Fixing the numbers.
+ The questionHow can a number fixed by vote define a kilogram?
A value chosen, not measured
The number 6.626 070 15, divided by 10³⁴, has been exact since 20 May 2019: it is the value of the Planck constant in joule seconds. Written out in full, it has 33 zeros after the decimal point before its first digit. Until that day laboratories measured the constant, and its value carried an uncertainty like any other measured value. Since then it carries none, because the General Conference on Weights and Measures, the CGPM of Chapter 3, decided in November 2018 that the constant has exactly this value, and that decision defines the kilogram.
The Planck constant, written h, comes from quantum physics, where it links the energy carried by light to the light's frequency. Its unit, the joule second, equals a kilogram multiplied by a square meter and divided by a second, as the SI Brochure states. The CGPM had already defined the second through the frequency of a caesium atom and the meter through the speed of light. The kilogram was therefore the one part of that unit not yet defined, so fixing the number for h fixed the kilogram. The definition in the SI Brochure says exactly this: the kilogram is defined by taking the numerical value of h as fixed at that number when h is expressed in joule seconds. An experiment that measured h in joule seconds against a known kilogram can therefore be run the other way: given h, it measures a mass (Chapter 20).
The second and the meter had each been defined by fixing a number. In 1967 the 13th CGPM defined the second as the duration of 9 192 631 770 periods of the radiation from a transition between two energy levels of the caesium-133 atom. In 1983 the 17th CGPM defined the meter as the length of the path that light travels in vacuum in 1/299 792 458 of a second. The CGPM thereby made the speed of light exactly 299 792 458 meters per second, because a meter became whatever length light covers in that fraction of a second. In 2018 the CGPM kept those two numbers unchanged, restating them as fixed values of two constants, the caesium frequency and the speed of light. Until 20 May 2019 the kilogram was the last unit still defined by an object, the cylinder of Chapter 18.
Seven constants, one system
Since 20 May 2019 the whole SI rests on seven constants whose values are fixed, listed below as the first table of the SI Brochure gives them. Each of the seven base units is defined through one or more of them: the second, meter, kilogram, ampere, kelvin, mole and candela. Fixing the values did not make them more accurate: laboratories no longer measure them at all, and the experiments that used to determine them now determine the units instead.
| Constant | Exact value |
|---|---|
| Hyperfine transition frequency of caesium-133 | 9 192 631 770 Hz |
| Speed of light in vacuum | 299 792 458 m/s |
| Planck constant | 6.626 070 15 ÷ 10³⁴ J s |
| Elementary charge | 1.602 176 634 ÷ 10¹⁹ C |
| Boltzmann constant | 1.380 649 ÷ 10²³ J/K |
| Avogadro constant | 6.022 140 76 × 10²³ per mole |
| Luminous efficacy of radiation at 540 THz | 683 lm/W |
The CGPM chose the values so that every unit kept its size on the day the new definitions took effect: the values "have been chosen to be consistent with the earlier definitions in so far as advances in science and knowledge allow", in the words of the SI Brochure. For the kilogram, the CGPM needed a value of h for which the international prototype still had a mass of 1 kg on 20 May 2019, within the uncertainty with which h was then known. Before any vote, the CCM, the committee that advises the CIPM of Chapter 3 on mass, set the conditions that the measurements of h had to meet, and a task group of CODATA calculated the value and its uncertainty. In 2017 CODATA was the Committee on Data for Science and Technology of the International Council for Science.
Conditions set before the vote
Before the CIPM asked CODATA for values to fix, the measurements of h had to meet four conditions, which the CCM set in 2013. At least three independent experiments had to give consistent values of h, so that the value would not rest on one method alone. They had to include work with watt balances, the instruments now called Kibble balances, which weigh a mass against an electrical force, and with silicon spheres, whose atoms are counted (Chapter 20). Each needed a relative standard uncertainty, the standard uncertainty of Chapter 11 expressed as a fraction of the value, no larger than 5 parts in 100 million, which is 50 parts per billion. At least one of them had to reach 2 parts in 100 million, or 20 parts per billion.
