+ Part VII · The field in 2026

Who builds the analyzers.

A cobas pro integrated solutions line joins clinical chemistry modules, which measure substances such as glucose, with immunoassay modules in one machine. In figures Roche supplied to the trade magazine CAP TODAY in July 2026, the largest configuration is rated at up to 6,400 tests an hour. Roche first sold one in the US in 2019 and reported more than 1,000 installed there and more than 5,000 elsewhere. Those lines belong to a business that Roche calls Core Lab and defines as immunoassays, clinical chemistry and its CustomBiotech unit. Core Lab sold 7.6 billion Swiss francs' worth in 2025, 55 percent of the company's diagnostics sales. This part is a single chapter, dated October 2026. It sorts the makers of such machines, and of the other analyzers and strips in use, by the way each reads its label.

22 — Who builds them

Analyzers and strips, by technique and maker.

+ The questionWho makes the analyzers and strips in use in 2026, and which way of reading a label does each rely on?

Four ways of reading, many makers

Two hundred tests an hour is a common rating for one immunoassay module in 2026, and the makers reach it with the different labels of Part II. The figures below come from the makers' own product pages unless a trade survey is named. Each is a best case for one mix of tests, as Chapter 10 warned.

Among the central-laboratory makers in this chapter, electrochemiluminescence belongs to Roche: light switched on at an electrode, the method of Chapter 9. Roche's cobas e 801 module is rated at up to 300 immunoassay tests an hour and its smaller e 402 at up to 120, both with reaction times of 9, 18 or 27 minutes.

Abbott's Alinity i, rated at up to 200 tests an hour per module, reads a flash from an acridinium label, the first family of Chapter 7. The FDA's review of one Alinity assay describes it as a chemiluminescent microparticle immunoassay with acridinium-labeled antibodies. Abbott states that its Alinity assays are free from biotin interference. Siemens Healthineers uses acridinium esters on its Atellica IM 1300 and 1600, rated at 220 and 440 tests an hour. It also uses them on the ADVIA Centaur XPT, rated at 240 tests an hour with a first result in 18 minutes.

Glow from an enzyme, the second family of Chapter 7, is the widest group. Beckman Coulter's Access analyzers use alkaline phosphatase with a light-producing substrate. For its DxI 9000, announced in 2023, the company moved to a faster substrate, which its 2019 poster described as producing its light within a minute instead of about six. Fujirebio rates its LUMIPULSE G1200 at 120 tests an hour. According to the FDA's decision summary for Fujirebio's Alzheimer's ratio test, the two assays of that test use alkaline phosphatase and a dioxetane substrate. Mindray states the same principle for its CL-2000i, which is rated at 240 tests an hour. QuidelOrtho's VITROS systems measure a peroxidase label with luminol, a method the maker calls enhanced chemiluminescence.

Snibe's MAGLUMI X8 is rated at up to 600 tests an hour per module and 2,400 with four modules. Snibe's product page describes chemiluminescence without naming the label.

Fluorescence, the method of Chapter 8, is still used in smaller systems. bioMérieux's VIDAS KUBE runs single-dose tests read by fluorescence at up to 36 an hour per module, and up to six modules can be stacked. Thermo Fisher Scientific's KRYPTOR analyzers use the energy-transfer method TRACE, which needs no washing. Quanterix sells single-molecule counting, the subject of Chapter 17, on analyzers labeled for research use only. In July 2025 the company completed its purchase of Akoya Biosciences.

49 — Who reads which label
ELECTROCHEMILUMINESCENCE ACRIDINIUM FLASH ENZYME GLOW CHEMILUMINESCENCE, LABEL NOT STATED FLUORESCENCE SINGLE-MOLECULE COUNTING Roche cobas e 801 300 Roche cobas e 402 120 Abbott Alinity i 200 per module, 800 per system, 1,550 with chemistry Siemens Healthineers Atellica IM 1600 440 Atellica IM 1300 220 ADVIA Centaur XPT 240 Beckman Coulter DxI 9000 450 maker gives 215 per m² Beckman Access 2 100 Fujirebio LUMIPULSE G1200 120 Mindray CL-2000i 240 QuidelOrtho VITROS 3600 189 Snibe MAGLUMI X8 600 per module bioMérieux VIDAS KUBE 36 per module Thermo Fisher KRYPTOR no hourly figure given Quanterix HD-X research use only makers' own figures unless marked; each is a best case for one test mix 0 100 200 300 400 500 600 TESTS PER HOUR, MAKER'S RATING vendor data, CAP TODAY Jul 2026
ReadoutFocal detail
Plate 49 — Central-laboratory makers split by the way they read a label: electrochemiluminescence at Roche, acridinium flash at Abbott and Siemens Healthineers, enzyme glow at Beckman Coulter, Fujirebio, Mindray and QuidelOrtho. Throughput ratings count different things and are best cases.
Solved example: what a throughput figure counts

Beckman Coulter gives the DxI 9000 a throughput of 215 tests an hour per square meter of floor. In a 2026 trade survey the vendor gave it up to 450 tests an hour, which at 215 per square meter implies about 2.1 square meters. Roche's cobas e 402 is rated at 120 tests an hour on 1.2 square meters, which works out at 100 per square meter. Abbott gives 200 tests an hour for one Alinity i module and 1,550 for an integrated system that includes chemistry, and the trade survey gives up to 800 for a four-module system. Three honest numbers for the same platform count three different things, so a comparison has to fix the unit first: per module, per system, per square meter, immunoassay alone or with chemistry.

