+ Part VI · Proving the measurement

In a product.

Design verification asks whether a product meets its requirements, and every answer it gives is a measurement. A device maker measures its products to verify them, and when the product is itself a measuring instrument, such as the blood-pressure monitor, the maker's test station measures how well the product measures. The station's numbers then need proof of their own before they can decide anything. In Chapter 22 the engineer tests the production pressure test itself and separates the scatter that comes from the monitors from the scatter that the station and its technicians add. In Chapter 23 the engineer sets out the evidence that makes the test method valid, and in Chapter 24 follows the monitor's accuracy claims to its label, including the claim that no calibration can prove, the one for the blood-pressure estimate.

22 — Gauge study

Is the gauge good enough?

+ The questionBefore trusting a measuring system's numbers, how do engineers test the system itself?

Three sources of scatter

Three technicians test the same ten monitors three times each, which gives 90 results, and the engineer who plans the study wants to know how much of their spread comes from the test itself. At the end of the line each monitor enters its service mode, a technician connects it to a rigid vessel and a reference manometer, and the station holds a series of steady pressures (Chapter 14). At each pressure the technician records the monitor's error, its reading minus the reference's, and the decision rule of Chapter 13 turns that error into a pass or a fail. Each recorded error is itself a measurement made by the station, so it carries the station's own scatter as well as the monitor's error, and the engineer must know the station's share before the station sorts any monitor.

Part of the spread of the 90 results is real, because the ten monitors truly differ from one another: each sensor keeps some hysteresis and some residual error after calibration (Chapter 11). That part-to-part variation is what the test exists to see. When one technician tests one monitor three times, the three results still differ a little, through small changes in how the pressure settles, in the reference's last digit and in the monitor's rounding. That spread is the station's repeatability, measured under the repeatability conditions of Chapter 2. When three technicians test the same monitor, their averages differ as well, because each connects, waits and reads in a slightly different way, and that spread is the station's reproducibility.

Engineers in manufacturing call any measuring instrument or test system a gauge. They call repeatability and reproducibility together the gauge R&R, the variation that the measuring system adds to what it measures, and the study that measures it a gauge study. The Automotive Industry Action Group, AIAG, publishes the method that separates the three sources in its Measurement Systems Analysis reference manual, and engineers call the whole discipline by those words, measurement-system analysis, or MSA. The manual's fourth edition, of June 2010, is the most recent that the American National Standards Institute, ANSI, lists in its store as of October 2026. A pressure test fits the manual's primary focus, which ANSI describes as "systems where readings can be replicated on each part": at a pressure station a technician can test the same monitor again at the same pressure.

54 — The crossed study: ten monitors, three technicians, three trials
10 monitors × 3 technicians × 3 trials = 90 results, in random order SERVICE 150 mmHg reference manometer T-piece result = the monitor's reading minus the reference's rigid vessel A B C −2 −1 0 +1 +2 RESULT BPI tutorial data, no unit 1 2 3 4 5 6 7 8 9 10 MONITOR (PART) 1 TO 10 part-to-part variation (what the test exists to see) repeatability: one technician, three trials reproducibility: the technicians' averages differ Sources: McNeese, BPI Consulting, 2015 (company tutorial); AIAG MSA-4 method as BPI reproduces it. Station schematic.
The quantity being measuredReferences and standardsNoise and errorReadingFocal detail
Plate 54 — A crossed study lets every technician test every part. The spread between parts is what the station must see; the scatter of one technician's repeats and the gap between technicians are what the station adds.

Bias, linearity and stability come first

Before measuring the scatter, the engineer asks whether the station is right on average, because a station that repeats well can still repeat the same error. Engineers who follow MSA estimate the systematic error of Chapter 2 as the station's bias: the average difference between the station's results on a part and that part's known value. That bias belongs to the station, not to the monitor. Linearity is the way the bias changes across the range, and stability is the way it changes over time.

The bias can differ from one point of the range to another. In a worked example of a linearity and bias study by Minitab, which makes statistical software, five parts span the range, and a separate inspection gives each a known value, its master value. The bias is present at three of the five master values and absent at the other two, so a station can be right at one point of its range and wrong at another, and the engineer checks the bias across the whole range.

