§ Glossary
Glossary.
Terms are defined as they are used in this guide.
- 510(k)
- A premarket notification to FDA. FDA clears the device when it finds it substantially equivalent to a legally marketed predicate; it is the route for a neonatal incubator, product code FMZ.
- accreditation
- The recognition of a testing laboratory or a certification body by an accreditation body, under ISO/IEC 17025 for testing or ISO/IEC 17065 for certifying products. Its scope names the kind of work and the standards it covers.
- air mode
- The incubator's mode in which the controller holds the air at the set temperature.
- alarm condition
- The state of an alarm system when it has determined that a potential or actual hazardous situation exists that needs the operator's awareness or response, in the terms of IEC 60601-1-8.
- alarm flood
- A burst of alarms faster than an operator can handle: more than 10 new alarms in 10 minutes, the definition that two control-system makers, Siemens in 2014 and Emerson in 2016, give from the alarm-management guides EEMUA 191 and ISA-18.2.
- alarm signal
- The signal an alarm system generates to announce an alarm condition, such as a tone or a light, in the terms of IEC 60601-1-8.
- ALARP
- As low as reasonably practicable: HSE's expression of the duty in UK law to reduce risk so far as is reasonably practicable. A measure that would reduce the risk has to be taken unless its cost is grossly disproportionate to the reduction.
- AND gate
- A gate of a fault tree whose output event occurs only when all of its inputs occur. For independent inputs its probability is the product of theirs, and for dependent inputs it may be significantly greater.
- angle of attack
- The angle at which an aircraft's wing meets the oncoming air. The 737 MAX measured it with a sensor on each side of the fuselage.
- application program
- In IEC 61511, the program that a safety instrumented system's logic solver runs; its own lifecycle, from design through implementation to review and testing, sits inside the lifecycle of the system.
- architectural constraints
- Requirements on the least hardware fault tolerance of each subsystem of a safety function, which a design must satisfy as well as its target for the average probability of failure on demand. IEC 61508-2:2010 sets two routes to them, 1H and 2H.
- architecture
- A safety function's channels and the vote that joins them, such as 1oo1, 1oo2, 2oo2 or 2oo3.
- ASIL decomposition
- Splitting a requirement between two or more elements that are each redundant with respect to it and sufficiently independent of each other, so that each carries a lower ASIL, such as ASIL B(D) twice for an ASIL D requirement. The letter in brackets records the ASIL before the split, and the pair meets the original level only together.
- auditable
- Of a protection layer: such that its performance can be checked by review, testing and records, so that its claimed PFD can be shown.
- automatic tank gauge
- An instrument that measures the level in a storage tank and shows it in a control room, usually with alarms set at high levels. Tank 912's gauge stuck on the night of 10 to 11 December 2005, and its three high-level alarms, which depended on its reading, could not sound.
- automotive safety integrity level (ASIL)
- The grade, from ASIL A, the least demanding, to ASIL D, that ISO 26262 gives a hazard in a driving situation from its severity, exposure and controllability classes. The worst situation sets the function's ASIL, and the letter then grades the methods and the independence asked of the work on the function.
- average probability of failure on demand (PFDavg)
- The probability, averaged over time, that a safety function has failed dangerously and would not answer a demand. For one channel with a perfect test and a small probability it is about λDU·T/2, half the probability just before a test.
- basic safety
- In IEC 60601-1, freedom from unacceptable risk caused directly by physical hazards, in normal condition and in single fault condition.
- bathtub curve
- The idealized curve of a part's failure rate against its age: high and falling for early failures, flat through the useful life, and rising again at wear-out.
- boundary value analysis
- Testing at each boundary in a requirement and on either side of it, such as the incubator's set temperatures of 36.9, 37.0 and 37.1 °C, where its cutout changes from one limit to the other.
- branch coverage
- A structural coverage measure met when every decision in the program has been taken both ways under test.
- broadly acceptable
- In HSE's framework, the band of individual risk below 1 in a million a year, for workers and the public alike.
- bund
- A wall or earth embankment around a storage tank, meant to contain fuel lost from the tank. At Buncefield, Tank 912 stood in Bund A, where the overflowing petrol and its vapor collected.
- bypass
- A deliberate means of stopping a safety function from acting while it stays installed, so that a test, a repair or a start-up can go ahead without a trip; PSLG calls it an override or an inhibit. While a bypass lasts the function would fail any demand, so the hours in bypass add directly to its PFDavg.
- capable
- On a functional safety certificate, as in "SIL 3 Capable": the highest integrity level of a safety function that the element's systematic capability allows, judged from the element alone. It is a ceiling against systematic failures only, since the element's hardware sets a ceiling of its own, and a function built from the element must still meet its own target with all its parts, arrangement and tests.
