25 — The analyzer's own standards

Safety, EMC, risk and software.

+ The questionAn analyzer never touches a patient. What can it harm, and which standards cover that?

A medical device with laboratory hazards

An immunoassay analyzer is a medical device, yet the standard that governs its electrical safety is written for laboratory equipment. That standard is IEC 61010-1, for laboratory equipment in general, and IEC 61010-2-101 of 2018 is its particular part for in vitro diagnostic equipment, including equipment for self-testing. A particular part adds requirements for one kind of equipment and is used together with the general standard. IEC 60601-1, the standard for medical electrical equipment, belongs to a different family. The choice makes sense once the question is asked the other way round: what could this machine harm?

An analyzer can harm the people who run it. It holds mains electricity, moving pipetting arms, heated incubators, magnets, blood that may carry infection, and chemicals such as the acidic peroxide and sodium hydroxide that trigger a flash label (Chapter 7). Those are the hazards of laboratory equipment. An analyzer can also disturb other equipment, or be disturbed by it, through electromagnetic interference. And it can harm a patient it never meets, through a wrong number. In its overview of IVD regulation, the FDA links the safety of an IVD to its false-negative and false-positive results, and that third route to harm is the one most particular to a diagnostic analyzer.

The wrong result as the hazard

Risk management for medical devices follows ISO 14971, whose 2019 edition was confirmed in 2025. Applied to an analyzer, risk management traces each fault forward to a wrong result, then to a situation in which the result is acted on, then to harm. The PCR guide in this library works through that chain for a thermal cycler, the heating and cooling instrument of PCR. For an immunoassay analyzer the faults are often the ones earlier chapters described, and real cases show the chain from end to end.

Carry-over is one such fault. Chapters 11 and 23 described the 2021 Beckman Coulter recall; by the firm's account in the FDA's recall record, the reason was that a sample above 270,000 ng/L could carry over through the probe into the particle well of the reagent pack, and the record gives the root cause as undetermined. The fault as described is mechanical, the wrong result a falsely raised troponin in a later patient, and the harm a needless cardiac workup or a missed alternative diagnosis. The remedy was a procedure: after any such result, the open packs are discarded, a new one is loaded, low-level quality control is run on every pipettor and the affected positive results are repeated.

Biotin is another. In assays such as the original version of Roche's troponin T assay, the fault was a design tolerance for biotin as low as 20 ng/mL, the wrong result a falsely low troponin, and the harm of the kind the FDA reported in 2017, a death. The biotin risk was controlled first by a warning in the instructions and later by a change to the reagent, the route Chapter 19 described.

Solved example: a hazard counted in results

In 2021 authors who included Roche staff published an evaluation of the reformulated troponin T assay. They estimated that the original assay's susceptibility to biotin would misclassify about 0.026 percent of samples, about 1 in 3,800, in the population they modeled. A hospital that measures 100 troponins a day runs 36,500 a year, so at that rate it would expect about ten misclassified results a year: rare for any one patient, routine for the laboratory. Risk management works with such counts, set against the severity of a missed heart attack, rather than with the impression that a rare interference is negligible.

55 — From a fault to a harm
FAULT WRONG RESULT RESULT ACTED ON HARM CARRY-OVER (2021 RECALL) probe carries a sample above 270,000 ng/L into the reagent pack falsely raised troponin in a later patient acted on needless cardiac workup CONTROL procedure: discard open packs, new pack, low-level QC on every pipettor, repeat positives BIOTIN (2017 WARNING) reagent tolerance 20 ng/mL falsely low troponin acted on missed heart attack; a death reported CONTROLS first: warning in the instructions then: reagent changed (scavenger antibody, longer linker; 1,500 ng/mL) a design control does not rely on a reader COUNTED IN RESULTS 1 in 3,800 samples misclassified (2021 estimate, maker's staff among authors) × 36,500 tests a year ≈ 10 results a year risk management per ISO 14971:2019
InterferentReadoutFocal detail
Plate 55 — An analyzer harms a patient through a wrong number, so risk management traces each fault forward to a result and a harm. The biotin risk was first controlled by a warning and later by changing the reagent; at 1 in 3,800 samples, a busy laboratory would still meet about ten such results a year.

Safety and compatibility, as of October 2026

The general laboratory safety standard, IEC 61010-1, stands at its third edition of 2010 with an amendment of 2016. A second amendment would align it with the standard for audio, video and information technology equipment. According to a test house, a company that tests equipment against standards, that amendment was at the committee draft vote in April 2025. Its publication could not be confirmed as of October 2026. The FDA recognizes IEC 61010-1 with US national differences, in its consolidated version, which merges the standard with its amendment. The FDA's database did not list the IVD part, IEC 61010-2-101, which a maker can still apply as the state of the art, the generally accepted current practice. The part for automatic laboratory equipment in general, IEC 61010-2-081, excludes IVD equipment and refers it to Part 2-101.

