pTau217 and amyloid.
+ The questionWhy does the first cleared Alzheimer's blood test report a ratio of two proteins, with a middle zone that gives no answer?
Two numbers and a remainder
Of the positive results, 91.7 percent were confirmed by a brain scan or a spinal fluid test, and of the negative results, 97.3 percent. The FDA gave those figures on 16 May 2025, when it cleared what it called the first in vitro diagnostic device that tests blood to aid in diagnosing Alzheimer's disease. Fewer than 20 percent of the 499 patients in the study received neither a positive nor a negative result. The device was Fujirebio's Lumipulse G pTau217/β-Amyloid 1-42 Plasma Ratio. The FDA cleared it through a 510(k) about eight months after the application arrived. Its predicate was the company's spinal fluid test of 2022, the earlier device to which the FDA found it substantially equivalent (Chapter 26).
The test does not diagnose the disease. It looks for amyloid pathology: deposits of a protein fragment called beta-amyloid in the brain. In clinical practice, amyloid pathology is assessed by positron emission tomography, or PET, a brain scan that shows the deposits, or by a test of the fluid around the spinal cord. The FDA cleared the test only for adults aged 50 and older who are seen in a specialized care setting with signs and symptoms of cognitive decline. The cleared indications state that it is not intended as a screening or stand-alone diagnostic test. Its results must be interpreted with other clinical information about the patient.
Two assays, one number
The decision summary describes the test as two immunoassays run on one analyzer, Fujirebio's LUMIPULSE G1200, using 40 microliters of plasma. Each is a two-step sandwich with an alkaline phosphatase label and a dioxetane substrate that glows at 477 nanometers, the chemistry of Chapter 7. One assay measures pTau217: tau, a protein of nerve cells, carrying a phosphate group at amino acid 217. The pTau217 assay measures from 0.047 to 10 picograms per milliliter, with a limit of detection of 0.041 pg/mL. The other assay measures beta-amyloid 1-42, the 42-amino-acid form of the fragment, from 0.8 to 500 pg/mL. A picogram per milliliter equals a nanogram per liter, so the pTau217 range starts far below even the troponin limits of detection in Chapter 1.
The analyzer reports the two concentrations, and the ratio is then calculated by hand: pTau217 divided by beta-amyloid 1-42, rounded to five decimal places. The FDA's decision summary states that the ratio "must be manually calculated because the instrument does not report the final result". The maker adds that both values must come from the same sample on the same analyzer, and that results from other makers' assays must not be mixed in.
Neither plasma assay has an international reference of its own. The summary states that pTau217 is traceable only to an in-house standard, one the maker prepared itself. It also states that no international reference material or method exists for plasma beta-amyloid 1-42, and that the calibrators for that assay are tied to the maker's spinal fluid assay. Troponin I lacks a common reference in the same way, as Chapter 14 showed.
The decision summary does not say why the test divides one protein by the other. The predicate is itself a ratio of two amyloid fragments in spinal fluid, and the FDA's review of it did not explain that choice either. Arithmetic still shows what a ratio gains and what it costs. Any factor that scales both concentrations in one tube by the same proportion cancels when one is divided by the other. But each measurement brings its own scatter, and the ratio carries both.
The summary reports that scatter as coefficients of variation, each a standard deviation expressed as a percentage of the mean. Measured within one laboratory over many runs and days, the total coefficients of variation were 2.2 to 10.3 percent for pTau217, 2.5 to 3.8 percent for beta-amyloid 1-42 and 3.1 to 10.2 percent for the ratio, so at low concentrations the scarcer protein, pTau217, dominates the ratio's scatter. Between reagent lots, the ratio's coefficient of variation was 7.3 to 16.3 percent. The highest figure came from a low panel whose ratio lay close to the lower cut-off described below, so a result near that cut-off can cross it when the reagent lot changes, and the decision summary asks for caution with such results.
Two cut-offs and a middle zone
The test sorts each ratio into three results. A ratio of 0.00738 or more is positive, one of 0.00370 or less is negative, and anything between is indeterminate. A positive result therefore needs at least about 7.4 picograms of pTau217 for every 1,000 picograms of beta-amyloid 1-42, and a negative one no more than about 3.7. The cut-offs were set on 208 stored plasma samples from three research cohorts. The decision summary states that an indeterminate result "is consistent with patients who are uncertain to have amyloid pathology", and that those patients "should be considered for further testing", such as a PET scan or a spinal fluid test.
The clinical study tested 499 people aged 52 to 93 against a reference, the established test whose answer is taken as correct. For 152 of them the reference was a PET scan read by eye, and for the other 347 it was the spinal fluid test. Just over half, 51.1 percent, had amyloid pathology. The decision summary gives the results in a table.
