US, EU, India, and WHO prequalification.
+ The questionWhat does each regulator, and the program that UN buyers of rapid tests rely on, ask of a reagent and its analyzer, as of October 2026?
The European Union, from 26 May 2022
On 26 May 2022 the EU's In Vitro Diagnostic Medical Devices Regulation, the IVDR, began to apply, five years after its publication. Its central mechanism is the one the three regulators in this chapter share: the risk a wrong result poses decides who must check the evidence before a test is sold. The IVDR sorts devices into four classes, A to D, by rules set out in an annex. For class B and above, and for sterile class A devices, a notified body designated under the regulation must take part in assessing the maker's quality system and technical documentation, more fully for the higher classes.
Under the EU's guidance on classification, as revised in September 2026, the troponin tests, strips and analyzers of this guide fall into the following classes. The parts of the guidance read for this guide give no example for the Alzheimer's blood tests, and their EU class was not found. A troponin test for patients who arrive with acute symptoms is class C, as a cardiac marker. Tests for self-testing are class C, except for a short list, including pregnancy tests, that is class B. Instruments intended specifically for in vitro diagnostic procedures are class A, for which the maker declares conformity itself, and class B is the default for anything the rules do not place elsewhere. The analyzer and its troponin reagent therefore reach the market by different routes of assessment, as Chapter 24 showed.
The number of notified bodies has lagged behind the need for them. At the end of February 2026 a Commission survey counted 19 notified bodies designated under the IVDR, which together had received 3,418 applications and issued 2,318 certificates. A regulation of 2024 extended the transition period for devices that need a notified body for the first time. During the extension such devices may stay on the market under the old directive, on conditions and with deadlines that depend on the class.
Take a troponin reagent sold under the old directive that needs a notified body for the first time under the IVDR. Under the 2024 extension it may stay on the market until 31 December 2028, provided its maker had its quality management system in place by 26 May 2025, lodged its application with a notified body by 26 May 2026, and signed a written agreement with that body by 26 September 2026. By October 2026 that last deadline had passed. For a class D device the application and agreement deadlines and the end date each fall a year earlier: application by 26 May 2025, agreement by 26 September 2025, and the market until the end of 2027. For class B and sterile class A devices they fall a year later: application by 26 May 2027, agreement by 26 September 2027, and the market until the end of 2029. The quality management system was due by 26 May 2025 in every class.
The rules may change again. In December 2025 the European Commission proposed to amend the IVDR and its companion regulation for other medical devices. The proposal would replace the five-year expiry of certificates with periodic reviews based on risk, reduce notified-body involvement for lower-risk devices, and allow changes to be agreed in advance. As of October 2026 it had not been adopted. The Council of the EU, where the member states' governments sit, had issued a progress report in May. In July the European Parliament's committee rapporteur, the member who leads its work on the proposal, had published a draft report.
The United States
The FDA also sorts devices into classes, I to III, and the class decides the route. A device similar to one already on the market reaches it through a 510(k), a demonstration of substantial equivalence to that earlier device, which is called its predicate. A low- or moderate-risk device with no predicate goes through a De Novo request, in which the FDA sets up a new classification for that type of device. A class III device needs premarket approval, in which the FDA reviews evidence of the device's own safety and effectiveness. Roche's troponin T assay is in class II and was cleared through a 510(k). The first spinal fluid test for amyloid had no predicate and was granted through De Novo in May 2022. It then served as the predicate for the blood test of Chapter 21.
A second decision follows clearance. Under the Clinical Laboratory Improvement Amendments, or CLIA, the law that governs US laboratories, the FDA assigns each test a complexity: waived, moderate or high. Seven criteria, each scored from 1 to 3, separate moderate from high complexity. A test whose scores sum above 12 is high complexity, and one at or below 12 is moderate, while waived status is decided separately. Until it is categorized, a test is treated as high complexity. The category decides which laboratories may run the test, so a maker designs for it as much as for the clearance.
Tests developed and run within a single laboratory have a separate history. In May 2024 the FDA issued a rule that would have made explicit that such laboratory-developed tests are devices under the law. A federal court in Texas set the rule aside, or vacated it, on 31 March 2025, and in a rule published that September the FDA removed from its regulation the words the 2024 rule had added, "including when the manufacturer of these products is a laboratory". As of October 2026 the Alzheimer's blood tests that Quanterix sells as laboratory-developed tests remained outside FDA clearance. Quanterix submitted a 510(k) for one of them in early 2026, and no decision on it had been found.
