Controls, limits and false results.
+ The questionWhat turns a reaction that works into a test whose result can be trusted?
Every result carries its own evidence
A single PCR that gives a signal shows only that something amplified. A test result must say more: that the right target amplified, that nothing else did, that a negative is a real absence and not a failure, and that the test would have detected the target at the concentrations that matter. Each of those claims rests on a specific piece of evidence, some of it built into every run as controls and some established once, before the test is used, as the test's performance characteristics.
Controls: one question each
Controls are reactions, or targets within reactions, whose expected result is known, and each answers one question about the run. A positive control contains the target at a known, modest level (Chapter 8). If it fails, the reagents or the instrument failed, and no negative in that run can be believed. A no-template control contains everything but the sample (Chapter 5). If it amplifies, something in the run is contaminated, and no positive can be believed.
An extraction control is a specimen-like sample carried through extraction alongside the patients' specimens. In the CDC test it was a human specimen control, which had to be negative for the viral targets and positive for the human gene, a pattern that showed that extraction had worked and had introduced no viral target. An internal control is amplified from every patient's specimen (Chapter 16). When it shares the well with the target, as in a multiplex test, it also shows that amplification worked in that very well. In the CDC test it was the human gene RNase P, present in the cells that any good swab collects. All the CDC probes used the same dye, so RNase P ran in a well of its own (Chapter 22). It showed that the specimen contained human material, that extraction recovered it and that amplification worked in the control's own well.
Under the CDC's rules, the controls were read together. A run was valid only if the positive control was positive for all three targets, N1, N2 and RNase P, the no-template control showed no amplification at all and the human specimen control behaved as expected. Within a valid run, a specimen with both viral targets detected was positive, whether or not RNase P was detected. A specimen with neither viral target, but with RNase P detected, was negative. A specimen with only one viral target was inconclusive and was retested. A specimen with nothing detected, not even RNase P, was invalid, and it was to be extracted again or collected again.
The limit of detection
The most important performance characteristic of a detection test is how little target it can detect, its limit of detection (Chapter 6). Laboratory medicine defines this limit carefully. The Clinical and Laboratory Standards Institute, CLSI, sets out the definitions in its guideline EP17-A2, whose second edition it published in 2012 and reaffirmed in 2017. The guideline distinguishes the limit of blank from the limit of detection. The limit of blank is the highest result expected from a sample with no analyte, the substance the test detects, and the limit of detection is the lowest amount that can be reliably told apart from the limit of blank. For qualitative molecular tests the limit of detection is usually defined by hit rate: the lowest concentration at which at least 95 % of replicates are positive. In practice, developers test 20 replicates and require at least 19 positive.
The CDC established the test's limit by spiking a synthetic copy of the N gene's RNA into a suspension of human cells in transport medium, then extracting it and testing replicates at a series of concentrations. The test's instructions give the limit as about 3 RNA copies per microliter of the spiked sample with one extraction system, and 1 copy per microliter with another, while the CDC's paper describing the test gives it as about 5 RNA copies per reaction. The paper's figure comes from a separate experiment: synthetic RNA, diluted in buffer and added directly to the reaction, was detected in all 24 replicates at 5 copies. Five copies per reaction sits just above the sampling floor of Chapter 6, the best that any PCR can do. Below a test's limit, detection does not stop abruptly but becomes hit-or-miss, and hit-or-miss detection is where inconclusive results come from: one target is detected and the other is not.
Inclusivity, exclusivity and interference
A test must detect every version of its target and nothing else. Inclusivity is the first half of that demand: the primers and probe must match all known variants of the organism. During the pandemic, the FDA set out what it expected in a template for emergency submissions. Developers were to show that all published SARS-CoV-2 sequences would be detected, by aligning their oligonucleotides against those sequences, and they were to keep checking as new variants appeared.
Exclusivity, or cross-reactivity, is the second half: the test must not react with related organisms, or with others likely to be in the same specimen. The same template called for laboratory testing wherever an oligonucleotide showed a similarity above 80 % to another organism, and it listed organisms to test at high concentrations. The CDC's scientists reported no cross-reaction with the four common human coronaviruses or with MERS coronavirus. The test's third target, N3, had been designed to detect a broader group of related coronaviruses, and it did react with the 2003 SARS virus, as intended. Interference testing checks that substances likely to be in specimens, such as blood or mucus, do not change the results, and another CLSI document, EP07, guides it.
