+ Part III · The analyzer as a system
Holding the floor steady, test after test.
The cobas e 801 is the analyzer that now runs Roche's troponin T assay in many hospital laboratories, and its maker states that one module can report up to 300 immunoassay results an hour. Every one of those results comes from a reaction that must be timed, warmed, washed and read exactly like the calibrators it is compared with, which are samples of known concentration. The chemistry of Parts I and II sets how low a floor can be. This part is about the machine that has to reach that floor on every test: how a tube becomes a result, how liquids are moved without carrying one sample into the next, why minutes and tenths of a degree matter, how light becomes a number, and how the number is tied to a scale that another laboratory can share.
One test, many subsystems.
+ The questionWhat has to happen, and in what order, between a tube of blood and a reported number?
A tube's path through the machine
A blood tube arriving at an analyzer carries a barcode, and the barcode is the first input the machine reads. From the barcode, the laboratory's information system tells the analyzer which tests have been ordered on that sample. A sampling arm then fits a fresh disposable tip, finds the surface of the liquid, draws up a few microliters of plasma while sensors watch for clots, bubbles and foam, and dispenses it into a single-use reaction cup. On a cobas e 801 the sample volume per test can be set between 4 and 60 microliters, and every pipetting uses a new tip, according to the maker's specifications.
Reagents come next. Each test's antibodies, beads and buffers sit in packs in a refrigerated store on board. The e 801 holds up to 48 packs at 5 to 10 °C and identifies each by a radio-frequency tag. A reagent arm adds the right volumes to the cup, a mixer stirs it, and the cup moves into an incubator held at 37 °C, plus or minus 0.3 °C. At the end of the incubation the mixture goes to the measuring cell of Chapter 9, where the beads are captured, washed and read. The analyzer's software turns the light into a concentration with the calibration curve, checks the result against flags and limits, and sends it to the laboratory system.
Each of those steps belongs to a subsystem that an engineer can name: sample handling, liquid handling, reagent storage, incubation, separation, detection, and the control software that schedules them. None of them is clever on its own. The design problem is that all of them must act on each test at exactly the right moment, the same way for every test and for every calibrator, while dozens of other tests are at other stages in the same machine.
Many tests in flight
The e 801 runs assays of fixed lengths, 9, 18 or 27 minutes. An 18-minute assay cannot be run one test at a time if the analyzer is to report hundreds of results an hour. The answer is a pipeline, in which many tests are in progress at once, each at a different step. On the e 801, a new cup starts every 12 seconds, so 300 tests can begin each hour. Each spends its incubation in one of 94 positions on the incubator, then leaves for the measuring cell as the next cup arrives.
A new test every 12 seconds, each incubating for 18 minutes, means 18 × 60 ÷ 12 = 90 tests in the incubator at any moment, close to the e 801's 94 positions. A 9-minute protocol needs 45. The smaller cobas e 402 starts a test every 30 seconds, for 120 an hour, and holds 36 tests of 18 minutes in its 38 positions. The first result of a run arrives after the full assay time, and after that one result leaves every cycle. A 9-minute protocol shortens the wait for each result; it does not raise the number of results per hour, which the cycle time sets.
The same arithmetic shows where the limits lie. On the maker's own numbers, a run made only of 27-minute assays would need 135 positions at full speed, more than the incubator holds, so such a mix must slow down. For its smaller cobas e 402, the maker states that throughput "may differ based on the mix of test orders per sample". Consumables set another limit. The e 801 holds 15 magazines, each of 105 tips and 105 cups, enough for 1,575 tests. At 300 tests an hour that supply lasts 5 hours 15 minutes, the unattended running time that its specifications list.
A maker's "up to" figure assumes a favorable mix of assay times and few reruns, dilutions or calibrations. The real rate in a laboratory depends on its own test menu, and on how often samples need repeating. Comparing analyzers by their headline rate without the laboratory's own mix compares two assumptions.
The conversion chain
An analyzer never measures troponin. It measures something several steps removed and converts it back. Troponin molecules in the sample become sandwiches on beads, and the sandwiches put labels within a few micrometers of an electrode. The labels emit photons, a fraction of the photons release electrons in the detector, and the electrons become a count or a current. The count becomes a signal value for that test, the calibration curve turns the signal into a concentration, and software compares the concentration with limits and flags before it becomes a reported number.
Every conversion in that chain can change the result without any change in the troponin. A pipetting error changes how many molecules enter the cup; a warm incubator changes how many become sandwiches in nine minutes; a fouled electrode changes how many photons each label gives; a drifting detector changes the count; an out-of-date calibration changes the conversion. The rest of this part takes the conversions in turn. The quality control samples of Chapter 14 exist because no analyzer can check each conversion separately on every test.
Other makers, the same skeleton
Other analyzers make different choices within the same set of subsystems. Abbott rates its ARCHITECT i2000SR at up to 200 tests an hour. The analyzer draws samples with a fixed probe that is washed between samples, and the maker specifies the probe's carry-over, the fraction of one sample that passes into the next, at no more than 0.1 part per million (Chapter 11). Abbott's troponin assay on that analyzer has two incubations, each followed by a wash, and a flash label. The 2017 specification sheet for Siemens Healthineers' Atellica IM 1600 gives a rate of up to 440 tests an hour. That analyzer uses disposable tips, an acridinium label read by a photomultiplier, sample volumes of 10 to 100 microliters and assay times from 10 to 54 minutes, according to the same sheet. The subsystems are the same in each machine; what differs is where each maker spends the time, the space and the cost.
For an analyzer or a reader, list every conversion from molecule to reported number, then give each one an owner, a requirement and a check. The requirements of the subsystems follow from the chain: pipetting precision from the volume step, incubator tolerance from the binding step, detector noise from the photon step. The verification plan of Chapter 23 can then be traced to them.
- A tube's barcode starts a sequence of sampling, reagent addition, incubation, separation, detection and calculation, each a subsystem.
- Analyzers pipeline tests: on the e 801 a new test starts every 12 seconds, and an 18-minute assay keeps about 90 tests in flight.
- The analyzer converts troponin into sandwiches, photons, counts and finally a concentration, and every conversion can move the result.
- Makers choose differently among tips and probes, flash and electrode labels, one or two incubations, within the same set of subsystems.