+ Part VIII · The industry in 2026

The makers at each link.

More than four fifths of the monitor's uncertainty budget at 150 mmHg, counted in squares, rests on two figures that organizations outside the monitor maker state: OIML's limit for the reference manometer and Omron's hysteresis for the sensor (Chapter 29). The companies behind such figures change, and one company can hold several links of the chain at once, as maker, seller and tester. A&D Company held all three roles for scales until 14 July 2026, when Japan's Ministry of Economy, Trade and Industry, METI, revoked its designation as a body that verifies weighing instruments on the state's behalf. A&D makes balances, scales, load cells and blood-pressure monitors, and METI had designated it to verify non-automatic scales other than vehicle scales, and automatic catchweighers. In September 2025 A&D disclosed that it had repaired scales without removing their verification marks, which Japan's Measurement Act requires a repairer to remove. A&D's own site still lists verified scales among its products, but A&D may no longer verify scales itself. Part VIII names who holds each link of the monitor's chain, with every company's status dated to November 2026 and every company figure labeled as the company's own.

29 — The makers, 2026

Who builds the chain.

+ The questionWho makes the references, instruments, sensors, converters and calibration services the chain depends on, India included?

The sensor's maker changed owners

On 1 October 2026 Omron transferred the shares of Aratas Corporation to a company established by The Carlyle Group, and Aratas left the Omron Group. Aratas makes the 2SMPP-03, the pressure sensor at the start of the monitor's chain (Chapter 1). The sensor's maker had been part of Omron itself, as its Device & Module Solutions business, until 1 July 2026, when that business passed by a company split to a subsidiary and took the name Aratas. Omron kept a minority interest after the sale: it invested in the parent of Carlyle's company for 5 % of the voting rights, as Aratas states it.

For the engineer who designs the monitor, the sale changed a company's name and nothing in the part. The 2SMPP-03 is a small part of the business that was sold: Omron's notice of the sale gives the value of the business transferred as 81,000 million yen and states no price, and Aratas describes itself as a maker of relays, switches, connectors and sensors with about 6,000 employees. Aratas lists the 2SMPP-03 as in production, and the data sheet on Aratas's site still carries Omron's name and catalog number. Every sensor figure in Chapter 11's budget, from the span tolerance to the hysteresis, is therefore Omron's statement about a part that Aratas now makes and sells. A design file that names the source of each figure needs two names and a date: Omron as the author of the data sheet, and Aratas as the maker since 1 July 2026.

Other companies make such sensors, and several chip makers supply the links that follow. TDK Electronics makes the C32/2, a bare sensor die, and NXP the MPXx2050 family, and both state that their sensors' outputs are ratiometric, the property that lets a design do without a separate voltage reference (Chapters 4 and 6). Texas Instruments, TI, supplies two chips of the reference design in Chapter 1: the INA350 instrumentation amplifier, and the MSPM0L1306 microcontroller, whose 12-bit converter digitizes the bridge's output. Analog Devices makes the bridge converters and voltage references that Part II cites. Microchip Technology demonstrates its 24-bit MCP3564 converter in a weigh scale, and Renesas offers a blood-pressure evaluation kit built on its RL78/H1D microcontroller.

Two of these chip makers agreed in 2026 to buy other companies, and neither purchase includes a converter business. TI agreed on 4 February 2026 to buy Silicon Labs, a maker of wireless processors, with closing expected in the first half of 2027. Analog Devices agreed on 9 September 2026 to buy Alif Semiconductor, a maker of low-power microcontrollers, with closing expected in the last quarter of 2026.

Monitors, and the names on them

The monitor makers cited earlier state their own accuracy figures. A&D Company states ±3 mmHg for the cuff pressure of its UA-651, and Microlife AG states the same accuracy and a 1 mmHg display step for two of its monitors (Chapter 8). A&D also holds three layers of the chain at once, because its own range runs from load cells and balances to blood-pressure monitors. It is a subsidiary of A&D HOLON Holdings. When METI revoked A&D's verification designation in July 2026, the decision concerned its scales, not its monitors (Part VIII).

