FMEA, FMEDA, fault trees.
+ The questionHow do engineers find the single failure, or the pair, that leads to harm?
Two searches from opposite ends
A fault tree runs backwards from the harm to the part failures that could cause it, and the smallest sets of those failures matter most. The tree starts from one event that must not happen, such as an infant overheated in an incubator, and works down to the failures that could produce it, then to those that could produce each of them, until every branch ends at the failure of a single part. A set of part failures that together produce the harm is one route to it. Each failure on its own is rare, so a route that needs two at once is far less likely than either of them, while a route of one failure is exactly as likely as that failure. The shortest routes therefore carry most of the probability.
Chapter 5's hazard study found the deviations that could release a hazard, and Chapter 10 gave the rates at which parts fail. The step between the two is a search for the part failures, alone or in pairs, that lead to the harm. One method starts from the parts and follows each failure up to its effect on the function. Another, the fault tree, starts from the harm and works down to the parts. A third extends the first, giving each failure a rate and a class, so that the second can put a number on each route.
The incubator of Chapter 4 shows why both directions are needed. A welded contact in its cutout harms no one by itself: while the control loop holds the air at the set temperature, the cutout is never asked to open, and the welded contact shows nothing (Chapter 2). Harm needs a second failure, in the control loop, that drives the air past the cutout's limit while the contacts are welded, and only a search from the harm finds that pair.
One failure at a time, from the part up
A failure modes and effects analysis, or FMEA, lists the parts of a system and the ways each part can fail, one failure mode to a row. Each row follows its failure upward: what it does where it occurs, what that does to the assembly the part belongs to, and what it finally does to the function. The row also records how the failure would become known, if anything would show it. Chapter 1 gave a thermistor three failure modes: open, short and drift. Chapter 2 sorted the modes of the protective thermistor by the direction of the error and by whether a check could see the reading. Those modes are the first rows of an FMEA of the incubator's protective channel, and the comparison and the cutout add their own.
The method applies to hardware, software, processes that include human action and the interfaces between them, according to IEC's summary of IEC 60812:2018, the international standard for FMEA. An FMEA can serve in a safety analysis, but by the same summary the standard gives no specific guidance for safety applications. Ranked by criticality, which takes in at least the severity of each effect, the rows make an FMECA, a failure modes, effects and criticality analysis. A safety function needs more from each row: a failure rate, whether diagnostics detect the failure, the share that a proof test would reveal, and a class. An FMEA extended in that way is an FMEDA.
The FMEA's strength is that it is complete by construction. Every part and every mode gets a row, so the analysis finds failures that no deviation of a hazard study names, which is why Chapter 5 set it beside the HAZOP. Its limit follows from the same row-by-row rule. Each row takes one failure with the rest of the system working, so the row for the welded contact records that the cutout would not open at a trip but cannot say when a trip will be needed. The harm lies in a combination: a welded contact present when the control loop fails. A list of single failures holds such combinations only if someone adds them.
A rate and a class for every row
The FMEDA of Chapter 10 takes those four additions one by one. The first is a failure rate, taken from a prediction handbook or a database of component rates and adjusted by an analysis of the component's stresses. The second is whether the device's automatic diagnostics detect the failure, and the third is the share of the remaining dangerous failures that a proof test would reveal. The row's class is then one of Chapter 2's four: safe or dangerous by what the failure does to the device's function, detected or undetected by whether the diagnostics reveal it. The rates of the rows in each class add up to the device's λSD, λSU, λDD and λDU. Chapter 10's transmitter, for example, has 71 FIT of dangerous undetected failures, a figure from an analysis that its maker commissioned.
An FMEDA works out each row's effect by analysis where it can. Where it cannot, a failure is injected into a component of a real unit, and its effect is watched at the device's output. The Rosemount 2130 is a vibrating-fork level switch, and its maker commissioned a 2024 analysis of the switch from the assessor exida. From the results of fault injection tests in a 2010 report that the maker provided, the analysis takes the effects it could not work out.