Electromagnetic compatibility, or EMC, is the ability of equipment to work without disturbing other equipment or being disturbed by it. Its standards changed more recently. IEC 61326-2-6, the particular part for IVD equipment, reached its fourth edition in June 2025. The FDA recognized that edition in full on 28 July 2025, having recognized the 2020 edition only in part. A maker designing an analyzer in 2026 therefore tests against an edition little more than a year old.

56 — Hazard, standard and status
HAZARD STANDARD (EDITION) FDA RECOGNIZED EU HARMONIZED (LIST TO OCT 2024) electric shock, heat, moving parts, chemicals, infection (operator) IEC 61010-1 Ed 3.1; IEC 61010-2-101:2018 (IVD) 61010-1 yes (US differences); 2-101 not listed no electromagnetic disturbance IEC 61326-2-6 Ed 4.0 (Jun 2025) yes , in full (28 Jul 2025) no wrong result: risk management ISO 14971:2019 (confirmed 2025) yes yes software IEC 62304 Ed 1.1 (2015); Ed 2 still a draft yes no use errors IEC 62366-1 + A1:2020 yes no information and symbols ISO 18113 parts 1–5 (2022); ISO 15223-1 + A1:2025 18113 not listed ; 15223-1 yes 15223-1 yes ; labeling and symbols decisions 2026 quality system ISO 13485:2016 (confirmed Oct 2025) incorporated in the QMSR, 2 Feb 2026 yes met as state of the art, without presumption status as of October 2026; the EU list is the consolidated list of October 2024
ReadoutFocal detail
Plate 56 — An analyzer's operator hazards fall under IEC 61010 with its IVD part, its electromagnetic compatibility under IEC 61326-2-6, its software under IEC 62304. The FDA recognizes most of them; the EU's harmonized list leaves out electrical safety, EMC, software and usability.

Software, usability and labels

The software that schedules an analyzer's steps and computes its results is developed under IEC 62304, which in October 2026 was still its 2006 edition as amended in 2015, the version the FDA recognizes. A second edition, widened to health software in general, remained a draft. In July 2026 the German standards association VDE described it as a working draft, expected to replace the three software safety classes with two levels of process rigor. The safety classes set how rigorous development must be according to the harm a software failure could cause. One software firm reported that the new edition was published in August 2026, but the IEC's own catalog listed no newer version, so this guide treats it as unpublished.

Usability engineering, which addresses the errors users make, follows IEC 62366-1 as amended in 2020, and the FDA recognizes that standard. The five parts of ISO 18113 set out the information a maker must supply with its products. All five are in second editions of October 2022, with Part 3 for instruments used by professionals and Part 5 for self-testing instruments, and the FDA database does not list them. The symbols on labels follow ISO 15223-1 of 2021, whose 2025 amendment the FDA recognized in December 2025.

Under all of these sits the quality management system, the procedures by which a maker controls design, production and change. Its standard is ISO 13485 of 2016, confirmed again in October 2025. The FDA incorporated it by reference into its Quality Management System Regulation, which has been in force since 2 February 2026.

The EU's harmonized list leaves out the analyzer's electrical standards

In the EU, a standard listed as harmonized under the IVDR gives a presumption of conformity: a device that meets the standard is presumed to meet the requirements it covers. The list as consolidated to October 2024 held 15 entries, among them the standards for quality management, risk management, symbols, traceability and clinical performance studies, and none for electrical safety, electromagnetic compatibility, software or usability. Judging by their titles, decisions in 2026 added the labeling standard and the amended symbols standard. For the other requirements, a maker applies the IEC standards as the state of the art, but without the presumption, and has to argue its case in its technical documentation.

The regulators thus agree on what an analyzer must show, and differ on which documents carry formal weight. The FDA recognizes most of the IEC set. The EU's harmonized list, as consolidated to October 2024 and as far as the titles of the 2026 decisions show, includes none of it; a 2025 amendment was not read. Even so, regulators on both sides treat the set as the established way to demonstrate safety, compatibility and sound software.

Trace every standard to a hazard

Start the standards list from the hazard analysis, not from a template: for each hazard to the operator, to other equipment and to the patient through the result, name the standard and the edition that addresses it, and record its status with the FDA, in the EU's harmonized list and in India. Then record the date each edition was checked, since several documents in this chapter changed status between 2025 and 2026, among them the EMC part, the symbols standard, the EU's harmonized list and the FDA's quality regulation.

+ What this chapter established
  • An analyzer's own hazards are those of laboratory equipment, so its safety standards are IEC 61010-1 and the IVD part, IEC 61010-2-101.
  • Its most particular hazard is a wrong result, traced through risk management from fault to harm, as in the carry-over and biotin cases.
  • The EMC part for IVD equipment reached a fourth edition in 2025, and IEC 62304's second edition was still a draft in October 2026.
  • The FDA recognizes most of these standards, and the EU's harmonized list leaves out electrical safety, EMC, software and usability.