- category
- One of ISO 13849-1's five structures for a safety function, B, 1, 2, 3 and 4, from a basic single channel to redundant channels whose diagnostics detect at least 99 % of dangerous failures; each has its own range of MTTFD and, from Category 2, a minimum diagnostic coverage.
- Under India's Medical Devices Rules, 2017, the authority that licenses the manufacture of class C and D devices and every import. A State Licensing Authority licenses the manufacture of class A and B devices.
- certification
- An assessment of a product, a process or a person against a standard by a body outside the maker, stated in a certificate. Certification to IEC 61508 is voluntary.
- channel
- One path that can carry out a safety function, or one subsystem's share of it, by itself, such as a thermistor with its comparison and cutout.
- coding standard
- A set of rules that forbids or limits constructs of a programming language that are error-prone or that the language leaves to the compiler, such as automatic type conversion.
- collateral standard
- A standard of the IEC 60601 family that covers one subject for medical electrical equipment in general, such as IEC 60601-1-8 for alarm systems and IEC 60601-1-10 for physiologic closed-loop controllers.
- common cause
- One cause that leaves two or more channels failed at the same time, whether it strikes them together or one after another before either failure is found. It acts through something the channels share, such as a design, a make of part, a supply, a place or a procedure.
- conditional modifier
- In LOPA, a probability on which the harm depends but which no protection layer controls, such as whether escaping vapor ignites or whether anyone is near.
- continuous mode
- The mode of a safety function whose demand never stops, because it acts all the time. Its target is a PFH.
- controllability
- The third of the classes from which ISO 26262 sets an ASIL: it grades whether the harm could still be avoided once the malfunction shows, from C0, controllable in general, to C3, difficult or impossible to control. It brings into the grade what a process plant credits, if at all, as a separate operator layer.
- cross-check
- A check that sets one reading against a second measurement of the same quantity; a difference larger than two healthy sensors would show reveals that one of them has failed, though not which.
- cutout
- In the incubator, the relay at the end of the protective channel whose contacts sit in the heater's supply and open to cut it.
- dangerous detected failure (DD)
- A dangerous failure that a check reveals with an alarm; until it is corrected the safety function cannot act, so it calls for repair or the safe state within the time the process allows.
- dangerous failure
- A failure that can stop the equipment from reaching its safe state in time when the hazard arises, such as an incubator cutout whose contacts have welded shut. It does harm only if a demand comes while it lasts.
- dangerous undetected failure (DU)
- A dangerous failure that shows nothing until a test or a demand finds it, such as cutout contacts welded shut, which look like closed contacts in normal operation.
- DCavg
- In ISO 13849-1, the diagnostic coverage averaged over a safety function, graded none, low, medium or high.
- de-energize to trip
- A design in which the protective function holds energy in its working state and the trip is the removal of that energy, so that a lost supply, a broken wire or a dead output also trips.
- demand
- The occasion on which a safety function must act, such as the moment the air in an incubator whose control has failed reaches the protective channel's limit.
- demand rate
- How often demands reach a safety function, counted per year: the frequency of the initiating events, multiplied by the share of them that meet their enabling conditions and by the share that pass every barrier acting earlier. Multiplied by the probability that the function fails on a demand, it gives the frequency of the event the function exists to prevent.
- dependable
- Of a protection layer: such that it cuts the frequency of the scenario's consequence at least tenfold.
- designated standard
- In Great Britain's machinery law, a technical specification adopted by a recognized standards body or designated by the Secretary of State, used to achieve a presumption of conformity.
- detected failure
- A failure that becomes evident, as soon as it occurs, to the people who run and maintain the equipment, for example because a check built into the system raises an alarm. It needs a failed state that looks different from the working state, and a check that looks for the difference.
- development assurance level
- In aviation, a grade of how thoroughly a function or an item is specified, built and verified, lettered from A, the most rigorous, and set from the severity of the function's failure conditions. Like systematic capability, it states rigor, not a failure rate.
- deviation
- In a HAZOP, a departure from the design intent of a line, found by applying a guide word to it, such as a level higher than intended.
- diagnostic coverage (DC)
- The ratio of the dangerous failures that a system's own checks detect to all its dangerous failures, λDD/λD. Failures found by proof tests are not counted in it.
- diagnostic test interval
- The time from one run of a diagnostic check to the next. With the time taken to reach or keep the safe state, it must be less than the process safety time under IEC 61508; where demands come more often than once a year, a check that runs more than 100 times as often as they arrive may serve instead.
- diversity
- Making redundant channels different, by type of part, maker, design or technology, so that a fault in one design, one make of part or one program is not also in the other. It does nothing against a requirement that both designs follow.
- enabling condition
- A state that is neither a failure nor a protection but must be present for an initiating event to lead to harm, such as a batch large enough to overflow the tank.
- energize to trip
- A design in which energy must be applied to make the trip, so that a lost supply or a broken wire prevents it and must be watched for.