On a kilogram, the uncertainties that the CCM allowed amount to small masses, since each part per billion of a kilogram is 1 µg (Chapter 18). A relative standard uncertainty of 50 parts per billion is therefore 50 µg, and 20 parts per billion is 20 µg, less than the 35 µg offset that the BIPM's as-maintained mass unit had built up against the prototype by 2014.
The third condition concerned the masses rather than h: the BIPM's prototypes and reference masses, and the masses used in the experiments, had to be compared as directly as possible with the international prototype. Meeting it took the prototype out of its safe in December 2013, and the comparisons that followed found the 35 µg offset (Chapter 18). The fourth required the procedures for realizing the new kilogram, and for disseminating it to other masses, to be validated in accordance with the principles of the CIPM's arrangement for mutual recognition (Chapter 21). Under that arrangement, national institutes accept one another's results.
The CCM declared the conditions met in 2017, but it warned that the best results did not fully agree. The most recent results with relative standard uncertainties below 50 parts per billion "do not pass the standard chi-squared test of consistency", in the CCM's words. That test shows whether a set of results scatters around its common value by more than the results' stated uncertainties allow. The results each met the conditions on uncertainty, but together they spread a little more widely than those uncertainties predicted, which suggested that at least one budget missed or understated an effect. The CCM still recommended that the CIPM proceed with the planned redefinition at the next CGPM, and the task group that calculated h had to account for the disagreement.
One calculation, and a wider uncertainty
The value of h came from one calculation over all the data, made by CODATA's Task Group on Fundamental Constants, which David Newell of NIST chaired. The group combined every relevant result available by 1 July 2017 in a special least-squares adjustment, a calculation that finds the one set of values for the constants that best fits all the measurements together. It was to supply values of h, the elementary charge, the Boltzmann constant and the Avogadro constant for the CGPM of November 2018.
Three results from Kibble balances entered the calculation for h, and two of them met the CCM's tighter condition of 20 parts per billion. NIST-4, NIST's fourth balance, had a relative standard uncertainty of 13 parts per billion, and the balance at NRC, Canada's national metrology institute, had 9.1 parts per billion. The balance at LNE, France's national metrology institute, had 57 parts per billion, just above the CCM's wider condition of 50. The group left out a fourth result, from NIM, China's national metrology institute, which had measured h with a related instrument, a joule balance. At 240 parts per billion that result failed the task group's self-sensitivity criterion, a test that removes results too uncertain to change the adjusted value.
Silicon spheres, the second method that the first condition required, supplied results inside 50 parts per billion. Fujii and colleagues, at the national metrology institutes of Japan, Italy and Germany, put the relative standard uncertainty of the latest sphere result, of 2017, at about 12 parts per billion (Chapter 20). NRC's balance, NIST-4 and the spheres therefore met the first condition, while LNE's result entered the calculation without counting toward it.
The task group found the data for h "somewhat inconsistent", and it widened the uncertainty to cover the spread among the results that the CCM's test had flagged. The adjustment alone would have given h a relative standard uncertainty of 6 parts per billion, and the task group expanded it to 10 parts per billion. It then chose the exact digits so that, at the moment of the change, the international prototype still had a mass of exactly 1 kg on the new scale. Widening the uncertainty did not change the digits chosen for h, but it set the relative standard uncertainty of the prototype's mass after the change at 10 parts per billion.
The CGPM turned that measured value of h into a definition by a vote, without measuring anything. At its 26th meeting, in Versailles, it voted unanimously for the revision on 16 November 2018. In a resolution of that meeting, the CGPM repealed the definition based on the mass of the international prototype, in force since 1889, and fixed the seven constants with effect from 20 May 2019. From then on, the experiments that had measured h in terms of the prototype measure masses in terms of h.
What changed on 20 May 2019
On 20 May 2019 no mass in any laboratory changed. What changed was which quantity carried the uncertainty: before that day the international prototype was exact and h carried a relative standard uncertainty of 10 parts per billion, and after it h is exact. At the change the prototype's mass was 1 kg with a relative standard uncertainty of 1 part in 100 million, which is 10 parts per billion, or 10 µg. The uncertainty of h had passed to the prototype: the SI Brochure gives that figure as the standard uncertainty of the combined best estimates of h at that time. From that day the prototype's mass is a quantity to be determined by experiment, whose latest value Chapter 21 gives, and the cylinder that defined the kilogram for 130 years is a mass like any other, measured and stated with an uncertainty.
