+ Part VIII · Quality and regulation, as of October 2026
What a test system must show.
About 8 percent of the in vitro diagnostics on the EU market needed a notified body's assessment under the 1998 directive. Under the regulation that replaced it, around 80 percent do. The European Commission gave those figures in October 2021, when only six notified bodies had been designated to do the work; by February 2026 there were 19. In vitro diagnostics, or IVDs, are the reagents, instruments, software and other devices used to examine samples taken from the human body. A notified body is an independent organization that an EU country designates to check a maker's evidence before a device is sold. The four chapters of this part set out what a reagent and its analyzer must show before they are sold, as of October 2026. Each chapter explains why a requirement exists and what evidence it calls for, and gives the clauses with their status as of October 2026, because the clauses will change sooner than the reasons.
Proving the floor and the scatter.
+ The questionHow many measurements does it take to show a test's floor and precision, and why that many?
A definition that is also an experiment
The limit of blank is the highest result expected, 95 times in 100, from a sample that contains none of the target. Chapter 5 worked the arithmetic once. This chapter asks what that definition commits a maker to measure, because each of its terms leaves a choice open: which samples count as blank, on how many analyzers and reagent lots, over how many days, and how many results are enough to locate the 95th percentile of a spread of results.
The Clinical and Laboratory Standards Institute, a US standards body for laboratory medicine, answers in a series of documents whose names begin with EP. In its second edition of 2012, EP17 covers the limits of blank, detection and quantitation. It requires that they fall in that order: the limit of detection above the limit of blank, and the limit of quantitation at or above the limit of detection.
Roche's high-sensitivity troponin T assay followed EP17. According to the FDA's decision summary, the assay's limit of blank rests on 60 results from blank samples for each reagent lot on each type of analyzer, and its limit of detection on another 60 per lot from low-level samples. Its limit of quantitation rests on 84 results per sample per lot. That limit is the lowest concentration measured with a coefficient of variation of no more than 20 percent. A laboratory that only wants to check a maker's claim needs far fewer results. The implementation guide to EP17 specifies at least 20 blank results in all, with at least two blank samples, each measured at least twice a day, on at least three days. Two samples measured twice on each of five days give 20 results; five samples measured twice on each of three days give 30.
Why precision takes eighty results
A measurement's scatter has more than one source. Results within one run differ because pipetting, mixing and counting are never identical, and this part of the scatter is called repeatability. Runs on the same day differ a little more, because each run starts afresh with its own controls and handling. Days differ again, with new reagent packs, operators or calibrations. The within-laboratory precision is the total scatter a laboratory sees over weeks, and it combines all three. A standard deviation from one run alone understates it.
To separate the sources, the design has to repeat each one. In the third edition of CLSI's EP05, of 2014, each sample is measured in two replicates, in two runs a day, on 20 days. That gives 80 results per sample, and a statistical method called analysis of variance splits their scatter into its within-run, between-run and between-day parts. Roche's troponin T assay was tested to the previous edition's design: two replicates in each of two runs a day on 21 days, or 84 results per lot. For the Alzheimer's ratio of Chapter 21, five panels of samples were measured in two replicates, in two runs a day, on 20 days. Two further studies followed, both with two runs of three replicates: one on three reagent lots over five days each, and one at three sites over five days each.
A standard deviation estimated from a handful of results is itself uncertain. If the results are treated as independent, the relative uncertainty is about one divided by the square root of twice the number one less than the count of results. A nested design, in which replicates sit within runs and runs within days, only approximates that independence. From 5 results, the count less one is 4, twice that is 8, and the uncertainty is one divided by the square root of 8, about 35 percent, so a true scatter of 5 percent could easily be reported as 3 or 7. From 20 results it is about 16 percent, and from 80 about 8 percent. With 80 results, a coefficient of variation of 5 percent can at best be placed between roughly 4.6 and 5.4 percent, one standard error either side. When day-to-day scatter is large, the effective count is closer to the number of days, so 20 days leave an uncertainty of about 16 percent, the same as 20 results. Spreading those results over 20 days, rather than running them in one afternoon, is what makes the between-day part visible at all.
A laboratory, again, checks a claim rather than establishing one. Under CLSI's EP15, also of 2014, a laboratory verifies a maker's precision claim and estimates bias, a consistent offset of results from their true value, with at least 25 measurements over at least five days. CLSI's product page for EP15 states that such verification cannot establish precision. The difference matters in practice: the maker's 80 results support a claim, and the laboratory's 25 test whether its own analyzer meets that claim.
