§ Glossary
Glossary.
Terms are defined as they are used in this guide.
- 510(k)
- A US premarket submission showing that a device is substantially equivalent to a similar device already on the market, its predicate; the guide's troponin T assay and the first Alzheimer's blood test were cleared this way.
- 99th-percentile upper reference limit
- The troponin value below which 99 percent of a healthy reference population falls; above it, troponin is a sign of injury to heart muscle. For the guide's troponin T assay in the US it is 19 ng/L overall, 14 ng/L for women and 22 ng/L for men.
- Absorbance
- A logarithmic measure of the light a colored solution removes from a beam, read by a photometer or plate reader: 1 means a tenth of the light got through, 2 a hundredth and 3 a thousandth. Readers stay linear only to about 2 or 3.
- Acridinium ester
- A chemiluminescent label that reacts once with hydrogen peroxide in alkali and gives its light in a flash lasting a few seconds; Abbott and Siemens Healthineers analyzers use it.
- AEB
- Average enzymes per bead: the unit in which a digital ELISA reports its result, worked out from the share of beads that light.
- Amyloid pathology
- Deposits in the brain of the protein fragment beta-amyloid, assessed in clinical practice by a PET scan or a spinal fluid test. The cleared Alzheimer's blood tests aid in assessing it; they do not diagnose the disease.
- Analyte
- The substance an assay measures; this guide mostly calls it the target.
- Antibody
- A protein the immune system makes to bind one molecular shape. The kind used in immunoassays, immunoglobulin G, is a Y-shaped protein of about 146,000 grams per mole with an identical binding site at the tip of each arm.
- Assay
- A procedure that measures how much of a particular substance a sample contains.
- Assay migration
- Moving a cleared or approved assay to a new combination of instrument and software, supported by focused analytical studies instead of new clinical trials; the intended use must stay the same, and no claim can be added.
- Association and dissociation rate constants
- The rate at which two molecules bind, typically about a million per molar per second for proteins, and the rate at which a bound pair comes apart, between about 1 in 1,000 and 1 in 100,000 per second for good antibodies. Their ratio is the dissociation constant.
- Beta-amyloid
- A protein fragment that forms deposits in the brain; its 42-amino-acid form, beta-amyloid 1-42, is measured in plasma beside pTau217 in the first cleared Alzheimer's blood test.
- Biotin
- A small B vitamin attached to an antibody so that streptavidin will grab it. Free biotin from high-dose supplements occupies the streptavidin, so assays built on it read falsely low in a sandwich and falsely high in a competitive format.
- Blocker
- An inert protein, such as bovine serum albumin or milk protein, that coats any surface the capture antibody left bare; added animal serum or mouse antibody likewise soaks up a patient's interfering antibodies.
- Calibration curve
- The curve, drawn through calibrators, that converts an analyzer's signal into a concentration. For an immunoassay it is S-shaped, flat near zero and near the top, and in a competitive assay it slopes downward.
- Calibrator
- A sample with an assigned concentration, measured to draw or adjust the calibration curve; calibration is repeated with each new reagent lot, at set intervals and whenever quality control fails.
- Capillary flow
- The pull that draws liquid through a porous material, as water climbs a paper towel. On a lateral flow strip it moves the sample and sets the timing, slowing as the wetted length grows.
- Carry-over
- The transfer of material from one sample, or one reagent, into the next, reported as a fraction of the high sample, such as parts per million; routine quality control does not reveal it.
- CDSCO
- The Central Drugs Standard Control Organisation, India's regulator; its Central Licensing Authority licenses all imported IVDs, the manufacture of class C and D devices and the approval of new IVDs.
- CE mark
- The marking under which a device is sold in the EU; under the IVDR most IVDs need a notified body's assessment first, while the maker of a class A instrument declares conformity itself.
- Chemiluminescence
- Light from a chemical reaction whose product forms in an excited state. Measured against darkness, with no lamp, it reaches lower concentrations and a wider range than color.
- CLIA categorization
- The FDA's assignment of a test to waived, moderate or high complexity under the Clinical Laboratory Improvement Amendments, which decides which US laboratories may run it. Seven criteria are each scored 1 to 3, and a sum above 12 means high complexity.
- Clinical performance study
- A study that judges a test's results against a reference diagnosis of each patient, as the Basel troponin study did; in the EU it follows ISO 20916, and India requires a local clinical performance evaluation for new class B to D IVDs.
- Closed and open system
- An assay that runs only on its maker's analyzers is a closed system; one whose maker validates it on other instruments is open, at least to that maker. The FDA's 2022 instrument family policy applies only to open systems.
