+ Part VII · The measuring system

As a system.

The sensor in TI's blood-pressure design can meet every figure on Omron's data sheet with an offset 4.0 mV from the typical value, about 36 mmHg at the cuff, and the monitor built around it can still read within ±3 mmHg, because its firmware measures that offset and subtracts it (Chapter 11). The reverse also happens: parts that each meet their own data sheets can make a monitor that fails, because no data sheet says how its errors add to those of the next part. This part treats the cuff-pressure channel as one engineered system. Chapter 25 shares one accuracy requirement among the people who own the parts, and Chapter 26 follows what a maker must do when a reference or a part turns out to have been wrong. Chapter 27 asks how the monitor stays accurate in use, and Chapter 28 walks the whole chain from the Planck constant to the cuff, which answers the central question.

25 — Error budget

The error budget as a requirement.

+ The questionHow is one accuracy requirement shared among the sensor, the electronics, the code and the calibration, and who owns each share?

One line becomes seven numbers

One line in a monitor's requirements states that the cuff pressure shall read within ±3 mmHg or 2 % of the reading, whichever is greater. A design team divides that line into seven smaller numbers, and different people own them. The maker cannot loosen the limit, because bodies outside the firm set it. OIML sets the limit in its 2020 recommendation for automated sphygmomanometers, at any point of the range, with the pressure rising or falling, from 10 to 40 °C. India's Department of Consumer Affairs set the same figures for the same devices in a rule of July 2025 (Chapter 30). The limit's two branches meet at 150 mmHg, where 2 % of the reading is 3 mmHg. The shares are therefore worked out at that pressure.

The seven numbers are the rows of the uncertainty budget that the engineer drew up at 150 mmHg in Chapter 11. Five rows carry a number there: the reference manometer used at calibration, the sensor's hysteresis, the display step, the converter step and the front end's noise. Two more rows carry only a name: the residuals that the firmware leaves after it corrects the sensor's nonlinearity and its drift with temperature. Chapter 11's budget asked whether a finished design would meet the limit. Before anyone chooses the parts, a design team uses the same rows as limits that each part must stay within.

Engineers call a budget used as a requirement an error budget: a set of shares for one measured quantity whose combination stays within the instrument's limit, each with its size, its owner and its proof. Before the design exists, the team decides how large each row may be, names the person or the firm who will deliver it, and states how each share will be shown to be met. Dividing a requirement among the parts of a design is a step of systems engineering, "Allocate system requirements to functions and to next lower level" in the words of NASA's systems engineering handbook.

61 — One requirement, seven shares, each with an owner
cuff pressure within ±3 mmHg or 2 % of the reading (OIML R 149-1; India Part VII-B) sensor maker hysteresis limit 0.75 mmHg electronics noise 0.045 mmHg electronics converter step limit 0.0545 mmHg firmware team nonlinearity residual, named firmware team temperature residual, named firmware team display step limit 0.5 mmHg calibration engineers reference limit 0.8 mmHg share set by the design team ×50 MCU 150 mmHg calibration bench algorithm (estimate) owned apart: no row in this budget requirement → share → verification → record failed test → requirement threatened Rows and limits from Chapter 11 (TI TIDA-010266 design; Omron 2SMPP-03 data sheet; OIML R 149-1 and R 149-2). Owners are this guide's reading. Schematic.
References and standardsThe signal in the chainLimits and decisionsReadingFocal detail
Plate 61 — An error budget used as a requirement gives each share a size and an owner. Five shares carry Chapter 11's numbers; the two firmware residuals are sized by the design team, and the algorithm's estimate sits outside the budget.