Range, interference and carry-over
Linearity is the claim that results stay proportional to concentration across the measuring range. The second edition of EP06, published in 2020, dropped the curve-fitting of its predecessor. It recommends a weighted straight-line fit, with the deviation at each level judged against what would matter clinically. Roche's troponin T assay was cleared with a measuring range of 6 to 10,000 ng/L and no hook effect up to 100,000 ng/L, the hook-free limit that Chapter 13 explained.
Interference testing measures how much a result moves when a possible interferent is added to a real sample. The third edition of EP07, of 2018, describes the method, and a companion document, EP37, lists the concentrations at which common substances should be tested. Chapter 19 showed the cost of choosing too low a level: the 2017 version of Roche's troponin T assay had been tested for biotin only up to 20 ng/mL, and in its 2020 guidance the FDA asks for 3,500 ng/mL.
Carry-over is the part of one sample that reaches the next, the subject of Chapter 11. It is tested by running high and low samples in alternation and is reported as a fraction. In 2021 Beckman Coulter recalled its high-sensitivity troponin I assay, and the recall showed what a small fraction can cost. By the firm's account, the reason was that a sample above 270,000 ng/L could contaminate the reagent pack, and so the results that followed. As an illustration, take 14 ng/L, the 99th-percentile limit for women of Roche's troponin T assay. A carry-over of about 52 parts per million from such a sample would by itself turn a blank into a result at that limit.
A laboratory's verification is sized to catch a gross failure of the maker's claim, not to measure precision anew. Twenty-five results over five days leave a standard deviation uncertain by at least roughly 14 percent, and by much more when day-to-day scatter dominates. A laboratory that finds a coefficient of variation a little better than the maker's has therefore learned only that its analyzer is not far worse. Claims at a decision limit, such as precision at a 99th percentile, rest on the maker's full design.
Comparison, reference limits and the documents regulators recognize
A new assay is compared with an existing one on patient samples. The method is that of EP09, in its third edition of 2018, which estimates the bias between the two assays across the range. Reference intervals, the range of results expected in healthy people, follow EP28. EP28's minimum of 120 healthy people per group gives a nonparametric interval between the 2.5th and 97.5th percentiles, read directly from the ranked results without assuming the shape of their distribution. A 99th percentile needs many more people: with 120 it would be the single highest value. That is why the 99th-percentile limits of Roche's troponin T assay drew on 1,301 healthy people.
Regulators adopt these documents instead of writing their own. As of October 2026, the FDA's database of recognized consensus standards listed, among others, the third editions of EP05, EP07, EP15 and EP28, the second editions of EP06, EP17 and EP37, and the third edition of EP09 in part. The FDA recognized the second edition of EP37 in December 2025. The database also listed an older CLSI document on interference by the patient's own antibodies in immunoassays. Once the FDA recognizes a standard, a maker can cite it in a submission instead of justifying its method from first principles.
In the EU, the IVDR sets out the evidence for an IVD in three parts: scientific validity, analytical performance and clinical performance. Scientific validity is the link between the substance measured and a clinical condition. In 2022 the Medical Device Coordination Group, made up of experts appointed by the EU countries, issued guidance with a list of analytical characteristics. Beside the limits, linearity and precision of this chapter, the list names carry-over and comparison between instruments. Clinical performance studies follow ISO 20916 of 2019, which was confirmed in 2025. The PCR guide in this library explains how often a positive or negative result is right.
List every analytical claim the labeling will make, with its concentration: limits of blank, detection and quantitation, precision at each decision level, the linear range, the interferents and their test levels, the hook-free limit and carry-over. Assign each claim its recognized design, and decide the lots, analyzers, sites and days before the first run, since a study that did not repeat a source of scatter cannot later report it.
- The limits of blank, detection and quantitation each rest on a stated design, which for Roche's troponin T assay meant 60 results per reagent lot and analyzer type for the limit of blank, 60 per lot for the limit of detection and 84 per sample per lot for the limit of quantitation.
- Precision is built from repeatability, between-run and between-day scatter, and only a design that repeats each can separate them, as CLSI's EP05 does with 80 results over 20 days.
- Linearity, interference, hook and carry-over each have their own recognized methods, and testing an interferent at too low a level can let wrong results reach patients later.
- The FDA recognizes the CLSI documents used here, and the EU groups the same evidence under scientific validity, analytical performance and clinical performance.