- error
- A value that departs from the true one, such as a measured air temperature below the true air temperature. A fault can put an error into a system, and an error that reaches what the system must deliver becomes a failure.
- error-correcting code (ECC)
- Extra bits stored with each word of memory, computed from its contents, by which the hardware finds a flipped bit. A single-error-correcting, double-error-detecting code repairs one flipped bit and reports two.
- essential performance
- In IEC 60601-1, the performance of a clinical function, other than one related to basic safety, whose loss or degradation beyond the limits the maker specifies would bring unacceptable risk. The maker identifies it for its device.
- expected harm
- The product of a hazard's frequency and the harm it causes each time, for example deaths a year. It puts frequent small harms and rare large ones on one scale, and says nothing of how the harm arrives.
- F-N curve
- A plot of F, the frequency a year of accidents that kill N or more people, against N, both in tenfold steps. A line with a slope of 1 on it is a line of equal expected harm; a slope of 2 asks large accidents to be rarer still.
- fail-safe
- Arranged so that the likely failures of a design leave the equipment in its safe state, as a semaphore arm returns to danger under its counterweight when its wire breaks. A design is fail-safe only against the failures its designer named.
- failure
- The event in which an item loses the ability to perform as required. It is judged against the requirement of a function, so a function can fail while no part has broken, and the loss can come long before a demand or a test reveals it.
- failure condition
- In aviation, the effect that one failure or a combination of failures has on the aircraft and the people on board. FAA's guidance for transport airplanes sorts failure conditions into five classes, from no safety effect to catastrophic, each with an allowed average probability per flight hour.
- failure mode
- The way a part departs from what it should be, as seen at its terminals: for a thermistor, an open circuit, a short circuit or a drift in resistance.
- failure modes and effects analysis (FMEA)
- An analysis that lists each part of a system and each way it can fail, one failure mode to a row, and follows each failure up to its effect on the function, one failure at a time. Ranked by criticality, which takes in at least the severity of each effect, it becomes an FMECA.
- failure rate
- The fraction of the parts still working that fail in each hour of operation, written λ. A rate for one failure class carries the class as a subscript, as λDU for dangerous undetected failures.
- fault
- A state in which a part cannot perform as required. Within one part it is the state that the part's own failure leaves; for the system that contains the part, it is the start of a chain that can lead through an error to a failure of the system.
- fault tree
- A diagram that starts from one undesired event, the top event, and works down through OR and AND gates to basic events, such as the failure of one part, so that it shows the combinations of faults that produce the top event.
- final element
- The part of a safety function that acts on the process, such as a valve that closes or a contact that opens. In the incubator it is the cutout in series with the heater.
- FIT
- A unit of failure rate: one failure in a billion (10⁹) hours of operation, so 100 FIT is one failure in 10 million hours.
- FMEDA
- Failure modes, effects and diagnostic analysis: an assessor's analysis that gives each component of a device a handbook rate and sorts each way a component can fail by its effect and by whether the device detects it, giving the device's rates by failure class. Its rates are calculated, not counted.
- freedom from interference
- The property that software of lower integrity cannot disturb software of higher integrity running beside it, in memory, in timing and execution, or in what passes between them; IEC 61508 calls it independence of execution.
- functional safety assessment
- An investigation, based on evidence, to judge the functional safety that a safety function achieves, including whether its requirements are complete and sufficient. IEC 61508 grades the assessor's independence by SIL, and the 2016 edition of IEC 61511 places the assessment in five stages, from the specification to each modification.
- functional safety certificate
- An assessor's declaration that an element complies with the relevant parts of IEC 61508, stating its ceilings, the versions covered and the conditions of use. Certification is voluntary, and the PFDavg and the architectural constraints of a function built from the element are verified for each application.
- functional safety management
- The management that runs across the lifecycle: it names the person or organization responsible for each phase, plans the work, verifies each phase's output, assesses the functional safety achieved and audits whether the procedures are followed in practice. Under IEC 61508-1 it also requires that everyone involved has the competence for their duties.
- GAMAB
- Globalement au moins aussi bon: the French transport principle that a new system must offer a level of risk globally at least as good as that of the system it replaces. The European railway standard EN 50126 calls it GAME, globalement au moins équivalent.
- gross disproportion
- The test behind ALARP, set by the Court of Appeal in Edwards v National Coal Board (1949): a measure may be left out only if its cost is grossly disproportionate to the risk it removes, and the duty-holder must be able to show it.
- guide word
- One of the short words a HAZOP applies to a design intent to produce a deviation: No, More, Less, As Well As, Part Of, Reverse and Other Than, with Early, Late, Before and After for timing and order.
- hardware fault tolerance (HFT)
- The number of hardware faults that an architecture can have and still perform its safety function: 0 for 1oo1 and 2oo2, 1 for 1oo2 and 2oo3.
