14 — Calibration

Two numbers for one sample.

+ The questionTwo well-made troponin assays give different numbers for the same blood. Which one is right?

Ten times apart on the same samples

In 2015 a working group of the International Federation of Clinical Chemistry published a pilot study in which sixteen troponin I assays from six manufacturers measured the same 79 patient samples. When one assay's results were plotted against another's, the slope of the best-fit line ran from 0.33 to 2.87, so the lowest-reading and highest-reading assays differed by a factor of almost nine, which the authors describe as about tenfold. None of the assays was broken. Each did what its maker designed it to do, with its own antibodies, its own calibrators and its own definition of a nanogram per liter of troponin I.

That is the puzzle of this chapter. An immunoassay does not count the molecules in a sample. It reads its result off a curve that converts the analyzer's signal into a concentration, and the curve is only as true as the materials used to draw it. Chapter 2 showed that two antibody pairs can see different subsets of the troponin in blood. This chapter shows how each assay is tied to a scale, and why for troponin I there is still no single scale to tie them all to.

A curve made by the maker, adjusted in the laboratory

A calibration is the set of measurements that relates an assay's signal to concentration. Calibrators are samples with assigned concentrations. An immunoassay's curve through them is not a straight line. Plotted against the logarithm of concentration, it is flat at the bottom, rises through the middle and flattens again near the top. A standard S-shaped curve, the four-parameter logistic function, has been found to fit that shape well. A 1997 patent on immunoassay calibration describes the usual practice as five or six standards, each measured in duplicate.

Roche splits the work in two. For each reagent lot, Roche measures six master calibrators on many instruments and in many runs, and encodes the resulting master curve with the lot. In the laboratory, the analyzer measures two calibrators and uses the two results to adjust the master curve to that instrument and that reagent pack. For Roche's troponin T assay, the two calibrators sit at about 18 and about 4,200 ng/L. Abbott's earlier ARCHITECT troponin I assay worked differently. It was calibrated on the analyzer itself with six calibrators, a zero made from human serum and five containing recombinant cardiac troponin complex, as its FDA decision summary records.

Solved example: why the two-point adjustment matters

Suppose, for illustration, that a fresh electrode and a new reagent pack make one analyzer give 10 percent more light per label than the analyzers on which the master curve was made. Near the bottom of the range the signal is close to proportional to concentration, so without adjustment a patient sample whose true concentration is 18 ng/L would read about 10 percent high: about 19.8 ng/L, just above the US 99th-percentile limit of 19 ng/L. The low calibrator, assigned 18 ng/L, shows the same 10 percent excess, and the adjustment scales it away. A shift too small for anyone to notice in a single result moves a patient across the decision line.

29 — A master curve, adjusted on site
10 100 1,000 10,000 SIGNAL, RELATIVE TROPONIN T, ng/L (LOG) this analyzer: 10 % more light (illustration) master curve (per reagent lot, made by the maker) low calibrator about 18 high calibrator about 4,200 two-point adjustment scales the excess away a true 18 ng/L, read without adjustment US 99th percentile 19 ng/L true 18 19.8 unadjusted 17 18 19 20 21 an unnoticed 10 % crosses the decision line holds 12 weeks per lot, 7 days per pack on board (FDA decision summary) schematic curves
AnalyteReadoutFocal detail
Plate 29 — Roche makes a master curve for each reagent lot and each analyzer adjusts it with two calibrators. Without that adjustment, a 10 percent excess of light would read a true 18 ng/L as 19.8, across the US 99th-percentile limit.

Calibrations do not last. Chapter 12 gave how long a calibration of Roche's troponin T assay holds, as the FDA's decision summary states: 7 days while the same pack stays on the analyzer, and 12 weeks on the same reagent lot. Roche's method sheets also call for recalibration with each new lot and whenever quality control fails.

Lots, drift and quality control

Every new lot of reagent is a slightly different product: antibodies from a new production run, calibrators from a new batch, beads from a new coating. The maker assigns each lot its own curve, but the assignment can be wrong. In July 2020 Siemens Healthineers began a recall of three lots of its Dimension Vista high-sensitivity troponin I reagent because their results read lower than they should, a fault called negative bias. The remedy was a correction factor specific to each affected lot, according to the FDA's recall record. Laboratories check each new lot themselves with patient samples, following protocols such as the Clinical and Laboratory Standards Institute's guideline EP26.

Between lots, laboratories watch for drift with quality control samples of known concentration, run at intervals and plotted on charts. In 1981 James Westgard and colleagues published a set of rules for judging these charts, now known as the Westgard rules. A run is rejected, for example, when one control falls more than three standard deviations from its target, or when two in a row fall more than two standard deviations from it on the same side. Each rule trades catching real shifts against rejecting good runs. Westgard's own guidance gives the cost of a stricter rule: treating a single result beyond two standard deviations as a failure would reject about 9 percent of good runs that use two controls.

