Ruling out in an hour.
+ The questionHow does a number measured at a few nanograms per liter let an emergency department send a patient home within an hour?
An hour in a guideline
At the end of August 2015 the European Society of Cardiology released new guidelines for patients with a suspected heart attack whose electrocardiogram does not show one particular pattern, called ST-segment elevation. An electrocardiogram is the tracing of the heart's electrical activity. The guidelines recommended a decision from two blood samples taken an hour apart, as an alternative to waiting three hours, wherever a high-sensitivity troponin assay with a validated algorithm was available. High-sensitivity describes an assay's analytical performance at low concentrations, not the share of heart attacks it finds; the last section of this chapter gives the definition. The guidelines added that the cut-off levels in that one-hour algorithm are specific to each assay. In the 2020 edition the one-hour algorithm became the best option and two hours the second best, and the 2023 guideline on acute coronary syndromes kept both. Acute coronary syndromes are the conditions in which blood flow to the heart muscle is suddenly reduced, heart attacks among them.
An hour is not long for a protein to accumulate. Troponin leaks from dying heart muscle over hours, and a patient who arrives soon after the pain began may not yet have much of it in the blood. The opening scene showed the problem: in the Basel study, the standard assay separated heart attacks from other causes much less well in patients who had arrived within three hours. The guideline's answer uses two properties of the sensitive assays together: a limit set in healthy people and a change measured between two blood draws.
The 99th percentile
In its fourth version, from 2018, the international definition of a heart attack calls troponin above the 99th-percentile upper reference limit a sign of injury to heart muscle. That limit is the value below which 99 percent of a healthy reference population falls. An acute heart attack requires a rise or fall in troponin, with at least one value above that limit, together with clinical evidence that the heart muscle was starved of blood. The definition also requires the high-sensitivity assays to measure with a coefficient of variation of 10 percent or less at the 99th percentile, and it recommends separate limits for women and men. The coefficient of variation is the standard deviation of repeated measurements divided by their mean, expressed as a percentage.
For Roche's troponin T assay, the FDA's decision summary gives limits of 19 ng/L overall from 1,301 healthy people, 14 ng/L from 656 women and 22 ng/L from 645 men. A percentile at the extreme of a distribution rests on few people: 1 percent of 656 is between six and seven, so the women's limit is set by the handful of highest values among them.
For Abbott's ARCHITECT high-sensitivity troponin I assay, cleared in 2019, the FDA's decision summary gives an overall limit of 28 ng/L from 1,531 people, with a 90 percent confidence interval of 22 to 33. A confidence interval is the range within which the true value plausibly lies, at the stated level of confidence, and this one shows how loosely even a large reference study fixes the limit. The European label for Roche's troponin T assay, the same chemistry as the US version above, gave 14 ng/L overall, from a different reference population. Chapter 14 showed that the next generation of the assay has different limits again.
The FDA's review of Roche's troponin T assay shows why one number is not enough early on. Using the 718 Basel patients of the opening, the FDA found that a single sample on arrival, judged against 19 ng/L, identified 93.5 percent of the heart attacks, and a sample at three hours 98.3 percent. Judged against the 99th percentile alone, a single sample taken on arrival misses one heart attack in fifteen and cannot be used to send patients home.
From one number to a change
The one-hour algorithm adds the change between two draws. With Roche's troponin T assay, a heart attack is ruled out under the 2020 guideline's table when the first result is below 12 ng/L and the second, an hour later, differs from it by less than 3 ng/L. A single very low first result, below 5 ng/L, can also rule a heart attack out under conditions the guideline sets. A heart attack is ruled in when the first result is 52 ng/L or more, or when the change in one hour is 5 ng/L or more. Everyone else is observed and tested again. The same table gives a separate row of numbers for each assay. For Abbott's troponin I assay, for instance, a heart attack is ruled out below 5 ng/L with a change under 2, or below 4 ng/L on the first result alone, and ruled in at 64 or more, or at a change of 6 or more.
The Basel researchers of the opening led a prospective study, published in 2015, at six centers in Switzerland, Spain and Italy. Such a study enrolls patients first and then follows them, rather than looking back at records. It tested the algorithm with Roche's troponin T assay on 1,320 patients, 229 of whom had a heart attack. The study was funded in part by several assay makers, and Roche donated the assay. The algorithm ruled out 786 patients, 59.5 percent, with a clinical sensitivity of 99.6 percent and a negative predictive value of 99.9 percent. It ruled in 216, of whom 78.2 percent had a heart attack, and left 318 for observation.
According to its decision summary, Roche's troponin T assay has a coefficient of variation of 10 percent at about 11 ng/L, which is a standard deviation of about 1.1 ng/L. Independent scatters add as squares. Each measurement contributes 1.1 squared, about 1.2, so the difference between two measurements has a squared scatter of about 2.4, and its standard deviation is the square root of that, about 1.6 ng/L, or the square root of two times 1.1. A change of 3 ng/L is therefore about two standard deviations of the difference. In a patient whose troponin did not change at all, measurement noise alone would produce an apparent change of 3 ng/L or more, up or down, in roughly one pair of samples in twenty. The algorithm's smallest step sits only just above the assay's own scatter.
A negative predictive value of 99.9 percent depends on how many of the patients tested have the disease, here 17 percent. The PCR guide in this library works through how predictive values depend on prevalence, and the same arithmetic applies here unchanged. The same algorithm, used in a population with more heart attacks, would leave more of them among the patients ruled out.
The observation zone is not a near-negative. In the 2015 study, 18.6 percent of the 318 patients left for observation had had a heart attack. The observation group's 30-day mortality was 1.6 percent, close to the 1.9 percent of those ruled in, against none among those ruled out. An algorithm's middle zone is a decision to keep testing, and a test that sorts many patients into it has moved work, not removed it.
What "high-sensitivity" means
The word has a definition. In a 2017 paper, a task force of the International Federation of Clinical Chemistry restated two criteria it had first proposed in 2012. The criteria are a coefficient of variation of 10 percent or less at the 99th percentile, and results above the limit of detection in at least half of healthy people, reported in whole nanograms per liter. The authors of the task force's paper disclosed research funding or consulting for several troponin makers. In the healthy population the task force used, Roche's high-sensitivity troponin T assay measured troponin in 7 percent of women and 43 percent of men at its limit of detection of 5 ng/L. The assay met the precision criterion but not the second one, and the task force noted that its rates were lower than recommended.
As the opening noted, a trade report states that when the FDA cleared the assay in January 2017, it asked the maker to call it "next generation" rather than high-sensitivity. Assays that measure most healthy people do more than lower a limit. A value that can be measured in a healthy person can be compared with the same person's value an hour later, and the change the algorithm uses needs that comparison at the bottom of the range.
The cut-offs in a one-hour algorithm apply to one assay, and the smallest change depends on that assay's precision near 5 to 12 ng/L. A laboratory adopting the algorithm should confirm its own analyzers' scatter at those concentrations, report in whole nanograms per liter, and agree with the emergency department what happens to patients in the observation zone. A maker introducing an assay needs its own validated algorithm, not another assay's numbers.
- The 2015 European guideline let a heart attack be ruled out from two samples an hour apart, with cut-offs specific to each assay.
- The 99th-percentile limit comes from healthy people, separately for women and men, and its extreme position makes it uncertain.
- The algorithm combines a low first value with a small change; the smallest change sits near the assay's own scatter.
- High-sensitivity means precise at the 99th percentile and measurable in at least half of healthy people, which Roche's troponin T assay only partly met.