§ Lessons
Lessons.
The chapters reduce to thirteen working rules for people who design, verify, buy, regulate or write about immunoassays and the analyzers that run them. Each rule points to the chapters that explain why it holds.
- Write the target as a count of molecules before choosing a method. A concentration of a few nanograms per liter is a few million molecules in a drop, and the count that reaches the detector is smaller still after capture, washing and sampling (Chapters 1, 17 and 18).
- Name the epitopes and the circulating forms, not only the protein. Two antibody pairs against the same troponin can count different subsets of it, and assays with different epitopes and calibrators read the same blood differently (Chapters 2 and 14).
- Budget the minutes of binding against antibody concentration and distance. An excess of antibody speeds binding, beads spread through the sample shorten the distance each molecule must diffuse, and a strip gives each part of the sample only seconds over its line (Chapters 3, 12 and 15).
- Choose the format from the size of the molecule, and state the hook-free limit. Large proteins go between two antibodies and small molecules compete for one; a one-step sandwich can read falsely low at extreme concentrations, the hook effect, so the hook-free limit belongs in the specification (Chapters 4, 13 and 19).
- Treat the blank as the floor, and report every limit with the design that produced it. The signal from label that remains when no target is present sets the limit of blank, and limits of blank, detection and quantitation mean little without the replicates, lots, analyzers and days behind them (Chapters 5, 18 and 23).
- Judge a label by the background it leaves, not by its brightness alone. Color reads a small dip in a bright beam; chemiluminescence, delayed fluorescence and electrochemiluminescence each succeeded by measuring against less unwanted light, and an amplifying label also amplifies the signal from wrongly bound label (Chapters 6 to 9).
- Specify the analyzer as a chain of subsystems, and test the joins. Sampling, reagent handling, incubation, washing, detection and software each move the result, and carry-over and temperature drift can escape routine controls (Chapters 10 to 12).
- Specify the detector from both ends of the range. Dark count and light collection set the detector's share of the blank, counting statistics set the scatter of small signals, and saturation and dilution set the top (Chapter 13).
- Tie every result to its calibration chain and to the limits of its own assay. A master curve adjusted on site, a lot change or a new generation can move results, and troponin cut-offs cannot be carried from one assay to another (Chapters 14 and 20).
- Validate the reader with the test, whether it is a person or a machine. The same strips gave different clinical sensitivities when read by different people, and an instrument reader applies its threshold just as consistently to a strip that ran badly (Chapters 15 and 16).
- Ship the interference checks with the assay, and run them when a result disagrees with the patient. Heterophile antibodies, biotin, macrocomplexes and the hook effect produce wrong results that look plausible, and a mismatch with the patient's condition, or between methods, exposes them (Chapter 19).
- Count the middle zone and the prevalence in every performance claim. An observation zone or an indeterminate result moves the work of reaching an answer to a later test instead of removing it, and predictive values change with the share of patients who have the condition (Chapters 20 and 21).
- Plan the evidence once for the whole system, then file it by each route. Regulators judge the reagent, calibrators, analyzer, software and labeling together, sort them by risk in different ways, and change their rules on calendars a maker does not control (Chapters 23 to 26).