The brand on a monitor's box does not always name the company that made it. Under the EU's device regulation the manufacturer issues the device's EU declaration of conformity, its statement that the device meets the regulation's requirements (Chapter 31). The declaration therefore names the company that the regulation treats as the manufacturer, even where the box carries another brand. Joytech Healthcare of Hangzhou issues that declaration for several of Beurer's arm and wrist monitors, among them the BM 92, and Ya Horng Electronic of Taiwan makes Garmin's Index BPM. An engineer who buys a monitor as a reference device, or audits a supplier, therefore reads the declaration and the certificate behind it.

Monitors are tested with instruments of their own. A simulator reproduces the oscillations that a cuff picks up from an arm, so it tests the monitor's estimate of blood pressure as well as its pressure reading, and FDA recognizes a technical specification for the repeatability of such simulators (Chapter 31). Fluke Biomedical of Everett, Washington, makes the BP Pump 2, a simulator and tester for non-invasive blood-pressure monitors, and the ProSim 8, another simulator. Fluke Biomedical belongs to Fortive. Rigel Medical makes another such instrument, the Uni-SiM, whose calibration certificate the engineer reads in Chapter 12.

A name on a data sheet or a box dates from its printing

The 2SMPP-03's data sheet carries Omron's name, and a Beurer monitor's box carries Beurer's. Neither tells who makes the part or the device today. The data sheet was written before the business became Aratas, and the box names a brand whose declaration of conformity another company issues. A supplier record that copies a name from a document therefore records who wrote the document, on the day it was printed. The current maker comes from a dated record of its own: an owner's filing, a release or a declaration of conformity.

The reference layer changed owners too

Above the monitor's calibration station stands the reference manometer, and one maker of such references also changed owners in 2026. On 1 January 2026 Crane Company completed its purchase of Baker Hughes's precision sensors and instrumentation business, which included Druck, Panametrics and Reuter-Stokes. Druck, which now calls itself "A Crane Company brand" on its site, makes instruments that generate, control and measure pressure for calibration. Druck also sells 4Sight2, software for managing calibrations, which places it in the software layer as well.

India's national institute uses a Druck gauge as the reference of a rig that calibrates 20 monitors together. Kumar and four colleagues at CSIR-NPL described the rig in a 2023 dataset (Chapter 23): a Proportion-Air regulator driven by a National Instruments card sets the pressure, and a Druck DPI-104 digital gauge reads it as the reference. The authors cite OIML's Recommendations R 148 and R 149 (Chapter 30), but the public description of their dataset gives neither the test points, nor the limits, nor the DPI-104's accuracy. The rig belongs to a national institute, not to a factory.

70 — One reference gauge for twenty monitors
only a better reference, one link up, can show it camera reads each display (image processing) 20 MONITORS AT ONCE DRUCK DPI-104 REFERENCE Proportion-Air regulator set by an NI card one manifold: the same air pressure at every monitor one offset, if present, e.g. +0.5 mmHg (illustrative) each result: the monitor's reading minus the gauge's, mmHg 0 −0.5 −0.5 −0.5 the same error in every result: comparing the monitors with each other cannot show it Rig: Kumar et al. (CSIR-NPL), IEEE DataPort dataset, 2023; its public description gives no test points, limits or gauge accuracy. The 0.5 mmHg is illustrative. Drawing schematic; display digits drawn as outlines.
The quantity being measuredReferences and standardsNoise and errorReadingFocal detail
Plate 70 — CSIR-NPL's rig compares twenty monitors at once against one Druck gauge. Any error in that gauge is shared by all twenty results, so only the gauge's own calibration against a better reference can show it.

