When a line becomes a number.
+ The questionWhat does a reader add to a test a person can read by eye, and what changes when the person does the reading?
The same strips, read twice
From November 2020 the city of Liverpool, home to about 498,000 people, offered its residents rapid antigen tests for the virus that causes COVID-19, tests that detect a protein of the virus. The city used the Innova lateral flow test, in a program run as a pilot. As part of the pilot's evaluation, people without symptoms at 48 testing sites took two swabs minutes apart under supervision. One swab went to the strip and one to a laboratory PCR test, which detects the virus's genetic material after copying it many times.
On 23 December 2020 the University of Liverpool published an interim evaluation that reported 5,869 paired results. After void results were set aside, the strips read on site had found 28 of the 70 people whose PCR test was positive. That is a sensitivity of 40 percent, where sensitivity in this clinical sense is the share of infected people a test finds. Among the people with higher viral loads, meaning more virus in their samples, the strips found about two thirds.
The evaluators then had reviewers who did not know the PCR results re-read photographs of the same strips. On re-reading, the sensitivity rose to 53.4 percent overall, and among the higher viral loads from 66.7 to 83.3 percent. The specificity, the share of people without the infection whose strip read negative, was unchanged at 99.9 percent. Most of the discordant readings had come in the first week of the pilot. The report linked that pattern to the testing staff's inexperience, and it cautioned that the numbers were small. The strips had not changed. What changed was who looked at them, and how.
A separate national evaluation was carried out at the government laboratory at Porton Down and the University of Oxford, and it found that the strip's result also depended on who performed the test. A briefing for journalists on 11 November 2020 quoted its preliminary report. The report gave the strip's sensitivity as about 79 percent when laboratory scientists performed the test, 73 percent with trained health workers and 58 percent with members of the public who had trained themselves from the instructions and swabbed and tested themselves. The groups were not formed at random, so the figures are not a like-for-like comparison.
Leaving aside void results, the paired study had 31 positive strips, and PCR confirmed 28 of them, so a positive strip was right about 90 percent of the time. Of 5,434 people whose PCR test was negative, 5,431 had a negative strip: a specificity of 99.9 percent. But 42 of the 70 PCR-positive people had a negative strip. In this population, where 70 of the 5,504 people with valid results, about 1.3 percent, were infected, a negative strip was right 5,431 times in 5,473, or 99.2 percent, because almost everyone was negative, yet the strip missed three in five of the infections it was meant to catch. The share of positive results that are right is called the positive predictive value, and the share of negative results that are right the negative predictive value; both depend on prevalence, the share of the people tested who have the condition. The PCR guide in this library works through the same arithmetic for test performance and prevalence.
What an instrument reader measures
An instrument reader replaces the eye with a light source and a detector. The FDA cleared the BD Veritor system for influenza on 23 March 2012. Its battery-powered bench reader measures the light reflected from each line of the strip. The decision summary records that the reader subtracts the signal at a negative-control line from the signal at each test line and scores the difference against a cut-off. The negative-control line is not the control line of Chapter 15, which catches labeled antibody to show that the strip has run. Its signal measures the strip's nonspecific color, the color that appears whether or not the target is present. The decision summary also states that the strip "cannot be interpreted visually" and that the subtraction is one that "the human eye is unable to accurately perform".
The subtraction is the reader's real contribution. The eye judges a faint line against the color of the membrane around it, under whatever light is in the room, and different people set their own threshold for "a line". A reader measures both the line and the background in fixed geometry, with its own light, and applies the same threshold every time. It can also save the reading, flag a strip that did not run, and send the result to a record.
Readers also allow labels that the eye cannot see. The FDA cleared Quidel's Sofia analyzer on 24 October 2011. According to its decision summary, the analyzer scans the whole strip and reads beads dyed with a europium chelate, excited at 365 nanometers and glowing at 618. A 2012 study compared europium nanoparticles, read by time-resolved fluorescence, with colloidal gold in two model lateral flow assays. The fluorescent label detected concentrations about 7 times lower than gold could in one assay and about 300 times lower in the other. In a 2017 study, academic developers of a thermal reader reported that it detected concentrations about ten times lower than reading by eye could. Their reader detects the heat that gold particles release when lit by a laser.
Phones as readers
A phone carries a camera, a light and a computer, and has become a reader in its own right. A 2023 study tested one smartphone reader, built around a 3D-printed holder and an app that works in real time. At a low antigen concentration, volunteers judged only 20 percent of test strips positive, while the app classified all of them correctly. The study used a single phone model. In August 2021 BD and Scanwell Health received emergency use authorization in the US for a home COVID-19 test read by a phone's camera, which BD described as the first of its kind. Through an emergency use authorization, the FDA allows a test to be used during a declared public health emergency without clearance or approval.
India used the phone differently. On 19 May 2021 the Indian Council of Medical Research issued an advisory on home testing with rapid antigen tests, naming one approved kit. It required users to photograph the strip with the phone they had used to register in a mandatory app, and the app recorded the result and sent it to the national testing portal. The same council had validated 95 rapid antigen kits by April 2021. Its threshold for use at the point of care was a sensitivity of at least 50 percent and a specificity of at least 95 percent.
Troponin at the bedside
The readers that now bring high-sensitivity troponin to the bedside are not strips read by eye but cartridges read by instruments. The FDA cleared the PATHFAST high-sensitivity troponin I test on 20 March 2024, and its distributor called it the first such test cleared for use at the point of care in the US. Abbott's handheld i-STAT high-sensitivity troponin I cartridge followed on 3 January 2025. According to its FDA decision summary, it is an assay with electrochemical detection that uses about 22 microliters of whole blood, gives a result in about 15 minutes and has 99th-percentile limits of 13 ng/L for women and 28 for men. Siemens Healthineers states that its Atellica VTLi gives a result from a fingerstick in eight minutes. The company announced the analyzer's CE mark in April 2021, and the analyzer was not found among US clearances as of October 2026.
A reader removes the variation of human judgment, but it cannot recover target that the strip failed to capture, and it can apply its threshold with great consistency to a strip that ran badly. Its cut-off is a design decision that has to be validated with the strips, samples and users it will meet, as Chapter 23 describes.
For a reader-based test, run the clinical study with the reader, software version and intended users the product will have, and include strips that are faint, late or partly run. For a visually read test, test reading by the intended users under realistic light, since the Liverpool figures show that the reader, human or electronic, is part of the test, and the Porton Down figures show that the operator is too.
- Re-reading photographs of the same Liverpool strips raised sensitivity from 40 to 53.4 percent, so the reader changed the result.
- Instrument readers measure line and background in fixed geometry, subtract nonspecific color and apply one threshold every time.
- Fluorescent labels, and thermal reading of gold labels, lowered detection limits severalfold to a few hundredfold in published comparisons.
- High-sensitivity troponin at the bedside comes from instrument-read cartridges, two of them cleared in the US by January 2025.