Single fault, risk and software.
+ The questionWhere does a device's safety sit between IEC 60601-1, ISO 14971 and IEC 62304, and why do the device standards use no SIL?
One fault at a time
IEC 60601-1 asks that a medical electrical device stay safe with any one fault present. It calls that state the single fault condition: a single means of reducing a risk is defective, or a single abnormal condition is present. Each fault is taken on its own, in a device otherwise working as designed, and the device must show no unacceptable risk with that fault in place. A hazard that one failure can release therefore needs either one means of reducing the risk whose probability of failure is negligible, such as reinforced insulation, or a second means that the same failure does not reach. In the incubator two faults drive the air too hot: a control thermistor that reads low and a heater driver stuck on. The cutout, on its own sensor, is the second means for both (Chapter 4). In the terms of Chapter 18, a device that relies on second means has a fault tree with no cut set of one: no single failure leads to the harm on its own.
A second means and the control it stands behind form a pair. The pair protects only while its second member works, and in normal use the second member does nothing, so a failure in it gives no sign (Chapter 11). A first fault is acceptable on one of two terms: it will be detected during the device's expected service life, or the probability that the second means of reducing the risk fails is negligible. The first term is met by finding the fault, through the automatic checks of Chapter 12 or a periodic test that drives the protective sensor past its limit, as Chapter 11 asks of the yearly service. The second term is a judgment of probability, recorded in the maker's risk management file, on which the standard's clauses for single faults and for risk management work together.
Basic safety is freedom from unacceptable risk caused directly by physical hazards, in normal condition and in single fault condition. Essential performance is the performance of a clinical function, other than one related to basic safety, whose loss or degradation beyond the limits the maker specifies would bring unacceptable risk, and each maker identifies it for its device. The simplest test is whether the absence or degradation of a performance would bring such risk, as FDA's guidance on electromagnetic compatibility suggests. For the incubator, the temperature it holds passes that test. A note to the general standard adds that the performance of a risk control measure might well become part of the essential performance. The cutout's action can therefore count among what the device must keep.
The probe off the skin
Skin mode puts the infant inside the loop. The controller heats the air until the probe on the skin reads the set value (Chapter 4). A probe can come off the skin and stay connected. It then measures whatever it lies against, and the loop goes on holding that reading at the set value. If the reading is lower than the temperature of the infant's skin, the controller heats the air further. The skin alarm is set 1 °C either side of the set value and reads the same probe, so it shares the fault, as the alarms on Tank 912 shared the stuck gauge (Chapter 7).
The fault has been reported in service. A maker's report to FDA of 2010, which Chapter 5 cites, describes one of its incubators in skin mode with a set value of 36 °C and the probe off the infant. By its account, the air read 38.4 °C, and the infant was febrile, at 102.9 °F, or about 39.4 °C. The facility reported that the incubator did not alarm, and the maker found the probe unattached and named it as the cause. The report records the temperature of the air, not that incubator's limit for it. A cutout on the air answers the faults that drive the air past its limit, and it leaves a fault in the measurement of the infant unanswered for as long as the air stays below that limit.
In the incubator as Chapter 4 sets it up, nothing registers a probe that has left the skin. An alarm system can treat a disconnected sensor as a technical alarm condition, one that arises from the equipment rather than from the patient. A probe that stays connected, though, goes on giving a plausible reading. Applied to the probe off the skin, the single fault condition thus finds a fault that the protective channel was not built to answer, because the channel guards the air and the fault lies in the measurement of the patient.
Risk control in ISO 14971's order
Controls differ in how many things must go right for them to act. A design that removes the hazard leaves nothing to fail. A protective measure acts only while it works, and a failed one can stay hidden until it is needed (Chapter 11). Information for safety acts only when someone reads it and follows it. That is the order of preference and effectiveness in which ISO 14971:2019 (Chapter 5) lists the options for risk control: inherent safety by design, then protective measures, then information for safety. In 2025 FDA cleared an incubator that has no skin mode, which its maker left out "to avoid the risk of masking clinical conditions", and without the mode there is no probe to come off the skin.
The cutout and the alarms are the incubator's protective measures. An infusion pump's protections against air in the line and against free flow are of the same kind, and so is a surgical generator's single-fault design; Chapters 13 and 14 of Every drop accounted for and Chapter 20 of Where the heat goes follow them. A protective measure against the probe off the skin would have to register the probe's contact or compare its reading with a second measurement, as the cross-check of Chapter 12 compares two sensors. Information for safety, such as an instruction to fix the probe to the skin, comes last: the foil of Chapter 5 was laid over the air vents against a warning in the instructions for use.