A rig that shares one reference among twenty monitors cannot find that reference's error by itself. The reference's error enters every monitor's result at the same moment and in the same direction, so it is one shared cause, not twenty independent errors (Chapter 11). If the DPI-104 read 0.5 mmHg high, all twenty monitors would appear 0.5 mmHg low, and comparing their results with each other would never show it. Only a comparison of the gauge with a better reference, at the link above the rig, can find that error. The gauge's own certificate therefore carries more weight than any number of monitors that agree with it.

Other makers supply the same reference layer, with instruments, calibration software and the masses for dead-weight testers. Beamex of Finland makes the MC6 family of field calibrators, a bench calibrator and the POC8 pressure controller, with calibration-management software of its own, CMX and LOGiCAL. Fluke, a Fortive business, makes calibrators and offers CalStudio, a cloud platform for calibration procedures. Nagman Instruments and Electronics of Chennai makes temperature, pressure and signal calibrators, hand pumps and dead-weight testers. A dead-weight tester generates a pressure the way Chapter 15's pressure balance does, with masses on a piston. Such masses come from makers of reference weights, among them Troemner of Thorofare, New Jersey, a unit of Mettler-Toledo since 2016. Troemner's calibration laboratories are accredited to ISO/IEC 17025, by its own statement (Chapter 16).

Institutes and laboratories

At the top of the chain stand the BIPM and the national institutes, which are public bodies: among them NIST in the United States, PTB in Germany, NPL in the United Kingdom and CSIR-NPL in India. The institutes realize and disseminate units, and they take part in the CIPM's arrangement for mutual recognition rather than holding accreditation as a laboratory does (Chapter 21). CSIR-NPL states that it keeps a quality system to ISO/IEC 17025:2017 and to ISO 17034:2016, the standard for producers of reference materials. CSIR-NPL also calibrates for customers directly: its calibration and testing service, by its own count, serves nearly 4,200 of them from industry, laboratories and government.

Between the institutes and the factories stand the calibration laboratories, and two groups among them grew by purchase around 2026. Trescal, headquartered in Paris, has been a portfolio company of EQT Infrastructure V since 2023. In January 2026 Trescal announced 14 acquisitions. A trade publication reported that one of them took Trescal into India, through a domestic laboratory in Bengaluru and Hyderabad offering NABL-accredited services, but did not name the laboratory. Transcat, a US company listed on Nasdaq, reports 33 calibration service centers with ISO/IEC 17025:2017 scopes of accreditation at the end of its fiscal year on 28 March 2026. It bought SCM Metrology and Laboratories of Costa Rica with effect from 9 April 2026.

Layer of the chainCompanies named in this chapterOwner, as of November 2026
SensorAratas (2SMPP-03); TDK Electronics; NXPAratas: a Carlyle company, since 1 October 2026
Amplifier, converter, referenceTexas Instruments; Analog Devices; Microchip; RenesasTI and Analog Devices: listed, each with a purchase pending
MonitorA&D Company; Microlife; Joytech for Beurer; Ya Horng for GarminA&D: A&D HOLON Holdings
Monitor testerFluke Biomedical; Rigel MedicalFluke Biomedical: Fortive
Pressure reference and calibratorDruck; Beamex; Fluke; NagmanDruck: Crane Company, since 1 January 2026
Reference weightsTroemnerMettler-Toledo, since 2016
Calibration serviceTrescal; Transcat; NABL-accredited laboratoriesTrescal: EQT Infrastructure V, since 2023
Calibration softwareDruck 4Sight2; Beamex CMX and LOGiCAL; Fluke CalStudioDruck: Crane Company; Fluke: Fortive

Whose figures the reading rests on

The owners in that table matter to the monitor's maker because its reading's uncertainty is built from their figures. In Chapter 11 the engineer's budget gave a combined standard uncertainty of 0.70 mmHg for the corrected reading at 150 mmHg, and each of its rows rests on a figure that some organization states. With each row set beside its owner, the maker can see how much of that uncertainty it controls in its own design and how much it obtains from suppliers and laboratories.

