08 — Fluorescence

Waiting for the glow.

+ The questionFluorescence works well enough to count single cells in a flow cytometer; why must an immunoassay wait before reading it?

Light in, light out, and light everywhere

A europium ion can be held in an organic chelate, a complex in which an organic molecule, the ligand, binds the metal ion at several points at once. The ligand absorbs ultraviolet light near 340 nanometers and passes the energy to the ion, which gives back red light near 615 nanometers. The ion keeps giving that light back for close to a millisecond after the exciting flash has ended. That long emission, unusual among labels, is the subject of this chapter, because it solved a problem that ordinary fluorescence could not.

Fluorescence is the obvious way to read a label. A dye absorbs light of one color and returns light of a longer wavelength, and filters separate the two, as the flow cytometry guide in this library explains in detail. But an immunoassay reads its label in a plastic well or cuvette full of serum and buffer, not in a narrow stream passing a focused laser, as a flow cytometer does. The exciting light scatters off the plastic and the liquid. Serum contains proteins and other molecules that fluoresce on their own. And some of the exciting light leaks through any filter. All of that unwanted light arrives together with the label's light, from the moment the lamp fires.

Timing can separate the label's light from the unwanted light. The unwanted fluorescence and the scattered light die away in about 10 nanoseconds, according to Revvity, which makes the DELFIA system described below. Ordinary fluorescent dyes fade just as fast, so their light and the background arrive together and cannot be told apart by timing. A label whose glow lasts tens of thousands of times longer can be read after the background has gone, and reading a label after such a delay is what time-resolved fluorescence means.

A label that glows for a millisecond

The lanthanides, a row of metals at the bottom of the periodic table, give off light when their electrons jump between energy levels, and those jumps are slow by atomic standards. Europium, samarium and terbium emit light in sharp, narrow bands of wavelength when they are held in a molecule that absorbs light and passes the energy on. Their decay times run from tens of microseconds to about a millisecond, where the decay time is how long the emission takes to fall to about 37 percent of its starting strength. In 1983 E. Soini and H. Kojola described a time-resolved fluorometer for such labels, built around an ordinary xenon flash lamp. Its one-second measurements could detect as little as the radioactive methods then in use.

Revvity now sells a time-resolved method called DELFIA, and its user guide gives the timing: an excitation flash, a wait of 400 microseconds, a counting window of 400 microseconds, and a thousand such cycles a second. Before reading, an acidic enhancement solution releases the europium from the label and binds it into a new, brightly fluorescent complex. The maker says that this step multiplies the fluorescence by one to ten million times.

Solved example: what a 400-microsecond wait keeps

Take a europium decay time of 730 microseconds and a background decay time of 10 nanoseconds. After a 400-microsecond wait, the europium has given out about 42 percent of its light and keeps about 58 percent; the window from 400 to 800 microseconds collects about 24 percent of its total. The background, in the same 400 microseconds, has passed 40,000 of its decay times, and what remains of it is far too small to measure. The wait gives up about two fifths of the label's light and almost all of the background's, and closing the window at 800 microseconds leaves about a quarter of the label's total counted.

15 — Wait, then count
WAIT 400 µs COUNT 400 µs 1 1 in 10 1 in 100 1 in 1,000 0 200 400 600 800 1,000 LIGHT, RELATIVE (LOG) TIME AFTER THE FLASH, µs excitation flash scatter and serum fluorescence: about 10 ns europium decay 730 µs only the label's light is left europium keeps 58 % of its light after 400 µs; the window collects about 24 % of the total 1,000 cycles a second (DELFIA user guide)
Label and signalTriggerInterferentReadoutFocal detail
Plate 15 — Background fluorescence and scattered light die away in about 10 nanoseconds; a europium label glows for close to a millisecond. Waiting 400 µs before counting gives up three quarters of the label's light and almost all of the background.
A delayed read gives up most of the label's light

Waiting 400 microseconds and then counting for 400 collects only about a quarter of a europium label's emission, and labels that decay in microseconds would lose almost all of theirs. Time-resolved fluorescence works only with long-lived labels and only because the background it rejects is far larger than the light it discards.

The gain is the same in kind as chemiluminescence's: the detector reads its label against near-darkness. But the darkness here is made by timing, with a lamp in the instrument. The optical background therefore comes from the lamp's own afterglow, the plate's phosphorescence and the detector's dark counts. Phosphorescence is a slow glow that some materials give off after light has fallen on them, and dark counts are the counts a detector registers when no light reaches it (Chapter 13). In an assay, labeled antibody stuck without target and stray europium usually set the floor. The DELFIA guide gives typical plate backgrounds of 100 to 1,000 counts, depending on the plate.

