18 — Beyond one label

What sets the floor now.

+ The questionOnce single labels can be counted, what limits a test?

Four cleared tests, none of them digital

By October 2026 the FDA had cleared at least four blood tests that aid in assessing the amyloid plaques of Alzheimer's disease, and none of them ran on a digital analyzer. The count comes from this guide's search of the FDA's records, and the search may have missed a clearance. Amyloid plaques are deposits of the protein fragment beta-amyloid in the brain. Three of the tests were immunoassays on conventional analyzers of the kinds described in Part III, and two of those measured tau protein that carries a phosphate group at amino acid 217, called pTau217. The fourth used mass spectrometry, which identifies and measures molecules by their mass. Quanterix, whose method Chapter 17 described, had submitted its own test for clearance, and no decision on it had been found.

The puzzle is that pTau217 circulates at very low concentrations. In a small 2024 study on one analyzer, people with amyloid plaques had on average about 0.66 picogram per milliliter in their plasma and people without them about 0.11. A picogram per milliliter is the same as a nanogram per liter, so these concentrations sit well below the troponin limits of Part III. Yet the conventional assays reach them. Fujirebio's pTau217 assay runs on its Lumipulse analyzer as a chemiluminescent enzyme immunoassay of the kind in Chapter 7. According to the FDA's decision summaries, that assay has a limit of detection of 0.041 pg/mL, and Roche's electrochemiluminescent pTau217 assay has a limit of 0.056 pg/mL. Roche's cut-offs sit at 0.223 and 0.378 pg/mL, four to seven times above its limit of detection.

The makers of the digital analyzer of Chapter 17, Quanterix and Stratec, reported limits of about 0.002 to 0.02 pg/mL for other proteins. Those limits are lower again, although limits for different proteins and antibodies are not directly comparable. Whatever the commercial reasons for the outcome, the arithmetic shows that the conventional assays had already reached the concentrations at which patients differ. That finding changes the question this part set out to answer. Counting single labels removed the detector as the limit. What remains is the label that lights without a target, the share of the sample the assay actually looks at, how strongly the antibody holds its target, and the number of molecules in the drop.

The blank, one bead at a time

In a digital assay the blank is not a faint glow but a count. The 2016 paper on the HD-1 analyzer reported that backgrounds were generally between 0.005 and 0.01 enzymes per bead. At those levels, 0.5 to 1 percent of the beads light in a sample with no target, about 125 to 250 of 25,000 beads read. The paper does not say which molecules light them. By definition they are beads carrying label that no target holds, the background of Chapter 5 seen one molecule at a time.

Solved example: where a digital assay's floor sits

Take a blank of 0.005 enzymes per bead and 25,000 beads read, so that about 125 beads light with no target. By the square-root rule of Chapter 13, repeated blanks scatter by at least about 11 beads. Real blanks scatter more, since the 2016 paper notes that counting noise dominates only below about 50 background beads. The paper set its limit of detection at 2.5 standard deviations above the background, so a sample must light about 28 more beads than the blank to be detected, an extra 0.0011 enzyme per bead, and more when the blank scatters more than this minimum. The analyzer read 25,000 of the roughly 500,000 beads in the test, so those 28 beads stand for about 560 labeled molecules or more, spread over all the beads. Where counting noise dominates, halving the blank lowers the threshold only by the square root of two, to about 20 beads.

The example shows why a more sensitive camera would not help: every lit bead in the blank is already seen. What sets the floor is how many beads light without a target and how much that number scatters from one blank to the next. A 2012 theoretical paper by the company's scientists notes that lower concentrations of label reduce the background. But binding is a balance that shifts with concentration, as Chapter 3 showed, so less label also binds fewer of the captured molecules. The trade between signal and background that Chapter 5 described therefore returns at the scale of single molecules.

Molecules in the drop

Counting also exposes a limit that a glow hides: the number of molecules there are to count. On the HD-1, only about 5 percent of the beads in a test are sealed into wells and read. Researchers in the laboratory of David Walt, a co-author of the 2010 method, gave that figure in a 2020 paper. At 10 attomolar, a hundredth of a femtomolar, a 100-microliter sample holds about 600 molecules of the target. Even if every one were caught and labeled, reading one bead in twenty would see about 30 of them, with a counting scatter near 18 percent.