- CLSI
- The Clinical and Laboratory Standards Institute, a US standards body for laboratory medicine; its EP documents, such as EP05 for precision and EP17 for detection limits, set study designs that the FDA recognizes.
- Coefficient of variation
- The standard deviation of repeated results divided by their mean, as a percentage; a high-sensitivity troponin assay must reach 10 percent or less at the 99th percentile.
- Colloidal gold
- Gold particles a few tens of nanometers across, the most common label on lateral flow strips; particles of 20 to 40 nanometers absorb most strongly in the green, at 524 to 530 nanometers, so a line of them looks red.
- Competitive assay
- An immunoassay in which a labeled copy of the target competes with the target in the sample for a scarce antibody, so the signal falls as the concentration rises. Small molecules, too small to hold two antibodies, are measured this way, as was insulin in the first radioimmunoassay.
- Coreactant
- The chemical, tripropylamine in the guide's troponin assay, that sits in large excess beside an electrochemiluminescent label; oxidized at the electrode, it breaks into short-lived fragments that excite the label.
- Cross-reactivity
- Binding of an antibody to a molecule other than its intended target, such as skeletal troponin T in a cardiac assay or a drug's breakdown products in a competitive assay.
- Cut-off
- A value that divides results into answers, such as positive and negative, or rule-out, observation and rule-in; troponin cut-offs belong to one assay and cannot be carried to another.
- Dark count
- Pulses a detector gives with no light falling on it, from electrons released by heat in the cathode; about 50 a second in one photon-counting module, it is the detector's own floor.
- De Novo
- The US route for a new low- or moderate-risk device with no predicate; the first spinal fluid test for amyloid was granted this way in May 2022.
- Decision summary
- The FDA's published summary of the review behind a clearance, the source of most of this guide's figures for cleared assays.
- Digital ELISA
- An immunoassay that seals beads into wells of tens of femtoliters, so that a single enzyme lights its well, and counts the lit wells instead of measuring a glow. Its analyzers are labeled for research use only.
- Dioxetane
- A light-producing substrate for alkaline phosphatase: the enzyme removes a phosphate group, and the unstable remainder falls apart and emits light, glowing for minutes.
- Dissociation constant
- The concentration of target at which half of an antibody's binding sites are occupied at equilibrium; smaller is tighter. With antibody at twenty times this constant, more than 90 percent of a scarce target is bound.
- Electrochemiluminescence
- Light from a label, a ruthenium complex in the guide's troponin assay, driven by a voltage at an electrode together with a coreactant. Each label returns to its starting form and can emit many photons, and only labels within about 3 micrometers of the electrode light.
- ELISA
- Enzyme-linked immunosorbent assay: an immunoassay read by the color an enzyme label makes from its substrate, first published in 1971 and now most often a sandwich in the wells of a plastic plate.
- Energy transfer (TRACE)
- A method in which a europium label on one antibody passes its energy to an acceptor on a second antibody only when both bind the same target molecule, so bound and free label are told apart without washing; Thermo Fisher Scientific's KRYPTOR analyzers use it under the name TRACE.
- Enzyme label
- An enzyme, usually horseradish peroxidase or alkaline phosphatase, attached to an antibody so that each bound antibody turns a substrate into many molecules of colored, fluorescent or light-giving product.
- Epitope
- The patch of a target's surface that an antibody touches, typically about 20 amino acids; a protein the size of troponin has room for several antibodies, each at its own epitope.
- HAMA
- Human anti-mouse antibodies: antibodies made against mouse proteins, for example after treatment with mouse antibodies, which bind them strongly and specifically; they matter because many reagent antibodies are made in mice. In practice the name is often used interchangeably with heterophile antibody.
- Heterophile antibody
- A patient's antibody, of no clearly defined origin, that binds the antibodies of two or more other species. Bridging the capture and labeled antibodies, it gives a falsely high result; binding only one of them, it can give a falsely low one.
- High-dose hook
- The fall in signal at very high target concentrations in a one-step sandwich, when free target fills the capture and labeled antibodies separately, so an enormous concentration can read as a modest one. Makers state a hook-free limit, 100,000 ng/L for the guide's troponin T assay.
- High-sensitivity troponin assay
- A troponin assay with a coefficient of variation of 10 percent or less at the 99th percentile and results above its limit of detection in at least half of healthy people, as defined by a task force of the International Federation of Clinical Chemistry in 2017.
- IEC 61010
- The safety standard for laboratory equipment, which analyzers follow instead of the medical electrical standard IEC 60601-1: Part 1 covers laboratory equipment in general, and Part 2-101, of 2018, in vitro diagnostic equipment.
