One test at national scale.
+ The questionWhat happened when one PCR test had to be designed, made and shipped across a country in weeks?
Weeks, not years
The first genome sequence of the new coronavirus was posted publicly on 10 or 11 January 2020, on behalf of a consortium led by Yong-Zhen Zhang of Fudan University. Which of the two dates applies depends on the time zone. Everything in this guide follows from what happened next: a sequence is all a PCR test needs to begin. Within days, a group at the Charité hospital in Berlin circulated a protocol, and on 23 January the group published a workflow in the journal Eurosurveillance with targets in three viral genes. The Berlin group had designed the targets from the sequence and from related coronaviruses, because no samples of the new virus were yet available to it. One of the paper's authors, Olfert Landt, was the chief executive of TIB Molbiol, a company that sells PCR assays and by November 2023 was a Roche subsidiary (Chapter 24).
The US Centers for Disease Control and Prevention designed its own test. It chose two targets in the N gene, called N1 and N2, and designed both to be specific to the new virus. A third target, N3, was designed to detect a broader group of related coronaviruses, and a fourth reaction detected the human gene RNase P as a control.
On 4 February 2020 the Secretary of Health and Human Services declared that circumstances justified the emergency authorization of diagnostic tests. The same day, the FDA issued an Emergency Use Authorization, or EUA, to the CDC for its panel. An EUA allows a test that has not been cleared or approved to be used while such a declaration is in force. Shipping began the next day. About 200 kits were to go to US laboratories at first, and a similar number to selected laboratories abroad, each kit enough for roughly 700 to 800 specimens. The laboratories qualified to order the kits numbered 115 in the US and 191 abroad.
About 200 kits were to go to US laboratories at first, each enough for roughly 700 to 800 specimens. At the low end, 200 kits × 700 specimens = 140,000 specimens. At the high end, 200 kits × 800 specimens = 160,000 specimens. In the first kits, each specimen took one well for each of the panel's four reactions, N1, N2, N3 and RNase P. The specimens alone therefore needed between 140,000 × 4 = 560,000 and 160,000 × 4 = 640,000 reactions, before the controls in every run. The first US kits were enough for roughly 140,000 to 160,000 specimens, which meant more than half a million reactions.
The test as designed
The panel used everything earlier chapters describe. It was a one-step real-time RT-PCR, in which reverse transcription and real-time PCR run in the same tube (Chapter 20). Each reaction held 20 microliters: 15 microliters of mix and 5 of extract. Each target had a hydrolysis probe carrying the dye FAM and a quencher (Chapter 12). All the probes used the same dye, so each specimen needed a separate well for each target. As designed, each specimen therefore took four wells, for N1, N2, N3 and RNase P. Once N3 was dropped, in February and March 2020, each specimen took three. The profile was a reverse-transcription step, a polymerase activation step and 45 cycles of 3 seconds at 95 °C and 30 seconds at 55 °C, with fluorescence read at 55 °C (Chapter 4).
The panel ran on one named instrument, with one version of its software, in its standard mode and with the passive reference set to none (Chapter 17). A target was detected if it crossed the threshold before cycle 40.00 (Chapter 13). Controls in every run and the internal control in every specimen governed the calls (Chapter 21). The panel's instructions gave the limit of detection as about 3 RNA copies per microliter of spiked sample with one extraction system and 1 with another, and the CDC's paper reported about five copies per reaction from a separate experiment (Chapter 21). It was authorized for laboratories certified for high-complexity testing, and the CDC estimated that the process took about four hours from the start of sample processing to a result.
What went wrong
Within days of the first shipments, public health laboratories reported problems: no-template controls were giving signals, mainly for N3 and to a lesser extent N1, and runs had to be called inconclusive. The CDC's own scientists later analyzed the first lot of kits and found two separate faults. They published the analysis in the journal PLoS ONE in December 2021.
The first was contamination. About 2 % of the no-template controls for N1 from the initial lot showed false reactivity. Sequencing traced the false signal to a synthetic template that the CDC had made at about the same time as the kits, and the template had reached the bulk reagent during quality assessment. The problem did not recur in later lots.
The second was design. The 3′ end of the N3 probe could pair with the 3′ end of the N3 reverse primer. In a tube with no target, the two paired, and the CDC's sequencing found duplex and larger products containing the N3 probe sequence, which gave fluorescence with no template (Chapter 5). False signals appeared in about 97 % of the N3 no-template controls made with the authorized kits' reagents, against 0.5 to 2 % with oligonucleotides from earlier batches and from commercial suppliers. The false signals became more frequent as the reagents aged. The flaw was present in all three sources of the oligonucleotides, but how often it showed depended on the batch. The authors concluded that flaws in both the assay's design and the handling of bulk material caused the first lot's problems.
