+ Part IX · Quality and regulation, as of November 2026
What the law asks.
The BIPM's Member States agree on the units together, but each state writes its own rules for the instruments its people buy and the devices its doctors use. The units are agreed in the General Conference on Weights and Measures, whose 28th meeting, at Versailles from 13 to 15 October 2026, adopted seven resolutions, none of which changed the SI (Chapter 19). Two of the resolutions concern how measurements cross borders. One gives States that take no part in the BIPM's work a status as Observers, and the CIPM's President expects that up to 85 States could take it. The other encourages the CIPM to extend the SI's reference point into a digital framework and invites national institutes to advance digital certificates. Part IX maps those rules for the monitor in the United States, the European Union and India as of November 2026. For each, it explains how the system works before it cites the clauses that apply to the monitor, and it says which texts are still drafts.
Measurement across borders.
+ The questionWho governs the chain across borders, and how does a measurement made in one country count in another?
A state enforces some measurements
A state can make a measurement a matter of law, and that use of its power is called legal metrology. OIML, the intergovernmental body that writes model rules for measuring instruments, defines it in its 2022 vocabulary as the "practice and process of applying statutory and regulatory structure and enforcement to metrology", and a note to the definition adds that legal metrology's scope "may be different from country to country". The state names in law the instruments that must meet a limit, and an official, or a body that the state designates, checks each instrument and marks it. Institutes and laboratories build the chain of Parts IV and V from comparisons that they agree to recognize, and the state adds its power to that agreement. The state can also take back a task it has delegated: in July 2026 Japan withdrew A&D's designation as a body that verifies scales on the state's behalf (Part VIII).
A state checks an instrument's design before the instrument is sold: an authority tests samples of it, a step that OIML calls type evaluation and India's law calls approval of the model. After that, an official or a designated body checks each instrument of the approved type, and OIML calls that check verification, a "conformity assessment procedure (other than type evaluation) which results in the affixing of a verification mark and/or issuing of a verification certificate". The mark records that the instrument met the limit when it was checked. In legal metrology the word therefore has a third sense, apart from verification in the VIM's sense (Chapter 14) and from the design verification of a device's development (Chapters 23 and 25).
A measurement made in one country counts in another through three arrangements, each for a different link of the chain: the national institutes, the accredited laboratories and the instruments themselves. At the top, national institutes recognize each other's calibration and measurement capabilities, or CMCs (Chapter 16), under the CIPM MRA (Chapter 21). In May 2025 the arrangement had 258 participants, and its key comparison database, the KCDB, listed 26,135 CMCs in the report that the BIPM gave in June 2025.
One level down, an accredited laboratory's certificate can count in other countries because the bodies that accredit laboratories assess each other, under the Global ACI arrangement, which took over ILAC's on 1 January 2026 (Chapter 16). Global ACI lists two US accreditation bodies, A2LA and ANAB, and NABL, India's accreditation body, states that it is a signatory. A certificate for a maker's reference manometer from a laboratory that one of these bodies accredits is therefore recognized where the others work.
For the instrument itself, the route across borders is an OIML certificate, and it counts only in a state that accepts it. OIML describes its Certification System, OIML-CS, as "a system for issuing, registering and using OIML certificates and their associated OIML type evaluation/test reports" for types of measuring instruments. The system holds 37 OIML Recommendations, OIML's model rules for particular kinds of instrument, as NIST's office of weights and measures reported in 2026. On 14 April 2026 OIML and Global ACI signed a memorandum under which Global ACI endorses part of the system, which OIML names Scheme A, as one of its own schemes. A certificate still binds no state, because OIML writes model rules and only a state makes them law. Under India's Act, for one, the authority may approve a model already approved abroad without a test, or with whatever test it thinks fit.
OIML's model rules for sphygmomanometers
An automated monitor built to OIML's model rule must read the cuff pressure within ±0.4 kPa (±3 mmHg) or ±2 % of the reading, whichever is greater. The limit holds at any point of the measuring range, with the pressure rising or falling, between 10 and 40 °C and at 15 to 85 % relative humidity. OIML sets it in R 149, one of the two Recommendations for blood-pressure instruments that it published in 2020: R 148 is for non-automated sphygmomanometers, read with a stethoscope or by another manual method, and R 149 for automated ones. Each comes in three parts, for the requirements, the test procedures and the format of the test report, so R 149-1 holds the requirements. The two replaced R 16-1 and R 16-2 of 2002, which older studies in Chapters 2, 10 and 13 refer to.
In R 149-1, OIML also holds the monitor to that limit through its working life. The monitor must zero itself at least when it is switched on, and zero again, or switch itself off, when its zero drifts by one scale interval, 1 mmHg. Under electromagnetic disturbance it must stay within the limit, or signal an abnormality and resume normal operation within 30 s. After 10,000 simulated measurement cycles its reading may have changed by no more than 3 mmHg. Verification covers at least the cuff-pressure limit, the repeatability and the air-leakage test, and may be carried out in any room climate between 10 and 40 °C without a climatic chamber. OIML sets no wider limit in R 149-1 for a monitor in service, so the monitor must meet the same limit at every verification as when it was new.
