Light in the dark.
+ The questionWhy does light from a chemical reaction reach lower concentrations than color does?
A measurement with no lamp
An absorbance reader cannot work in the dark. It needs a lamp, and it reports the target as a small shadow: a slight dimming of the lamp's light as it passes through the sample. If instead the label itself makes the light, and the detector sits in a light-tight box, the measurement works on a different principle. With no label present, the detector sees nothing but its own small electrical noise. Every photon that arrives was made by a reaction in the tube, so the instrument counts upward from near zero instead of subtracting one large reading from another.
A label that makes its own light relies on chemiluminescence: a chemical reaction whose product is formed in an excited state, holding more energy than its stable form, and releases that extra energy as light. Fireflies and some deep-sea animals make light this way with enzymes; immunoassays do it with small organic molecules. Each such label often gives more signal than a color label, but the gain over color does not come from that. It comes from measuring the signal against darkness. A shadow of 0.005 absorbance units is a change of about 1 percent in a bright beam, whereas a light of a few hundred photons a second is unmistakable to a detector that registers a few dozen in the dark.
The label still has to be attached to an antibody and separated from free label by washing, and it still adds to the signal when it sticks where it should not. What changes is how small a quantity of label the instrument can see, and over how wide a range. Two families of chemistry dominate analyzers today: labels that flash once, and enzymes that keep glowing.
Flash: one label, one burst
In 1983 Ian Weeks and colleagues described acridinium esters, small organic molecules that give off light when they react, as labels for immunoassays. They attached up to about 2.8 labels to each monoclonal antibody, and the paper reported that the ester could be detected at about 0.8 attomole, which is roughly 480,000 label molecules. Their two-site assay incubated for 30 minutes and read each sample in 5 seconds.
The light comes when the ester meets hydrogen peroxide in an alkaline solution. In the scheme used on one maker's analyzers, the reagents are added in two steps: an acidic peroxide solution and then sodium hydroxide, with a detergent that speeds and strengthens the emission. Chemists at Siemens Healthcare Diagnostics, a company that uses these labels, published a study in 2014. They found that with the detergent at least 95 percent of the light from four esters arrived within 1.2 to 5.3 seconds, against up to a minute without it. Each acridinium molecule reacts once and gives at most one photon, so the whole flash must be caught as it happens: the trigger solutions that start the reaction are injected in front of the detector, and the light is counted in a few seconds.
Analyzers report chemiluminescence in relative light units, a count scaled by each instrument's optics and electronics. In the 2014 Siemens study, proteins joined to acridinium labels gave about 4.3 to 6.1 × 10¹⁹ light units per mole of label on one laboratory luminometer. That is about one unit for every 10,000 to 14,000 label molecules. The number links label to signal only on that instrument; it is not a photon count and does not transfer between analyzers.
Two makers' analyzers use acridinium labels, according to FDA decision summaries and the makers' own pages. Abbott's ARCHITECT and Alinity analyzers use acridinium-labeled antibodies on paramagnetic microparticles, tiny beads that are drawn to a magnet but do not stay magnetized, in a method the company calls chemiluminescent microparticle immunoassay. Siemens Healthineers' Atellica and ADVIA Centaur analyzers use acridinium esters too.
Glow: enzymes that make light
The other family keeps the enzyme label of Chapter 6 and changes its substrate. In 1983 researchers in Birmingham reported that firefly luciferin multiplied the light from luminol oxidized by horseradish peroxidase. In 1985 the same group found phenol compounds that raised that light more than a thousandfold and kept it glowing for several minutes. Light-producing substrates for alkaline phosphatase followed in the late 1980s, molecules called 1,2-dioxetanes: the enzyme removes a phosphate group, and the unstable remainder breaks apart and emits light. A 1991 review gave the detection limit for alkaline phosphatase with such a substrate as 0.001 attomole, about 600 enzyme molecules, with emission lasting more than 30 minutes.
As in Chapter 6, the enzyme works as an amplifier, but now of light instead of color. Each enzyme molecule keeps converting substrate molecules for minutes, each of which can give a photon, so a few hundred enzymes can be seen. Fujirebio's Lumipulse analyzers use alkaline phosphatase with the dioxetane substrate AMPPD, whose light peaks at 477 nanometers. The FDA's 2025 decision summary describes that chemistry for the company's blood test for Alzheimer's disease, one of the tests in the guide's second case. QuidelOrtho's VITROS analyzers measure peroxidase with luminol.
The detection limit in the 1983 acridinium paper works out to about 480,000 label molecules. The limit in the 1991 review, for alkaline phosphatase with a dioxetane substrate, works out to about 600 enzyme molecules. The ratio is about 800: an amplifying enzyme can be seen at far fewer copies than a label that reacts once. The comparison is of labels, not of assays. Both figures were measured with label alone, and in an assay both labels sit on the same wrongly bound antibodies of Chapter 5, so an enzyme also amplifies the light from wrongly bound label.
Glow has costs of its own. The light rises over minutes, so the analyzer must wait before reading, and that wait sets a minimum time per test. A 2019 poster by Beckman Coulter and its substrate subsidiary states that the older substrate for the company's Access analyzers needed about 6.3 minutes to generate its signal. Beckman's DxI 9000 analyzer uses a newer acridan-based substrate that gives its light in about a minute, and the time to the first troponin result fell from 17 to 11 minutes, by the company's account.
What light changes, and what it does not
Measuring light against darkness widens the range as well as lowering the floor. A detector that counts single photons can register a few dozen per second in the dark and millions per second before it saturates and can count no faster, a span of five or more factors of ten against less than three for absorbance (Chapter 13). A sample can then be measured from near the floor to far above the clinical decision point without dilution, which saves reruns and reagent.
What light does not change is the label that stays behind without a target. An acridinium ester stuck to the wall of a cuvette, the small vessel in which the reaction is read, flashes as brightly as one on a sandwich. An enzyme on the plastic glows as long as one on a bead. Chemiluminescence removed the lamp, the scratches and the bubbles from the noise of Chapter 6, so the floor that remains is much closer to the one set by wrongly bound label. The next two chapters describe two more ways to read a label against near-darkness. In Chapter 8 a lamp remains, and the reading waits until the scattered light and the sample's own glow have faded; in Chapter 9 the light is made at an electrode, so label that is not held against the electrode contributes little light.
A flash label needs its trigger injected in front of the detector and the whole burst captured in seconds; a glow needs an incubation before reading and a stable temperature while the light builds. The choice fixes the analyzer's mechanics, its cycle time and where its detector sits, so it belongs in the first architecture decisions, together with the label (Chapter 10).
- Chemiluminescence measures light made by the label against darkness, not a shadow in a lamp's beam.
- Acridinium esters flash once per label within a few seconds of an alkaline peroxide trigger.
- Enzymes with light-producing substrates glow for minutes; alkaline phosphatase with a dioxetane can be seen at a few hundred copies.
- Light widens the measuring range and removes lamp noise, but label stuck without target still shines.