Two channels, three channels.
+ The questionHow do 1oo2 and 2oo3 trade a missed trip against a false one?
Either channel can cut the heater
A second sensor, wired so that either sensor can cut the heater, makes the incubator safer and less available at the same time. The second protective channel has its own thermistor, its own comparison and its own cutout, and the two cutouts sit in series in the heater's power path, so either one opening removes the power. The heater then stays on through a demand only if both channels have failed dangerously. It goes off with no demand, and stops warming the infant, if either channel calls for a trip falsely, and two channels can now make that call.
A trip with no demand behind it is a spurious trip. It can follow from a safe failure that acts, of the kind Chapter 2 described. It can also follow from a detected failure on which the channel trips, the first of the two reactions that Chapter 12 set out. Rausand's 2014 textbook on safety-critical systems counts a third cause in a function's trip rate beside these two: false demands, which are either signals mistaken for a demand or real demands that no one intended, such as a welding flame seen by a fire detector. Whatever the cause, the cutout opens, the incubator enters the safe state of Chapter 4, and the infant cools until the staff take over. A spurious operation counts as a safe failure under IEC 61508, so no integrity level limits how often a function trips without a demand. The rate of spurious trips is set as a requirement of its own, in the safety requirements specification (Chapter 17).
Redundancy is described in channels and votes. A channel is one path that can carry out the safety function, or one subsystem's share of it, by itself, such as a thermistor with its comparison and cutout. The vote is the rule by which several channels decide to trip: a function voted two out of three, written 2oo3, trips when at least two of its three channels call for a trip. The incubator of Chapter 4 is 1oo1, one out of one, and with its second channel it becomes 1oo2. A 2oo2 function trips only when both of its channels call. The channels and their vote together make the function's architecture.
The vote sets at once both ways in which the function can go wrong. A function misses a demand when too few working channels remain to make the call, and it trips spuriously when enough channels call without a demand. In 1oo2 one channel's call is enough either way, so 1oo2 is the hardest vote to defeat and the easiest to trip. A 2oo2 vote is the reverse: either channel's dangerous failure defeats it, and a false trip needs false calls from both. The number of hardware faults that an architecture can have and still perform its safety function is its hardware fault tolerance, or HFT. It is 0 for 1oo1 and 2oo2, and 1 for 1oo2 and 2oo3.
Both channels must have failed
A 1oo2 function misses a demand only if both channels have failed dangerously before it arrives. If the two fail independently, each for its own reasons, as Chapter 3's two thermistors would, the chance that both have failed is the product of the two chances, as at the AND gate of a fault tree (Chapter 18). After a proof test that leaves both working, each chance climbs from zero along a nearly straight line (Chapter 11). Their product therefore climbs as a square, slowly at first: halfway through the interval it is a quarter of its final value, not a half.
A demand at a random moment meets the average of that climb, as in Chapter 11. A straight climb from zero averages half its peak, but a square climb spends most of the interval low and averages a third of its peak. For a pair tested together, therefore, PFDavg ≈ (λDU·T)²/3, where λDU·T is one channel's probability of having failed just before the test. The form keeps the conditions of Chapter 11's shortcut and adds two: the channels are identical and fail independently, and both are tested at the same time.
The other votes follow from counting the ways to fail. A 2oo3 function fails when any two of its three channels have failed, and three different pairs can be drawn from three channels, so its average is three times that of a 1oo2 pair: (λDU·T)², the square itself. A 2oo2 function fails when either channel has failed, so its average is the sum of two single channels' averages, λDU·T, twice that of 1oo1. The fuller formulas of IEC 61508-6 Annex B add the repair times and the detected failures, as for one channel in Chapter 11. They also add a term for the share of failures that strikes more than one channel at once, which Chapter 14 takes up.
