 104 to 105 °C. Without it, water would evaporate from the hot liquid and condense on the cooler cap, changing the volume and concentration of the reaction a little more in every cycle. The lid also presses the tubes down into the block, because a tube that sits loosely in its hole has an air gap around it and heats more slowly than its neighbors.
The temperature sensor is in the block, not in the liquid. The design described in the patent mentioned above has an electronic sensor in a hole drilled into the block, next to a column of wells. The controller reads the sensor many times a second and adjusts the current in the modules so that the block follows the programmed profile. The liquid sits behind a plastic wall and runs a few seconds behind the block, and Chapter 10 shows that handling this lag is the central problem of temperature control.
Not every block instrument uses Peltier modules. The design in the same early patent heated its block with a thin electrical heater and cooled it with chilled liquid from a refrigeration unit, pumped through channels in the block. The Peltier module was adopted in most later designs for its compactness and its ability to both heat and cool, not because nothing else works.
Other ways to change temperature
Changing the temperature of a metal block is one solution, but every change must heat or cool the block's mass along with the sample. Other architectures avoid the block.
In 1989 Carl Wittwer and colleagues described a cycler that put the reaction in thin glass capillaries and heated them with a stream of hot air, venting room air in to cool them. Glass capillaries hold little liquid and have a large surface for their volume, and air has little mass to heat, so the sample changes temperature quickly. Roche's original LightCycler instruments used this design. For the later of them, Roche specified heating a 20-microliter capillary from 40 to 95 °C in under 15 seconds. Another design spins its tubes in a rotor inside a chamber of moving air. Because every tube passes through the same air, its maker claims a temperature uniformity of ±0.02 °C, measured as a standard deviation.
Microfluidic designs can move the sample instead of changing the temperature. In a 1998 demonstration, a glass chip carried the reaction mixture along a channel that wound back and forth across three zones held at fixed temperatures, so that each pass through the zones was one cycle. Twenty cycles took between 90 seconds and about 19 minutes, depending on the flow rate.
Other demonstrations have let natural convection carry the liquid between a hot bottom and a cooler top. Another heated a thin gold film under the wells with blue light and completed 30 cycles in 5 minutes. The authors of a review of microfluidic PCR sort these designs into two groups: in one, the sample stays put and the temperature cycles, so the profile can be changed in software, and in the other, the sample moves through fixed zones, so the cycle is built into the channel.
For anyone specifying, buying or servicing an instrument, three things follow from this chapter. The Peltier modules and the fan have finite lives, so the service plan should state when they are checked or replaced. The block, the lid and the tube or plate form one thermal system, so a change of plate type or seal changes the sample's temperature. And the controller knows only the block's temperature, so the instrument's real performance must be measured in the wells, a subject Chapter 10 takes up.
- Heat flows from the liquid through a plastic wall, a metal block, thermoelectric modules and a heat sink to the room and back, and the resistance and stored heat of each layer shape how the instrument behaves.
- Peltier modules heat or cool by the direction of their current, with resistive losses that make cooling slower, and they wear out under thermal cycling, failing in tests after tens of thousands of cycles.
- The block spreads heat, the lid at about 105 °C stops condensation and presses tubes into contact, and the sensor sits in the block, not in the sample.
- Air-heated capillaries, rotors, flow-through chips, convection and light are other ways to cycle, and they avoid the block's mass at the price of different limits.