Light switched on at an electrode.
+ The questionWhat does starting the light with a voltage gain over starting it with a reagent?
A label the electrode re-arms
On 12 June 1996 the FDA cleared an assay for the pregnancy hormone hCG that Boehringer Mannheim had submitted. It was one of the first tests for an analyzer the company called Elecsys. Its label was a metal complex that gives off light only while an electric voltage drives a reaction at an electrode, a method called electrochemiluminescence. The technology came from IGEN, a company in Maryland, which had licensed it in 1992 to the business that Roche later acquired. Twenty years on, the same method reads the label of Roche's high-sensitivity troponin T assay, the assay of this guide's opening story.
The label is a ruthenium atom held by three ring-shaped organic molecules. The complex weighs about a thousand grams per mole, small enough that several can be attached to one antibody. In the measuring solution the label sits with a large excess of a second chemical, tripropylamine, called the coreactant. When the electrode is raised to a positive voltage, it takes electrons from the tripropylamine, which breaks into short-lived, highly reactive fragments. A 2020 study describes the sequence that follows: one kind of fragment hands the ruthenium complex an extra electron, and another then takes an electron away, which leaves the complex in an excited state. The complex relaxes by emitting a photon of orange-red light near 620 nanometers and returns to exactly the form it started in.
The label's return to its starting form is the point of the method. An acridinium ester reacts once and is spent. A ruthenium label can be excited again after it emits, as long as the electrode keeps the voltage on and fresh coreactant reaches it. University chemists carried out the 2020 study of the mechanism with researchers from Roche Diagnostics and Hitachi High-Tech. The study set out the reaction steps and applied 1.4 volts to drive them. Roche's own patent on the measuring method describes the label reaction as preferably circular, so that one label emits "a plurality of photons".
The measuring cell
The light is made in a small flow cell, a chamber that liquids are pumped through, not in the reaction vessel. A patent by Roche Diagnostics and Hitachi High-Technologies describes the arrangement. The reaction mixture holds streptavidin-coated magnetic beads, typically 2 to 3 micrometers across, and is drawn into the cell over a working electrode, the electrode whose voltage drives the reaction. A permanent magnet under the electrode pulls the beads, with the sandwiches they carry, onto the electrode surface. A buffer containing the coreactant then flows through, washing away unbound label. The voltage is applied, and a photomultiplier tube records the light through the electrode assembly for about two seconds. A photomultiplier is a light detector sensitive enough to register single photons (Chapter 13). The magnet is then moved away so that the beads can be flushed out and the cell cleaned for the next test.
Roche's troponin T assay follows this sequence. The FDA's 2017 decision summary describes the steps: a biotin-labeled antibody and a ruthenium-labeled antibody form a sandwich around troponin T, streptavidin-coated microparticles collect the sandwich, the microparticles are captured magnetically on the electrode, and a voltage stimulates the emission that a photomultiplier measures. The cobas e 801, an analyzer that now runs the assay, holds its measuring cells at 28 °C, separate from the 37 °C of its incubator, according to its maker's specifications.
Why the thin layer matters
The reactive fragments that excite the ruthenium last only a fraction of a millisecond. The 2020 study estimated their half-life at about 200 microseconds. It concluded that they are not expected to travel more than about 3 micrometers from the electrode before they decay. Light is made only in that thin layer.
A bead 2.8 micrometers across that lies on the electrode has every part of its surface within 2.8 micrometers of the electrode, inside the 3-micrometer layer where light is made. Every label on the bead can therefore emit, though the labels nearest the electrode give most of the light. A label floating 30 micrometers away in the liquid is ten times farther than the layer reaches and stays dark. And the two-second reading lasts about 10,000 times the fragments' 200-microsecond half-life, so the electrode keeps making fresh fragments throughout the read.
Making the light at the electrode has three consequences. The method adds a second separation on top of the wash: label that is not on a bead held against the electrode contributes little light, whatever the wash left behind. There is no exciting light, so, as in chemiluminescence, there is no scatter and no autofluorescence, the sample's own glow, to subtract. And because each label can emit many photons during the read, the signal per label is larger than a single flash would give. A 1991 paper by IGEN's scientists reported that electrochemiluminescent labels could be measured over six factors of ten, with a detection limit of 200 femtomolar for the label. A femtomolar concentration is a millionth of a billionth of a mole per liter. Troponin T at 5 ng/L is about 140 femtomolar, so the label alone could be measured at about the concentration of troponin in Chapter 1.
The same 2020 study measured how much the light depends on geometry. Beads of 0.3 micrometer gave about eight times more light than beads of 2.8 micrometers, consistent with more of their labels sitting close to the electrode. A new coreactant raised the signal of commercial cobas e 801 assays by up to 47 percent. Both findings come from a research study, not from a product's specifications. Both show that the few micrometers next to the electrode are a design variable.
In electrochemiluminescence, the light per label depends on the electrode surface, the coreactant reaching it and the voltage applied. A fouled electrode, a stale buffer or a drifting temperature in the cell changes the signal without any change in the sample. That is why the measuring cell is cleaned after every read, why its buffers are counted as consumables, and why its temperature is controlled separately from the incubator's.
Roche's troponin T assay, measured
The 2017 assay has a 9-minute protocol, a limit of blank of 2.5 ng/L on one analyzer and 3 ng/L on the other, and a measuring range that reaches 10,000 ng/L without dilution, more than three factors of ten above its limit of quantitation of 6 ng/L. Its limit of detection had fallen to 3 or 5 ng/L, depending on the analyzer, from the 10 ng/L of Roche's earlier, fourth-generation troponin T assay, but that fall came from changes to the assay itself, within the same detection method. Other companies use electrochemiluminescence differently. Meso Scale Discovery builds carbon electrodes into the bottoms of plastic plates, so that a research laboratory can read several spots, each with its own antibody, in one well. Only labels near the electrode surface are detected, according to the company.
An analyzer built on electrochemiluminescence carries a requirement set for the cell alone: electrode material and cleaning cycle, magnet travel and timing, voltage waveform, cell temperature, buffer consumption, and the photomultiplier's sensitivity to red light, since the light is near 620 nanometers. Each item needs its own verification before the assay is run on it.
- Electrochemiluminescence uses a ruthenium label that emits near 620 nanometers when it is excited by fragments of a coreactant, which a voltage at an electrode creates.
- The label returns to its starting form after emitting, so it can give many photons while the voltage is on.
- Beads are held on the electrode by a magnet, and light comes only from a layer about 3 micrometers thick, which adds a second separation to the wash.
- Roche's troponin T assay forms its sandwich on streptavidin beads and reads it in this cell in 9 minutes.