For the pressure station, a known pressure serves as the part with a known value. A calibration laboratory compares the station's reference manometer with a better reference and finds the station's bias at each test point, from the lowest test pressure to the highest (Chapter 14). Its certificate reports the bias point by point, and so shows the linearity as well. The engineer then follows the stability of the bias between calibrations with the intermediate checks of Chapter 17. No laboratory reports what the technicians add, so the engineer measures that share with a gauge study.

The crossed study

Engineers call a gauge study crossed when every technician tests every monitor, as in the design that opens this chapter: three technicians each test the same ten monitors three times, in random order. The engineer judges the station against the spread of the monitors in the study, so the ten monitors must resemble those that the line produces. Ten monitors chosen to be nearly identical would make any station look poor, and ten chosen to differ widely would make any station look good.

The engineer can analyze the 90 results by the simplest method, the average-and-range method, which needs nothing beyond differences, products and square roots. For each technician and each monitor, the engineer takes the range of the three results, the highest minus the lowest, so 3 technicians and 10 monitors give 30 ranges. A factor from a table turns an average range into a standard deviation. The factor depends on how many results each range covers, because the range of more results tends to be wider for the same scatter. The engineer multiplies the average of the 30 ranges by the factor for three trials and obtains the repeatability as a standard deviation, which MSA calls the equipment variation, EV.

The engineer then multiplies the largest difference between the technicians' averages by the factor for three technicians, squares the result, subtracts a small correction for the repeat scatter that each average already holds, and takes the root. The result is the appraiser variation, AV, appraiser being MSA's word for the person who measures. From the range of the ten monitors' averages and the factor for ten parts, the engineer obtains the part variation, PV, in the same way. The three are standard deviations of independent sources, so the engineer combines them by the root-sum-of-squares of Chapter 11: EV and AV into the gauge's R&R, written GRR, and GRR and PV into the study's total variation, TV. The engineer's main result is %GRR, the gauge's R&R divided by the total variation and expressed as a percentage.

Engineers who follow AIAG's manual judge %GRR by three bands, as McNeese of BPI Consulting, a company that sells a statistics add-in for spreadsheets, reproduces them in a 2015 tutorial. Under 10 % the tutorial's verdict is "Generally considered to be an acceptable measurements system", from 10 to 30 % it is "May be acceptable for some applications", and over 30 % it is "Considered to be unacceptable". Minitab gives the same bands in its help pages and makes the middle one depend on the application, on the cost of the measuring device and on the cost of repair. The manual itself is sold rather than published openly, and both companies cite its fourth edition.

A variance is the square of a standard deviation, and Minitab also states the rule in variances, as the percent contribution to the study's total variance: under 1 % acceptable, 1 to 9 % depending on the application and costs, and over 9 % not acceptable. The two scales are one rule: a ratio of 10 % or 30 % between standard deviations is a ratio of 1 % or 9 % between their squares, since 0.1 squared is 0.01 and 0.3 squared is 0.09. A station that takes a quarter of the spread as a standard deviation therefore takes only a quarter squared, a sixteenth, as a variance. An engineer who reports a share must say which of the two it is.

The number of distinct categories, or ndc, measures how finely the station can sort monitors: it estimates how many groups of parts the station can tell apart within the spread of the parts. Minitab computes it from the ratio of the part variation to the gauge's R&R: it multiplies that ratio by a fixed factor, 1.41, the square root of 2, and cuts the result down to a whole number. Minitab writes that AIAG recommends five or more categories, and that fewer than two are "of no value for controlling the process". A station with one category can only report that every monitor looks the same, because its own scatter covers the real differences between monitors.

Solved example: a gauge study of 90 results by the average-and-range method

The data are the tutorial set that BPI published in 2015, with the design that opens this chapter: 10 parts, 3 operators and 3 trials. The values carry no unit, so they stand for monitors' errors only in their arrangement. The arithmetic is the same in any unit.