A false call trips the pair
Spurious trips depend on the vote in the opposite way to missed demands. In 1oo1 every false call trips the function, so its spurious trip rate is the channel's rate of false calls. In 1oo2 a false call from either channel trips the function, so the rate doubles, and each channel added to a one-out-of vote adds its own false calls. A 2oo3 function trips falsely only if a second channel calls falsely while the first is still under repair. Any of three channels can make the first call and either of the other two the second, so its spurious trip rate is about six times the square of one channel's rate of false calls, multiplied by the repair time.
How often a design may trip without a demand is a requirement, and the rate computed against it depends on what the method counts. For a single channel, the trip rate in the textbook's method includes the channel's detected failures, while the rate in SINTEF's PDS method includes only the channel's spurious operations. ISA-TR84.00.02 writes the rate of each vote in forms of its own. A trip rate is therefore stated with its method and with what the design does on a detected failure, and the example below counts each detected failure as a trip.
Every channel uses Chapter 11's stand-in sensor, a transmitter whose resistance-thermometer input has a λDU of 71 FIT and a λDD of 203 FIT at an average of 40 °C. Both are maker's figures, from the analysis that the assessor exida made for the transmitter's maker, PR electronics. Each channel trips on its own detected failures, the first reaction of Chapter 12, so its 203 FIT are false calls and only its 71 FIT can leave it unable to act. All channels are tested together once a year with a perfect test, and a channel that has called falsely is repaired in 24 hours, the repair time of Chapter 12's example. One channel's probability of having failed just before the test is 71 × 8,760 ÷ 10⁹, about 6.22 in 10,000. A single channel averages half of that, about 3.11 in 10,000, and trips falsely 203 × 8,760 ÷ 10⁹ times a year, about once in 560 years. The pair averages a third of the square of 6.22 in 10,000, about 1.3 in 10 million, so one channel's average is about 2,400 times the pair's. The pair trips falsely twice as often as one channel, about once in 280 years. Three channels voting 2oo3 average the square, about 3.9 in 10 million, and trip falsely 6 × 203 × 203 × 24 × 8,760 ÷ 10¹⁸ times a year, about once in 19 million years. The figures cover the sensors alone and independent failures only.
The pair gains its protection against missed trips at the cost of twice the false trips. A 2oo3 vote averages three times the pair's and in return removes nearly all of the false trips. For one incubator a false trip once in 560 years is rare, so the choice between the votes matters most where channels call falsely more often or where each trip costs more. Of the three votes in the example, the repair time matters only to 2oo3: a channel left failed for twice as long doubles that function's spurious trip rate, because the second false call has twice as long to arrive.
Two out of three keeps most of both
A 2oo3 vote keeps most of the pair's protection and, from its channels' own failures, trips falsely far less often than a single channel. Like 1oo2, it still trips on a demand with one channel failed dangerously, and like 2oo2, it ignores a false call from one channel. Its average is three times the pair's, because three pairs of channels can defeat it where the pair has one, but in the example that average stays about 800 times below one channel's. Its spurious trips from independent failures need two false calls within one repair, so they nearly vanish. What remains is the share of trips that one cause sets off in two channels at once (Chapter 14). In the textbook's table comparing the methods, both the textbook's short form for 2oo3 and the PDS form keep only that share, while the full form that the textbook works through for 2oo3 keeps the term for independent failures as well.
A third channel also shows which channel has failed. Two readings that disagree show only that one of them is wrong, as in Chapter 12's cross-check, and a pair must then trip or run on without knowing which reading to trust. Among three channels, the one whose call or reading differs from the other two identifies itself as the failed one, so the design can flag it for repair while the other two carry on the function.
The 2oo3 vote has costs of its own. It needs a third sensor with its wiring, a voter through which every channel passes, and a third channel to test. In the incubator the three thermistors would sit in one air stream and report to one voter. The air stream, the power supply and the enclosure that Chapter 4 found shared by the control loop and the protective channel would then be common to all three, and Chapter 14 weighs them. Chapter 20 of the Academy guide Where the heat goes describes single-fault design in a surgical generator, a second example of protection built to survive one failure. Chapter 25 of this guide sets out the single fault condition for medical devices.