At the bedside

Chapter 16 described the high-sensitivity troponin tests that reached the bedside: Abbott's handheld i-STAT, cleared in January 2025, the PATHFAST system of PHC Corporation of Tokyo, cleared in March 2024, and Siemens Healthineers' Atellica VTLi, CE-marked in 2021. Other point-of-care analyzers carry broader menus of tests. Radiometer's AQT90 FLEX runs up to 30 tests an hour from whole blood or plasma, with results in 11 to 21 minutes. Its menu includes troponin I and T, natriuretic peptide, D-dimer, procalcitonin, C-reactive protein and hCG, among others. In December 2023 Roche agreed to buy the parts of LumiraDx tied to that company's point-of-care platform, which runs immunoassays and clinical chemistry on one instrument. The agreed price was 295 million US dollars, and the sale was to be made out of administration, a UK insolvency procedure.

India

Judged from their own pages, India's makers sell mostly rapid tests, ELISA kits and chemiluminescence technology used under license. In June 2020 India's medical research council recommended a rapid antigen test for point-of-care use, SD Biosensor's Standard Q, which was made at a unit in Manesar, Gurugram. Agappe Diagnostics of Kochi signed a supply and license agreement with Fujirebio in March 2023. Under it, Fujirebio supplies the Mispa i60 and i121 analyzers and reagent raw materials, and Agappe makes the cartridge reagents in India and sells them under its own name. Agappe's 2024 brochure also lists acridinium-ester reagents and two chemiluminescence analyzers rated at 200 and 180 tests an hour, without saying where the analyzers are made.

J. Mitra & Co. of New Delhi sells rapid tests, ELISA kits, two bench-top chemiluminescence analyzers and an analyzer developed with a technology center at the Indian Institute of Technology Madras. Again, no place of manufacture is stated for the two chemiluminescence analyzers. Transasia Bio-Medicals sells ELISA kits and an automated ELISA processor that reads absorbance. Mindray sells its chemiluminescence analyzers in India from an office in Gurugram. No maker's page opened for this guide described a central-laboratory immunoassay analyzer designed in India.

Every "first" in a press release carries a qualifier

Snibe calls the MAGLUMI 2000 the first chemiluminescence analyzer from China to be cleared by the FDA. Siemens Healthineers called the VTLi the first high-sensitivity troponin I test from a fingerstick. Fujirebio and Agappe called Agappe the first local Indian company with a full chemiluminescence solution, and Roche called its pTau217 assay the first and only FDA-cleared single-biomarker blood test for both ruling in and ruling out amyloid pathology across primary and specialty care. Each claim is the maker's own, none was checked against a regulator's records for this guide, and each can hold only within its qualifier.

A field in motion

The field changes from one quarter to the next. Reporting their own counts to the same 2026 trade survey, the makers put about 3,625 of Abbott's Alinity ci-series systems in the US and about 16,600 elsewhere, and more than 950 of Diasorin's LIAISON XL analyzers in the US and more than 6,500 elsewhere. These counts are self-reported and unaudited. Siemens Healthineers said in 2026 that it was preparing to separate its diagnostics business, after a quarter in which that business's reported revenue fell 12 percent, according to a trade report. At constant currency, with the effect of exchange rates removed, the same report put the fall at 7 percent. Fujirebio has developed reagents for Sysmex's HISCL analyzers since 2020, and the two companies announced a basic agreement on wider collaboration around the start of 2024.

The products change too. By the company's account, Roche's sixth-generation troponin T assay was CE-marked by September 2025, with new reference limits (Chapter 14). The FDA cleared QuidelOrtho's VITROS high-sensitivity troponin I assay on 29 October 2025, with a claimed limit of detection of 0.59 ng/L. Four blood tests for amyloid pathology were cleared between May 2025 and August 2026, three of them immunoassays on conventional analyzers (Chapter 21).

50 — A field in motion
2023 2024 2025 2026 Mar 2023 Agappe–Fujirebio supply and license agreement (India) around start of 2024 Fujirebio–Sysmex basic agreement Jul 2025 Quanterix completes Akoya purchase 29 Oct 2025 VITROS hs troponin I cleared, LoD 0.59 ng/L claimed 19 Aug 2026 two more amyloid blood tests cleared Dec 2023 Roche agrees to buy LumiraDx point-of-care platform May 2025 first US blood test for amyloid (Ch 21) by Sep 2025 Roche troponin T Gen 6 CE mark 2026 Siemens Healthineers prepares diagnostics carve-out (trade report) THIS CHAPTER: OCTOBER 2026 every statement here is dated hollow: approximate date
ReadoutFocal detail
Plate 50 — Between 2023 and 2026 makers bought, merged and licensed platforms, and regulators cleared new troponin and Alzheimer's tests. The chapter's picture of who builds what is dated to October 2026.
Choose a platform by its menu and approvals before its speed

For a laboratory, the deciding facts are which cleared or CE-marked assays the platform runs in its own jurisdiction, how its assays are protected against biotin and heterophile antibodies, its throughput measured on the laboratory's own test mix, and the supply of reagents and service where the laboratory is. For a maker entering a market, the same list defines the competition: a faster analyzer without the assays a laboratory orders every day does not displace an installed one.

+ What this chapter established
  • Central-laboratory makers split by label: electrochemiluminescence at Roche, acridinium flash at Abbott and Siemens Healthineers, enzyme glow at Beckman Coulter, Fujirebio, Mindray and QuidelOrtho.
  • Throughput figures count different things, per module, per system or per square meter, and need a common unit before comparison.
  • India's makers sell mostly rapid tests, ELISA kits and chemiluminescence technology used under license, and no central-laboratory analyzer designed in India was found.
  • Companies and clearances changed through 2025 and 2026, so every statement in this chapter is dated to October 2026.