Worst case or root-sum-of-squares

The team first decides how the shares are to add, because the method decides how large each share may be. The worst case adds every share at its limit, as if every error sat at its edge in the same direction at once. Eicke and colleagues of NASA's metrology and calibration working group compared three methods in the agency's 1995 guidelines for measurement assurance. They wrote that the linear method, the worst case, "assumes that all errors are additive in one direction". The quadrature method, the root-sum-of-squares, "yields the smallest uncertainty of the three", by their account, and the estimates "can differ significantly (sometimes as much as a factor of 2)". For the monitor, that difference decides how much of the 3 mmHg is left for the firmware's corrections.

An engineer may use the root-sum-of-squares where the errors are independent, because independent errors seldom all push the reading the same way at the same moment (Chapter 11). None of the monitor's five rows shares a cause with another. The hysteresis belongs to the sensor, the display and converter steps come from rounding, the noise is random, and the reference manometer is a separate instrument with errors of its own. The worst case still has a use beside the root-sum-of-squares, because it describes a single bad unit in which every part sits at its limit, and so gives a bound that no unit built from parts within their limits can exceed.

Solved example: what the two methods leave for the firmware's corrections at 150 mmHg

The five numbered rows of Chapter 11 enter as limits: the reference manometer 0.8 mmHg, the hysteresis 0.75 mmHg, half the display step 0.5 mmHg, half a converter step 0.0545 mmHg and half the front end's peak-to-peak noise 0.045 mmHg.

  • Worst case: 0.8 + 0.75 + 0.5 + 0.0545 + 0.045 = 2.15 mmHg. The limit of 3 mmHg leaves 3 − 2.15 = 0.85 mmHg for the two residuals together, or 0.85 ÷ 2 = 0.425, about 0.43 mmHg each, if they share it equally.
  • Root-sum-of-squares: the five standard uncertainties square to 0.485 mmHg² in all (Chapter 11). The expanded uncertainty, here twice the combined standard uncertainty (Chapter 12), must stay within 3 mmHg. The combined value may therefore reach 1.5 mmHg, whose square is 2.25 mmHg². That leaves 2.25 − 0.485 = 1.765 mmHg² for the residuals.
  • Shared equally, each residual may take 1.765 ÷ 2 = 0.88 mmHg². Its square root, 0.94 mmHg, is the residual's standard uncertainty. As a flat distribution, that is a half-width of 0.94 times the square root of 3: 0.94 × 1.732 = 1.63 mmHg.

By the root-sum-of-squares, each correction may leave a band of ±1.63 mmHg, about 3.8 times the ±0.43 mmHg that the worst case allows, since 1.63 ÷ 0.43 is 3.8. The choice of method therefore decides how good the firmware's table of linearity corrections and its temperature correction must be. Neither figure can go to the firmware team in full, because the same 3 mmHg must also contain the guard band of the production test (Chapter 13).

62 — Worst case against root-sum-of-squares, at 150 mmHg
WORST CASE: LIMITS ADDED IN A LINE every row at its limit, the same way LIMIT 3 mmHg reference 0.8 hysteresis 0.75 half display step 0.5 converter 0.0545 noise 0.045 0.85 mmHg left: about ±0.43 mmHg for each residual five limits end to end: 2.15 mmHg ROOT-SUM-OF-SQUARES: SQUARES ADDED the five standard uncertainties, squared 3 ÷ 2 = 1.5 mmHg allowed: (3 ÷ 2)² = 2.25 mmHg² 1.765 mmHg² left: about ±1.63 mmHg for each residual five rows: 0.485 mmHg² ±1.63 against ±0.43: about 3.8 times Sources: inputs from Chapter 11; method as Eicke et al., NASA RP-1364 (1995).
UncertaintyLimits and decisionsFocal detail
Plate 62 — Adding limits in a line leaves the firmware's two corrections about ±0.43 mmHg each; adding standard uncertainties as squares leaves about ±1.63 mmHg each. The method chosen at design sets how good the corrections must be.