India
India regulates IVDs under the Medical Devices Rules, 2017, which took effect at the start of 2018 and use the same four classes, A to D. India's national regulator for drugs and devices, the Central Drugs Standard Control Organisation or CDSCO, houses the Central Licensing Authority. That authority licenses all imports and the manufacture of class C and D devices, approves new IVDs and their clinical performance evaluation, and issues test licenses, which allow a device to be made or imported in small quantities for testing, evaluation, clinical investigation or training. State licensing authorities license the manufacture of class A and B devices and the sale of devices. An importer applies on Form MD-14 and receives a license on Form MD-15. Analyzers, instruments and software came under licensing in two phases, a year apart: the lower classes on 1 October 2022, and classes C and D on 1 October 2023. The regulator's 2023 list puts a chemiluminescent immunoassay analyzer in class A.
India's distinctive requirement is local evidence. The rules call an IVD new until the Central Licensing Authority has approved it for manufacture or import in India. A new IVD in class B, C or D needs a clinical performance evaluation in India, a study of its performance on patient samples, whatever its approvals in the EU, the US or elsewhere. For imported kits, the evaluation uses three batches made from three lots of key raw materials and takes place at central or registered testing laboratories. The regulator updated its lists of those laboratories three times between June 2025 and June 2026. In September 2025 the regulator and the medical research council published standard evaluation protocols and a portal for kit validation. The class of a troponin reagent in the regulator's lists was not found for this guide.
WHO prequalification
The fourth route is not a regulator. The World Health Organization says of itself that it "is not a regulatory authority". It assesses products for prequalification, a benchmark that UN and other international procurement agencies use when they buy, and that national regulators can rely on. Its program for IVDs became operational in 2010, growing out of a scheme for evaluating test kits begun in 1988. An assessment has three parts: a review of the dossier, an inspection of the manufacturing site and a review of the labeling. The dossier is the maker's file of evidence, and an abridged one is allowed where a recognized regulator has already assessed the product stringently. WHO has also made the laboratory evaluation of performance a separate procedure that comes first, done by designated evaluation laboratories.
A prequalified rapid test still needs registration in each country that uses it. Prequalification shortens national review where a regulator chooses to rely on it.
Eligibility is limited to priority diseases, mainly infections such as HIV, hepatitis B and C, malaria, syphilis and tuberculosis, and a few point-of-care tests. There is no category for cardiac markers or for Alzheimer's disease, so neither of this guide's two cases has a route through prequalification. In October 2026 the list held 125 prequalified IVDs. The entries from makers whose names indicate Indian companies were all rapid tests on strips, for HIV, hepatitis C, syphilis and malaria. Through prequalification, the strip technology of Chapter 15 reaches international procurement.
The analytical studies of Chapter 23 and a well-designed clinical study serve every route. Plan them once, with the strictest requirement of each in view: the US classification and CLIA category, the EU class, notified body and transition dates, an Indian clinical performance evaluation on three batches from three raw-material lots, and prequalification only where the disease is eligible. Start the Indian evaluation and the notified-body agreement early, since both run on calendars a maker does not control.
Back to Basel
The Basel-coordinated study drew blood from 718 patients between 2006 and 2008 and sought nothing from a regulator. The blood was frozen and measured blind in batches, and the results were judged against a diagnosis made from the records, which is the design of a clinical performance study. The paper appeared in August 2009, and Roche's sensitive troponin T assay reached laboratories in Britain and Ireland that year. The European guideline based a one-hour decision on such assays in 2015, and in January 2017 the FDA cleared the assay for the US, using the same 718 patients as part of its evidence. Nearly seven and a half years separated the paper from the clearance.
Most chapters of this guide describe part of what had to hold in that interval. The antibodies and the label set how few molecules the test could see; the analyzer held that floor from run to run; the calibration tied it to a scale; the studies showed its blank, its scatter and its interferents; and the regulators decided, each in its own way, what evidence was enough. A few molecules in a drop became a number a physician could act on in an hour only when all of those held at once.
- The IVDR has applied since 26 May 2022, and under it troponin reagents are class C and need a notified body while analyzers are class A, with transition deadlines running to 2029.
- The FDA routes devices by class through a 510(k), De Novo or premarket approval and then assigns a CLIA complexity, and a court vacated its 2024 rule on laboratory-developed tests in 2025.
- India licenses IVDs under the Medical Devices Rules, 2017, and requires local clinical performance evaluation for new class B to D tests.
- WHO prequalification is not regulation but a procurement benchmark for priority infections, and it covers neither troponin nor Alzheimer's tests.