- harmonized standard
- A European standard developed by a recognized European standards organization following a request from the European Commission, whose reference is published in the Official Journal of the European Union. Its use is voluntary, and following it gives a presumption of conformity with the requirements it covers.
- hazard
- A potential source of harm, such as the hot air that builds up in an incubator while its heater stays on.
- hazardous situation
- In ISO 14971, the circumstance in which people, property or the environment are exposed to one or more hazards; it stands between the sequence of events that begins with a hazard and the harm.
- HAZOP
- The hazard and operability study. A team states the design intent of each line of a drawing and applies guide words to it to find deviations, recording for each its causes, consequences and safeguards. IEC 61882:2016 is its international application guide.
- high-demand mode
- The mode of a safety function called on more than once a year. Its target is a PFH.
- human error probability
- The probability that a person fails to carry out a task correctly, estimated by methods such as THERP, HEART and SPAR-H from a nominal value for the kind of task and factors for its conditions.
- identical adoption
- A national standard whose text is the international standard's, identical under single numbering: its number is the IEC's or ISO's with IS in front, as in IS/IEC 61508. A function built to the international text meets the Indian Standard on the same evidence.
- independence (in verification)
- The separation between the author of a piece of work and the person who reviews or tests it, which the software standards grade by level, so that a misreading of a requirement is not shared by the code and its test.
- independent
- Of a protection layer: such that nothing that causes the initiating event and nothing that defeats another layer can also defeat it. Only the PFDs of independent layers multiply.
- independent high-level switch
- A level switch on a storage tank, separate from the tank's gauge and set above the gauge's alarms, meant to sound a final alarm and close the valves on the incoming pipelines automatically if the tank is overfilled. Tank 912's switch could not act on the night of the explosion.
- independent protection layer
- A device, system or action that can stop a scenario before it reaches its consequence, without being affected by the initiating event or by any other protection layer in the same scenario, in the definition of the Center for Chemical Process Safety.
- individual risk
- The yearly probability that a particular person, such as a worker in a given job or a resident at a given address, is killed by an activity. It adds up every hazard of the activity that can reach that person.
- inertial reference unit
- A computer that senses a vehicle's motion. Ariane 501 carried two, with identical hardware and software, one active and one on hot stand-by, and its on-board computer steered by the guidance and attitude information they supplied; on 4 June 1996 both failed almost simultaneously.
- initiating event
- The smallest combination of failures or errors that sets an accident sequence going, such as a gauge that stops reading during a fill.
- integrity
- Of a safety function: how rarely it fails to do its job when called on, from random and systematic causes together.
- LAH
- A tank's high level alarm, set below the LAHH and above the normal fill level; a fill should end before it sounds, so it is not used to control routine filling.
- LAHH
- A tank's high-high level alarm, set at or below the tank rated capacity and driven by a level sensor of its own; where it trips the transfer, the response time it needs sets its margin below the overfill level.
- latent fault metric (LFM)
- One of ISO 26262's hardware metrics: a percentage concerning faults that violate nothing alone and stay unnoticed until a second fault lets them act, such as a failed cross-check; the higher it is, the fewer such faults stay hidden. Its target is at least 90 % at ASIL D, as chip makers tabulate it.
- layer of protection analysis (LOPA)
- A method that multiplies the frequency of one initiating event by the probability of each enabling condition, the PFD of each independent protection layer and each conditional modifier, and compares the frequency of the consequence with the frequency judged tolerable. Its values are usually rounded to tenfold steps.
- light curtain
- A row of light beams between an emitter and a receiver whose outputs switch off when any beam is interrupted, so that the machine's hazard is stopped; a guard that blocks nothing.
- lining up
- Setting each valve on the route a product will take, in a fixed order, before the product moves. At Jaipur in 2009 the procedures for it were not observed, the immediate cause of the fire in the Ministry's account to Parliament.
- lockstep
- An arrangement of two processor cores that run the same program on the same inputs while a comparator checks their outputs on every clock cycle, so that a difference reveals a hardware fault in one of them.
- logic solver
- The part of a safety function that compares the sensor's measurement with the limit and decides whether to act. It can be a programmable controller, a dedicated circuit or a single comparator.
- low-demand mode
- The mode of a safety function called on no more than once a year. Its target is a PFDavg.
- management of change
- A management system that identifies, reviews and approves each modification of equipment, procedures, raw materials or processing conditions before it is made, so that the modification is analyzed for adverse effects, documented and made known to those it affects. A replacement in kind is exempt.
- MC/DC
- Modified condition/decision coverage: a structural coverage measure met when each condition in a decision has been shown to change the decision's outcome on its own, with the other conditions held fixed. It takes at least one test more than the decision has conditions.
- MCAS
- The Maneuvering Characteristics Augmentation System of the Boeing 737 MAX, a function of its speed trim system that trimmed the horizontal stabilizer nose down when the angle of attack exceeded a threshold in manual flight with the flaps up. Before the changes required in November 2020, it read one angle-of-attack sensor.