Of 219 positive results, 201 were amyloid-positive by the reference, 91.8 percent by this arithmetic; the FDA's announcement gave 91.7 percent. Of 182 negative results, 177 were amyloid-negative, 97.3 percent. The 98 indeterminate results split exactly in half, 49 and 49, so for those patients the chance of amyloid pathology after the test, one in two, was close to the 51.1 percent of the whole study, and the blood test added almost no information. Leaving them aside, the test found 201 of 206 amyloid-positive patients, a clinical sensitivity of 97.6 percent, and 177 of 195 amyloid-negative ones, a clinical specificity of 90.8 percent. Counting each indeterminate result as a failure in either direction, sensitivity falls to 201 of 255, 78.8 percent, and specificity to 177 of 244, 72.5 percent. Counted simply as negative, the indeterminates would leave specificity at 226 of 244, 92.6 percent. The middle zone holds the results among which the test's errors would otherwise have fallen.
Two cut-offs answer fewer patients, and answer them more reliably. A single cut-off would have given every patient a yes or no. If the ratio separates the patients inside the middle zone no better than the zone does as a whole, roughly half of them would have been sorted wrongly. With two cut-offs, a fifth of patients go on to the scan or spinal fluid test they would have needed without the blood test, and the other four fifths get an answer that matches the reference most of the time. The algorithm of Chapter 20 makes the same trade, with an observation zone in place of an indeterminate one. The predictive values also depend on how common amyloid pathology is among those tested, which in this specialist population was about half.
The same 499 results give a clinical sensitivity of 97.6 percent or 78.8 percent, depending on whether indeterminate results are set aside or counted as negative. In its 2025 clinical practice guideline, the Alzheimer's Association asks blood tests used for triage to reach at least 90 percent sensitivity and 75 percent specificity, and tests used to confirm amyloid pathology to reach at least 90 percent of both. A triage test decides who goes on to a confirmatory test. Whether a two-cut-off test meets those numbers depends on how its indeterminate results are counted. With them set aside, the Fujirebio ratio's 97.6 and 90.8 percent meet the thresholds for confirmation; with each counted as a failure, its 78.8 and 72.5 percent meet neither set. A performance claim should state the share of indeterminate results and how they were counted.
What the clearance established, and what came after
The clearance established that the ratio, run as specified on the named analyzer, agrees with amyloid PET or the spinal fluid test closely enough for the FDA to find the test substantially equivalent to its spinal fluid predicate. That finding covers symptomatic adults in specialist care. The clearance did not establish use in people without symptoms, for screening, or in primary care, and it did not make a positive result a diagnosis. The handling of samples matters too. The summary states that specimens "degrade rapidly at room temperature", so a result depends on how the tube was kept before it reached the analyzer, as well as on the analyzer's limit of detection.
Three more blood tests for amyloid pathology were cleared within fifteen months. The FDA cleared Roche's Elecsys pTau181 assay on 8 October 2025. It uses a single cut-off to rule out amyloid pathology in primary care. In 312 people, of whom 13.1 percent were amyloid-positive, a negative result was right 97.9 percent of the time and a positive one only 22.4 percent of the time.
On 19 August 2026 the FDA cleared two more. C2N's PrecivityAD2 measures by mass spectrometry (Chapter 18). Roche's Elecsys pTau217 assay uses a single protein and two cut-offs. According to the decision summary for that assay, 19.9 percent of results in Roche's study of 958 people fell between its two cut-offs. The share is about the same as for the Fujirebio ratio. In the EU, Fujirebio announced a CE mark under the IVDR for its single-protein pTau217 plasma assay on 11 May 2026. The IVDR is the EU's regulation for in vitro diagnostics, explained in Chapter 26. No approval of such a test by India's regulator was found as of October 2026.
For a test with two cut-offs, state three things in the specification and the instructions: the share of results expected between the cut-offs in the intended population, the confirmatory test those patients go to, and the performance both with indeterminate results set aside and with them counted. Then confirm the share in the clinical study, since a larger middle zone than planned moves cost and waiting time onto the confirmatory test.
- The first cleared blood test for amyloid pathology divides a pTau217 result by a beta-amyloid 1-42 result, both from one analyzer and one tube.
- The FDA's review does not explain why the test uses a ratio, which by arithmetic cancels factors shared by both measurements but carries the scatter of both assays.
- Two cut-offs leave about a fifth of patients indeterminate, where the test carries no information, and make the other answers more reliable.
- The clearance covers symptomatic adults in specialist care, later tests use other designs, and the test's clinical sensitivity depends on how the middle zone is counted.