A standard that writes the shares down

OIML has written such a division into its rule for weighing instruments, because firms often build a commercial scale from modules that other firms make. In R 76-1, its 2006 recommendation for non-automatic weighing instruments, OIML lets three modules each take a fraction of the instrument's maximum permissible error. They are the load cell, which senses the load, the electronic indicator, which shows the result, and the connecting elements between them. OIML requires the fractions to satisfy an equation, which appears to limit the sum of their squares to 1. Its default fractions are 0.7 for load cells, 0.5 for electronic indicators and 0.5 for connecting elements. Squared, they are 0.49, 0.25 and 0.25, which add to 0.99 and meet the rule, while the fractions themselves add to 1.7.

Three modules may therefore claim 170 % of the instrument's limit between them, added directly, because OIML expects their errors to combine as squares, as independent errors do. OIML also bounds each share. Where more than one module contributes, a fraction may not exceed 0.8 or fall below 0.3, a purely digital device may take 0, and a module that is already a complete weighing module may take 1. The firm that assembles the scale then checks that the modules fit each other. Detecto, a scale maker, records such checks on its worksheet for one of its platform scales: the sum of the fractions, the number of scale intervals, or display steps, that each module allows, and the load cell's signal for each interval at the indicator's input.

For a blood-pressure monitor the maker writes the shares itself, because OIML wrote this allocation for weighing instruments only. OIML and India set one limit for a whole sphygmomanometer and divide none of it among its parts. A maker can still take two things from the weighing rule that any allocation needs: a stated method by which the shares add, and one agreed number that both firms check wherever one firm's module meets another's.

63 — OIML R 76-1: three modules share one limit
platform scale, exploded into three modules kg scale intervals each module allows signal per interval at the indicator's input LOAD CELL 0.7 CONNECTING ELEMENTS 0.5 INDICATOR 0.5 sum of fractions 0.7 0.5 0.5 1 1.7 the shares added in a line: 1.7 allowed because errors add as squares the instrument's limit, squared 0.7² = 0.49 0.5² = 0.25 0.5² = 0.25 0.01 unfilled 0 1 squares add to 0.99 Sources: OIML R 76-1:2006, 3.10.2.1 and Table 7 (the sum-of-squares condition as this guide reads the text); check names from Detecto's composition-of-modules worksheet (company document). Weighing instruments only.
The signal in the chainUncertaintyLimits and decisionsReadingFocal detail
Plate 63 — OIML's weighing rule lets three modules from different firms claim 0.7, 0.5 and 0.5 of one limit, because their squares add to 0.99. The firm that assembles the scale checks the joints between them.

Each share has an owner

Someone owns each of the monitor's seven rows. Engineers call the boundary between two owners an interface, the place where one part's output becomes another's input. The sensor's maker answers for its hysteresis and its nonlinearity before correction, as Omron's data sheet states them. The electronics team answers for the converter step and the front end's noise, which TI's design sets. The interface between these two owners is the bridge's output of about 112 µV for each millimeter of mercury, a number that the sensor must deliver and the amplifier must accept. The firmware team answers for the display's rounding and the two residuals, because the coefficients and the correction tables are its work (Chapter 9). A separate team owns the algorithm that computes systolic and diastolic pressure, since its error belongs to the estimate, which no row of this budget covers (Chapter 24).

Three more owners stand outside the monitor: the calibration chain, the production test and the user's surroundings. The maker's calibration engineers answer for the reference manometer's row, which OIML caps at ±0.8 mmHg, and the laboratory that calibrates the manometer answers for the uncertainty on the manometer's certificate. How small that uncertainty must be follows from the four-to-one ratio of Chapter 12. Eicke and colleagues stated in the same 1995 guidelines that a calibration's uncertainty "must be no greater than 1/4 of the tolerance of the instrument or standard being calibrated". For a manometer with a tolerance of ±0.8 mmHg, that is 0.8 ÷ 4 = 0.2 mmHg at most.

The production test team answers for the uncertainty of the check that each monitor meets the limit, which Chapters 22 and 23 examine. The user's surroundings set the conditions: OIML's limit holds from 10 to 40 °C, and Omron states its temperature figures from 0 to 85 °C, so the sensor's figures cover the whole band. No single published source names all these owners. The maker therefore draws up the list from its own chain.