The risk that remains after the controls is the residual risk of Chapter 6, and a benefit–risk analysis justifies why the benefits outweigh it. The risk analysis behind that weighing follows each hazard along ISO 14971's chain of hazard, sequence of events, hazardous situation and harm. In the incubator the hazard is the heater (Chapter 5). In skin mode the probe leaving the skin begins a sequence of events, and the infant lying in air that the loop keeps heating is the hazardous situation, the circumstance in which the infant is exposed to the hazard. The overheated infant is the harm. Skin mode holds the infant's own temperature rather than the air's, and the chain comes with the mode, so a maker that keeps skin mode weighs the chain's residual risk against what the mode gives the infant.
The maker's record of the whole analysis is the risk management file. For the incubator the file holds the hazard study of Chapter 5, the FMEDA and the fault tree of Chapter 18, the controls chosen for each hazard, and the residual risks with their weighing. The same file serves IEC 60601-1. The conditions and the single faults to be tested are documented there during design, before testing begins, so the tests check the analysis rather than stand in for it.
The standards as a stack
A device's safety requirements come in layers: what every medical electrical device must meet, what one subject such as alarms adds for all devices, and what one kind of device needs. The general standard, IEC 60601-1, is in force as Edition 3.2, the 2005 text with amendments of 2012 and 2020. It sets requirements for basic safety and essential performance that apply to medical electrical equipment generally. A particular standard sets the requirements for one kind of device and is used with the general standard, and where a particular standard exists, the general standard should not be used alone. For infant incubators the particular standard is IEC 60601-2-19, in force as Edition 3.1, the 2020 text with its 2023 amendment. Transport incubators and infant radiant warmers have particulars of their own.
Between the general standard and the particulars sit the collateral standards, each covering one subject for medical electrical equipment in general. IEC 60601-1-8 is the collateral standard for alarm systems, in force as Edition 2.2, and an alarm system that a maker has fitted must comply with it. Chapter 9 sets out the standard's alarm priorities. On Chapter 4's reading, closed-loop controllers such as skin mode come under IEC 60601-1-10, in force as Edition 1.2. That standard sets requirements for their development, from analysis and design to verification and validation.
Beneath the device standards lie three that govern how a device is made. ISO 14971:2019 governs risk management, and the general standard's clauses draw on it. IEC 62304:2006, with its 2015 amendment, governs the life cycle of the software, and ISO 13485:2016 the quality system in which design, risk management and the software's development run. IEC 62304's software safety classes rest on the severity of the possible harm (Chapter 16). Chapters 3 to 8 of The next patient, the next update take IEC 62304's activities in turn. How each market's law takes up the stack is the subject of Chapters 28 to 30.
Decomposition on the left, integration on the right
Edition 3.2 of IEC 60601-1 rests on the 2005 text, Edition 3. In that text, clause 14 adds a development life cycle for programmable electronics to the single-fault tests. The clause covers programmable electronics that bear on basic safety or essential performance, unless the risk analysis shows that their failure brings no unacceptable risk. The development life cycle must be documented, and at each milestone the activities to be finished and the methods that verify them must be defined. Every function that implements basic safety, essential performance or a risk control measure must be verified, and no member of the design team may be responsible for validating their own design.
An informative annex of the same text draws this life cycle as decomposition on the left and integration on the right. Each layer of design has its layer of integration and verification, and validation comes at the end. That is the arrangement of Chapter 19's V, although the annex does not call it a V and requires no particular life cycle.
IEC 62304 and ISO 14971 sit beside the V. The V is this book's frame for the two standards, not a frame of their own. IEC 62304 prescribes no life cycle model, yet it asks that software requirements be traced to the system requirements and to the tests that cover them, and that a trace be recorded for each software hazard. ISO 14971 sets out a risk process. Its risk management file gives traceability for each identified hazard, and each risk control is verified both as implemented and as effective.
In the United States a device's design falls under design controls, requirements on the steps of the design's development, among them planning, inputs, outputs, review, verification, transfer to production and change. The Quality System Regulation required them, and since 2 February 2026 the Quality Management System Regulation, or QMSR, has set those requirements through ISO 13485:2016. FDA's guidance of June 2023 on software in premarket submissions recommends no life cycle model either, and among the verification activities it lists traceability analysis, which links requirements to design and design to code in both directions, and tests to both.
