+ Part II · Making a bond visible
Four ways to read a label.
In 1971 two groups independently published ways to replace the radioactive label of the first immunoassays with an enzyme, a catalyst that keeps working for as long as it has something to work on. One group worked at Stockholm University and the other in the Netherlands. Their idea opened a sequence of labels, each giving more signal per bound molecule against less unwanted light. This part follows four of them in order: an enzyme read by the color it makes, molecules that give off light in a chemical reaction, labels that glow long after a flash of exciting light, and a label switched on by a voltage at an electrode, the method of Roche's troponin T assay. By the end of this part, the floor of Chapter 5 can be stated exactly: a label's light measured against everything else that reaches the detector.
Color from an enzyme.
+ The questionHow does an enzyme turn a few bound antibodies into a color a photometer can read, and where does color run out?
An enzyme as an amplifier
Horseradish peroxidase is the enzyme most often attached to antibodies in color-based immunoassays. A single molecule of it can convert a few hundred to a few thousand molecules of a colorless substrate, the molecule the enzyme acts on, into colored product every second. Measurements put the rate at about 240 per second near neutral pH and room temperature, while figures more than ten times higher are often quoted for the acidic, warmer conditions in which the enzyme works fastest. Over a half-hour incubation, even the lower rate lets each enzyme molecule make some 400,000 colored molecules.
That multiplication is why enzymes replaced radioactivity. A radioactive iodine atom on an antibody gives at most one decay, and an iodine-125 label decays slowly, so in a counting time of a minute only a small fraction of the labels on a tube decay at all. Radioactive labels also needed shielding, licenses and disposal, and they lost activity on the shelf. An enzyme label, by contrast, is stable and safe, and it keeps adding signal for as long as substrate remains. The signal of each bound antibody is no longer one event but an amount of colored product that grows with time.
The amplification has a cost. The enzyme makes product at the same rate whether it is attached to an antibody that holds a target or to one stuck on the plastic. Enzyme amplification therefore raises the signal from wrongly bound label exactly as much as the signal from target. That is the argument of Chapter 5, met here in its first practical form.
The first ELISA
The enzyme-linked immunosorbent assay, ELISA, was first worked out by Eva Engvall, then a graduate student of Peter Perlmann at Stockholm University. In a retrospective published in 2010, she recalled obtaining the first calibration curve early in 1970. The assay was a competitive one, adapted from a standard radioimmunoassay, with antibodies coupled to particles of cellulose; bound and free label were separated by repeated centrifugation and washing. The enzyme was alkaline phosphatase. It was chosen because its substrate dissolved readily and could reveal small amounts of the enzyme, and because the product could be measured. The target was rabbit immunoglobulin G. "We were continually asked why anyone would want to measure rabbit IgG!" she wrote.
The paper appeared in 1971 in the journal Immunochemistry. In the Netherlands, B. K. van Weemen and A. H. W. M. Schuurs published an enzyme immunoassay the same year, and a 2005 history describes the two methods as developed independently and simultaneously. Most laboratories now picture an ELISA as a sandwich assay in the wells of a plastic plate. That form came later, as plates, monoclonal antibodies and plate readers arrived. The principle did not change: catch, wash, add substrate, and measure the color.
Reading color: absorbance
A colored solution absorbs part of the light shone through it. A photometer measures how much light of one wavelength passes through the well and how much passed through a blank, and reports the absorbance, a measure of the loss in which each whole unit means a tenfold drop in light. A plate reader is a photometer that holds many wells at once. An absorbance of 1 means a tenth of the light got through, 2 means a hundredth, and 3 a thousandth. Absorbance is proportional to the concentration of the colored molecule, to the length of liquid the light crosses, and to how strongly that molecule absorbs at that wavelength, its molar absorptivity.
The common peroxidase substrate, tetramethylbenzidine, turns blue as the enzyme works, and yellow when the reaction is stopped with acid. Its yellow product absorbs at 450 nanometers with a molar absorptivity of about 59,000 per molar per centimeter, high for a small molecule. The product of alkaline phosphatase, para-nitrophenol, absorbs at about 400 nanometers with 18,380 per molar per centimeter, a value measured in 1980 carefully enough for the compound to serve as a reference material. Plate readers made for this work resolve a thousandth of an absorbance unit, and their makers specify that readings repeat within about half a percent plus 0.005 units.
Take a reader's repeatability of 0.005 absorbance units as the smallest change it can trust, and assume a 200-microliter well in which light crosses about 0.6 centimeter of liquid. That depth is an assumption, because it depends on the plate. The colored product then has to reach 0.005 divided by the product of 59,000 and 0.6, about 0.14 micromolar, or roughly 17 million million molecules in the well. Assume one enzyme per captured molecule and 1 million product molecules per enzyme over the incubation, a round figure of the same order as the 400,000 that the section above gives for half an hour. That takes about 17 million enzyme molecules, some 28 attomoles. A troponin sample of Chapter 1 holds 4.2 million target molecules in 50 microliters, fewer than the enzymes this estimate needs.
Laboratory sandwich ELISAs reach low concentrations by putting more sample in, waiting longer and catching more of the target. One maker states that its commercial kit for interleukin 6, an inflammatory protein, reached a sensitivity of 0.7 picogram per milliliter with 100 microliters of sample in an assay that took four and a half hours. Sensitivity here means the lowest concentration the kit can detect. A picogram per milliliter equals a nanogram per liter, so the maker's figure is 0.7 ng/L, lower than the 5 ng/L of Roche's troponin T assay in Chapter 1, although the two figures are for different proteins and may not be defined the same way. Analyzers must answer in minutes, so they cannot lower their floor that way.
Where color runs out
Color fails at both ends. At the low end, a trace of product changes the light leaving the well by a small fraction of a large beam. The reader has to measure that beam twice, through the well and through a blank, and subtract; any flicker of the lamp, any scratch on the plastic and any bubble enters both measurements and survives the subtraction as noise. The enzyme stuck on the plastic makes its own color in the meantime. At the high end, absorbance stops following concentration. Readers report up to 4 absorbance units, but their makers specify linearity only to about 2 or 3, because very little light gets through a dark well and stray light inside the reader starts to dominate.
A plate reader that shows 3.8 does not mean twice the product of a well showing 1.9. Above about 2 absorbance units the reading compresses, so high calibrators and high samples crowd together and a dilution is needed to place them. Checking the reader's specified linear range, not its display range, belongs in setting up any color assay.
Between a noise floor near 0.005 and a ceiling near 2 lies a factor of about 400, less than three factors of ten. A sample whose concentration falls outside that window must be diluted and run again, or read at a different time. The light-emitting labels of the next three chapters improve both ends at once, because they measure light where there should be darkness instead of a dip in a bright beam.
In a color assay the enzyme keeps working until the reaction is stopped, so the same plate can be too pale or too dark depending on minutes. Fix the substrate time so that the highest calibrator stays inside the reader's linear range and the lowest stays clearly above the blank, and record that window as part of the method, not as a habit of whoever runs it.
- An enzyme label turns each bound antibody into hundreds of thousands of colored molecules, and began replacing radioactivity from 1971.
- The first ELISA was a competitive assay on cellulose particles with alkaline phosphatase, measuring rabbit IgG.
- Absorbance measures a small dip in a bright beam, so at the low end lamp and plate noise and stuck enzyme set the floor.
- Readers stay linear only to about 2 or 3 absorbance units, giving color a working range of less than three factors of ten.