13 — From light to number

From a few photons to a few million.

+ The questionHow does one detector count a few hundred photons and also a few million, and what limits the count at each end?

A detector that multiplies electrons

A photomultiplier tube is a glass vacuum tube a few centimeters long with a thin metal-alkali film, the photocathode, on the inside of its window. A photon striking the film can knock out one electron. Inside the tube sits a chain of electrodes called dynodes, each at a higher voltage than the last. The voltage accelerates the electron into the first dynode, where it knocks out several more, and each dynode down the chain repeats the step, so that one electron at the cathode becomes a pulse of a million to ten million electrons at the end. Hamamatsu, which makes such tubes, gives that range in its handbook as the tube's gain, the number of electrons out for each electron in.

Not every photon starts a pulse. The fraction of photons that release an electron is called the quantum efficiency, and it depends on the cathode material and the wavelength. Hamamatsu quotes up to 43 percent at 350 nanometers for its best blue-sensitive cathodes. A further loss occurs because only 60 to 90 percent of released electrons reach the first dynode. For one photon-counting module, Hamamatsu gives 380,000 counts a second per picowatt of 400-nanometer light, which means that the module counts about 19 percent of the photons that arrive. At 600 nanometers the same module counts only about 3 percent. That loss matters for the orange-red light of electrochemiluminescence near 620 nanometers, and an analyzer reading that light is better served by a cathode chosen for red light.

26 — One photon, a million electrons
photon PHOTOCATHODE DYNODES, EACH AT HIGHER VOLTAGE ANODE quantum efficiency up to 43 % at 350 nm (best blue cathodes) 60–90 % of electrons reach the first dynode one pulse: 1 to 10 million electrons (maker's handbook) schematic SHARE OF ARRIVING PHOTONS COUNTED, ONE MODULE 0 % 10 % 20 % 300 400 500 600 700 WAVELENGTH, nm maker's figures 477 nm dioxetane glow 620 nm electrochemiluminescence 19 % about 3 % schematic between points orange-red light needs a red-sensitive cathode
Label and signalReadoutFocal detail
Plate 26 — A photomultiplier turns one photon into a pulse of millions of electrons, but only a fraction of photons start a pulse. One maker's module counts about 19 percent at 400 nm and about 3 percent at 600 nm, so the cathode must suit the label's color.

The tube also produces pulses when no light falls on it, from electrons released by heat in the cathode. The rate of these pulses is called the dark count. It is small, about 50 a second typical for that module, but it is never zero, and it is the detector's own floor. A light-tight chamber and a label that makes its own light let a blank reading approach that floor.

Counting statistics

Counting light is counting random events. Photons from a steady source arrive at random moments, so a count made over a fixed time scatters from one reading to the next even when nothing changes. The scatter follows a simple rule from probability: the standard deviation of the count is the square root of its average. A reading of 100 counts carries a scatter of about 10, or 10 percent; 10,000 counts carry 100, or 1 percent; and a million counts carry 1,000, or 0.1 percent. The more light a measurement collects, the more precisely it knows the light.

Solved example: the smallest signal a 2-second read can see

Read for 2 seconds, a detector with 50 dark counts a second collects on average 100 dark counts, with a scatter of the square root of 100, which is 10. A signal has to add about three times that scatter, roughly 30 counts, before it stands clearly above the dark reading. If 19 percent of photons are counted, 30 counts need about 160 photons to reach the tube in those 2 seconds. In a real assay the blank also contains light from wrongly bound label, which raises the count and its scatter, so the practical floor sits higher (Chapter 5).

27 — Counts scatter by their square root
A 2-SECOND READ HOW OFTEN 60 80 100 120 140 160 COUNTS IN 2 s dark: mean 100, scatter 10 plus 30 counts from label about three times the dark scatter about 160 photons at 19 % counted give the extra 30 counts SCATTER SHRINKS WITH THE COUNT 0.1 % 1 % 10 % 100 10,000 1,000,000 RELATIVE SCATTER (LOG) AVERAGE COUNT (LOG) 100 counts: 10 % 10,000: 1 % 1,000,000: 0.1 % scatter = 1 ÷ √N
InterferentReadoutFocal detail
Plate 27 — Light arrives as random photons, so a count of N scatters by about √N. A dark reading of 100 counts scatters by 10, and a signal must add about 30 counts to stand clearly above it; more counts mean proportionally less scatter.