- Essential performance, which the maker identifies for its device, may include the performance of a risk control measure, such as the cutout's action.
- The single fault condition covers a single abnormal condition as well as a defective means of reducing a risk, and the risk analysis decides which failures are tested. On this book's reading, a probe that leaves the skin while the air stays below the cutout's limit is one such condition.
- The single faults to be tested are documented in the risk management file during design, before testing begins.
- Programmable electronics are verified function by function, no member of the design team is responsible for validating their own design, and the risk management file keeps a trace for each hazard.
Where the method of IEC 61508 fits
Most of this book's method carries over to the device. The hazard study of Chapter 5 begins ISO 14971's analysis, and the protective channel of Chapter 4 is a protective measure in the standard's order. The FMEDA and fault tree of Chapter 18 find the single faults that the tests apply. The proof tests and diagnostics of Chapters 11 and 12 meet the single fault condition's first term by finding a first fault in service. The software methods of Chapter 16 have their counterpart in IEC 62304's activities. What does not carry over is the integrity level.
A SIL grades a function by a band of failure probability, set from the risk reduction needed to reach a tolerable frequency of harm (Chapter 8). The device standards state their requirements in other terms: tests passed one fault at a time, and a residual risk weighed against the benefit to the same patient (Chapter 6). A weighing against one patient's benefit fixes no tolerable frequency in advance, so no band follows from it. Chapter 8 took its target of 1 in 10,000 a year from HSE's framework for members of the public, so the SIL that this book gives the incubator's channel is IEC 61508's reading of a device. The arithmetic of Chapter 8 can still put a number on the single fault condition's second term.
The faults in the control loop that drive the air past the cutout's limit are the causes of a demand, and they come about 0.1 times a year, the stand-in figure of Chapter 8. The second means is the protective channel, whose sensor alone averages a probability of failure on demand of about 3.1 in 10,000 when a yearly test drives it past its limit (Chapter 11). The air then overheats with an infant inside about 0.1 × 0.00031 = 0.000031 times a year, or once in about 32,000 years for each incubator, inside Chapter 8's target of 1 in 10,000 a year. Chapter 11 assumed a mission time of 10 years. Left untested for that long, the same sensor at 71 FIT averages 71 × 87,600 ÷ 10⁹ ÷ 2, about 3.1 in 1,000. The air then overheats with an infant inside about 0.00031 times a year, about three times the target. Measured against that target, the second means stays negligibly likely to have failed only while the test keeps finding its hidden failures. Both figures cover the protective sensor alone, since the comparison and the cutout add their own shares.
Each single-fault test applies one fault to a device whose other parts work, so a pass shows that the second means acts on a device fresh from the bench. A pass cannot show that the second means is still working years later, and in service the faults can arrive in the other order: the hidden failure of the second means first, then the fault it was there to answer. The solved example puts a factor of 10 on that difference. The service documents therefore carry the test that finds the hidden failure, and the risk management file records the test interval that its judgment of negligible probability assumed.
Take each failure mode from the FMEDA of Chapter 18 and apply it as a single fault in every mode of operation, skin mode included, recording what bounds it and how a failure that gives no sign will be found in service. List faults of measurement, such as a probe that leaves the skin and stays connected, as well as faults of parts. For each hazard that remains, choose controls in ISO 14971's order, and record why a protective measure or information was chosen where a design could have removed the hazard. Name the essential performance from the limits the device must hold, and count in it any risk control whose loss would bring unacceptable risk. Weigh the residual risk of each mode against what the mode gives the patient, and complete the analysis before the tests begin.
- IEC 60601-1 asks that a device stay safe with any one fault present, so each hazard that one failure could release needs either one means of negligible failure probability or a second means. With a second means, a first fault must be detected in service, or the second means must be negligibly likely to fail.
- In skin mode a probe that leaves the skin misleads the loop and the skin alarm alike. A cutout on the air leaves that single fault unanswered for as long as the air stays below the cutout's limit.
- ISO 14971 orders risk control from inherent safety by design through protective measures to information for safety, and the residual risk is weighed against the benefit to the patient in the risk management file.
- The general standard, its collaterals and the incubators' particular rest on ISO 14971, IEC 62304 and ISO 13485. Together they state their requirements in single-fault tests, residual risk weighed against the patient's benefit and software classes set by severity, not in a band of failure probability.