- mean time between failures (MTBF)
- The average operating time between failures of a part that is repaired and returned to service; it applies to repairable items only.
- mean time to failure (MTTF)
- The average operating time before a part fails. With a constant failure rate it is the reciprocal of the rate, so it describes a population on the flat of the bathtub, not how long one part lasts.
- MEM
- Minimum endogenous mortality: a railway criterion that takes as its reference the lowest natural death rate, 2 in 10,000 a year for children aged 5 to 15 in developed countries, and allows no single technical system to add more than 1 in 100,000 a year to a person's risk of death.
- minimal cut set
- A smallest combination of failures that, if all of them occur, causes a fault tree's top event. A cut set of one is a single failure that causes the top event, and a cut set of two is a pair.
- MISRA C
- A coding standard for the C language. Its current edition, MISRA C:2025, was published in March 2025, and each of its guidelines is mandatory, required or advisory.
- mission time
- The period over which a safety function's equipment is expected to operate without major overhaul or replacement. The failures a proof test cannot reveal accumulate over it.
- mode of operation
- The class of a safety function by how often it is called on: low-demand, high-demand or continuous. The mode decides whether its integrity is judged per demand or per hour.
- MTTFD
- A channel's mean time to dangerous failure: the MTTF counted over dangerous failures only and stated in years; with a constant rate, the reciprocal of the dangerous failure rate.
- OISD
- The Oil Industry Safety Directorate, a technical directorate under India's Ministry of Petroleum and Natural Gas that writes standards for oil and gas installations and describes them as self-regulatory measures. An OISD standard becomes mandatory through its inclusion in a statute such as the Petroleum Rules, 2002.
- OR gate
- A gate of a fault tree whose output event occurs when one or more of its inputs occur. For independent inputs of low probability its probability is close to the sum of theirs, slightly high.
- output override
- An override that holds a safety function's final element whatever its sensors say, and so removes the whole function. PSLG's guidance for overfill protection rules it out, and allows overrides on inputs only where justified.
- overfill level
- A tank's maximum capacity: the highest level consistent with avoiding an overflow or damage to the tank, at the top of the stack of tank levels in the guidance for fuel storage sites.
- overshoot
- An excursion of the air past its set temperature under normal control, before the loop settles. An overshoot of up to 2 °C is acceptable in a 2019 white paper from a maker of incubator test equipment.
- partial stroke test
- A test in which a valve's controller moves the valve through part of its travel while the plant runs, enough to detect a stuck valve without disturbing the process. It exercises only part of the travel, so it reveals only some of the failures that a full stroke would.
- particular standard
- A standard of the IEC 60601 family that sets the requirements for one kind of device and is used with the general standard, IEC 60601-1; for infant incubators, IEC 60601-2-19.
- performance level (PL)
- ISO 13849-1's grade of a safety function, PL a to PL e: a band of the frequency of dangerous failure per hour, from between 1 in 100,000 and 1 in 10,000 hours for PL a to below 1 in 10 million for PL e.
- performance shaping factor
- In SPAR-H, a condition that multiplies a task's nominal error probability, among them the time available and stress; barely adequate time for a diagnosis multiplies it by 10.
- PESO
- The Indian regulator, formerly the Department of Explosives, that administers the Explosives Act, 1884 and the Petroleum Act, 1934 and the rules made under them, and licenses petroleum storage tanks under the Petroleum Rules, 2002.
- PFH
- The frequency of dangerous failures of a safety function per hour, the measure of its integrity in high-demand and continuous modes.
- physiologic closed-loop controller
- A controller whose loop closes through a measurement taken on the patient, the subject of the collateral standard IEC 60601-1-10. A 2017 article from a test-engineering site treats the incubator's skin mode as one.
- plausibility check
- A check that asks whether a reading fits what the process can do, such as how fast a temperature can change or whether a level can stand still while a tank fills.
- point of no return
- The moment after which no action can prevent the hazardous event. The time an operator has runs from the alarm to this point, and every step of the response, with the travel of the final element, must fit inside it.
- presumption of conformity
- In EU product law, the rule that a product built to harmonized standards whose references are published in the Official Journal is taken to meet the essential requirements that those standards cover. A maker that chooses another technical solution shows conformity by its own analysis and tests.
- prior use
- IEC 61511's term for a justification, from use in a similar application and environment, that a device without certification to IEC 61508 may serve in a safety instrumented system.
- probabilistic metric for random hardware failures (PMHF)
- ISO 26262's rate per hour for random hardware failures of a function, set beside the PFH of high-demand mode; its target is at most 10 FIT, one failure in 100 million hours, at ASIL D, and at most 100 FIT at ASIL B and C.
- probability of failure on demand (PFD)
- The probability that a protection layer fails to act when a demand reaches it. A layer with a PFD of 1 in 10 lets one demand in 10 pass.