Reading without washing: energy transfer

Long-lived labels allowed a further step, which Chapter 5 mentioned as one of the few ways around the wash: telling bound label from free label without washing at all. In 1993 a chemist at the French company CIS bio international described a europium label for this purpose. The europium is held in a cage-like molecule called a cryptate and attached to one antibody as an energy donor. A second antibody carries an acceptor molecule. When both antibodies bind the same target molecule, donor and acceptor sit a few nanometers apart, and the europium passes its energy to the acceptor, which then emits at its own wavelength. Because the acceptor receives its energy from the long-lived europium, its light is delayed as the europium's is. Free antibodies are scattered through the liquid, too far apart for the transfer.

Thermo Fisher Scientific's B·R·A·H·M·S KRYPTOR analyzers use this method under the name TRACE. The company's white paper gives the sequence: a laser pulse at 337 nanometers excites the europium, the instrument waits 50 microseconds, and it then measures light at 665 nanometers from the acceptor and at 620 nanometers from the europium itself. The ratio of the two signals corrects for a cloudy or colored sample, and the paper states that no washing or separation step is needed. The 1993 paper reported that its assay measured the hormone prolactin down to 0.3 microgram per liter in human serum.

16 — Light from the acceptor only when two antibodies meet
BOTH ANTIBODIES ON ONE TARGET target energy transfer: a few nm transfer only when they sit nanometers apart laser 337 nm 620 nm 665 nm europium cryptate (donor) acceptor FREE ANTIBODIES, FAR APART 620 nm donor still emits no transfer: no 665 nm measure after a 50 µs wait: ratio 665 / 620 nm corrects for cloudy or colored samples 1993: prolactin to 0.3 µg/L in serum no washing or separation (maker's white paper)
AnalyteAntibodyLabel and signalTriggerReadoutFocal detail
Plate 16 — In TRACE the label is a pair: a long-lived europium donor and an acceptor that lights only when both antibodies hold the same molecule. Reading the ratio of the two wavelengths after a short wait lets the analyzer skip the wash.

Outside the central laboratory

Long-lived labels suit small instruments, because the timing that rejects background depends less on a dark, precisely built optical chamber than on a fast switch between flash and count. The FDA cleared Radiometer's AQT90 FLEX analyzer in January 2012 with a test for myoglobin, cleared a CK-MB test for it under a separate decision, and in 2017 cleared its myoglobin and CK-MB test kits. The analyzer reads a europium chelate by time-resolved fluorescence, counting single photons, in a test cup that holds the antibodies, according to the FDA's decision summary. The maker lists troponin among its tests and gives results in 11 to 21 minutes.

A fluorescence reader can also measure an enzyme label instead of a lanthanide. bioMérieux's VIDAS analyzers use what the company calls an enzyme-linked fluorescent assay: an enzyme on the detection antibody turns a substrate into a fluorescent product, so the amplification of Chapter 6 is read as light rather than color. The company lists a high-sensitivity troponin I test on these analyzers that takes 20 minutes and covers 1.5 to 40,000 pg/mL, a span of more than four factors of ten.

Fluorescent beads put fluorescence to a different use. Authors from Luminex described the company's bead arrays in 1997. In these arrays, tiny beads are color-coded so that dozens of assays run in one sample and are read in a flow cytometer, a method the flow cytometry guide in this library covers.

Set the gate from two decay times

The gate of a time-resolved reader is its delay and counting window, and both are specified from the label's decay time and the background's. The delay must be long compared with the background's decay time and short compared with the label's, and the counting window must collect enough of the label's light to make the count precise. A homogeneous energy-transfer assay, one run without any wash, adds a third requirement, a ratio channel that corrects for sample color and turbidity, or cloudiness. The timing and the ratio channel both belong in the reader's design inputs, the requirements that its design must meet.

+ What this chapter established
  • Ordinary fluorescence in serum competes with scattered light and the sample's own glow, which die away in about 10 nanoseconds.
  • Europium and other lanthanide labels glow for tens of microseconds to about a millisecond, so a delayed read keeps the label and drops the background.
  • Energy transfer between two labeled antibodies on one target lets an assay tell bound from free without washing.
  • The delayed read discards most of the label's light, so the method depends on long-lived labels and stable timing.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.