For everyday measurement the change was far too small to see, because an uncertainty of 10 parts per billion at the top of the chain is tiny beside the uncertainties added further down. A pressure balance generates a pressure equal to a downward force divided by the area of its piston, and its masses under gravity supply that force, as NIST describes it (Chapter 15). A relative change in those masses therefore changes the pressure by the same fraction. At a cuff pressure of 150 mmHg, 10 parts per billion is 0.0000015 mmHg. The channel's combined standard uncertainty at that pressure is 0.70 mmHg (Chapter 11), about 470,000 times as large. Electrical laboratories, by contrast, had a shift to apply on that day, in the conventional values of the electrical units, for reasons that Chapter 20 explains.
As of October 2026, the seven constants and the kilogram's definition stand as the 26th CGPM fixed them. The 28th CGPM met in Versailles from 13 to 15 October 2026 and adopted seven resolutions, none of which changes the SI or the kilogram.
An exact constant does not make the kilogram exact. The definition fixes the number for h, but a laboratory still has to realize a kilogram from it with an instrument, and every realization carries an uncertainty budget of its own. Even the best balances in the adjustment carried an uncertainty. Until the day of the change the international prototype was the one mass in the world without an uncertainty, and on that day it received one. What the redefinition removed was the dependence of the whole system on a single object that nobody could check, not the uncertainty of measuring a mass.
A kilogram holds a billion micrograms, so a relative standard uncertainty of 1 part per billion on a kilogram is 1 µg (Chapter 18).
- Before the change, h carried a relative standard uncertainty of 10 parts per billion, after the task group expanded it from 6, a factor of 10 ÷ 6 = 1.67. The prototype was exact.
- After the change, h is exact, and at the change the prototype's mass was 1 kg with a relative standard uncertainty of 10 parts per billion, which on a kilogram is 10 µg.
- The same relative uncertainty at the cuff: a pressure balance's pressure changes by the same fraction as its masses, so 10 parts per billion of 150 mmHg is 150 × 10 ÷ 10⁹ = 0.0000015 mmHg.
- The cuff channel's combined standard uncertainty at 150 mmHg is 0.70 mmHg (Chapter 11), and 0.70 ÷ 0.0000015 is about 470,000.
On the day of the change the prototype's mass was given a standard uncertainty of 10 µg, and h, which had carried the same relative uncertainty, became exact. Carried down through a pressure balance, that uncertainty is 0.0000015 mmHg at the monitor's cuff, and the channel's own 0.70 mmHg is about 470,000 times as large, so the redefinition changed nothing that a calibration of the monitor could detect.
- Cite the SI by its current document when a file states the units its measurements use: the SI Brochure, 9th edition (2019), version 4.01 of June 2026.
- Leave the definition out of the monitor's uncertainty budget. The definition contributes 0.0000015 mmHg, and the reference manometer's row of 0.46 mmHg is about 310,000 times as large (Chapter 11).
- When a reviewer asks how the cuff reading traces to the SI, name each link up to the national institute and its key comparisons, the tests in which national institutes show whether their results agree (Chapter 21). Then end the chain with one line: the kilogram, defined since 20 May 2019 by a fixed value of the Planck constant.
- Expect the definition to stay fixed while realizations improve. Changes at the top now come as revised consensus values for the kilogram, which the CCM computes from the key comparisons, and as new comparison results (Chapter 21), not as new definitions.
- Since 20 May 2019 the Planck constant has the exact value 6.626 070 15 ÷ 10³⁴ joule seconds, and because the meter and the second were already fixed through two other constants, that number defines the kilogram.
- Seven constants with exact values now define the whole SI, as fixed numbers had already defined the second since 1967 and the meter since 1983.
- The CCM set conditions in 2013, declared them met in 2017 although the best results failed a test of consistency, and CODATA's task group widened h's relative standard uncertainty from 6 to 10 parts per billion before the CGPM's vote of 16 November 2018.
- The vote made h exact and gave the international prototype's mass a standard uncertainty of 10 µg at the change. Carried down to the cuff, the uncertainty of 2019 is so small that the monitor's own uncertainty at 150 mmHg is about 470,000 times as large.