06 — Conditioning

Amplify, filter, refer.

+ The questionWhat does the electronics between the sensor and the converter do to the number?

Every count is a fraction of the reference

The monitor's converter turns the amplified signal into a number by comparing it with a steady voltage, so each count it reports is a fixed fraction of that voltage. The part that holds the comparison voltage is the voltage reference, and in Texas Instruments' reference design for a blood-pressure monitor it is the ATL431LI, which holds 2.5 V. The 12-bit converter divides the 2.5 V into 4,096 steps, twelve doublings (Chapter 1), and 2.5 V divided by 4,096 is about 0.61 mV a step (Chapter 8). The number on the display is therefore the signal's voltage divided by the reference's voltage, and the reference acts as the instrument's internal standard: whatever it does, every reading follows in proportion.

In TI's design the bridge current and the reference come from separate sources, so each must hold its value on its own between calibrations. TI drives the bridge with a constant current of 100 µA, set by holding 0.249 V across a 2.49 kΩ resistor whose value may differ by up to 0.1 %, its tolerance. The sensor's output follows the current and the converter's count follows the 2.5 V reference, so a drift in the current moves the output, and a drift in the reference moves the scale against which the output is counted.

The amplifier measures a difference

The bridge's output is a difference of a few millivolts between two terminals that both sit about 1 V above the bridge's lower end (Chapter 4). An instrumentation amplifier, or in-amp, is built to measure exactly such a difference and to multiply it by a set factor, its gain. It is a precision gain block that must reject volts of common-mode signal, the voltage that both inputs share, while it amplifies signals of microvolts, in Analog Devices' description. TI's design uses the INA350 at a gain of 50, so the 16.8 mV that a cuff pressure of 150 mmHg adds at the bridge becomes 0.84 V at the amplifier's output.

How well an in-amp ignores the shared voltage is its common-mode rejection ratio, or CMRR, stated in the decibels of Chapter 5. The INA350 reaches at least 85 dB and typically 95 dB by TI's figures. At 20 dB for each tenfold step, 10 dB is a factor of the square root of 10, 3.16, and 5 dB the square root of 3.16, 1.78, so 85 dB is 10,000 times 1.78, about 17,800. A change of 1 V in the shared voltage therefore looks like 1 V divided by 17,800, about 56 µV, between the inputs, which at the sensor's 112 µV for each mmHg is about 0.5 mmHg. A steady common-mode voltage gives a steady error, which the zero reading taken before each measurement removes (Chapters 2 and 17). A changing one passes into the reading as noise in the same proportion.

The amplifier adds two errors of the kinds the sensor also has. A gain error multiplies every reading by a factor slightly off the nominal one. For the INA350 it is at most 0.6 %, and typically 0.082 % at a gain of 50, by TI's figures. At 150 mmHg these come to 0.6 % of 150, 0.9 mmHg, and about 0.12 mmHg. The INA350's offset adds the same voltage to every reading, at most 1.2 mV and typically 0.2 mV. Referred to the input (Chapter 5), 1.2 mV divided by the sensor's 112 µV for each mmHg is about 11 mmHg, and 0.2 mV is about 1.8 mmHg. Both look large beside a ±3 mmHg limit, but each is fixed for a given amplifier at a given temperature, so the zero reading removes the offset, and the span coefficient that calibration stores for each unit removes the gain error (Chapter 9).

Calibration does not remove the change of either error with temperature, its temperature coefficient. The INA350's offset drifts by 0.37 µV for each degree on average, the mean of 140 amplifiers whose drift TI plots in its data sheet. Across the 30 °C from 10 to 40 °C in which a monitor must meet its limit (Chapter 1), that is 0.37 µV times 30, about 11 µV, or about 0.1 mmHg. The zero reading takes out the offset's drift as long as the temperature holds during the measurement, but the gain's drift stays in the reading until the next calibration. Such residual drifts, and the voltage reference's drift, the drift of the scale itself, still belong in the monitor's uncertainty.

One sensor, two paths

No single gain serves both parts of the oscillometric signal of Chapter 4, which the amplifier's output carries together. At 150 mmHg the output holds about 0.84 V of cuff pressure and oscillations of 5 to 15 mV from peak to peak, the 100 to 300 µV at the amplifier's input multiplied by 50. One mmHg of oscillation, 112 µV at the input, becomes 5.6 mV, so on the converter's steps of 0.61 mV the shape of a 1 mmHg oscillation would rest on only about 9 counts. More gain on the whole signal would not help. TI sets 70 dB as the oscillations' target, three and a half tenfold steps, or 1,000 times 3.16, about 3,200 times, and at that gain the 16.8 mV of cuff pressure at 150 mmHg would become about 53 V, about twenty times the converter's 2.5 V.

Designers therefore feed two paths from one sensor. A filter passes some frequencies and weakens others, and its corner frequency marks where the weakening begins. In Chua and Siew's 2005 meter for Freescale, the cuff pressure, which changes at under 0.04 Hz, goes straight to the converter. The oscillation, at about 1 Hz, passes a two-stage high-pass filter, which weakens the slow cuff pressure and keeps the faster oscillation, and then an amplifier with a gain of 150. A band-pass filter also weakens what is faster than the oscillation. TI's oscillation path is a band-pass filter with corner frequencies at 1.7 Hz and at 5.9 Hz.