IEC 61508's four classes distinguish only safe and dangerous failures, detected and undetected, and not every effect the analysis finds falls in one of them. The 2024 analysis keeps two more groups. One holds the failures that leave the safety function unaffected. The other holds annunciation failures, which do not affect safety directly but impair the switch's ability to detect a later fault, as a fault in its own diagnostic circuit would. The analysis keeps the two groups apart for reliability models more detailed than IEC 61508 requires.
An FMEDA's rates hold only within the assumptions of its analysis. In the level switch's analysis every rate is constant, so failures early in life and at wear-out are excluded. On errors of operation and maintenance the analysis makes two statements. One section leaves such failures out as specific to each site, while the management summary states that the predictions include site-specific failures due to human events, for the site safety index that the analysis specifies, a figure for the site of use (Chapter 21). The analysis also takes one component failure at a time. That last assumption is the FMEA's rule, carried over from its rows to the totals.
For the incubator's cutout, the FMEDA writes Chapter 2's classes down as rows. Contacts stuck open and a coil burnt open are safe undetected, and contacts welded shut are dangerous undetected, found only by a proof test (Chapter 11). Each row then takes a rate for its mode from a prediction handbook, at the temperature inside the incubator's enclosure (Chapter 10). The welded row's rate becomes the cutout's share of λDU.
A fault tree works down from the harm
Drawn as a diagram, a fault tree has one undesired event at its root, the top event. Its branches end in basic events, such as the failure of one part, and the FMEDA supplies their rates. Between the two, the tree models the combinations of faults that produce the top event. IEC 61025:2006 is the international standard for the method.
The causes join the event above them through gates. At an OR gate the output event occurs when one or more of the inputs occur, and at an AND gate only when all of them occur. Read upward, an OR gate lists alternatives, any one of which is enough, and an AND gate requires several things to go wrong together. A working barrier stops the branch it guards, so protection enters a tree as an AND gate: the harm then needs the barrier's failure as well as the cause that the barrier answers.
The incubator's tree has the overheated infant as its top event and one AND gate beneath it. The gate's two inputs are the demand, in which the control loop drives the air past the cutout's limit, and the failure of the protective channel to cut the heater. Each input is itself an OR gate. Under the demand stand two of the causes that Chapter 5 named: a control thermistor reading low and a heater driver stuck on. Under the protective failure stand the channel's failures that give no sign: its thermistor reading low, its comparison never deciding to trip, and the cutout's contacts welded shut. These three failures are the dangerous undetected rows of the channel's FMEDA, so the analysis of the parts supplies the bottom of the tree.
The smallest cut sets carry the probability
Reducing a tree turns it into a list of its routes to the top event. A minimal cut set is a smallest combination of failures that, if all of them occur, causes the top event: no member can be removed and the rest still cause it. An OR gate passes up each input as a set of its own, and an AND gate joins one set from each input into a larger one. The incubator's AND gate pairs each of the two causes of a demand with each of the three protective failures, so the tree reduces to six minimal cut sets, all of them pairs. A cut set of one would be a single failure that causes the harm. The incubator of Chapter 4 keeps its protective channel apart from the control loop, and that leaves the tree without one.
The same structure carries the arithmetic. For independent inputs, the probability of an AND gate's output is the product of its inputs' probabilities, and that of an OR gate is close to their sum. The US Nuclear Regulatory Commission's Fault Tree Handbook of 1981 calls the sum the rare event approximation. The sum errs slightly and always on the high side, by no more than about 10 % when each probability is below 0.1. By the product rule, two independent failures at 1 in 1,000 each come together about once in a million. That is the handbook's figure for two amplifiers working in parallel, where a count that treated either failure as failing the system would give 2 in 1,000. A pair is as rare as the product of its members, so among parts of similar reliability the smallest cut sets carry nearly all the probability.
