12 — Time and temperature

Minutes at 37 °C.

+ The questionWhy do the analyzer's clock and thermostat shape the result, and how closely must they hold?

A tolerance of three tenths of a degree

Roche specifies the incubator of its cobas e 801 at 37 °C, plus or minus 0.3 °C, and the module's measuring cell at 28 °C, plus or minus 0.5 °C. Those numbers are requirements on the chemistry as much as on the heaters. Every rate in an immunoassay rises with temperature: molecules diffuse faster in warmer liquid, antibodies bind and release faster, and enzymes convert substrate faster. In a 9- or 18-minute protocol the reaction is usually stopped and read before it is complete, and a warmer reaction then gives more signal for the same amount of target.

How much more depends on the assay. An application note from Revvity, the maker of DELFIA reagents, gives one measured case. In a two-step assay for thyroid-stimulating hormone, a second incubation of two minutes reached 67 percent of its maximum signal at 35 °C, against 25 percent at 25 °C. From such data the note derives a rate that roughly triples for every 10 °C. That rise corresponds to an activation energy of about 84 kilojoules per mole, where the activation energy is the energy barrier the reaction must climb. The measurement comes from one assay on one platform, but the order of magnitude is common: chemistry slows sharply as it cools.

Solved example: what three tenths of a degree is worth

With an activation energy of 84 kilojoules per mole, the reaction rate near 37 °C changes by about 11 percent for each degree, from the standard relation between rate and temperature. A tolerance of plus or minus 0.3 °C therefore allows the rate to vary by about plus or minus 3 percent between one incubation and another. About 3 percent either way is the most the rate can vary within the tolerance, not the error in a result. Calibrators and patient samples run in the same incubator under the same protocol, so a steady offset largely cancels, and a reaction close to completion responds less to a change in rate than one far from it. A fluctuation between the calibration run and the patient's run does not cancel.

24 — What three tenths of a degree is worth
0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 25 30 35 37 40 REACTION RATE, RELATIVE TO 37 °C TEMPERATURE, °C a steady offset cancels; a fluctuation does not 37 ± 0.3 °C → rate ± about 3 % 0.97 1.00 1.03 36.7 37.0 37.3 °C about 11 % per degree near 37 °C computed for an activation energy of 84 kJ/mol (from one maker's application note): rate roughly triples per 10 °C
AntibodyReadoutFocal detail
Plate 24 — For a reaction whose rate roughly triples every 10 °C, each degree near 37 °C changes the rate by about 11 percent, so a tolerance of ±0.3 °C allows about ±3 percent. Calibrators share a steady offset with patient samples; a fluctuation between runs they do not share.

The measuring cell has its own temperature for a related reason. Its light comes from reactions at the electrode surface, which are also faster when warmer. In one patent example the cell is read for about two seconds, so a difference in the cell's temperature from one test to the next changes the signal per label directly. Holding the cell at a fixed temperature, separate from the incubator's, keeps that signal the same from test to test.

Temperature tolerance is a requirement on the result

A heater that holds 37.0 °C at its sensor can still leave a reaction cup a few tenths of a degree away, through the cup's position, the reagent's temperature as it is added, or the room. What matters is the temperature of the liquid during the reaction, so the tolerance has to be verified in the cup, across positions and room conditions, not only at the controller.

The same minutes for every cup

Timing matters for the same reason as temperature. A reaction that has not reached equilibrium gives a signal that depends on how long it ran, so every cup must run exactly as long as the calibrators did. Analyzers therefore run on a fixed cycle, a new cup every 12 seconds on the e 801, and their protocols come in fixed steps of 9, 18 or 27 minutes. The scheduler can then guarantee that each cup spends its incubation in the same number of cycles, whatever else is happening in the machine.

The fixed protocol also makes a useful check possible. Chapter 3 showed that the fraction of target caught after a given time depends on how fast the target binds. A target that binds unusually slowly, because it is attached to something else in the blood, can give different results in a 9-minute and an 18-minute version of the same assay. Chapter 19 describes laboratories that used exactly that difference to find an interference, a distortion of the result caused by something in a patient's sample other than the target.