The average and the architecture
The overfill loop of Chapter 11 shows what duplication buys in a whole function. In PSLG's second version of the loop, the level transmitter and the valve are duplicated and voted 1oo2, while the logic solver stays single. PSLG states that the example's values are fictitious, as Chapter 11 noted. Tested once a year, the duplicated transmitters average about 3.8 in 10,000 against 3.0 in 1,000 for a single transmitter, and the valves fall from about 32 to about 6 in a million. These results include a share of failures common to both channels, set at 2 % of the undetected failures and 1 % of the detected ones (Chapter 14). The whole loop averages about 1.1 in 1,000 against 3.8 in 1,000 in the first version, and the logic solver's fixed 7.1 in 10,000 is now about two-thirds of that total.
PSLG plots both loops against the proof-test interval. The duplicated loop rises from about 1.1 in 1,000 at one year to about 5.5 in 1,000 at 10 years and stays inside the SIL 2 band throughout, while the single loop leaves that band at about three years. Tested yearly, both loops meet SIL 2's requirement for the average. A design must also satisfy the architectural constraints, as PSLG's example adds. The constraints set the least hardware fault tolerance of each subsystem, whatever its average.
IEC 61508-2:2010 gives two routes to the constraints. On route 1H the highest integrity level that a subsystem may claim depends on its HFT, on its type and on its safe failure fraction, the SFF of Chapter 12. An element is type A if its failure modes are well defined, its behavior under fault can be fully determined and its failure data are dependable, and type B if any of these falls short. With the stand-in sensor's SFF of 74 %, one channel may claim at most SIL 2 if it is type A and SIL 1 if it is type B, and a second channel raises each by one level. One channel averages 3.1 in 10,000, inside the SIL 3 band, yet on this route it cannot claim SIL 3 alone.
On route 2H the least HFT follows from the integrity level alone, without the SFF. The failure rates on this route must come from field feedback on equipment in use in a similar application and environment, with 90 % confidence in the data, as a 2020 blog post from exida describes the route. A subsystem at SIL 3 then needs an HFT of 1 and one at SIL 4 an HFT of 2, while one at SIL 1, or at SIL 2 in low demand, may have an HFT of 0. On either route, the integrity level of Chapter 8 sets both the redundancy that an architecture must have and the requirement for the average. Chapter 11 computes that average, and this chapter extends the computation to channels that back each other up.
Write the integrity level and the maximum allowable spurious trip rate into the safety requirements specification (Chapter 17) before choosing an architecture. For each candidate vote, compute the average probability of failure on demand and the spurious trip rate from the same failure rates, test interval and repair time, and name the method used for the trip rate. Decide what a channel does when it detects its own failure, whether it trips or leaves the function in a reduced vote, and count that choice in both numbers. Check each subsystem's hardware fault tolerance on route 1H or route 2H as well as its average. For the incubator, give a second channel its own thermistor, comparison and cutout, put the two cutouts in series, and test both channels end to end every year (Chapter 11).
- The vote is how many of a function's channels must call for a trip, and it sets at once both ways in which the function can go wrong: 1oo2 is the hardest to defeat and the easiest to trip, 2oo2 the reverse, and 2oo3 tolerates one failure of each kind.
- For identical channels that fail independently and are tested together, a pair's average is a third of the square of one channel's probability of having failed just before the test, so the stand-in sensors at 71 FIT tested yearly average about 1.3 in 10 million as a pair against 3.1 in 10,000 alone.
- Each channel in a one-out-of vote adds its false calls, so the pair trips spuriously twice as often as one channel. A 2oo3 vote averages three times the pair's, and its spurious trips from independent failures need two false calls within one repair.
- A design must satisfy the architectural constraints as well as its average: route 1H sets the least hardware fault tolerance from the integrity level, the safe failure fraction (SFF) and the element's type, and route 2H from the level alone, on field data.
Let's Talk