+ Part IV · Systematic failure
Confidence without a rate.
A systematic failure is built in, so it is prevented by how the work is done, not by spares. A second copy of a wrong limit fails on the same demand as the first, and no count of past hours puts a number on a fault that waits for its conditions. What a standard can ask for instead is evidence about the work: the methods used to specify, design, code and check a part, how many of them each integrity level calls for, and what a record of use can show about the faults those methods missed. Chapter 15 sets out how IEC 61508 grades that evidence, and what becomes of it when a part, a tool or a program is carried into a new use. Chapter 16 applies it to software, whose failures Chapter 3 found to be systematic.
Earned by method.
+ The questionIf systematic failures have no rate, how does a standard ask for confidence in them?
A grade for systematic faults, and three routes to it
A systematic fault has no rate, as Chapter 3 found. It sits in every copy and causes a failure whenever its conditions occur, so no count of parts or hours says how rare its failures are. The integrity levels of Chapter 8 are set for the random share of a function, so a standard that sets them needs a second measure for the systematic share, one that rests on how the function was specified, designed and checked. IEC 61508:2010 calls that measure systematic capability and grades it from SC 1 to SC 4. Systematic capability is a claim about how a part was made, graded like an integrity level but earned by method.
The grades match the integrity levels number for number. SC 3 expresses confidence that a part's systematic safety integrity meets the requirements of SIL 3, while its random failures are counted separately, as Part III counted them. The standard's definition adds a condition: the confidence holds when the part is applied as its safety manual instructs (Chapter 21). A grade is therefore a statement about a part in a described use, and it says nothing about uses that the manual does not describe.
IEC 61508-2 sets three routes by which a part can reach its grade. Route 1S calls for the part to be developed with the techniques and measures that the standard sets for avoiding and controlling systematic faults, and for that development to be documented. Route 2S rests on the part's record in service. Route 3S applies to software that already exists, and it asks that the software meet the requirements IEC 61508-3 sets for such software. The first route draws its evidence from how the part was made and the second from its record, and both rest on assumptions about the use.
A grade can also be built from parts. Two parts in parallel, with sufficient independence between them, may claim one grade above the lower of the two. Two SC 2 parts may thus together serve as SC 3, and no number of parts takes the claim more than one step higher. The independence has to be justified by an analysis of common cause failure, the subject of Chapter 14.
The tables grade the method by level
No single technique finds every systematic fault, because each looks for faults of one kind. A coding standard keeps out constructs that are easy to get wrong, static analysis finds faults without running the code, an inspection compares the code with its requirement, and structural coverage shows which parts of the code the tests ran. The standard asks for more of these techniques at a higher integrity level, so that a fault missed by one method meets another. Annexes A and B of IEC 61508-3 set out the techniques in tables and grade each for each SIL. The techniques that follow are graded highly recommended, recommended or with no recommendation, as Parasoft, a maker of software-testing tools, maps them.
SIL 3 is the higher of the two levels Chapter 8 found for the incubator's cutout. In the SIL 3 column the tables highly recommend a strongly typed programming language and a restricted subset of it. A strongly typed language mixes kinds of value only where the code converts them explicitly. The tables also highly recommend a coding standard, defensive programming, formal inspections, and tests that run every statement and every branch. Coverage to MC/DC, a stricter measure that Chapter 16 explains, is recommended at SIL 3 and highly recommended only at SIL 4. Diverse redundancy, two programs written to one specification, is recommended only at SIL 4. Two such programs still share any fault of the specification they follow, as Chapter 14 showed.
A record counts only what the use contained
A part developed without the tables' techniques can still earn a grade from its history. A part is proven in use when records of its operation show it free of dangerous failures for long enough, in a use close enough to the new one, to support the claim. The minimum evidence includes a formal system for failure data that separates safe from dangerous failures, each device's hours in use, a clearly comparable application and the firmware revision, all reviewed regularly. The 61508 Association, an industry body, sets that minimum in its 2023 policy.
In the process sector the same idea is called prior use, the term of IEC 61511. One reading of IEC 61511 comes from exida, an assessment and certification company, in a 2025 post that Gandy wrote for its blog. On that reading, the standard accepts a device for a safety instrumented system by "two (and only two) means": certification to IEC 61508 for the level required, or a justification from prior use. The same reading holds that previous use alone is not sufficient: the justification must show use without dangerous failures in a similar application and environment, a maintained system of field-failure data, and the same hardware and software versions.