Shorter protocols, higher floors

When a protocol is shortened, binding has less time to approach equilibrium, so fewer sandwiches form for the same amount of troponin, and the signal from target shrinks while the background does not. The floor can rise. The International Federation of Clinical Chemistry publishes a table of troponin assay characteristics supplied by their makers. Its 2019 version gives two limits of detection for the high-sensitivity troponin T assay on one family of Roche analyzers: 2.05 ng/L for the 18-minute protocol and 2.85 ng/L for the 9-minute one. Both figures are below the 5 ng/L that a 2017 trade report gives as the limit on the assay's European label, a reminder that a limit depends on how it was measured. On the newer cobas e 801, both protocols were listed at 3 ng/L.

In the US clearance of 2017 the same trade-off was made a different way. On the cobas e 411 the 9-minute assay adds the beads in a second incubation, after the sandwich has formed, and its limit of detection was 5 ng/L. On the cobas e 601 the beads are present from the start, and its limit was 3 ng/L, according to the FDA's decision summary. The reagent was the same, but the two analyzers differ both in how they arrange the minutes and in their measuring hardware. The Federation's table also lists a higher limit for the e 411 than for Roche's other analyzers in the 18-minute protocol. The e 411's floor is therefore higher in both protocols, and the timing of the bead step explains at most part of the difference.

25 — The same reagent, arranged in time
ONE STEP cobas e 601 TWO STEPS cobas e 411 sample biotin antibody ruthenium antibody beads first incubation second incubation beads added 0 3 6 9 min schematic timelines 0 1 2 3 4 5 6 LIMIT OF DETECTION, ng/L one Roche family, 18-min protocol (IFCC 2019) 2.05 same, 9-min protocol 2.85 cobas e 801, both protocols 3 FDA 2017, cobas e 601, one step, 9 min 3 FDA 2017, cobas e 411, two steps, 9 min 5 same reagent, two analyzers, different floor limits of detection as listed by makers (IFCC table) and in the FDA decision summary
AnalyteAntibodyLabel and signalReadoutFocal detail
Plate 25 — A shorter protocol leaves binding less time and can raise the floor, from 2.05 to 2.85 ng/L in one maker-supplied table. The arrangement matters too: in the 2017 US clearance the same reagent reached 3 ng/L in one step and 5 ng/L in two.

Reagents on board and on the shelf

Time acts on the reagents too. Antibodies slowly lose activity, labels decay or degrade, and the beads can clump, so every reagent pack has a shelf life in the refrigerator and a shorter life once opened on the analyzer. The e 801 keeps its packs at 5 to 10 °C on board, and its maker gives on-board stabilities of up to four months, depending on the assay. Abbott gives 14 to 30 days on board for its ARCHITECT i reagents, and Siemens Healthineers gives 4 to 90 days for the Atellica IM.

A calibration loses its validity as its reagent ages. The FDA's decision summary for Roche's troponin T assay states that a calibration remains valid for 7 days while the same reagent pack stays on the analyzer, and for 12 weeks with packs from the same lot. After that the analyzer requires a new calibration, because the reagent's response may have drifted. In 2023 the Clinical and Laboratory Standards Institute published the second edition of its guideline EP25, which describes how makers establish reagent shelf-life and in-use stability, by real-time and accelerated studies. Accelerated studies store reagents at raised temperatures so that aging shows sooner.

Write time and temperature as one budget

Measure how the assay's signal changes with incubation temperature and with time around the chosen protocol, then set the tolerance on each so that their combined effect stays inside the precision the assay must meet. The same data show how far a protocol can be shortened before the floor moves, and they belong in the design file as the rationale for both numbers.

+ What this chapter established
  • Every rate in an immunoassay rises with temperature, and in one measured assay the rate roughly tripled per 10 °C.
  • A tolerance of plus or minus 0.3 °C at 37 °C allows the rate to vary by about plus or minus 3 percent, and that variation cancels only when calibrators and samples see the same conditions.
  • Fixed cycles and fixed protocol times make every cup run as long as the calibrators did.
  • Shorter protocols and different step arrangements can raise the floor, and reagents and calibrations age on board.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.