- IEC 61326-2-6
- The electromagnetic compatibility standard for IVD equipment, covering how an instrument disturbs other equipment or is disturbed by it; its fourth edition, of June 2025, was recognized in full by the FDA on 28 July 2025.
- IEC 62304
- The standard for developing medical device software, such as the software that schedules an analyzer's steps and computes its results; in October 2026 its 2006 edition, amended in 2015, was still current.
- Immunoassay
- An assay in which the part that recognizes the substance being measured is an antibody.
- Indeterminate result
- A result between two cut-offs, which gives neither a positive nor a negative answer and sends the patient to a confirmatory test; about a fifth of results in the first cleared Alzheimer's blood test.
- Interference
- A change in a result caused by something in the sample other than the target, such as biotin, heterophile antibodies, rheumatoid factor or molecules related to the target; the substance responsible is an interferent.
- International Federation of Clinical Chemistry (IFCC)
- The international body whose working groups and task forces compare troponin assays, publish a table of their characteristics supplied by the makers, and in 2017 defined a high-sensitivity troponin assay.
- ISO 13485
- The international standard for the quality management systems of medical device makers, in its 2016 edition; the FDA's Quality Management System Regulation (QMSR), in force since 2 February 2026, incorporates it by reference.
- ISO 14971
- The standard for risk management of medical devices, in a 2019 edition; applied to an analyzer, it traces each fault to a wrong result, to a situation in which that result is acted on, and to harm.
- IVD
- In vitro diagnostic device: a reagent, calibrator, analyzer or other product used to test samples, such as blood, outside the body.
- IVDR
- The EU's In Vitro Diagnostic Medical Devices Regulation, applied since 26 May 2022; it sorts IVDs into classes A to D and requires a notified body for around 80 percent of them, against about 8 percent under the 1998 directive it replaced.
- Label
- The part of an immunoassay that makes a bound antibody visible: an enzyme that makes color, a molecule that gives off light, or a particle that shows as a colored line.
- Laboratory-developed test
- A test developed and run within a single laboratory. A 2024 FDA rule that would have made such tests explicitly devices was vacated by a federal court in March 2025.
- Lateral flow test
- An immunoassay run inside a strip of porous material: the sample moves by capillary flow, dissolves a dried labeled antibody and carries it past a printed test line of capture antibody and then a control line.
- Limit of blank
- The highest result expected from samples that contain none of the target, set so that only 5 percent of blank results exceed it; in its simplest form, the mean of the blanks plus 1.645 standard deviations.
- Limit of detection
- The lowest concentration whose results exceed the limit of blank 95 percent of the time; 3 to 5 ng/L, depending on the analyzer, for the guide's troponin T assay in the US.
- Limit of quantitation
- The lowest concentration measured with an agreed precision, often a coefficient of variation of no more than 20 percent; 6 ng/L for the guide's troponin T assay in the US.
- Macrocomplex
- A target bound to the patient's own antibodies, such as macroprolactin, which is prolactin bound mostly to immunoglobulin G, or troponin bound to antibodies; it can read high on one assay and normal on another.
- Magnetic microparticles
- Beads a few micrometers across, typically coated with streptavidin, that collect sandwiches throughout the reaction volume and are held by a magnet while the liquid around them is exchanged.
- Master curve
- A calibration curve the maker measures for each reagent lot, from master calibrators on many instruments, and encodes with the lot; the analyzer adjusts it with a two-level calibrator set.
- Measuring range
- The span of concentrations an assay reports without dilution, across which results are claimed to stay proportional to concentration; 6 to 10,000 ng/L for the guide's troponin T assay in the US.
- Medical Device Coordination Group (MDCG)
- The EU group that issues guidance under the IVDR; its 2022 guidance divides the evidence for an IVD into scientific validity, analytical performance and clinical performance.
- Medical Devices Rules, 2017
- The rules under which India regulates IVDs, in effect since the start of 2018, with four classes, A to D, and licensing divided between the CDSCO and state authorities.
- Mole
- An amount of substance holding about 6.02 × 10²³ molecules; an attomole, a billionth of a billionth of a mole, is about 602,000 molecules. One mole per liter is one molar, so picomolar means trillionths of a mole per liter.
- Monoclonal antibody
- An antibody made by one line of cells, a hybridoma, that binds one epitope. Unlike a polyclonal antibody, the varying mixture in an immunized animal's serum, it can be supplied unchanged for decades, which made reagent lots consistent.
- Multiplex assay
- An assay that measures several targets in one sample. Each added target adds unintended antibody pairs that can cross-react, so a floor measured for one target alone does not carry over to a panel.
- Notified body
- An independent organization, designated by an EU member state, that assesses a maker's quality system and technical documentation before an IVD from class B upward can be sold; 19 were designated under the IVDR at the end of February 2026.