The same group sealed beads into droplets instead of wells and read up to 60 percent of 100,000 beads. For two signaling proteins of the immune system, limits of detection fell to 30 attomolar for interferon gamma, against 350 attomolar on the commercial analyzer, and to 20 attomolar for interleukin 2, against 550. At the lowest limit, that is about 1,200 molecules in the whole sample. The method was a research technique, not a product.

The droplet study's authors also noted that the antibodies' dissociation constants limit how low a concentration the assay can detect. As Chapter 3 showed, the share of target an antibody holds depends on how the antibody's concentration compares with that constant, and a complex that comes apart during the washes takes its label with it. In their 2012 theoretical paper, the company's scientists put the efficient range for digital ELISA at antibodies with dissociation constants from about 10 picomolar to 1 nanomolar.

38 — Where the molecules go
IN THE SAMPLE 10 aM in 100 µL CAUGHT AND LABELED BEADS READ about 5 % (HD-1) READOUT about 600 molecules not caught or lost in washes a share set by the antibodies (Chapter 3) about 30 counted if all were caught the result rests on a few dozen molecules 30 ± about 5 scatter near 18 % droplet method (research, 2020): up to 60 % of beads read; limits 20–30 aM blank: about 125 of 25,000 beads lit (AEB 0.005); +28 needed to detect arithmetic from the 2016 and 2020 papers; schematic proportions
AnalyteAntibodyLabel and signalInterferentReadoutFocal detail
Plate 38 — At 10 attomolar, a 100 µL sample holds about 600 molecules, and an analyzer that reads one bead in twenty sees at most about 30 of them. Below the detector's limit, the floor is set by molecules lost in capture and sampling and by beads that light without target.

Sample volume is the last factor that can be raised. A larger sample holds proportionally more molecules, and the tests in the 2016 paper used from 25 microliters of sample for most proteins up to 144 microliters for one viral protein. The limit on volume is practical: more sample means more time to capture from it, more beads to read, and more of everything else in blood that can stick.

Two antibodies that must meet

Another way to lower the background is to demand more before a signal counts. In a sandwich, a detection antibody stuck to the surface reads as if a target held it in place. In 2002 a group at Uppsala University in Sweden described a proximity ligation assay. In this assay, two binders, each carrying a short strand of DNA, must attach to the same molecule before their strands lie close enough to be joined into a new piece of DNA. The joined piece is then copied and measured, and a binder on its own leaves nothing to copy. The group used short DNA molecules called aptamers as binders and detected a growth factor at 40 zeptomoles, about 24,000 molecules, "without washes or separations". A zeptomole is a thousandth of an attomole.

In 2014, authors mostly from the company Olink described a version of the proximity extension assay that they called 96-plex. It used pairs of antibodies carrying matched DNA strands, which an enzyme extends into a new piece of DNA when both antibodies bind the same molecule, and it ran 92 protein assays together in 1 microliter of sample. The omics guide in this library explains how the resulting DNA barcodes are read. Running several assays in one sample is called multiplexing, and the Olink authors argued that conventional multiplex sandwich assays stop at about ten targets, because the chances of a wrong antibody pairing multiply with every target added. In their format a signal needs both antibodies on the same molecule and DNA strands that match each other.

Each added target adds ways to be wrong

With ten capture antibodies and ten detection antibodies in one well, there are ten intended pairs and ninety unintended ones, each a possible bridge if one antibody cross-reacts. The authors of the HD-1 paper expected cross-reactivity between antibodies to limit the analytical sensitivity of multiplex tests. Their multiplex tests read about 10,000 beads per target, against 25,000 to 40,000 in a single test, so the counting scatter grows as well. A floor measured for one target alone does not carry over to a panel.