- process safety time
- The period from a failure that can give rise to a hazardous event to the moment by which action must be complete to prevent the event, as IEC 61508-4 defines it.
- proof test
- A periodic test of a safety function that reveals the dangerous failures its automatic diagnostics cannot find, exercises the function and shows that it works as intended. PSLG requires it for overfill protection, end to end so far as is reasonably practicable.
- proof-test coverage
- The share of the dangerous failures, among those the automatic diagnostics miss, that a proof test reveals; 100 % is a perfect test. The failures outside it accumulate until the equipment is overhauled or replaced.
- proof-test interval
- The time from one proof test to the next, the T of PFDavg ≈ λDU·T/2. PSLG sets it at design from the calculation of the average probability of failure on demand and revises it from operating data.
- protection layer
- Also barrier. Anything that can stop a developing event before it becomes harm: the process design, the basic control system, an alarm with someone to answer it, an automatic safety function, a relief device, containment or emergency response. The inner layers prevent the event and the outer ones mitigate it.
- proven in use
- A claim that a part's records of operation show it free of dangerous failures for long enough, in a use close enough to the new one, to support an integrity level. The records count only the conditions the use contained.
- QM
- The grade below ASIL A, given under ISO 26262 to a hazard whose classes sum to 6 or less, or have any class at zero; no ASIL is set for it.
- QMSR
- The Quality Management System Regulation: FDA's regulation of the quality system that designs and builds a medical device for the US market, 21 CFR Part 820. Since 2 February 2026 it has incorporated ISO 13485:2016 by reference.
- RAGAGEP
- Recognized and generally accepted good engineering practices: the quality that OSHA's process safety rule asks the equipment of a covered process, and its inspection and testing, to meet. The rule names no standard; the employer documents the practice it follows.
- random hardware failure
- A failure of hardware that occurs at a random time and results from one or more degradation mechanisms, such as drift, cracking or wear, whose pace the part's stress sets. Among many parts of one kind such failures can be described by a failure rate.
- range check
- A check that compares each reading with the band of values the measured quantity could really take, and treats a reading outside it as a failure of the sensor or its wiring. It sees an open circuit or a short in a thermistor, but not a drift or a stuck reading.
- ranking reversal error
- The fault of a risk matrix that rates the smaller of two risks above the larger, because the two fall on opposite sides of a band boundary.
- readback
- A check on an output that compares what the logic commanded with what the output does. It sees a failure only when the output is commanded to change, so a cutout's welded contacts show only at a trip or a test.
- recognized consensus standard
- A standard that FDA enters, edition by edition, in its database of recognized standards, with a date of entry and an extent of recognition, complete or partial, and sometimes a condition such as US national differences.
- replacement in kind
- For a safety instrumented system under IEC 61511-1, an exact duplicate of an element, or an approved substitute that needs no modification of the system as installed; any other replacement is a change.
- required performance level (PLr)
- The performance level a machine's safety function needs, read from ISO 13849-1's risk graph of the severity of the injury (S), the exposure to the hazard (F) and the possibility of avoiding the harm (P).
- response time
- The time a safety function takes from the demand to the safe state: the sensor's lag, the logic solver's decision and the final element's action.
- risk
- The combination of how often a deviation leads to harm and how severe that harm is. To compare hazards the combination must be a single number, and the natural choice is the product of frequency and consequence, the expected harm. ISO 14971:2019 defines risk as the combination of the probability of occurrence of harm and the severity of that harm.
- risk control
- In ISO 14971, the measures that reduce a risk, listed in order of preference and effectiveness: inherent safety by design, then protective measures, then information for safety.
- risk graph
- A method that sets an integrity level from four questions about a hazard, its consequence, the exposure to it, the possibility of failing to avoid the hazardous event and the probability of the unwanted occurrence without the function, each answered by choosing a class. It gives a sound answer only when each class stands for a number calibrated against the tolerable frequency.
- risk management file
- The maker's record of its risk management for a device: for the incubator, the hazard study, the FMEDA and fault tree, the controls chosen for each hazard and the residual risks with their weighing. The single faults that IEC 60601-1's tests apply are documented in it before testing.
- risk matrix
- A grid with bands of frequency on one side and bands of consequence on the other, in which each cell carries a verdict. It sorts risks quickly, but one cell can hold risks far apart.
- risk reduction factor (RRF)
- The factor by which a new layer must cut the frequency of a consequence: the frequency with every credited layer in place divided by the frequency judged tolerable. The layer may then fail on no more than one demand in that factor.
- route 1H
- The route to the architectural constraints on which the highest integrity level a subsystem may claim depends on its hardware fault tolerance, its safe failure fraction and whether its element is of type A or type B.
- route 2H
- The route to the architectural constraints on which the least hardware fault tolerance follows from the integrity level alone, with failure rates from field feedback on equipment in similar use, held with stated confidence.