Who decides what a nanogram is

For a measurement to mean the same in every laboratory, each assay's calibrators must be linked to a reference that everyone accepts through an unbroken chain of comparisons, a property called traceability. The reference is ideally a reference measurement procedure and a certified reference material, both tied in turn to the International System of Units. In laboratory medicine the international standard for that chain is ISO 17511. Its 2020 edition is still current as of October 2026: ISO reviewed it in February 2026 and kept it without revision.

30 — An unbroken chain, or a missing top
THE IDEAL (ISO 17511) TROPONIN T (ONE MAKER) TROPONIN I (MANY MAKERS) SI unit reference measurement procedure certified reference material maker's master calibrator product calibrator patient result none none none Roche's earlier troponin T assay master calibrator product calibrator patient result agreement by design, within one generation no accepted reference procedure no common scale to tie them to SRM 2921a, NIST, 10 Mar 2026: for validating methods maker A calibrator result maker B calibrator result maker C calibrator result 2015 IFCC pilot: 16 assays, slopes 0.33–2.87
ReadoutFocal detail
Plate 30 — A result is comparable between laboratories only if its calibrators trace back to a common reference. Troponin T agrees because one company makes it; troponin I has no complete chain, so its assays read up to about tenfold apart on the same samples.

Troponin I has no such chain from end to end. In 2006 a committee of the American Association for Clinical Chemistry found that the scatter between 15 assays ran from 82 to 97 percent before harmonization. Adjusting them with shared serum pools cut that to 9 to 23 percent. But the two purified troponin materials tested behaved like patient samples in only 45 percent and 39 percent of the assays respectively, too few for either to serve as a common calibrator. The 2015 pilot that opened this chapter recalibrated its assays to common samples, which cut the scatter between them from 40 to 22 percent at low concentrations. Its authors concluded that troponin I results were not yet traceable to a higher-order reference. On 10 March 2026 the US National Institute of Standards and Technology issued a new certified troponin complex for validating methods, SRM 2921a, and its certification report notes that the assays still need to be standardized.

Troponin T is in a different position, because one company makes the assay. Its results agree between laboratories by design, and the FDA's decision summary states that the 2017 assay is traceable to Roche's own earlier troponin T assay. Even that agreement has limits across generations. By Roche's account, its sixth-generation troponin T assay carried the CE mark by September 2025. The CE mark allows a test to be sold in the European Union. Roche's own reference study gives the new assay 99th-percentile limits of 27 ng/L overall, 18 for women and 32 for men, against 19, 14 and 22 ng/L for the fifth-generation assay cleared in the US. Its results therefore cannot be read against the older limits.

Traceability to one material does not make assays agree

An assay can be traceable to a reference material, as the FDA's record says of Abbott's high-sensitivity troponin I assay, and still read differently from another assay traceable to something else, or to the same material if that material does not behave like patient blood in both. Agreement between assays is shown with patient samples, not inferred from a traceability statement.

The practical consequence runs through every later chapter. In the International Federation's 2019 table of high-sensitivity assays, the makers' 99th-percentile limits for troponin I ran from 8.67 to 58.2 ng/L. The European guideline's one-hour algorithm gives a separate set of cut-offs for each assay. A troponin result has meaning only against the limits of the assay that produced it.

31 — Each assay its own limits
IFCC 2019 table: troponin I 99th percentiles 8.67–58.2 ng/L about sevenfold between troponin I assays 5 10 20 50 100 TROPONIN, ng/L (LOG) Roche troponin T Gen 5 (US) 14 · 19 · 22 ◆ 52 Roche troponin T Gen 6 (CE, 2025) 18 · 27 · 32 Abbott ARCHITECT troponin I 17 · 28 · 35 ◆ 64 Abbott i-STAT troponin I 13 · 21 · 28 QuidelOrtho VITROS troponin I 9 · 11 · 12 ◆ 40 99th percentile (women · overall · men) ESC 2020 rule-in cut-off makers' figures and FDA decision summaries; Gen 6 not cleared in the US
AnalyteReadoutFocal detail
Plate 31 — Each troponin assay carries its own 99th-percentile limits and its own algorithm cut-offs; troponin I limits ran from 8.67 to 58.2 ng/L in the 2019 IFCC table. A number means something only against the limits of the assay that produced it.
Name the calibration chain in every specification

For a new assay, state what the calibrators are traceable to, how the master curve is made and transferred, which on-site calibration adjusts it and how often, and which patient-sample studies show agreement with the assays already in use. For a laboratory, the same list decides whether results from two analyzers can be pooled or must be reported against separate limits.

+ What this chapter established
  • An immunoassay's number comes from a curve through calibrators, and is only as true as the calibrators and the curve's adjustment.
  • Roche ships a master curve per lot and adjusts it on site with two calibrators; a 10 percent drift left uncorrected can push a result across a decision limit.
  • Lot changes and drift are caught with patient-sample checks and quality control rules, which balance missed shifts against false rejections.
  • Troponin I has no complete reference chain and its assays see different forms of the protein, so they differ severalfold, and even successive troponin T generations need separate limits.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.