A protein that recognizes a shape.
+ The questionWhat is an antibody, and how can one of them pick a single protein out of thousands?
A Y-shaped protein with two tips
Nearly every immunoassay in this guide uses one kind of antibody, immunoglobulin G. An immunoglobulin G molecule weighs about 146,000 grams per mole and is built from four protein chains: two identical heavy chains of about 50,000 grams per mole each and two identical light chains of about 25,000. The chains fold into a Y. Each of the two arms is formed by one light chain and the upper half of one heavy chain, and the arms are the fragments that bind antigen, the substance an antibody recognizes. The stem, joined to the arms by a flexible hinge, is the part the immune system uses to signal what the antibody has caught. Measured by how it moves through water, the whole molecule behaves like a sphere about 13 nanometers across.
The binding happens at the tip of each arm. There, six loops of the protein chain, three from the heavy chain and three from the light, come together to form a small surface whose shape and chemistry vary from one antibody to the next. The rest of the molecule is nearly the same in every antibody of its class. An immune system can make enormous numbers of different tips because the genes for those loops are assembled from pieces and then mutated as the immune response matures, so that cells making tighter binders are favored.
For an immunoassay the important consequence is that one antibody carries two identical binding sites, and that the binding site is small compared with the antibody itself. The surface of the target that the antibody touches, its epitope, is only a patch of the target, so a protein the size of troponin offers room for several different antibodies to bind at once, each at its own epitope. Chapter 4 shows why that room makes possible the format that can detect the smallest amounts.
Recognition is shape and chemistry over a small patch
A 2008 survey of 53 crystal structures of antibodies bound to protein targets found that a typical epitope contains about 20 amino acids of the target and covers about 790 square ångströms. An ångström is a tenth of a nanometer, so that is an area about 3 nanometers on a side. Three in four epitopes fell between 15 and 25 amino acids and between 600 and 1,000 square ångströms. Most epitopes on proteins are not a continuous stretch of the chain but amino acids, or residues, from different parts of it, brought together by the way the protein folds. And only a fifth to a third of the residues in the antibody's loops actually touch the target.
Binding over such a patch is the sum of many weak contacts: hydrogen bonds, attractions between opposite charges, and the close packing of surfaces that fit. Each contact alone would break apart in an instant at body temperature. Together, when the shapes match, they hold the two molecules for minutes to hours, and a mismatch of a few atoms removes enough contacts to weaken the hold by a large factor. That sharp loss is the basis of an antibody's analytical specificity, meaning which shapes it binds and which it does not. Recognition is not a lock that admits one key but a fit whose strength falls off quickly as the shape departs from the one the antibody was selected against.
The strength of the fit is measured by the dissociation constant: the concentration of target at which half of the antibody's sites are occupied at equilibrium, when binding and release balance. Smaller is tighter. In 1980 the company Hybritech filed the founding patent for two-antibody assays built from single-clone antibodies, which Chapter 4 describes. The patent required each antibody to bind with an affinity, or binding strength, equivalent to a dissociation constant of 10 nanomolar or less. A 2010 analysis of single-molecule assays assumed antibodies a thousand times tighter, at 10 picomolar. Chapter 3 shows what such numbers mean for how much target an assay can catch.
One clone, one epitope
An animal immunized with a protein makes many different antibodies against it, each from a different line of antibody-producing cells and each binding its own epitope. Serum is the liquid left when blood clots, and serum collected from the animal holds this mixture, a polyclonal antibody. The mixture binds strongly because its antibodies hold the target at many sites, but no two animals make the same mixture, and a supply ends when the animal's serum runs out.
Working at the Medical Research Council's Laboratory of Molecular Biology in Cambridge, Georges Köhler and César Milstein fused antibody-producing cells from an immunized mouse with tumor cells that grow without limit. Each fused cell line, a hybridoma, produced one antibody indefinitely. Nature received their paper on 14 May 1975 and published it on 7 August, and in 1984 Köhler and Milstein received shares of the Nobel Prize in Physiology or Medicine for the work. A monoclonal antibody binds one epitope, and the same cell line can supply the same antibody for decades. That consistency, more than any gain in strength, is what turned immunoassays into products whose results could be compared from one reagent lot to the next.
Choosing what the antibody sees
Troponin shows how much rests on the choice of epitope. The troponin complex that controls muscle contraction has three proteins, T, I and C, and both heart and skeletal muscle make forms of T and I. The heart's forms differ from the skeletal ones, notably at one end of the chain; human cardiac troponin T has 298 amino acids and cardiac troponin I has 210. An assay for heart damage must bind the cardiac forms and ignore the skeletal ones, which injured muscle elsewhere in the body releases.
The first generation of the troponin T assay did not fully manage that: one of its antibodies cross-reacted with skeletal troponin T, binding it in addition to the cardiac form. The second generation, described in 1997, replaced that antibody with a different one specific for the cardiac form. Three of the 1997 paper's nine authors worked at the Boehringer Research Center in Tutzing. The 1997 paper reported no cross-reactivity with purified skeletal troponin T at 1,000 micrograms per liter, two hundred thousand times the 5 ng/L of Chapter 1, and no false results in 43 marathon runners.
The cardiac forms also do not circulate as one molecule. A 2021 review whose authors work at an antibody supplier describes how the circulating forms change over time. In the first hours after a heart attack, much of the released troponin travels as the three-protein complex; later, more of it circulates as smaller complexes and fragments cut by enzymes. Troponin I is mostly bound to troponin C in the blood, and an international task force on cardiac biomarkers has advised that troponin assays should recognize every circulating form equally. In practice each pair of antibodies recognizes a subset, defined by where its two epitopes sit on the chain and whether those regions survive cutting.
Assays that both report "troponin I" can bind different epitopes, and a fragment that keeps one assay's epitopes may have lost the other's. The two assays then measure overlapping but different populations of molecules, and each converts its signal into a concentration against a different calibration material. Chapter 14 shows how far apart their numbers can be.
Roche's troponin T assay uses two monoclonal antibodies against cardiac troponin T. One is made in mouse and the other is a mouse and human hybrid, and the FDA's decision summary lists each at 2.5 milligrams per liter in its reagent. Those amounts make the antibody far more plentiful than the target.
The decision summary gives no molecular weight, so assume the 146,000 grams per mole given at the start of this chapter for immunoglobulin G, or IgG. Then 2.5 milligrams per liter is 0.0025 ÷ 146,000 moles per liter, about 17 billionths of a mole per liter, or 17 nanomolar. Troponin T at 5 ng/L is 0.14 picomolar (Chapter 1). Dividing 17 nanomolar by 0.14 picomolar gives about 120,000. Each troponin molecule is surrounded by about 120,000 antibody molecules of each kind, so the target never runs short of binders; what limits the catch is how fast the two meet, the subject of Chapter 3.
A reagent specification that names "anti-troponin I" is incomplete. It should name the epitopes, the circulating forms each antibody pair recognizes, the cross-reactivity panel covering skeletal forms, fragments and complexes, and the antibody's source and clone, since a change of clone is a change of the test. Chapter 24 shows how regulators treat such a change.
- An IgG antibody is a 146,000 g/mol Y-shaped protein with two identical binding sites, one at each arm's tip.
- Binding is the sum of many weak contacts over an epitope of about 20 amino acids, so a small change in shape weakens it sharply.
- Monoclonal antibodies bind one epitope and come from one cell line, which made reagent lots consistent.
- For troponin the choice of epitopes decides which circulating forms are seen and whether skeletal muscle troponin is ignored.