03 — Binding takes time

Hold and let go.

+ The questionIf an antibody grips its target so tightly, why does the time an assay allows change how much of the target it catches?

Binding is a balance

Placed in a small tube with an excess of antibody, each troponin molecule of Chapter 1 is close to thousands of binding sites, the fit is good, and the bond, once made, lasts for minutes or hours. It seems the target should be caught almost at once and kept for good. Neither is true. The reason governs every incubation, the period in which sample and reagents are left to react, and every wash and every protocol in the analyzers of Part III.

Two molecules bind only when diffusion brings them together in the right orientation, and a bound pair comes apart whenever the jostling of the surrounding water happens to break enough of its weak contacts at once. Both events happen all the time. The rate of binding is the association rate constant: for proteins meeting proteins, antibodies included, it is typically about a million per molar per second, a hundred to a thousand times slower than the limit that pure diffusion would allow, because most collisions arrive at the wrong angle. The rate of coming apart is the dissociation rate constant: for good antibodies, between about one in a thousand and one in a hundred thousand per second.

The dissociation rate constant divided by the association rate constant is the dissociation constant of Chapter 2. An antibody that binds at a million per molar per second and lets go at one in ten thousand per second therefore has a dissociation constant of 0.1 nanomolar. The dissociation rate also gives the lifetime of a single bond. Half of a set of bound pairs comes apart in about 0.7 divided by the dissociation rate constant: about 12 minutes at one in a thousand per second, about two hours at one in ten thousand and about 19 hours at one in a hundred thousand. Binding is never permanent; it is a balance that shifts with concentration and time.

How much is caught at equilibrium

When the target is scarce and the antibody plentiful, the fraction of target bound at equilibrium depends only on the antibody concentration measured against the dissociation constant. It equals the antibody concentration divided by the sum of the antibody concentration and the dissociation constant. Roger Ekins spent much of his career on the limits of immunoassays. He worked out what this rule gives in numbers: antibody at a hundredth of the dissociation constant binds about 1 percent of the target, at a tenth about 9 percent, at the dissociation constant itself about half, and at twenty times the dissociation constant more than 90 percent.

In a sandwich assay, the format of Chapter 4, two antibodies hold each target molecule between them. Sandwich assays are designed to sit at the top of that range: they use far more antibody than the dissociation constant, so that almost all the target can be caught. Roche's troponin T assay carries its antibodies at about 17 nanomolar in the reagent, the figure worked out in Chapter 2, and at several-fold less once sample, reagents and beads are mixed. For an antibody with a dissociation constant of 0.1 nanomolar, even a few nanomolar is many times the constant, and more than 95 percent of the troponin would be bound at equilibrium. The question is how long equilibrium takes to arrive.

How long equilibrium takes

With the antibody in excess, the approach to equilibrium follows a simple rule. The observed rate is the association rate constant times the antibody concentration, plus the dissociation rate constant, and half of the remaining distance to equilibrium is covered in 0.7 divided by that rate. After five such half-times the binding is about 97 percent complete, a rule of thumb from a 2020 guide to measuring binding. The more antibody, the faster the approach, which is a second reason for the large excess.

Solved example: minutes to equilibrium

Take an association rate of one million per molar per second and a dissociation rate of one in ten thousand per second. With antibody at 1 nanomolar, the association rate constant times the concentration is 0.001 per second, and adding the dissociation rate of 0.0001 gives an observed rate of 0.0011 per second. Half-way to equilibrium then takes 0.7 ÷ 0.0011, about 630 seconds or 10.5 minutes, and 97 percent takes five times as long, about 53 minutes. With antibody at 17 nanomolar, the concentration in the reagent of Roche's assay before mixing, the observed rate is 0.017 per second, the half-time is about 41 seconds, and 97 percent arrives in about three and a half minutes. Seventeen times more antibody makes binding in a well-mixed solution about fifteen times faster. Once the antibody is diluted several-fold in the reaction, binding takes several times longer, which is why the minutes of a protocol are budgeted with care.

05 — More antibody, faster capture
0 % 25 % 50 % 75 % 100 % 0 10 20 30 40 50 60 FRACTION OF TARGET BOUND TIME, min 9 min 18 min equilibrium at 1 nM: 91 % ANTIBODY 17 nM (ROCHE'S REAGENT) ANTIBODY 1 nM half-time 10.5 min at 9 min, 1 nM antibody has caught about 41 % computed: association 10⁶ per molar per second, release 1 in 10,000 per second
AnalyteFocal detail
Plate 05 — With an association rate of a million per molar per second, antibody at 17 nanomolar binds almost all of the target within about three minutes, while antibody at 1 nanomolar needs most of an hour. Excess antibody is bought for speed as much as for completeness.