Clinical performance, and why prevalence changes everything
Analytical performance describes the test on prepared samples. Clinical performance describes it on patients. Against a reference standard, a method that establishes who truly has the condition, a test's sensitivity is the fraction of true cases it calls positive, and its specificity the fraction of non-cases it calls negative. Often no perfect reference exists, and the new test is compared with another test instead.
In its statistical guidance of 2007, the FDA recommends that such comparisons be reported as positive percent agreement and negative percent agreement, not as sensitivity and specificity. These are the shares of the comparator's positive and negative results that the new test matches, and the reason for the different names is that agreement with an imperfect comparator is not accuracy. The FDA also recommends that each figure be reported with a 95 % confidence interval. In its emergency template for SARS-CoV-2 molecular tests, the FDA asked for at least 30 positive and 30 negative clinical specimens, with agreement of at least 95 % in each.
A clinician, though, needs the probability that a positive result is right, which is called the positive predictive value. It depends on how common the condition is among the people tested, its prevalence.
These figures are illustrative, computed for this guide. A test with 95 % sensitivity and 99.5 % specificity is used on 10,000 people. Where 1 % are infected, 100 people are infected, and 95 % of them, 95, test positive. The other 9,900 are uninfected, and 0.5 % of them, about 50, test falsely positive. Of about 145 positives, only 95 are real, so the positive predictive value is 95 out of 145, about 66 %. Where 20 % are infected, 2,000 people are infected and 1,900 test positive. Of the 8,000 uninfected, 0.5 %, or 40, test falsely positive. Of 1,940 positives, 1,900 are real, about 98 %. The test did not change. The population did, and the share of positives that are right rose from about two in three to about 98 in 100.
Where false results come from
Every false result has a mechanism, and each mechanism has a defense described somewhere in this guide.
False positives come from contamination, by synthetic material made in the same place as the reagents or by old product (Chapter 20). They also come from oligonucleotides that react with each other in every tube, as the CDC's N3 oligonucleotides did, and from cross-reaction with a related organism (Chapter 5). Mix-ups of specimens or wells are a further source. The no-template control catches some of these, and design and laboratory practice must prevent the rest.
False negatives come from a poor specimen: a swab that collected little, or one taken too early or too late in the infection. They also come from a target concentration below the limit of detection, and from losses in extraction, degraded RNA and inhibitors (Chapter 20). A change in the organism's sequence under a primer or probe causes them too, which is why the FDA's special controls for SARS-CoV-2 multi-target tests require ongoing monitoring of variants. Instrument faults that leave wells too cool, or that misread them, are a further source (Chapters 10 and 19). The internal control catches several of these by turning them into invalid results.
A negative PCR result means that the target was not detected in the material that reached the reaction, above the test's limit, with a valid internal control. It does not exclude infection, because the swab may have missed the virus, or the specimen may have been taken before the virus multiplied or after it cleared. Instructions for use state this limitation, and a test's design can reduce such misses but cannot remove them.
Laboratories that run such tests work to their own standards. The international standard for medical laboratories is ISO 15189:2022, and ISO 17822:2020 is a quality practice guide written specifically for nucleic acid amplification tests for microbial pathogens. CLSI documents cover the validation of multiplex molecular tests and the setting up of molecular testing environments.
The evidence that makes a result believable has five parts: controls, a limit of detection established by hit rate, inclusivity and exclusivity checks by computer and in the laboratory, interference testing, and clinical agreement with confidence intervals. Plan them from the intended use, run them on the complete test system, and keep monitoring inclusivity for as long as the target keeps evolving.
- Each control answers one question: positive control for reagents and instrument, no-template control for contamination, extraction control for extraction, internal control for the specimen and, when it shares the target's well, for that well.
- The limit of detection of a qualitative test is the lowest concentration detected at least 95 % of the time, and the CDC's paper detected about 5 copies per reaction, near the sampling floor, in an experiment separate from the instructions' limits.
- Inclusivity, exclusivity and interference testing establish what the test detects, and clinical agreement is reported as sensitivity and specificity only against a reference standard.
- Predictive value depends on prevalence: in the illustrative example, the same test gets about 66 % of its positives right at 1 % prevalence and about 98 % at 20 %.
Let's Talk