  • Repeatability: the average range of the 30 groups of three trials is 0.342, and the factor for three trials, 0.5908, gives EV = 0.342 × 0.5908 = 0.202.
  • Reproducibility: the operators' averages differ by at most 0.445, and the factor for three operators, 0.5231, gives 0.445 × 0.5231 = 0.2328, whose square is 0.0542. The correction is EV squared divided by the 30 results of each operator, because averaging 30 results divides the repeat scatter by the square root of 30, and its square by 30 (Chapter 5): 0.0408 ÷ 30 = 0.00136, about 0.0014. Then 0.0542 − 0.0014 = 0.0528, and its square root gives AV = 0.230.
  • The gauge: EV squared is 0.0408, and AV squared is the 0.0528 of the step above. Their sum is 0.0936, and its square root gives GRR = 0.306.
  • The parts: the ten parts' averages span 3.511, and the factor for ten parts, 0.3146, gives PV = 3.511 × 0.3146 = 1.105.
  • The total: 1.105 squared is 1.2210, the sum 0.0936 + 1.2210 is 1.3146, and its square root gives TV = 1.147.
  • %GRR = 0.306 ÷ 1.147 = 26.7 %, inside the band from 10 to 30 %. BPI's table gives 26.68 %.
  • ndc = 1.41 × 1.105 ÷ 0.306 = 5.09, cut down to 5, which just meets AIAG's five.

The gauge takes about 27 % of the study's spread as a standard deviation and about 7 % as a variance, since 0.0936 ÷ 1.3146 is about 0.071. AV, at 0.230, exceeds EV, at 0.202, so the operators differ from one another more than any one of them repeats. If these were errors from a monitor station, the first fix would lie in what the technicians do differently, such as how long each waits before reading, and only then in the equipment.

55 — The study's spread, added as squares
DRAWN × 4 0.0528 0.0408 AV 0.230 reproducibility EV 0.202 repeatability 0.0408 + 0.0528 = 0.0936 GRR 0.306 0.0936 the gauge's square at true scale GRR 0.306 PV 1.105 part variation 1.2210 TV 1.147 1.3146 total variation TRUE SCALE 0.0936 + 1.2210 = 1.3146 number of distinct categories ndc = 1.41 × 1.105 ÷ 0.306 = 5.09, cut to 5 0 10 % 30 % 40 % generally acceptable may be acceptable for some applications unacceptable 26.7 % share of the standard deviation (%GRR) 0 1 % 9 % 12 % 7.1 % share of the variance Sources: BPI tutorial data, no unit; average-and-range method; bands after AIAG MSA-4 as BPI and Minitab give them.
The quantity being measuredNoise and errorLimits and decisionsFocal detail
Plate 55 — The station's two kinds of scatter combine as squares, and so do the station and the parts. Taken as a standard deviation, the gauge is 26.7 % of the spread; taken as a variance, it is about 7 %.

Statistics software usually analyzes the same results by the analysis of variance, which splits the sum of the squared deviations directly into its sources. BPI ran that analysis on the same 90 results and found the gauge at 7.76 % of the total variance, in the same middle band as the average-and-range method's 7 %. Because the two methods give slightly different numbers, the engineer reports which one produced each figure and never mixes them in one table.

A study near a band's edge can pass by one method and fail by another

By BPI's analysis of variance, the part variance is 1.086 and the gauge variance 0.0914, and their square roots are 1.042 and 0.302. Minitab's rule then gives 1.41 × 1.042 ÷ 0.302 = 4.86, cut down to 4 distinct categories, where the average-and-range method gave 5. The same station therefore meets AIAG's recommended five by one method and misses it by the other. A result that close to an edge shows that the station is marginal, and the useful response is to reduce its scatter, not to report the method that passes.

A monitor that shows whole millimeters of mercury can hide the station's repeatability, as Chapter 10 showed for repeated readings: three trials that all display 150 give a range of zero, although the pressure behind them scattered. The study then reports a repeatability that belongs to the display step, not to the station. The engineer therefore records the monitor's value before rounding, from a service mode or a data link where the maker provides one, and otherwise reports the display step beside the result.