That calculation assumes that every antibody and every troponin molecule are free in a well-mixed liquid. In many immunoassays they are not: one antibody is fixed to a surface, the wall of a plastic well or the surface of a bead, and the target has to travel to it. Then the time the target takes to reach the surface starts to matter as well.

Distance is time

How fast a molecule spreads through water by diffusion is measured by its diffusion coefficient. For a protein the size of troponin T it is about 0.0000007 square centimeters per second, a figure scaled from the measured value for albumin. The distance a molecule wanders grows only with the square root of time: about 0.7 millimeter in an hour, and four times that distance takes sixteen times as long. In a standard plastic well holding 100 microliters, the liquid stands about 3.2 millimeters deep. A troponin molecule near the top of an unstirred well needs most of a day to reach antibodies coating the bottom, so in an hour's incubation the antibody on the walls catches target only from a thin layer of the liquid next to it.

06 — Distance is time
ANTIBODY ON THE FLOOR OF A WELL CAPTURE BEADS SPREAD THROUGH THE LIQUID 100 µL, about 3.2 mm deep from the top: most of a day (about 19 h) an hour's diffusion reaches only this layer reached in about 1 hour: about 0.7 mm beads a few micrometers across, everywhere in the sample diffusion of a troponin-sized protein: about 0.0000007 cm² per second; distance grows with the square root of time
AnalyteAntibodyFocal detail
Plate 06 — A protein the size of troponin diffuses about 0.7 mm in an hour, so antibody coating the floor of a 3.2 mm-deep well samples only a thin layer of the sample in that time. Beads spread through the liquid bring the capture surface to the target instead.

Laboratory assays in plates make up for slow diffusion with long incubations, shaking and warmth. Analyzers shorten the distance instead. In Roche's troponin T assay, the two antibodies find the troponin while all three molecules are dissolved, which is fast because the antibody concentration is high. Beads coated with streptavidin then collect the complexes. Streptavidin is a bacterial protein that binds biotin, a small vitamin, almost irreversibly, and one of the two antibodies carries biotin. Depending on the analyzer, the beads are added after the first incubation or are present from the start. The beads, a few micrometers across, are spread through the whole reaction volume, so no molecule is ever more than a short distance from one. A 2008 analysis of surface sensors made the general point: transport of the target to the surface "can play as critical a role as the chemical reaction itself".

A complex can come apart during the wash

Time works in both directions. A bound pair with a dissociation rate of one in a thousand per second loses about 6 percent of its target in each minute of washing, because the wash removes free target and nothing rebinds. Antibodies for washed assays are therefore chosen for slow release as much as for tight binding, and wash steps are kept short.

Roche's analyzers show the trade-off between time and floor in their protocols. Each test on them takes 9, 18 or 27 minutes, and the troponin T assay comes in a 9-minute version for emergencies and an 18-minute version for routine work. The International Federation of Clinical Chemistry compiled a table of figures supplied by the makers. In that table the 18-minute version on the cobas e 601 analyzer family had a limit of detection of 2.05 ng/L and the 9-minute version 2.85 ng/L, while on the newer cobas e 801 both were listed at 3 ng/L. These are the figures Roche supplied for the Federation's 2019 table, and they are lower than the 5 ng/L of the assay's European label in Chapter 1. Shorter incubations can raise the floor, by an amount that depends on the whole system.

Budget the minutes

An assay protocol is a time budget split between binding, capture, washing and reading. When a project shortens it, the first questions are how close each step was to equilibrium, which step was limited by diffusion rather than chemistry, and how much target a wash of the new length will strip. Answering them with measured binding curves, not with the old protocol's habits, saves a failed verification later (Chapter 23).

+ What this chapter established
  • Binding and release both happen continuously; for good antibodies, release half-times run from minutes to many hours.
  • With antibody far above the dissociation constant, nearly all the target is bound at equilibrium, and more antibody gets there faster.
  • When an antibody sits on a surface, diffusion over millimeters can take hours, so analyzers bind in solution and spread capture beads through the sample.
  • Shorter protocols can raise the floor, and washes strip some target, so incubation times are a design trade, not a habit.