How long is long enough follows from the arithmetic of a record with no failures. Such a record does not show that the chance of failure is zero. It shows that a high chance would have been unlikely to leave so clean a record. If a part failed on 1 demand in 1,000, a run of 3,000 demands would bring three failures on average, and a run with none would occur only about 1 time in 20. The same rule sets the record for every integrity band, in the tables of a 2015 statistical paper by Ladkin and Littlewood: about 3 times the inverse of the target at 95 % confidence, and 4.6 times at 99 %.
Chapter 3's fleet holds 1,000 incubators of one design, all on one firmware version. Each is called on 0.1 times a year, the stand-in demand rate of Chapter 8, so the fleet sees 1,000 × 0.1 = 100 demands a year. The upper edge of SIL 3 is a probability of failure on demand of 1 in 1,000. To support a figure below it at 95 % confidence, the record needs about 3,000 demands without a dangerous failure: 3,000 ÷ 100 = 30 years. At 99 % it needs about 4,600 demands, or 46 years, and for SIL 2, below 1 in 100, about 300 demands, or 3 years. A demand counts only if it is recorded with how the cutout answered it. By use alone, the fleet could support SIL 2 in about 3 years and SIL 3 only after decades, since each level up multiplies the record by 10.
A record is also evidence only for the conditions its demands contained. A fault that causes a failure on every demand, such as Chapter 3's typed limit, shows at the first one, but a fault that waits for a rarer condition is tested only by the demands that meet it. The cutout uses its 39.5 °C limit only for set temperatures above 37 °C (Chapter 4). A fault in that branch of its firmware is therefore exercised only by demands that arise at such settings. A fleet that seldom runs above 37 °C gathers little evidence about that branch, however many demands it records at lower settings.
A fleet of 1,000 incubators runs about 8.76 million hours a year, which looks like evidence for every part inside. Those hours are no evidence for the cutout's dangerous failures, because in normal operation the cutout does nothing, and only demands and tests call on its firmware (Chapter 11). In the view of an Analog Devices employee writing on the chip maker's blog in 2021, the records must also tell systematic failures apart from random ones. The same author adds that a failure in the field may disappear quickly and be hard to reproduce, which makes that sorting difficult.
Tools that write the code are part of the method
A part's method includes the tools that turned its design into code and checked that code. A compiler translates every line of a program, so a fault in the compiler can place a systematic fault in code whose source is correct, where no review of the source will find it. A coverage tool or a static analyzer cannot add a fault, but a fault in either tool can hide one by reporting a test as complete or a defect as absent. A text editor produces nothing that reaches the executable code at all. IEC 61508-3:2010 sorts off-line support tools by these differences. T1 tools produce no output that can contribute to executable code, T2 tools can fail to reveal defects but cannot directly create errors in that code, and T3 tools, an optimizing compiler among them, produce output that contributes to it.
For a T3 tool the standard asks for evidence that the tool conforms to its specification or documentation, records of its validation with the result of each test, and its version held in the project's configuration baseline. That work is tool qualification. Its evidence can come from validating the tool, from confidence gained in its use, from an evaluation of how it was developed, or from its development under a safety standard. The standard also asks that each new version of an off-line support tool be qualified again, as UL Solutions, a certification body, quotes the requirement in slides for the 61508 Association. The techniques tables grade the trust placed in tools as well: certified tools and translators are recommended at SIL 1 and highly recommended from SIL 2, and tools and translators with increased confidence from use are highly recommended at every level.
ISO 26262-8:2018, for road vehicles, asks two questions of a tool: whether its malfunction could introduce an error or fail to detect one, and how surely such an error would be prevented or detected. Under ISO 26262-8, a tool needs no qualification if its malfunction could neither introduce nor miss an error, or if its errors would be prevented or detected with high confidence. Engineers at a design-tool maker and a chip maker, and a consultancy that sells tool qualification, describe the rule in these terms. UL Solutions' slides, dated June 2024, also describe a draft of IEC 61508's third edition. That draft classed a tool by its impact, TIC1 to TIC3, by the detection of its errors, TD1 to TD3, and by its use for safety, and it exempted TIC1 tools from qualification. The classes in force are the 2010 edition's (Chapter 8).