The handbook works one tree through to its end: the rupture of a pressure tank after its pump starts. The contacts of relay K2 close to run the pump motor, and a pressure switch opens at the threshold pressure to stop the motor. With the handbook's illustrative probabilities, the tree reduces to five minimal cut sets. Two are single: the tank itself at 5 in a million and relay K2 at 3 in 100,000. Three are pairs. Each joins the pressure switch, at 1 in 10,000, with one of the parts that back it up: relay K1 at 3 in 100,000, a timer relay at 1 in 10,000, or a switch at 3 in 100,000. Multiplied, the pairs give 3 in a billion, 1 in 100 million and 3 in a billion. Summed, the five give about 35 in a million, or 3.5 in 100,000, and the pairs make up only about 1 part in 2,000 of that total. K2 alone carries about 86 % and the tank 14 %, the split the handbook gives. Halving K2's probability would lower the total by about 43 %, and halving the pressure switch's would lower it by less than 1 part in 4,000. The probability of rupture is set almost entirely by the tank and relay K2, and a change to any part that appears only in a pair barely moves it.
A single cut set is removed by adding a part that must fail as well, which puts an AND gate above the single failure. The incubator's protective channel is such a part. The incubator tree's pairs join events of two kinds. Each cause of a demand arrives at a moment, at some frequency per year, and the protective failure must already be present, hidden, when it arrives, with the probability that Chapter 11 averaged as PFDavg. Each pair therefore occurs at its cause's frequency multiplied by its protective failure's probability. The six products add up to the demand rate from these causes multiplied by the channel's PFDavg.
The demand rate and the PFDavg are both sums at the tree's OR gates. The demand rate adds the causes of a demand, each with a rate from an analysis of the control loop's parts. In this way the parts' rates of Chapter 10 become the demand rate of Chapter 5, from these two causes. The PFDavg adds the channel's three failures, as Chapter 11 added the averages of the transmitter, logic solver and valve of an overfill loop. With Chapter 11's stand-in sensor at about 3.1 in 10,000, the channel's PFDavg is at least that figure before its comparison and cutout add their shares. The product of the two is the one that Chapter 7 applied to each layer of protection.
An AND gate's product holds only for inputs that fail independently. The handbook warns that for dependent inputs the output's probability may be significantly greater than the product. The incubator's two loops share the air stream, the power supply, the enclosure and the setting that Chapter 4 listed, and they may share the software build. A cause acting through any of these can sit under both inputs of the AND gate at once. When the tree is reduced, such a cause becomes a cut set of one, as likely as the cause itself. In a tree of Tank 912's overfill, the stuck gauge of Chapter 7 sits under both the continuing fill and the alarms set on its reading, so the alarms add no member to any cut set. Chapter 14 follows such shared causes.
Write the FMEDA of each protective channel from its parts list, with each mode's rate, the conditions the rate holds for and the diagnostics that decide its class. Draw the fault tree from the harm that the hazard study of Chapter 5 named, take its basic events from the FMEDA's rows, and reduce it to minimal cut sets. List every cut set of one and remove it with a part that must fail as well. Check that the two members of every pair share nothing: no power supply, air stream or other medium, enclosure, software build or setting (Chapter 14). Record the cut sets and the rates behind them with the safety requirements specification of Chapter 17. For a device such as the incubator, record them also in the risk management file that Chapter 25 describes.
- An FMEA follows each failure mode of each part up to its effect on the function, one failure at a time, so it lists every single failure and holds a combination of failures only if someone adds it.
- An FMEDA gives each row a rate and one of the four classes and adds the rows into λSD, λSU, λDD and λDU. Its rates hold only within the assumptions of its analysis: the level switch's analysis, for example, takes every rate as constant and one component failure at a time.
- A fault tree works down from the harm through OR and AND gates, and its minimal cut sets name every smallest combination of failures that causes the harm, single failures and pairs among them: the incubator's tree has six pairs and no single.
- An AND gate multiplies the small probabilities of independent failures, so among parts of similar reliability the smallest cut sets carry nearly all the probability, as the two single failures of the handbook's pressure tank do. A shared cause turns a pair into a single failure, and the product no longer holds.