Moving microliters, washing beads.
+ The questionHow does a machine pipette a few microliters, wash particles a few micrometers across, and leave nothing behind for the next sample?
A drop that should not travel
On 9 August 2021 Beckman Coulter began a recall of more than 290,000 kits of its Access high-sensitivity troponin I reagent, according to the FDA's recall record. The firm attributed the problem to carry-over: after a sample with an extremely high troponin result, traces could pass "through the probe into the particle well" of the reagent pack and contaminate later tests. Its instructions to laboratories were concrete, and their trigger was any result above 270,000 picograms per milliliter, which is 270,000 ng/L, the unit used for troponin everywhere else in this part. After such a result, laboratories were to discard the open reagent packs, load a new one, run a low-level control sample on every pipettor, and repeat any positive results that might have been affected.
Carry-over is the transfer of material from one sample, or one reagent, into the next. A troponin assay is especially exposed to it because its range is so wide: Roche's troponin T assay, for example, was tested and found free of the high-dose hook up to 100,000 ng/L (Chapter 13). Troponin I assays meet samples above 270,000 ng/L, as the trigger value in the Beckman recall shows, while their makers' decision limits sit at about 9 to 58 ng/L (Chapter 14). A sample ten thousand times above the decision limit needs to leave behind less than one part in ten thousand of itself, and in practice far less, to keep the next result accurate.
Suppose a sample of 100,000 ng/L is followed by one containing almost no troponin. A carry-over of 1 part per million adds 0.1 ng/L to the second result, below the limit of blank. Abbott specifies the carry-over of the fixed probe on its ARCHITECT i analyzers at no more than 0.1 part per million. In the FDA's 2019 review of Abbott's high-sensitivity troponin I assay, a sample of at least 500,000 ng/L left about 0.3 ng/L in the next, at most 0.6 part per million. A 2014 letter reported an older troponin I assay, measured on another analyzer, that carried over 41 parts per million: after a sample of about 484,000 ng/L, a sample of about 60 ng/L read 80. At 41 parts per million, the 100,000 ng/L sample would add 4.1 ng/L to a blank, above the limit of blank of Roche's troponin T assay.
One common protocol measures carry-over with a fixed sequence: a high sample run twice, then a low sample three times. If the first low result is higher than the third, the difference, divided by the high concentration, estimates the fraction carried. The test has to be planned as part of the design, because routine quality control cannot reveal the problem. The Clinical and Laboratory Standards Institute's guideline on statistical quality control excludes random errors that affect individual samples, such as carry-over, from its scope, because control samples run between patients will not reveal them.
Control samples measured between patients test whether the whole system has shifted. A carry-over event affects one sample, the one after an extreme result, and leaves the next control untouched. Carry-over has to be tested by a dedicated sequence at verification and guarded against by design, by probe washing, tips, flags on extreme results and rules for repeating the following samples.
Tips, probes and pressure
The first defense is the pipetting hardware. A disposable tip used once and thrown away cannot carry sample on the probe to the next sample. Carry-over can still arise through reagent probes, wash stations or a shared measuring cell. Roche describes the cobas e 801 as using "one disposable AssayTip per pipetting". Siemens Healthineers and QuidelOrtho both say that their disposable tips eliminate sample carry-over, which is the makers' claim, not a measured value. A fixed probe is instead washed inside and out between samples. It saves the cost and waste of tips and can be cleaned to a specified carry-over, as Abbott's figure shows. If the firm's account of the cause is right, the Beckman recall shows what happens when sample can take a path that the cleaning does not reach.
The second defense is to watch the liquid as it moves. A probe that touches a clot, a bubble or foam draws the wrong volume, and the result is wrong with no visible sign. A 2000 patent from Bayer describes one way to catch a wrong draw, by recording the vacuum inside the probe while it aspirates, or draws in, the liquid. Bayer's diagnostics business later passed to Siemens. In the patent, a normal draw of about 100 microliters shows a smooth rise and decay within about half a second. A clot blocks the tip and drives the pressure higher and the draw longer, foam lets the pressure collapse early, and a second rise before the decay means that the tip lost contact with the liquid. Analyzers also sense the liquid surface before aspirating, and the e 801 adds a camera for foam.
Small volumes make every error larger in relation to the volume drawn. A 4-microliter draw that loses a tenth of a microliter to a droplet on the tip is 2.5 percent short, and the troponin result is 2.5 percent low with it. Pipetting precision at such volumes does not appear in the makers' public specifications. It is included, however, in the total precision of the assay, which the studies of Chapter 23 measure.
Magnets and beads
Most analyzers separate bound from free label with magnetic microparticles. The reagent of Roche's troponin T assay contains streptavidin-coated particles at 0.72 milligram per milliliter, according to the FDA's decision summary. Patents describe the Elecsys beads as typically 2 to 3 micrometers across, and research products use beads of that size too, such as one supplier's 2.8-micrometer streptavidin beads, which bind about 500 to 750 picomoles of biotin per milligram. Beads that small stay suspended long enough to collect the sandwiches throughout the reaction volume, as Chapter 3 required, yet a permanent magnet can still pull them out of suspension.
The wash then depends on the magnet, which holds the beads against a wall or an electrode while the liquid around them is drawn off and fresh buffer is added. How much unbound label survives depends on the volume of liquid left clinging to the pellet of beads, the number of exchanges, and whether the beads are released and resuspended between them. Those figures do not appear in public sources, but each exchange works as a dilution: two exchanges that each leave 1 percent behind leave one part in ten thousand.
Magnets attract more than beads. In October 2019 Roche issued a field safety notice for one lot of its troponin T reagent on the cobas e 801, after laboratories had reported high results that could not be reproduced. A field safety notice is a maker's formal warning to the users of a product. Roche issued the final version of the notice in January 2020, and that version gave the cause as contamination with magnetic particles that were not beads, introduced during filling of the reagent for that analyzer only. Until the lot was replaced, laboratories were told to repeat any result of 14 ng/L or more, not to shake the reagent packs, and to discard each pack after its first 200 tests.
Run the carry-over sequence with the high sample at or above the hook-free limit, not at the top calibrator, and express the result in the units of the decision limit. Then write the instruction for laboratories as the Beckman recall had to: which result triggers which action, on which pipettors, for which following samples.
- Carry-over moves traces of one sample into the next; for troponin it matters because results span more than four factors of ten around a low decision limit.
- Disposable tips and washed probes limit carry-over to a claimed or specified level, and pressure monitoring and level sensing catch wrong draws.
- Magnetic microparticles a few micrometers across collect sandwiches throughout the sample and are held still while the liquid is exchanged.
- Carry-over and stray particles affect single samples, and routine controls do not reveal them, so they must be tested by design.