OIML holds the non-automated instruments to the same figures in R 148-1, with one difference in temperature. An aneroid or electro-mechanical gauge must read within ±3 mmHg between 15 and 25 °C, and within ±3 mmHg or ±2 % of the reading, whichever is greater, between 10 and 40 °C, a wider band that does not apply to mercury manometers. A digital display must step in 0.1 kPa, or 1 mmHg, and at least three samples of a new type must be tested for its approval.
OIML also sets limits in R 149-1 for the automated monitor's estimate: a mean error of no more than ±5 mmHg and a standard deviation of no more than 8 mmHg. A maker shows that the estimate meets them by a clinical investigation, not on the bench (Chapter 24).
The three checks of verification under R 149-1 all test the reading, while the estimate's limits are met once, by the clinical investigation behind the approved type. A monitor that carries a verification mark has therefore been shown to read the cuff pressure within its limit, and the mark says nothing about whether its algorithm still estimates blood pressure as the validated type did. PTB made the same point in its 2021 proposal to OIML for a guide on simulators (Chapter 1). Three countries intend to apply the guide, the Czech Republic, Portugal and Germany, as the proposal records.
India writes the limits into its rules
In India a maker or importer must obtain approval of the model before manufacturing or importing a weight or measure, and a user must have each one verified before using it in a transaction. Both duties come from the Legal Metrology Act 2009, in force since 1 April 2011, whose purpose, in its long title, is "to establish and enforce standards of weights and measures". The Union, India's central government, sets the standards, the Union and the States share their enforcement, and each State frames its own enforcement rules. The General Rules of 2011, made under the Act, cover about 40 types of instrument, the sphygmomanometer and the clinical thermometer among them, by the account of the Department of Consumer Affairs, DoCA. Two Parts of the rules' Eighth Schedule hold the sphygmomanometers: Part VII-A the non-automated instruments and Part VII-B the automated monitors.
India's limits for the automated monitor are the same as OIML's. Under Part VII-B, the maximum permissible error of the cuff pressure is "±0.4 kPa (±3 mmHg) or ±2 % of the reading, whichever is greater", from 10 °C to 40 °C and for rising and for falling pressure. The monitor must zero itself at least at power-on, and zero again or switch off when its zero drifts by "one scale interval (0.1 kPa or 1 mmHg) or more". The estimate's limits are in the same Part, as limits for a clinical investigation, and for the reading and the estimate alike the figures are those of R 149-1. DoCA substituted this text for the Part on 28 July 2025, in force from publication.
The rule's cuff-pressure test, Test 1, follows OIML's procedure of Chapter 14. The tester replaces the cuff with a 500 mL metal vessel and compares the monitor with a reference manometer whose maximum permissible error is within ±0.1 kPa (±0.8 mmHg), stepping the pressure by no more than 6.7 kPa (50 mmHg) at a time. For model approval the authority tests at least three samples of each new model, and an officer or an approved test center may then verify each monitor in any climate from 10 to 40 °C, without a climatic chamber. Seals must guard the enclosure and the measuring parts against tampering, and the maker must mark each monitor with its model approval number, serial number, year of manufacture and country of origin.
India holds the non-automated instruments to the figures of R 148-1, in Part VII-A, which DoCA replaced on 7 January 2026, in force from publication. Such an instrument must read the cuff pressure within ±3 mmHg between 15 and 25 °C, and within ±3 mmHg or ±2 % of the reading from 10 to 40 °C, a band that excludes mercury manometers. The reference manometer is held to ±0.8 mmHg, the steps to 50 mmHg and model approval to three samples. These figures come from a reproduction of the text by TaxGuru, a website that reprints Indian notifications.
No Indian rule is known to name a period of re-verification for the monitor. Neither Part sets one, and rule 27 of the General Rules sets the periods for the whole Act. On 18 December 2025 DoCA substituted the rule's first clause, which now gives twenty-four months for weights, capacity measures, length measures, tapes, beam scales, counter machines and fuel dispensers, a list without the sphygmomanometer. By the accounts of law firms and a State regulator, the second clause gives sixty months for storage tanks and the third twelve months for every other weight or measure. Neither the new clause nor these accounts names sphygmomanometers, so twelve months under the third, catch-all clause is an inference from its wording, not a stated rule.
By DoCA's account, officers of the States, private test centers and a layer of regional laboratories share the checking. Privately run test centers, which the Central Government approves under the Act, may verify sphygmomanometers and clinical thermometers among other instruments. Five Regional Reference Standard Laboratories, at Ahmedabad, Bangalore, Bhubaneswar, Faridabad and Guwahati, link CSIR-NPL, the national institute, to the States' weights and measures laboratories, and two more are being set up at Nagpur and Varanasi. All five are accredited by NABL, so in DoCA's account the standards with which a State verifies instruments in use trace to CSIR-NPL through a regional laboratory.