- One-hour algorithm
- The European Society of Cardiology's rule for ruling a heart attack in or out from two troponin samples an hour apart, using the first value and the change between them, with cut-offs specific to each assay; patients in neither group are observed.
- PET
- Positron emission tomography: a brain scan that, with a spinal fluid test, is how amyloid pathology is assessed in clinical practice.
- Photomultiplier tube
- A vacuum-tube detector in which a photon striking the photocathode releases an electron, and a chain of dynodes multiplies it into a pulse of a million to ten million electrons.
- Poisson distribution
- The pattern of counts of random events, which fixes, for example, the share of beads carrying no enzyme from the average number per bead. A count that follows it scatters by its own square root: 100 counts by 10 percent, 10,000 by 1 percent.
- Positive and negative predictive value
- The share of positive results that are right and the share of negative results that are right; both depend on prevalence among the people tested.
- Premarket approval
- The US route to market for a class III device, the highest of the FDA's three classes.
- Prevalence
- The share of the people tested who have the condition; 17 percent had a heart attack in the 2015 study of the one-hour algorithm.
- Proximity ligation and extension assays
- Methods in which two binders carrying DNA strands must attach to the same molecule before their strands yield a piece of DNA that is copied and measured; a binder on its own leaves nothing to copy, so background falls and many assays can share one sample.
- pTau217
- Tau protein carrying a phosphate group at amino acid 217, in forms that circulate as fragments, at fractions of a picogram per milliliter of plasma; measured to aid in assessing amyloid pathology.
- Quality control
- Samples of known concentration run at intervals and plotted on charts to catch drift; a run is rejected under rules such as one control more than three standard deviations from its target, or two in a row more than two on the same side.
- Quantum efficiency
- The fraction of photons striking a detector's cathode that release an electron; up to 43 percent at 350 nanometers for the best blue-sensitive cathodes, and much lower in the red.
- Reader
- An instrument that reads an assay's signal. A plate reader measures absorbance in many wells at once; a strip reader measures each line against the strip's own background with its own light and applies the same threshold every time.
- Recall
- A maker's correction or removal of a product already on the market, recorded by the FDA; the remedies in this guide include new instructions for laboratories and lot-specific correction factors.
- Reference material
- A material whose value is known well enough to anchor a calibration chain; it serves only if it behaves like patient samples in the assays that use it.
- Rheumatoid factor
- An antibody that binds the stem of other antibodies; it can bridge reagent antibodies like a heterophile antibody and cause false positives in troponin assays.
- Sandwich assay
- An immunoassay in which a capture antibody and a labeled antibody bind two different epitopes of the same target, so the signal rises with concentration from near zero; also called a two-site assay.
- Sensitivity
- Clinically, the share of people with the condition whose test is positive, such as 40 percent for the Liverpool strips read on site. Analytically, as in a more sensitive assay, how low a concentration a test can detect.
- Solid phase
- The surface that holds the capture antibody, such as the inside of a tube or well or the surface of beads, so that bound label stays while free label is washed away.
- Specificity
- Clinically, the share of people without the condition whose test is negative, 99.9 percent for the Liverpool strips. For an antibody, its selective binding to one shape, which weakens quickly as a molecule departs from that shape.
- Standard deviation
- The usual measure of the scatter of repeated results around their mean.
- Streptavidin
- A bacterial protein that grips biotin almost irreversibly; coated on beads, it collects the biotin-carrying capture antibody together with the sandwich it holds.
- Time-resolved fluorescence
- Reading a long-lived label such as europium, which glows for tens of microseconds to about a millisecond, after a delay long enough for scattered light and the sample's own fluorescence, which fade in about 10 nanoseconds, to disappear.
- Traceability
- The link, through an unbroken chain of comparisons, from an assay's calibrators to an accepted reference, ideally a reference measurement procedure and a certified reference material; troponin I still lacks such a chain end to end.
- Troponin
- A complex of three proteins, T, I and C, that controls muscle contraction; the cardiac forms of troponin T and I leak into the blood when heart cells die and are measured to rule a heart attack in or out.
- Wash
- Rinsing away everything that did not bind while the surface keeps what did, so that only label held by the target is left to measure; it works by repeated dilution, not as a filter.
- WHO prequalification
- The World Health Organization's assessment of an IVD as a benchmark for UN and other international buyers, limited to priority diseases such as HIV, hepatitis B and C, malaria, syphilis and tuberculosis; WHO is not a regulator, so registration stays national.
- Within-laboratory precision
- The total scatter of results a laboratory sees over weeks, combining repeatability within a run with scatter between runs and between days; CLSI's EP05 estimates it from 80 results per sample over 20 days.