In October 2023 the Inspector General of the Department of Health and Human Services reported on the circumstances that allowed these faults. The Inspector General found inadequate control of documents, about four different versions of quality-control procedures in use, a lack of separation of duties, viral material and test reagents made in the same core laboratory, and weak oversight. The Inspector General made six recommendations, and all of them had been closed as implemented by January 2025. In November 2020 NPR reported that, according to a CDC internal review, a final quality-control test had indicated that the kit could fail about a third of the time, and that the kit was released anyway. The Inspector General also found that the CDC had a viable test within about two months of the genome's publication.
About two months from the sequence to a working, accurate test is fast by any historical standard. The faults in the first kits were not caused by speed alone. They came from gaps that would have mattered at any speed: an oligonucleotide interaction that development testing did not catch, and a manufacturing environment that let a synthetic target reach the reagents. According to the CDC's internal review as NPR reported it, a third gap was a release that went ahead despite a failing quality-control result. Speed shortened the time in which those gaps could have been found and acted on.
The fix
The repair came quickly once the causes were understood. On 26 February the CDC validated a protocol that left out the third target and used only N1 and N2. By 28 February it was distributing a working kit to public health laboratories. On 29 February the FDA issued a policy that allowed laboratories certified for high-complexity testing to use tests they had validated themselves while their requests for authorization were reviewed, so other laboratories could bring their own tests into service. On 15 March the FDA reissued the CDC's authorization without N3, and it accepted probes with an additional internal quencher and reagents made by commercial suppliers. In their paper describing the test, the CDC's scientists reported that testing with N1 and N2 alone agreed completely with the original three-target interpretation.
Further revisions followed as the test was used: three more master mixes on 30 March 2020, more extraction systems in June and July, pooling of up to four specimens in December 2020, and later home-collected nasal swabs. In pooling, several specimens are combined and tested together as one sample. On 2 July 2020 the CDC received authorization for a multiplex test that detected influenza A, influenza B and SARS-CoV-2, with the human control, in one well with four dyes. In their 2021 analysis, the CDC's scientists noted that the CDC had used a stricter review process for that test, with outside experts and a pilot in three public health laboratories.
Afterward
In July 2021 the CDC told laboratories that it would withdraw its request to the FDA for emergency use authorization of its original panel after 31 December 2021, and it encouraged them to move to a multiplex test that could also detect influenza. In August the CDC clarified that there were no concerns about the panel's performance and that its primer and probe designs would remain public. In August 2026 the FDA still listed the panel among authorized tests, with the most recent letter for it dated 7 March 2023. On 29 June 2026 the Secretary of Health and Human Services terminated the 2020 declaration that justified emergency authorization of COVID-19 diagnostic tests. The termination takes effect on 26 December 2026, and from that date the remaining COVID-19 test authorizations cease to be in effect.
What the case shows
Nothing in the first weeks went wrong with PCR's chemistry. The cycle copied, the probes reported, the instrument heated and read. What failed was the system around the chemistry. The design check of how the oligonucleotides interacted with each other failed (Chapter 5). So did the separation of the work that made synthetic targets from the work that made reagents (Chapter 20). Quality control at release was not strong enough to stop a faulty lot. Even the conditions of the emergency authorization concerned those layers: the FDA waived most of its manufacturing requirements for the kit but kept those for acceptance activities, for nonconforming product and for statistical techniques, the parts that decide whether a lot is fit to ship.
The CDC test also shows how much a test system includes (Chapter 17). The authorization named one instrument and particular reagents and extraction methods, so laboratories with other equipment could not simply run the test, and each extension of the list took data and a revised letter. Speed came from the flexibility of the emergency route. Safety had to come from the same design controls, risk management and quality system that apply to any test, applied under pressure (Chapters 16 and 19).
For any team that may need to build a test quickly, the lessons of February 2020 concern the steps after design. Check every oligonucleotide pair for interaction, by computer and in no-template reactions at scale. Separate synthetic-template work physically from reagent manufacture. Define lot-release criteria before the first lot is made, with enough no-template controls per lot to see a failure rate of a few percent.
- The CDC designed a one-step RT-PCR test with three N-gene targets and an RNase P control, and the test was authorized on 4 February 2020 and shipped from 5 February.
- The first lot failed for two reasons: contamination of N1 reagent by a synthetic template, and an N3 design in which the probe's and reverse primer's 3′ ends paired, giving false signal in about 97 % of the N3 no-template controls made with the kits' reagents.
- A working N1 and N2 kit was distributed by 28 February, the authorization was reissued without N3 on 15 March, and later revisions added more mixes, extraction systems, pooling and specimen types.
- The reactions copied what their primers started, as every PCR does, while the design check, the manufacturing separation and the quality control at release failed, and naming one instrument in the authorization limited which laboratories could run the test.