+ Part VI · Reading the number
When a result is right, and when it is not.
On 28 November 2017 the FDA issued a safety communication warning that biotin can make laboratory tests read falsely high or falsely low. Biotin is a B vitamin, sold in high doses in supplements for hair, skin and nails. The agency reported that one patient taking high levels of biotin had died after falsely low troponin test results. Biotin harms neither a reagent nor an analyzer; it works through the chemistry many tests are built on. The three chapters of this part read results: what makes a well-run assay wrong for one patient, how the numbers from Roche's troponin T assay become a decision within an hour, and why the first blood test cleared to aid the diagnosis of Alzheimer's disease reports a ratio, with a zone that gives no answer.
Hook, heterophiles and biotin.
+ The questionWhat can make a well-run assay give one patient a wrong number?
Seven samples in 21,000
A 2004 review of interference in immunoassays cites a study of hospital samples run on immunoassays that contained blocking reagents. In that study, a patient's own antibodies interfered in seven samples in 21,000. The same review reports that human antibodies against mouse proteins, which can bind the mouse antibodies many reagents use, have been found in 30 to 40 percent of patient samples. Both numbers can be true at once, because the blocking reagents of that study and the other defenses that makers build into the reagent, described below, account for much of the gap between them.
Human antibodies that bind the antibodies of other animals take two forms. Heterophile antibodies have no clearly defined origin and bind the antibodies of two or more species. Human anti-animal antibodies are made against one animal's proteins, for example after treatment with mouse antibodies, and bind them strongly and specifically; the most common are human anti-mouse antibodies, or HAMA. In practice the two names are often used for each other. A 1999 review notes that they can arise from treatment with animal antibodies and from other causes, can reach high levels and can persist for years. Many immunoassay antibodies are made in mice, as Chapter 2 described, so a heterophile antibody can bind the capture antibody with one arm and the labeled antibody with the other. The analyzer then reads a sandwich with no target in it, and the result is falsely high. An antibody that binds only one reagent antibody can instead stop it catching the target, and the 2004 review offers that blocking as the likely route to a falsely low result. Rheumatoid factor, an antibody that binds the stem of other antibodies, can bridge in the same way, and the same review notes that it can cause false positives in troponin assays.
Makers build their defenses against all of these into the reagent. They add non-immune animal serum or mouse antibodies that bind the patient's antibodies before they reach the sandwich; they use antibody fragments that lack the stem; and they use antibodies built partly from human sequence, such as the labeled antibody of Roche's troponin T assay. These defenses are not complete. The 2004 review summarizes a study in which 66 laboratories in seven countries measured 74 analytes in the blood of ten donors, chosen because their blood was known to interfere. About 6 percent of the results were false positives, and the kits' blockers failed to prevent about two thirds of the interferences. Because the donors were selected for interference, the 6 percent is a measure of how hard the problem is, not of how often it occurs.
Biotin, the reagent's own glue
Biotin interferes for a different reason: many assays use it themselves. Chapter 4 described Roche's troponin T assay, in which the capture antibody carries biotin and the magnetic beads carry streptavidin, a protein that binds biotin very tightly. Free biotin in the patient's blood occupies the streptavidin on the beads, so the capture antibody has nowhere to attach and is washed away with the troponin and the label it holds. In a sandwich assay the result reads falsely low. In a competitive assay, where less label means more target, the same loss reads falsely high.
A 2017 study in JAMA measured the effect directly. Six healthy adults took 10 milligrams of biotin a day for a week. Their blood was drawn about two hours after the last dose and run on 37 immunoassays on four systems. Interference appeared in 9 of the 23 assays that used biotin and in none of the 14 that did not. Five of eight competitive assays read falsely high and four of fifteen sandwich assays falsely low; the result of one assay for thyroid-stimulating hormone fell by about 94 percent.
The FDA's warnings of 2017 and 2019 described supplements of up to 20 milligrams, then pills of up to 100 milligrams. They also described doses of up to 300 milligrams a day, recommended by physicians for some conditions, and put the resulting blood levels at up to 1,200 ng/mL. In its 2020 guidance the FDA asked makers to test biotin up to 3,500 ng/mL, about three times the highest concentration expected in patients. Roche's troponin T assay had been cleared in 2017 after testing biotin only up to 20 ng/mL. Its instructions said not to draw blood from patients taking more than 5 milligrams a day until at least 8 hours after the last dose.