The design team keeps each interface under control by writing its number, with its unit and its limit, into both owners' documents, so that a change by one owner shows up as a broken requirement for the other. That practice is "establishing and using formal interface management to maintain internal and external interface definition", in the words of NASA's systems engineering handbook. On the monitor, the budget assumes that the firmware's auto-zero step removes the sensor's offset (Chapter 11). A firmware engineer who changes that step therefore changes how much of the offset reaches the reading, and the team records the change in the budget before it reaches the code.

From requirement to evidence

A team that allocates shares must also prove that each one was met. The record of that proof gives the word "traceability" a second meaning. In metrology the word names the chain of calibrations from a result to the SI, metrological traceability (Chapter 15). Device makers use it for a different link, design traceability: the documented path from each requirement to the design output that answers it and to the verification that shows the output meets it. An engineer who keeps the two apart in the design file avoids a confusion, because the two meanings share the word and nothing else.

In the United States the duty to keep that record rests on FDA's quality management system regulation, the QMSR, in force since 2 February 2026. The duty itself is older, because the Quality System Regulation that the QMSR replaced had required design controls, its rules for design and development, since 1997. For the devices that the QMSR's design provision covers, FDA brings design and development under clause 7.3 of ISO 13485, the standard's clause on design.

For the budget, the record must hold each share's owner, the way its trace is kept, and a documented test with an acceptance limit (Chapter 23). In FDA's training slides on clause 7.3, Tartal of FDA's device center asks that design planning "Note roles and responsibilities, traceability methods, and resources needed", which covers the owner and the trace. Verification, Tartal states, must be "performed per documented plans and procedures to ensure outputs meet inputs".

For the monitor, the team writes each share as one line of that record. The maker traces the hysteresis row to Omron's data sheet and to its own incoming check of sample sensors. It traces the converter step to TI's design analysis, and the two residuals to a test at several pressures and temperatures, with an acceptance limit set by the method of the solved example. It traces the reference row to the manometer's certificate, and the firmware team's results from verifying its code under IEC 62304 (Chapter 9) enter the same record. Read downward, the record leads from each requirement to its evidence. Read upward, it tells the team which requirement a failed test threatens and which owner must act. A trace kept both ways is the "bidirectional traceability" that NASA's systems engineering handbook asks of a systems engineer.

Writing the budget as a requirement
  • Write the budget before choosing parts, one row per share, with its size, its owner, its distribution and the test that will show it was met.
  • State the method of combination and why it holds: root-sum-of-squares only for rows with no shared cause, and the worst case beside it as a bound.
  • Keep part of the limit back for the production test's guard band before sharing the rest among the design's owners.
  • Give every interface a number with a unit and a limit, such as the sensor's microvolts for each millimeter of mercury, and put it in both owners' documents.
  • Keep the trace in both directions, from each requirement to its evidence and from each failed test to the requirement it threatens.
  • Write "design traceability" for this record and "metrological traceability" for the chain of calibrations, and never use the bare word for either.
+ What this chapter established
  • An error budget divides one limit, here ±3 mmHg at 150 mmHg, into shares, each with its size, its owner and the test that proves it.
  • Adding Chapter 11's five numbered rows at their limits uses 2.15 mmHg and leaves each of the two firmware residuals about ±0.43 mmHg, while the root-sum-of-squares leaves each about ±1.63 mmHg, so the method chosen sets how good the corrections must be.
  • OIML writes such an allocation into R 76-1 for weighing instruments, giving load cell, indicator and connecting elements 0.7, 0.5 and 0.5 of the limit, whose squares add to 0.99.
  • Design traceability links each requirement to its design output and its verification, which FDA's training on ISO 13485's design clause asks makers to plan and document, and it is a different thing from the metrological traceability of Chapter 15.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.