39 — Two binders must meet
PROXIMITY LIGATION (2002) joined into new DNA binders: DNA aptamers in 2002 copied and measured alone: nothing to copy MULTIPLEXING (A) CONVENTIONAL MULTIPLEX SANDWICH A B C wrong pairing reads as signal each added target adds ways to be wrong (B) PROXIMITY EXTENSION A B tails do not match: no signal 10 capture + 10 detection antibodies: 10 intended pairs, 90 unintended conventional multiplexing stops at about ten targets (authors of the 2014 paper); omics guide, Chapter 9
AnalyteAntibodyLabel and signalInterferentReadoutFocal detail
Plate 39 — Proximity methods require two binders on the same molecule before their DNA strands can be joined or extended, so a binder stuck alone gives nothing. That also removes most wrong pairings, which multiply with every target added to a conventional multiplex.

What sets the floor

Chapter 4 traced the troponin T floor from 100 ng/L in 1992 to a quantitation limit of 6 ng/L in the 2017 assay. Abbott's high-sensitivity troponin I assay detected 1.7 ng/L in its 2019 US clearance. In 2007 scientists at the company Singulex described a single-molecule counting assay that detected 0.2 ng/L. Across those years, what set the limit changed. The early floors in immunoassays were set by the label and the way it was read: a radioactive count, a color in a photometer. Labels that give light from a chemical reaction or at an electrode, read in the dark, cut the reader's own noise until the detector stopped being the limit. The analyzer's washing, timing and calibration kept the floor steady from run to run.

With labels counted one by one, the floor is set by what the antibodies do and by what the sample offers: the labels left stuck without a target, the share of target caught and kept through the washes, and the number of molecules in the volume tested. The answer to the central question therefore comes in sequence. The label and its reader set the floor first, the instrument held it, and today the antibodies and the drop set it.

40 — Who set the floor
healthy people: about 1–50 ng/L 0.1 1 10 100 1990 1995 2000 2005 2010 2015 2020 2025 TROPONIN DETECTION LIMIT (OR BOTTOM OF RANGE), ng/L, LOG LABEL AND READER SET THE FLOOR THE INSTRUMENT HOLDS IT T, first assay (Boehringer Mannheim): range from 100 T, 2nd generation: 12 T, high-sensitivity (Europe): 5 I, Abbott ARCHITECT: 1.7 I, VITROS hs (maker's claim): 0.59 T, US clearance: LoD 3–5, reported from 6 I, single-molecule counting (Singulex staff): 0.2 ANTIBODIES AND THE DROP below the detector, the blank and the drop decide T = troponin T, I = troponin I
AnalyteReadoutFocal detail
Plate 40 — The troponin floor fell from about 100 ng/L in 1992 to a few ng/L and, in single-molecule counting, to 0.2 ng/L. The limit passed from the label and its reader to the analyzer that held it steady, and then to the antibodies and the number of molecules in the drop.

A lower floor is worth having only where the biology calls for it. An international task force, whose criteria Chapter 20 sets out, tabulated results from a 2012 study that measured several assays in one healthy population. At the limit of detection of 5 ng/L then on its label, Roche's high-sensitivity troponin T assay measured troponin in 43 percent of the men and 7 percent of the women. Abbott's ARCHITECT troponin I assay measured troponin in 80 percent of the men and 67 percent of the women, at the limit of detection of 1.9 ng/L that the task force's table gave it, against 1.7 ng/L in its US clearance. Chapter 20 shows why measuring healthy people matters for a decision made within an hour. For pTau217 the conventional analyzers already reach the concentrations that separate patients. Below that point a lower floor adds little to the decision the test supports, unless it also improves precision at the cut-offs, where about a fifth of patients in two of the cleared tests, which use two cut-offs, get an intermediate result that is neither positive nor negative.

Find the binding limit before paying for sensitivity

Before specifying a more sensitive assay or analyzer, work out which limit sets the floor at the decision concentration: the scatter of the blank, the share of molecules caught, the share of the sample read, or the number of molecules in the volume. A better detector helps only with the first, and only when the blank's scatter comes from the detector rather than from label stuck where no target holds it.

+ What this chapter established
  • On a digital analyzer the blank is a count of beads lit without target, and its scatter, not the camera, sets the floor.
  • Reading a twentieth of the beads, and catching only part of the target, can leave a result resting on a few dozen molecules.
  • Proximity methods require two binders on one molecule before a signal counts, which cuts background and lets many assays share a sample.
  • The floor passed from the label and its reader to the antibodies and the drop, and a test needs only the floor its biology sets.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.