No legal-metrology rule found for the monitor in the US or the EU
In the United States no weights-and-measures code is known to apply to the monitor, and its accuracy falls instead under FDA's device rules (Chapter 31). NIST publishes Handbook 44, which provides "the specifications, tolerances and other technical requirements" for commercial weighing and measuring devices. Its current edition, of 2026, is the one that the National Conference on Weights and Measures adopted. The 2022 edition lists codes from 1.10 to 5.60, for scales, weights, liquid and gas meters, timing devices and grain moisture meters among others, and none for sphygmomanometers, blood-pressure instruments, clinical thermometers or medical devices.
In the European Union, too, device law and not the law on measuring instruments sets the requirements for the monitor's accuracy. Under the Medical Devices Regulation, the maker of a device with a measuring function must state the device's limits of accuracy (Chapter 31). The EU regulates measuring instruments through two directives, the Measuring Instruments Directive, the MID, and the directive on non-automatic weighing instruments, the NAWI directive, and neither is known to place a blood-pressure device under it. The European Parliament and the Council amended the MID on 11 March 2026 without changing that: the amendment, in force from 9 April 2026, concerns charging points for electric vehicles, dispensers of compressed gas, and meters for electricity, gas and thermal energy. The NAWI directive has no known amending act.
The rules side by side, with their dates
| Jurisdiction | Rule for the monitor's reading | Limit | In force | Who checks |
|---|---|---|---|---|
| OIML (model rule) | R 149-1, clause 5.1 | ±3 mmHg, or ±2 % of the reading where greater | 2020 | states that adopt it |
| India, automated | Eighth Schedule, Part VII-B | as R 149-1 | 28 July 2025 | model approval; State officers and approved test centers |
| India, non-automated | Eighth Schedule, Part VII-A | ±3 mmHg at 15–25 °C; the 2 % branch from 10 to 40 °C | 7 January 2026 | as above |
| United States | none found in Handbook 44 | — | — | FDA, under device law (Chapter 31) |
| European Union | none found in the MID or the NAWI directive | — | — | under the Medical Devices Regulation (Chapter 31) |
Each of India's limits is printed twice, in kilopascals with millimeters of mercury in brackets, and the two units do not always give the same number. That difference explains a shortfall in Chapter 12, where the engineer set OIML's reference manometer against the monitor's limit at 3.75 to one, short of the four to one.
One millimeter of mercury is 133.322 387 415 Pa (Chapter 3), so 1 kPa is 1,000 ÷ 133.322 = 7.5006 mmHg. Each limit of Part VII-B converts from one unit to the other with that factor:
- Cuff-pressure limit: 0.4 kPa × 7.5006 = 3.0002 mmHg, printed as 3 mmHg, the same to within 0.01 %.
- Reference manometer: 0.1 kPa × 7.5006 = 0.7501 mmHg, printed as 0.8 mmHg, which is about 0.05 mmHg, or 7 %, more.
- Step between test points: 50 mmHg ÷ 7.5006 = 6.666 kPa, printed as 6.7 kPa; here the round figure is the one in millimeters of mercury.
- The estimate's mean error: 5 mmHg ÷ 7.5006 = 0.667 kPa, printed as 0.7 kPa, also round in millimeters of mercury.
The ratio of the cuff limit to the reference limit is 0.4 ÷ 0.1 = 4.0 in kilopascals, and 3 ÷ 0.8 = 3.75 in millimeters of mercury. Four to one would need a reference within 3 ÷ 4 = 0.75 mmHg (Chapter 12), and 0.75 mmHg is 0.1 kPa. In kilopascals the reference's limit is exactly a quarter of the monitor's, and the 3.75 comes from rounding 0.75 mmHg up to 0.8 mmHg. A laboratory that holds its reference within ±0.75 mmHg meets the limit in both units.
- Apply for model approval before the first monitor is made in or imported into India, and plan at least three samples of each new model for the tests of Part VII-B.
- Specify the production reference manometer to ±0.75 mmHg, which meets the ±0.1 kPa limit in either unit.
- Test the cuff pressure rising and falling, in steps of no more than 50 mmHg, across 10 to 40 °C, and record at each step the monitor's reading minus the reference's.
- Put the model approval number, serial number, year of manufacture and country of origin on the marking, and design the seals so that no one can reach the measuring parts without breaking one.
- Ask the State's legal-metrology department which re-verification period it applies to sphygmomanometers, because rule 27 is not known to name one for them.
- For the United States and the EU, plan the accuracy evidence under device law (Chapter 31).
- OIML defines legal metrology as applying statutory and regulatory structure and enforcement to metrology, and in its sense a verification ends in a mark or a certificate on each instrument of an approved type.
- Measurements cross borders through the CIPM MRA for national institutes, the Global ACI arrangement for accredited laboratories and OIML's certificates for instruments, which count only in a state that accepts them.
- India's Part VII-B, substituted on 28 July 2025, holds an automated monitor's cuff pressure to OIML's ±3 mmHg or 2 % of the reading from 10 to 40 °C, with a reference held to ±0.1 kPa, exactly a quarter of the monitor's ±0.4 kPa, and printed as ±0.8 mmHg.
- No Indian rule is known to name the monitor's re-verification period, and no legal-metrology rule is known to apply to the monitor in the United States or the EU, where device law sets the requirements for its accuracy.
Let's Talk