One instrument, believed.
In December 1978, opticians at the Perkin-Elmer Corporation in Wilton, Connecticut, began grinding a disc of glass 2.4 m across into the primary mirror of the Hubble Space Telescope. The mirror was to be a concave hyperboloid, and Perkin-Elmer tested its shape with optical instruments called null correctors. In its plan the company placed "complete reliance" on one of them, the reflective null corrector, which served as the template for the mirror: the opticians polished until the mirror matched what the instrument showed as correct.
Inside the null corrector, a lens called the field lens had to sit at an exact distance from the instrument's mirrors. The operator who set it took that distance from metering rods made of Invar, an alloy that hardly expands when it warms. The operator found the end of a rod by reflecting a focused beam of light off it, through a small hole in a cap that kept the rod centered on the beam. The light that came back was reflected instead by the cap, from a spot where its non-reflecting material was missing. Set by that false reflection, the lens had to go farther than its mount could adjust, so spacers were inserted to move it there. It came to rest about 1.3 mm too far from the instrument's lower mirror.
Perkin-Elmer polished the mirror against the reflective null corrector until April 1981. Two other test instruments disagreed with that null corrector. An inverse null corrector showed a pattern that could not be aligned out, and the pattern was attributed to spacing errors in that instrument's own lenses. A refractive null corrector also showed the error in the mirror, and it was set aside because it was believed to be less precise than the reflective one. Both warnings were discounted at the time as faults of the instruments that gave them. Perkin-Elmer did not verify the reflective null corrector's dimensions after its original assembly, and no one measured its spacings independently. After the final test of the coated mirror in February 1982, the instrument stood unused.
The Space Shuttle Discovery carried the telescope into orbit on 24 April 1990, and the telescope's images were blurred. On 21 June the Hubble project manager announced the failure, and on 2 July NASA formed a board of investigation chaired by Lew Allen, director of the Jet Propulsion Laboratory. The Board found the reflective null corrector unchanged at the top of its test tower and measured it where it stood. Its field lens sat about 1.3 mm out of place, and the Board traced the error to the stray reflection from the cap. The Board's analysis also showed that the refractive null corrector had been easily accurate enough to detect the error. In its report of November 1990, the Board wrote that the measured 1.3 mm "accounts in detail for the amount and character of the observed image blurring".
The opticians had polished the mirror with great care to the shape its one reference gave, and that reference was wrong. Nothing had tied the null corrector's critical length to a second, independent measurement. The length was set once and not verified after assembly. The Board could find no record that the instrument was certified, and the two instruments that disagreed were treated as the faulty ones. Perkin-Elmer had accepted agreement with one instrument in place of a chain of checks back to a known length.