Counting statistics apply at every step of the chain. A digital assay counts beads, each lit or dark according to whether it carries a labeled target molecule, and its bead counts obey the same square-root rule (Chapter 17). In any assay, the scatter of a low-level sample is set partly by how few labels, and how few photons, it produces. Collecting more of an analyzer's light, or reading for longer, lowers the scatter of the blank, and with it the limit of blank, when photon counting contributes much of that scatter. When label bound without target dominates the blank, as Chapter 5 described, those steps help little.

From dozens to millions

At the other end of the scale, the detector cannot count fast enough. Each pulse takes time to register, and if pulses arrive faster than the electronics can separate them, some are lost. The same Hamamatsu module keeps its count in proportion to the light up to about 5 million counts a second, at which point it loses about 10 percent. From 50 dark counts to 5 million is five factors of ten, a far wider span than the less than three factors of absorbance in Chapter 6.

Analyzers report the result of this counting as a signal value. Abbott's documents call it relative light units and state that there is a direct relationship between the troponin in a sample and the light units detected. Above the counting range, an analyzer can measure the tube's current instead of counting pulses, a mode called analog, or it can reduce the light reaching the tube, or dilute the sample. Roche's troponin T assay quantifies troponin from 6 to 10,000 ng/L without dilution. Its US clearance allows a tenfold dilution, which extends reporting to 100,000 ng/L. The cobas e 801 can dilute automatically by factors up to 27,000, according to its maker.

The hook, seen from the detector

A detector reports how much light it saw, not which side of a curve the light came from. In a one-step sandwich assay, a very high concentration of target occupies the capture antibodies and the labeled antibodies separately, so fewer complexes carry both antibodies and the light falls again after a peak (Chapter 4). A single reading cannot tell a moderate sample on the rising side of the curve from an enormous one past the peak.

A single reading cannot reveal a hook

Because the curve rises and then falls, one light value matches two concentrations, and the analyzer reports the lower one. The defense is in the design: a stated hook-free limit far above the measuring range, a two-step format where the target allows it, and dilution rules for samples that the clinical picture suggests are extreme.

28 — One light value, two concentrations
measuring range of Roche's assay 1 100 10,000 1,000,000 SIGNAL, RELATIVE TARGET CONCENTRATION, ng/L (LOG) no hook up to 100,000 ng/L (FDA 2017) two steps: levels off one step: rises, then falls moderate sample enormous sample reported as the lower one schematic: curve shapes, not the assay's data
ReadoutFocal detail
Plate 28 — In a one-step sandwich, enormous concentrations fill capture and labeled antibodies separately and the signal falls again, so one light value matches two concentrations. Makers state a hook-free limit far above the range; a two-step design levels off instead of falling.

Makers therefore test how far the signal keeps rising and state a limit. The FDA's decision summary for Roche's one-step troponin T assay reports no hook effect up to 100,000 ng/L, ten times the top of its measuring range. Abbott's high-sensitivity troponin I assay captures the target and washes before adding the labeled antibody, and no hook was reported for it up to 500,000 ng/L, a hundred times its top of range. A two-step design avoids the classic hook because excess free target is washed away before the label arrives. Its signal levels off at the capacity of the capture antibody instead of falling. Roche's method sheet for its insulin assay reports no hook up to 20,000 microunits per milliliter, twenty times the assay's range.

Specify the detector from both ends of the range

Write the detector's requirements from the blank and from the top of the range at once: dark count and collection efficiency set by the limit of blank, linear count rate or analog range set by the highest signal before dilution, and the cathode chosen for the label's wavelength. Then confirm the hook-free limit with real high samples, since the curve beyond the top calibrator is where the detector's numbers and the chemistry interact.

+ What this chapter established
  • A photomultiplier turns a photon into a pulse of millions of electrons, with about 19 percent of 400-nanometer photons counted in one module and far fewer in the red.
  • Photon counts scatter by the square root of the count, so a 2-second read with 50 dark counts a second needs about 30 extra counts to show a signal.
  • Counting spans about five factors of ten before saturating; analog modes and dilution extend the range.
  • A one-step sandwich can hook at very high concentrations, so makers state a hook-free limit well above the range.