Solved example: whose figures set the corrected reading's uncertainty at 150 mmHg

Chapter 11's five rows that carry a number, each as a standard uncertainty, its square, and the organization whose figure it is:

  • Reference manometer: 0.462 mmHg, squared 0.213 mmHg². The ±0.8 mmHg limit is OIML's. The monitor maker chooses a gauge that meets it, from a reference maker such as Druck, and an accredited laboratory calibrates the gauge.
  • Hysteresis: 0.433 mmHg, squared 0.187 mmHg². The 0.2 %FS limit is Omron's figure for a part that Aratas now makes.
  • Display step: 0.289 mmHg, squared 0.084 mmHg². The 1 mmHg step is the monitor maker's own design.
  • Converter step and front-end noise: 0.031 and 0.014 mmHg, squared 0.001 and 0.0002 mmHg². Both come from TI's design and data.

The squares add to 0.485 mmHg², and the square root of that sum is Chapter 11's 0.70 mmHg. Dividing each square by 0.485 gives the shares: 0.213 divided by 0.485 is 0.44, so the reference supplies 44 %, the sensor 39 %, the display 17 % and TI's two chips 0.2 %. The reference and the sensor together supply 0.400 mmHg², and since 0.400 divided by 0.485 is 0.82, that is more than four fifths of the total, from figures that two organizations outside the monitor maker state.

71 — The corrected budget, by whose figure each row is
side = standard uncertainty in mmHg, one scale for every square; the areas add 44 % REFERENCE MANOMETER 0.462 mmHg 0.213 mmHg² OIML's ±0.8 mmHg limit; the maker's gauge (e.g. Druck) 39 % SENSOR HYSTERESIS 0.433 mmHg 0.187 mmHg² Omron's data-sheet figure; part made by Aratas since 1 Jul 2026 17 % DISPLAY STEP 0.289 mmHg 0.084 mmHg² the monitor maker's own design 0.2 % CONVERTER, FRONT END 0.031, 0.014 mmHg 0.001, 0.0002 mmHg² TI's design and data areas add 44 % 39 % 17 % COMBINED 0.70 mmHg total 0.485 mmHg², root 0.70 mmHg (Chapter 11) 0.400 of 0.485 mmHg², more than four fifths: figures stated outside the monitor maker Sources: Chapter 11's rows; shares worked from the figures above. Sensor figures are Omron's; TI's figures are TI's.
References and standardsNoise and errorUncertaintyFocal detail
Plate 71 — More than four fifths of the corrected reading's uncertainty comes from two figures that organizations outside the monitor maker state: OIML's limit for the reference and Omron's hysteresis for a sensor that Aratas now makes.

The maker sets the display step itself, but it can change the two largest rows only through its choice of supplier and of laboratory. A reference whose certificate states an expanded uncertainty well below OIML's limit would shrink the largest row, as the patient simulator's 0.09 mmHg in Chapter 12 shows, while a sensor with less hysteresis would shrink the second. The owner changes of 2026, at Aratas and at Druck, altered none of these figures. They altered the name that the design file must record beside two of them, and the date from which each name holds.

India, as the records show

India's chain for the reading runs through Indian institutions. CSIR-NPL keeps the country's primary standard of pressure, an air piston gauge, and piloted a key comparison in gas pressure (Chapters 15 and 21). NABL accredits the calibration laboratories below the national institute, and its register held 1,305 of them in early September 2026 (Chapter 16). Five Regional Reference Standard Laboratories link the national institute to the States' legal-metrology laboratories (Chapter 30).

Indian makers build instruments and sensors of their own. Besides Nagman, ADI ARTECH Transducers of Vadodara makes strain-gauge load cells and force, torque and pressure sensors, Essae-Teraoka of Bengaluru makes weighing scales and balances, and Tempsens of Udaipur makes temperature sensors and calibrators. The pages of all four state neither their owners nor an accreditation body for a laboratory of their own.