- safe detected failure (SD)
- A safe failure that a check announces, such as a short circuit in a protective thermistor that the channel trips on and reports as a sensor fault.
- safe failure
- A failure that pushes the equipment toward its safe state. It can cost production or service, such as a train held at a signal or an incubator left to cool, but not safety.
- safe failure fraction (SFF)
- The share of all failures that are either safe or dangerous detected, (λS + λDD)/(λS + λD). IEC 61508's architectural constraints set the least redundancy a subsystem must have by one of two routes: on route 1H it follows from the SFF, the integrity level sought and whether the element is of type A or type B, and route 2H rests on failure data from the field and leaves the SFF out.
- safe state
- The state a safety function brings the equipment to, safe against the hazard it was built for. In the incubator it is the heater without power, held until someone resets the cutout.
- safe undetected failure (SU)
- A safe failure that gives no warning and, if it acts, shows itself later, as a trip without a demand.
- safety case
- A structured argument, supported by a body of evidence, that a system is safe for a given application in a given operating environment, in the definition of the UK Ministry of Defence's Def Stan 00-56, Issue 6.
- safety component
- Under the EU's Machinery Regulation, a physical or digital component, software included, that is designed to fulfill a safety function and is placed on the market on its own.
- safety function
- One job aimed at one hazard: a sensor, a logic solver and a final element in a row, which bring the equipment to its safe state when a demand comes and do nothing until then.
- safety instrumented function (SIF)
- IEC 61511's name for a safety function carried out by sensors, a logic solver and final elements; together they make up a safety instrumented system.
- safety integrity level (SIL)
- One of four bands, SIL 1 to SIL 4, each one tenfold step wide, in which IEC 61508 states how rarely a safety function may fail: a range of PFDavg in low-demand mode and of PFH in high-demand and continuous modes.
- safety lifecycle
- The ordered set of steps through which a safety function passes, from the first idea of the equipment to its disposal, with what each step must produce and how its result is checked. IEC 61508 divides its overall safety lifecycle into 16 phases.
- safety management system
- An organization's own process for managing the safety risk of its work. FAA's rule of 2024 extends its requirement for one to holders of a production certificate who hold or license the type certificate for the same product, and to type certificate holders who license their certificate out for production.
- safety manual
- The maker's document that an element needs in order to comply with IEC 61508, giving the conditions on which its stated figures hold: its failure modes and rates, diagnostics, fault tolerance, type, systematic capability, useful life, proof-test requirements and how to report failures.
- safety of the intended functionality (SOTIF)
- The subject of ISO 21448: in ISO's summary, the absence of unreasonable risk due to hazards that result from functional insufficiencies, where a function needs situational awareness and causes harm with nothing malfunctioning. A second sensor of the same design shares the same insufficiency.
- safety requirements specification
- The document that states, for each safety function, what it must do and how reliably: its functional requirements, such as the safe state, the trip point and the response time, and its integrity requirements, such as the integrity level, the target, the demand rate and the proof-test interval.
- separation
- Keeping redundant channels apart, so that one place, one supply or one stress does not reach both: sensors mounted apart, separate power supplies, separate wiring.
- set temperature
- The controller's set point for the incubator's air: the temperature at which the controller holds the air around the newborn. In skin mode the air has no set temperature.
- single fault condition
- In IEC 60601-1, the state in which a single means of reducing a risk is defective or a single abnormal condition is present. The device must stay free of unacceptable risk in it, so a hazard that one failure can release needs either one means of reducing the risk whose probability of failure is negligible or a second means that the same failure does not reach; with a second means, a first fault must be detected during the expected service life or the second means be negligibly likely to fail.
- single-point fault metric (SPFM)
- One of ISO 26262's hardware metrics: a percentage concerning faults that act on the function at once, which comes closer to 100 % the smaller the share of the hardware's failure rate made up of such faults with nothing to detect them or turn them safe. Its target is at least 99 % at ASIL D, as chip makers tabulate it.
- skin mode
- The incubator's mode in which the controller holds the infant's skin, measured by a probe fixed to it, at a set value instead of the air, and heats the air as much as it must to bring the skin to that value; the air then has no set temperature. IEC 60601-2-19 calls an incubator run this way "baby controlled".
- societal risk
- The relation between the number of people killed in one accident and how often accidents of that size occur, drawn as an F-N curve. It measures what individual risk and expected harm leave out: whether the harm arrives in large accidents.
- software control category
- In MIL-STD-882E, the grade of how much control a software function has over a hazard: autonomous, semi-autonomous, redundant fault tolerant, influential (only producing safety-related information for an operator) or no safety impact.
- software criticality index
- In MIL-STD-882E, a number from 1 to 5 given by a software function's control category and the severity of the hazard, which sets the level of rigor of its analysis and testing; 1 is the most rigorous.