A verdict needs its own study

Some stations record a verdict instead of a number. Rossmax's CalPro, the field system of Chapter 14, shows PASS or FAIL, and a production station that applies the decision rule of Chapter 13 may keep only that result. A verdict cannot be averaged, so the engineer judges it by attribute agreement: how often one technician gives the same verdict on the same monitor when repeating, how often different technicians agree, and how often the verdicts match the standard, the verdict known to be right for each monitor. Minitab reports consistency apart from correctness, because agreement alone does not show that a verdict is right: "Consistent ratings aren't necessarily correct ratings." Two technicians who pass the same faulty monitor every time agree perfectly and are wrong every time.

Minitab measures agreement with a statistic called kappa, which corrects the share of agreeing verdicts for the share that chance alone would produce. Minitab writes that AIAG suggests a kappa of at least 0.75 for good agreement and that larger values, such as 0.90, are preferred. For a pressure station, the hard cases are monitors whose true errors lie close to the limit. Technicians shown only clearly good monitors would agree almost every time and prove nothing, because every verdict would be easy.

Where the rules come from

A maker with several identical stations need not repeat the whole study on each one, and the rule that allows this comes from the car industry, as the method does. IATF, the International Automotive Task Force, owns the automotive quality standard IATF 16949, and in its list of frequently asked questions it answers that a complete study of each single instrument is not required. Instruments with the same characteristics "can be grouped", and "a sample instrument (representative of the gauge family) can be used", so one station can serve as the sample for the stations identical to it. IATF 16949:2016 is the edition in force as of October 2026, and IATF plans a revision for mid-2027, after which the answer may change.

Device makers encounter the gauge study in process validation, in which a maker shows with evidence that a manufacturing process consistently gives products that meet their requirements. The Global Harmonization Task Force, GHTF, an international body, published guidance on process validation in 2004. In an annex, GHTF states that "When studying variation, good measurements are required", and that "Many times an evaluation of the measurement system should be performed using a gauge R&R or similar study". GHTF describes the gauge study as a means of "evaluating the precision and accuracy of a measurement device". GHTF closed in 2012, and the guidance remains available in the archive of GHTF's documents that its successor, the International Medical Device Regulators Forum, IMDRF, keeps. In Chapter 23 the engineer follows the same guidance from the gauge to the whole test method.

The engineer can also judge a measuring process by a second route, set out in ISO 22514-7, on the capability of measurement processes. On that route the engineer builds an uncertainty budget for the process and judges the resulting uncertainty with a capability ratio, which sets it against the tolerance the process must check. ISO describes the standard as following the approach of the GUM, the method of Chapter 11, and as "primarily developed to be used for simple one-dimensional measurement processes", a class that includes a pressure reading. ISO 22514-7 is in force in its 2021 edition with an amendment of 2024, and its third edition is in the approval stage as of October 2026. A gauge study instead measures the scatter that actually appears, operators included, and the two answers check each other.

Running a gauge study on a production pressure test
  • Study the station before it sorts product, and again after any change to its reference, its fixture or its procedure.
  • Choose ten or more monitors that span the errors the line produces, and have each technician test each monitor at least twice, in random order.
  • Record the monitor's value before rounding where a service mode or data link provides it, and report the display step where it does not.
  • Report %GRR, the percent contribution and the number of distinct categories, each with the method that produced it.
  • Where the station records only PASS or FAIL, run an attribute study with monitors known to lie inside, outside and close to the limit, and report correctness against the standard apart from consistency.
+ What this chapter established
  • A production pressure test adds scatter of its own, and a gauge study separates that repeatability and reproducibility from the real differences between monitors.
  • The gauge's R&R and the part variation add as squares, and by AIAG's manual, as Minitab and BPI reproduce it, a %GRR under 10 % is generally acceptable, 10 to 30 % may be acceptable for some applications, and more than 30 % is unacceptable.
  • In BPI's tutorial set the gauge takes 26.7 % of the spread as a standard deviation and gives 5 distinct categories by the average-and-range method, while an analysis of variance of the same data gives 4.
  • A station that records only verdicts needs an attribute study, in which agreement among technicians is kept apart from agreement with a known standard.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.