For the monitor itself, India has an announced factory but no confirmed production. Omron Healthcare established a company in India on 14 July 2022 to run a factory of about 24,000 m² in Chennai. In June 2023 Omron Healthcare stated that the factory would begin making blood-pressure monitors in March 2025, with the date still to be confirmed. In July 2025 a trade publication quoted the managing director of Omron Healthcare India, Hiroshi Ogawa: "Currently we are sourcing BPMs from Vietnam," BPMs being blood-pressure monitors. No known record confirms that production has started, and the factory belongs to a Japanese company.

Indian companies sell monitors, but no known record shows one of them making home automated monitors in India. Morepen Laboratories stated in a 2021 filing to the stock exchanges that it led the domestic market for blood-pressure monitors, and it reported revenue of Rs 26 crore, about 260 million rupees, from them in the quarter from April to June 2024. Neither document says where its monitors are made. BPL Medical Technologies sells a fully automatic monitor whose product page names no maker or country of origin. Skanray Technologies of Mysuru holds an ISO 9001 certificate whose scope includes the contract manufacture of multi-parameter patient monitors. Such a monitor can carry a blood-pressure channel, but neither the certificate nor Skanray's About page names one, and a hospital patient monitor is a different device from the home monitor this book follows.

72 — India's part of the chain, as the records show
NATIONAL STANDARD CALIBRATION LABORATORIES REFERENCES, CALIBRATORS SENSORS FINISHED INSTRUMENTS CSIR-NPL, NEW DELHI air piston gauge (Chapter 15); 20-monitor rig with a Druck reference NABL 1,305 calibration laboratories, early Sep 2026 (GTAI's figure) FIVE REGIONAL REFERENCE STANDARD LABORATORIES linking CSIR-NPL to the States' laboratories (DoCA) NAGMAN, CHENNAI pressure calibrators, dead-weight testers ADI ARTECH, VADODARA load cells; force, torque, pressure sensors TEMPSENS, UDAIPUR temperature sensors, calibrators home automated monitor made in India by an Indian-owned company: no record found ESSAE-TERAOKA, BENGALURU scales, balances THE NEAREST RECORDS OMRON HEALTHCARE, CHENNAI FACTORY planned from Mar 2025; 'sourcing BPMs from Vietnam' (Jul 2025, as reported); owned by a Japanese company SKANRAY, MYSURU contract manufacture of multi-parameter patient monitors (ISO 9001 scope); blood pressure not named Status as of November 2026, from the records read. Company statements are the companies' own. Drawings schematic.
References and standardsThe signal in the chainReadingFocal detail
Plate 72 — India's chain for the reading runs through Indian institutions and makers of references, but no record read shows an Indian-owned company making home automated monitors in India.
Keeping a supplier record for a measuring chain
  • Write beside each row of the uncertainty budget the organization that states the figure, the document and its date, and the company that makes the part today.
  • Check the owners of the suppliers behind the largest rows at every design review; in 2026 alone the maker of the monitor's sensor and a reference maker changed owners.
  • Record for each reference manometer its maker, the laboratory that calibrates it, that laboratory's accreditation body and scope, and the certificate's date.
  • Take the maker of a bought device from its declaration of conformity or its regulator's record, not from the brand on the box.
  • For an Indian supplier, confirm a laboratory's accreditation and scope in NABL's own register before relying on it.
+ What this chapter established
  • In 2026 the maker of the monitor's sensor became Aratas, owned since 1 October 2026 by a Carlyle company, and Druck, a maker of pressure references, became a Crane brand, while the parts and their published figures stayed the same.
  • A data sheet or a brand names who wrote or sold it on its date, so a supplier record takes the current maker from a dated filing, release or declaration of conformity.
  • In Chapter 11's corrected budget, the reference manometer and the sensor's hysteresis supply more than four fifths of the total of squares, from figures stated outside the monitor maker, which sets only the display step's 17 % itself.
  • India's chain for the reading runs from CSIR-NPL through NABL's laboratories, while no known record shows an Indian-owned company making home automated monitors in India.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.