- software safety class
- IEC 62304's grade for medical device software, A, B or C. It follows from whether the software can contribute to a hazardous situation whose risk stays unacceptable once risk controls outside the software are in place, and then from the severity of the possible harm: A where it cannot, B where it can and the harm is not serious injury, C where it can and the harm is death or serious injury.
- specific
- Of a protection layer: designed to act against the scenario in question and its consequence. A bund is specific to a spill of liquid and not to a cloud of vapor.
- spurious trip
- A trip with no demand behind it, from a safe failure that acts or from a detected failure on which the channel trips. A 2014 textbook counts a third cause in a function's trip rate: false demands, which are either signals mistaken for a demand or real demands that no one intended. Under IEC 61508 a spurious operation counts as a safe failure, so its rate is set as a requirement of its own.
- statement coverage
- A structural coverage measure met when every statement of the program has run at least once under test.
- static analysis
- Examination of a program's code without running it, such as control-flow analysis, which follows every path the program can take, and data-flow analysis, which follows each variable from where it is set to where it is used.
- structural coverage
- The share of a program's structure, its statements, branches or conditions, that the tests written from the requirements ran.
- systematic capability
- A grade, SC 1 to SC 4, of the confidence that a part's systematic safety integrity meets the requirements of the integrity level with the same number, when the part is applied as its safety manual instructs. IEC 61508:2010 gives three routes to it: the techniques and measures of the standard, proven in use, and, for existing software, the requirements IEC 61508-3 sets for it.
- systematic failure
- A failure that follows in a deterministic way from a cause built into the specification, design, manufacture, installation, operation or maintenance of an item, or a later change, and occurs whenever the conditions that call on it occur. It is removed only by changing the design, the process, the procedures or the documents.
- systematic fault
- The defect behind a systematic failure, such as a limit typed in the wrong unit: built into a design, dormant until the conditions that call on it occur, and carried by every copy made to that design.
- tank rated capacity
- A level below a tank's overfill level, set low enough to leave time for whatever action is needed to stop the overfill level being reached; the high-high level alarm is set at or below it.
- technical alarm condition
- An alarm condition that arises from the equipment or the alarm system, such as a disconnected sensor, rather than from a monitored variable of the patient.
- thermistor
- A resistor whose resistance changes with its temperature, used as a temperature sensor. In an NTC thermistor, one with a negative temperature coefficient, the resistance falls as the part warms.
- tolerable
- Of a risk: lived with, not accepted, on condition that it is reduced as far as is reasonably practicable. In HSE's framework a risk between the upper and lower lines is tolerable only on that condition.
- tool qualification
- The evidence that an off-line support tool does what its specification says, required in IEC 61508-3:2010 for T3 tools, whose output becomes part of the executable code. Each new version of an off-line support tool is to be qualified again.
- top event
- The one undesired event from which a fault tree is drawn, such as an infant overheated in an incubator.
- trace data
- In DO-178C, a life cycle data item that shows the associations that traceability makes; in FAA's summary of the changes from DO-178B, it should show them in both directions.
- traceability
- The record that links each requirement back to the hazard it answers and forward to the design element that meets it and the tests that show it is met, so that a requirement without a test, and every item a change reaches, can be seen.
- trip
- The act by which a protective function brings the equipment to its safe state, such as a cutout opening the heater's supply. In a safety function it is the act in which the logic solver drives the final element to its protective position.
- type A and type B
- Type A: an element whose failure modes are well defined, whose behavior under fault can be fully determined and whose failure data are dependable. Type B: one that falls short of any of these. For the same SFF and HFT, route 1H allows a type A element the same or a higher level.
- undetected failure
- A failure that nothing makes evident when it happens. A safe one, if it acts, shows itself later by its effect, such as a trip without a demand; a dangerous one stays hidden until a test or a demand.
- useful life
- The middle stretch of the bathtub curve, in which a steady share of the parts still working fails in each hour, so that one constant failure rate describes the parts.
- validation
- Confirmation, by examination and objective evidence, that the finished, installed safety function meets its specification as a whole, for the use it was specified for; phase 13 of IEC 61508's lifecycle.
- verification
- Confirmation, by examination and objective evidence, that the output of a lifecycle step meets the requirements that step was given, such as a design against its specification or code against its design.
- vote
- The rule by which several channels decide to trip, written as how many must call out of how many there are: 1oo2 trips when either of two channels calls, 2oo2 only when both do, 2oo3 when any two of three do.
- watchdog
- A timer, running apart from a program, that the program must service within a set time, or the watchdog resets or interrupts the processor. A windowed watchdog also treats servicing that comes too soon as a fault. It checks timing only.
- β factor
- The fraction of a channel's failures that are common cause failures, failures that also fail the other channel of a redundant pair. IEC 61508-6 applies β to undetected failures and a second factor, βD, to detected ones, and its Annex D estimates β from a checklist score.