A 2021 study measured plasma biotin peaks of about 1,130 to 1,160 ng/mL one to one and a half hours after a 300-milligram dose; its authors included the maker's staff. Against the original tolerance of 20 ng/mL, 1,160 ng/mL is 58 times too much. In molecules the contrast is starker. Biotin weighs 244 grams per mole, so 1,160 ng/mL is about 4.7 micromolar. Troponin T at 14 ng/L, the US 99th-percentile limit for women, is about 0.39 picomolar, so the blood holds about 12 million biotin molecules for every troponin molecule. Each of those biotin molecules can take a place on the streptavidin.
Roche changed the reagent rather than the instructions. It added an antibody that captures free biotin and lengthened the linker that holds biotin on the capture antibody. The FDA cleared the reformulated assay on 21 September 2021, and the decision summary records no interference up to 1,500 ng/mL. Abbott states that its Alinity assays are free from biotin interference, a maker's claim. In June 2022 the FDA's list of troponin assays whose biotin risk had not been addressed still named the 2017 version of Roche's troponin T assay, because the list goes by clearance number.
Too much target, or target in disguise
The high-dose hook of Chapters 4 and 13 turns into a clinical error when an enormous concentration reads as a modest one. The 2004 review cites a patient whose prolactin, a pituitary hormone, read 1,500 milliunits per liter undiluted and 950,000 on dilution, about 630 times higher. A 2017 case report describes a woman with an advanced molar pregnancy, an abnormal growth of placental tissue, who had two negative urine pregnancy tests. Measured after proper dilution, her serum hCG was above 900,000 milli-units per milliliter. The defense is dilution: a laboratory that suspects an extreme value reruns the sample diluted, and an analyzer can do so automatically.
Some targets circulate bound to the patient's own antibodies. Macroprolactin, prolactin bound mostly to immunoglobulin G, is generally considered biologically inactive, yet assays detect it. The 2004 review reports it in 24 percent of sera with raised prolactin, and differences of 2.3- to 7.8-fold between assays. Troponin can do the same. In a 2022 case report, a patient's troponin I read 872 ng/L on one analyzer, against that assay's upper reference limit of 54 ng/L for women, and peaked at 3,165. Roche's high-sensitivity troponin T assay gave 12 ng/L, against its limit of 14, and a second troponin I assay 11 ng/L, against its own limit of 9. Polyethylene glycol is a polymer that makes antibodies and other large proteins fall out of solution. Precipitating the antibodies with it left only 31 percent of the troponin I, below the 40 percent threshold the authors cited for suspecting a complex of troponin and antibody. The patient had already been treated in error for inflammation of the heart, and the troponin I was still 130 ng/L a year later.
Small molecules bring a different problem. The antibody of a competitive assay for a drug or hormone can bind related molecules of similar shape, so the assay measures them too. The 2004 review cites immunoassays for the transplant drug cyclosporin that read up to 174 percent higher than a chromatography method, because they also counted the drug's breakdown products, which chromatography separates from the drug before measuring it.
An interference that produces an impossible value is caught at once. One that moves a result into the normal range looks like any other normal result: a falsely low troponin in a patient taking biotin, or a modest prolactin from an enormous tumor. Interference checks therefore have to be triggered by disagreement between the number and the patient's clinical picture, not only by the number itself.
What the analyzer and the laboratory can do
The reagent carries most of the defense: blockers, fragments, partly human antibodies, an antibody that captures free biotin, a two-step format that washes away excess target, and a hook-free limit far above the measuring range. The analyzer adds rules: it flags results above the measuring range and reruns them diluted.
The laboratory runs the last checks. Comparing results that should agree can expose a complex. In a 2021 case series, four patients with troponin T bound to antibodies were identified when the 9-minute and 18-minute versions of the same assay disagreed, with a ratio between the two results below 0.80. The 2004 review lists the other checks: diluting the sample to see whether the result falls in proportion, adding a blocking reagent, precipitating the antibodies, or measuring with a method that uses antibodies from another species. For biotin, the matching check is a method that uses no biotin. A result that changes by more than the test's own scatter under one of those checks was not a measurement of the target alone.
When specifying a new assay, specify its interference checks too: the biotin, heterophile and rheumatoid factor levels it tolerates, its hook-free limit, the dilution a laboratory should use to test linearity, and a second method with different antibodies for confirmation. Write the conditions that should prompt a check into the instructions, so that a laboratory can act on a result that disagrees with the patient.
- Patients' own antibodies can bridge or block reagent antibodies; blockers and partly human antibodies make that rare but not impossible.
- Free biotin occupies the streptavidin that biotin-based assays use, reading falsely low in sandwiches and falsely high in competitive assays.
- Extreme concentrations can hook, and targets bound to antibodies can read high on one assay and normal on another.
- Reagents carry most defenses; analyzers add flags and dilution; laboratories confirm with dilution, blocking and a second method.