Why it doesn't copy forever.
+ The questionIf every cycle doubles the product, why does every run flatten out?
The shape of a real reaction
If doubling continued, forty cycles would turn a single molecule into about a trillion, and fifty into a quadrillion. No reaction does this. Plotted on an ordinary scale, the amount of product in a PCR stays invisibly small for many cycles, then rises steeply, then bends over and levels off. The level at which it stops is called the plateau, and for a standard reaction it corresponds to roughly 0.3 to 1 picomole of product, or about 2 × 10¹¹ to 6 × 10¹¹ molecules.
The plateau does not arrive suddenly. Efficiency declines gradually as product accumulates, so the curve bends rather than breaks: a phase of near-constant doubling, then a phase in which each cycle adds less than the last, and finally a phase in which cycles add almost nothing. That gradual bend is what gives an amplification curve its familiar S shape.
The usual explanation, and what experiments found
The usual explanation is that the reaction runs out of something, primers or building blocks. The numbers from Chapter 3 make that explanation hard to sustain for a normal reaction.
In a 50 µL reaction, each primer at 0.2 µM comes to 10 picomoles, or about 6 × 10¹² molecules. The plateau holds about 2 to 6 × 10¹¹ product molecules, roughly 0.3 to 1 picomole, and each product molecule used one molecule of each primer. At most about a tenth of each primer has therefore been used, so at the plateau nine in ten primer molecules or more are still unused. Each dNTP at 200 µM comes to about 6 × 10¹⁵ molecules. A 100-base-pair product has 200 nucleotides in its two strands, about 50 of each of the four kinds, so the supply could make about 6 × 10¹⁵ ÷ 50, or about 10¹⁴ products, several hundred times more than the plateau. Primer and dNTP depletion can stop a reaction designed with very low concentrations, but they do not explain the plateau of a standard one.
Experiments that looked directly for the cause point elsewhere. The authors of a 1994 study ruled out several causes, each as a sufficient explanation: exhaustion of the enzyme, the dNTPs or the primers, the accumulation of pyrophosphate, and enzyme degradation of the product. Pyrophosphate is the small molecule released with each nucleotide the polymerase adds (Chapter 3). The authors proposed instead that the product strands themselves were responsible. Once product is abundant, the two strands of each product find each other quickly when the reaction cools, and re-pairing with a long, perfectly matched partner competes with binding a short primer. Every strand that re-pairs before a primer reaches it is not copied in that cycle.
The competition shifts as the reaction proceeds, because of how often the molecules meet. The rate at which product strands re-pair depends on how often a product strand meets its complement, so it grows with the square of the product concentration: twice the product, four times the re-pairing. The rate at which primers bind depends on how often a product strand meets a primer, and the primer concentration hardly changes during the reaction. So the fraction of product strands lost to re-pairing in each cycle grows in step with the product itself. While product is scarce, in the early cycles, almost every strand is primed and copied. Once it is abundant, a growing share re-pairs first, and the efficiency falls cycle by cycle.
A study published in 2000 found a second effect of abundant product. The polymerase binds double-stranded DNA in general, not only at primer ends, and as product accumulates the enzyme is increasingly held on product duplexes where it does nothing useful. In that study, short double-stranded DNA fragments inhibited the polymerase completely when they outnumbered its molecules about thirty to one, while the depletion of dNTPs and the loss of enzyme to heat were not the main limits.
Other factors contribute in particular reactions: nonspecific products and primer-dimers competing for enzyme and primers, the gradual loss of enzyme activity to heat, and incomplete separation of strands when the product is very concentrated. The common thread is that the reaction is limited by its own success. The more product there is, the more the product interferes with making more.
The most repeated explanation of the plateau is that the primers and nucleotides run out, and for a standard reaction it is the least supported. Experiments point first to product strands that re-pair instead of binding primers, and to the polymerase being taken up by the accumulated product.
The plateau forgets where the reaction started
The most important consequence of the plateau concerns measurement. The authors of the 1994 study also observed that reactions of the same target started from different amounts reached the same maximum. A sample with a thousand target molecules and a sample with a million both end in roughly the same amount of product, because both are stopped by the same self-limiting effects. The plateau carries almost no information about how much target there was at the start.
Those self-limiting effects depend on the amount of product in the tube, not on how many cycles it took to get there, so a reaction that starts with more target simply reaches the limiting amount sooner. A reaction with a million starting copies is about ten cycles ahead of one with a thousand all the way up the exponential phase, and then both stall at roughly the same height.
The first generation of PCR measured the product at the end of a run, by running it on a gel. That measurement can say whether the target was present and whether the product is the right size, but because of the plateau it cannot reliably say how much target was present. The information about the starting amount is found earlier, in the cycles where the reaction is still doubling. In those cycles, two samples that differ tenfold in their starting amount are separated by a steady gap of a little over three cycles: the sample with more target reaches any given amount of product sooner. Part IV builds on exactly that gap.
For a yes-or-no answer with a size check, an endpoint reaction of 25 to 35 cycles followed by a gel is cheap and adequate. It needs a positive control, a reaction that contains the target at a known level, to show that the run worked. A failure in one sample's own reaction, from a substance in the specimen that slows the enzyme, for example, shows only in a control amplified from that same sample (Chapter 21). For any question of quantity, the reaction must be observed during the exponential phase, which means reading fluorescence in every cycle. The other route is to divide the sample into many small separate reactions, called partitions, so that each partition gives a yes-or-no answer (Chapter 14).
- Every PCR levels off at a plateau, roughly 2 to 6 × 10¹¹ product molecules in a standard reaction, after a gradual decline in efficiency that gives the curve its S shape.
- In a standard reaction the primers and dNTPs are far from exhausted at the plateau, and experiments point to product strands re-pairing and to the polymerase binding accumulated product.
- Reactions started from very different amounts reach the same plateau, so the endpoint says little about the starting amount.
- Quantity must therefore be read in the exponential phase, where a tenfold difference in starting amount appears as a steady gap of a little over three cycles, or counted in many partitions that each give a yes-or-no answer.