Reuse carries the assumptions of the old use
Ariane 501's inertial reference units carried code from Ariane 4, and with it the conditions under which that code had been judged safe. Chapter 14 told how a conversion in the alignment function, still running in flight, stopped both units. The function was not required on Ariane 5. It had been added to Ariane 4 more than 10 years before, so that the count-down could resume quickly after a hold late in the sequence, since a fresh alignment takes 45 minutes or more. It had been used once, on Flight 33 in 1989. It was kept for commonality. In its report of 1996, the Inquiry Board presumed that choice to rest on the view that it was "not wise to make changes in software which worked well on Ariane 4".
The value that overflowed had been left unprotected on a judgment about Ariane 4. The Board was told that, to keep each unit's computer within a workload target of 80 %, protection was added to four of the seven variables at risk of the error and left off the other three. Those three were judged physically limited or to have a large margin of safety. For the horizontal bias that judgment held on Ariane 4's early trajectory and failed on Ariane 5's (Chapter 14). The source code recorded no justification for the judgment, and the Board found no evidence that trajectory data were used to check it. The units' specification declared no restriction on their use, and by joint agreement it left out Ariane 5's trajectory.
The code's record on Ariane 4 had been earned on Ariane 4's trajectory. The Board found that no test had checked the units against Ariane 5's count-down, flight sequence and trajectory, and that such a test, by the supplier or at acceptance, would have exposed the failure mechanism. It also saw a definite risk in assuming that critical equipment had been validated by its qualification alone or by its previous use on Ariane 4. Its recommendations included explicit restrictions on use, a review of flight software for implicit assumptions and for the ranges of its variables, and trajectory data in specifications and test requirements.
An installation instruction carries an assumption of the same kind from a part's maker to whoever fits it. Buncefield's switch was designed so that a padlock held its test lever in the operating position, and Chapter 11 tells how the switch was fitted without one. The padlock's role as a safety feature was a condition of the design, and that condition had to reach the fitter as an instruction. In its 2011 report the Competent Authority found that the information from TAV Engineering, the switch's maker, did not make the key aspects of the switch's design and use sufficiently clear. It also found that the ordering of the switch fell short on both sides, the maker's and the installer's (Chapter 17). The report states that TAV was aware its switches were used in high-hazard installations, and that clear guidance on the padlock's safety criticality should have passed to installers and users. In the report's view, a rigorous review of the design changes could have removed the fault in the design early.
Both cases break the condition written into the definition of systematic capability, that evidence about a part holds for a described use. A reused program carries the ranges it was checked against, and a switch carries the way it must be fitted and tested. The safety manual of Chapter 21 is where a maker writes such conditions for the user. Where the conditions are not written down and checked against the new use, the evidence from a part's method and from its record covers the old use and not the new one.
Read IEC 61508-3's techniques tables down the column for the function's SIL, and record for each technique whether and where the project applied it. Class every tool that touches the code as T1, T2 or T3, keep each T3 tool's version in the configuration baseline, and qualify every new version before use. Claim proven in use only from records of the same hardware and firmware versions, with every demand and every failure counted and sorted into safe and dangerous. Before reusing a part, list the conditions it was checked against, its ranges, its inputs and its installation and test instructions, and check each one against the new use. For medical device software, follow IEC 62304's activities and its handling of software of unknown provenance as The next patient, the next update sets them out in Chapters 3 to 8.
- Systematic failures have no rate, so IEC 61508 grades confidence in them as systematic capability, SC 1 to SC 4. The grade is earned by method, by a record of use or, for existing software, by meeting the requirements IEC 61508-3 sets for it, and it holds only for the uses a part's safety manual describes.
- IEC 61508-3's tables grade each technique by SIL. For SIL 3, the higher of the incubator's two levels in Chapter 8, they highly recommend a strongly typed language, a coding standard, formal inspections and tests of every statement and branch.
- A record without failure supports a claim in proportion to its size and only for the conditions it contained. A claim below 1 in 1,000 per demand at 95 % confidence takes about 3,000 demands, about 30 years for 1,000 incubators that together see 100 demands a year.
- Tools whose output becomes part of the executable code must be qualified, and each new version of an off-line support tool qualified again. Reused parts carry the assumptions of their old use: Ariane 501's alignment code carried Ariane 4's trajectory, and Buncefield's switch a padlock whose safety role appears not to have reached its fitter.