A filter also weakens the signal near its corner frequency, so the gain that follows must allow for the loss. Freescale's filter weakens the 1 Hz oscillation by 10 dB, a factor of 3.16. The meter's integrated sensor, which carries its own amplifier on the chip, gives 12 to 36 mV for 1 to 3 mmHg, and divided by 3.16 these leave the filter as 3.8 to 11.4 mV. The gain of 150 then brings them to about 0.57 to 1.7 V.

TI's design splits its gain between two amplifiers. The in-amp's gain of 50 is 100 halved, 40 dB less about 6 dB, or 34 dB, and the amplifier built into the microcontroller adds about 32 dB after it, so the two stated gains make about 66 dB. TI does not state the gain of the rest of the path, and it sets a total target of 70 dB or more across 0.5 to 7 Hz. That band reaches beyond the filter's corner frequencies at 1.7 and 5.9 Hz, and the oscillations reach below the lower one: a pulse rate of 40 to 180 beats a minute, the range A&D states for one of its monitors, divided by 60 seconds is 0.67 to 3 Hz. Each corner of the filter, set by one resistor and one capacitor, weakens frequencies gradually beyond it, so the filter weakens a slow oscillation without removing it, and the gain must allow for that loss as well.

Taken as a gain of 70 dB, about 3,200 times, the target brings the 100 to 300 µV oscillations to about 0.32 to 0.95 V, a large share of the converter's 2.5 V. One mmHg of oscillation, 112 µV at the input, then becomes about 0.35 V, or about 580 counts of 0.61 mV instead of 9.

On the cuff-pressure path a monitor can place a low-pass filter, which keeps slow signals and weakens fast ones, ahead of the converter. There it limits the noise bandwidth of Chapter 5 and guards against aliasing, the false slow signal that a converter makes from a fast one it samples too slowly (Chapter 7). TI does not describe the cuff-pressure path separately, so Chapter 11 works its budget on the assumption that the cuff pressure reaches the converter at the in-amp's gain of 50. One sensor feeds two paths, and each path is shaped for one of the two numbers the monitor reports: the cuff-pressure path for the reading, and the oscillation path for the estimate that the firmware derives (Chapter 9).

13 — One sensor, two paths
2SMPP-03 bridge ×50 IN-AMP INA350 gain 50 (34 dB) cuff pressure 0.84 V at 150 mmHg oscillation 5 to 15 mV peak to peak 12-BIT CONVERTER, 2.5 V reference 1 mmHg = about 9 counts the reading cuff pressure to the algorithm (Chapter 9) band-pass filter 1.7 to 5.9 Hz further gain, total target 70 dB or more over 0.5 to 7 Hz 0.32 to 0.95 V peak to peak same 12-bit converter 1 mmHg = about 580 counts THE READING PATH cuff pressure at the in-amp's gain THE ESTIMATE PATH filtered, then amplified 0.84 V + ripple ripple only removes the steady 0.84 V so the oscillation can be amplified about 3,200 times without exceeding 2.5 V; amplified whole, 150 mmHg would need 53 V TI TIDUF53 (2023). The cuff path's own circuit is drawn as Chapter 11's budget takes it, at the in-amp's gain. 70 dB read as a total gain target. Counts from 2.5 V ÷ 4,096 = 0.61 mV.
The signal in the chainThe algorithm's estimateReadingFocal detail
Plate 13 — The steady cuff pressure and the small oscillation need different treatment. A filter takes the steady part off the second path, so the oscillation can be amplified until each millimeter of mercury spans hundreds of counts.

A shared source cancels

A design can avoid depending on a separate reference by letting one source serve both the sensor and the converter. A measurement made this way is ratiometric: the converter compares the sensor's output with a reference derived from the sensor's own excitation, so a change in the excitation moves both sides of the comparison by the same fraction and the count stays the same. In a precision weigh scale that Analog Devices describes in a circuit note, the converter's reference inputs take their voltage from the load cell's sense lines, the pair of wires that carry the excitation voltage back from the cell. That connection creates "a ratiometric configuration that is immune to low frequency changes in the power supply excitation voltage", in Analog Devices' words. By Analog Devices' account the design then needs no precision reference.

Some blood-pressure designs and their sensors use the same principle. Chua and Siew's meter feeds its sensor and the pair of resistors that sets its converter's reference from one 5 V supply, so supply variations do not affect its accuracy, by their account. The sensitivity of TDK's C32/2 die and the output of NXP's MPX2050 are ratiometric to the supply, in their makers' descriptions: each scales with the supply, so a design that also derives its reference from that supply can cancel the change. TI's monitor design, with its separate sources, is not built this way.

A ratiometric design still needs a part that holds its value, but a different part. In a resistance-thermometer circuit that TI describes in a guide, the current that flows through the sensor also flows through a reference resistor, whose voltage serves as the converter's reference. Each voltage is the same current times a resistance, so the current, with its accuracy and drift, cancels from their ratio, and the result depends only on the two resistances, as TI states. The reference resistor's accuracy and drift do not cancel, so the stable part of that design is a resistor instead of a voltage.

14 — A separate reference, and a shared one
TI'S DESIGN: SEPARATE SOURCES 0.249 V across 2.49 kΩ, 0.1 % → 100 µA converter ATL431LI, 2.5 V current +0.1 % → reads 150.15 mmHg reference +0.1 % → reads 149.85 mmHg RATIOMETRIC: ONE SOURCE 5 V supply converter divider: reference for the converter supply +0.1 % → both sides rise → reads 150.00 mmHg one source on both sides of the comparison TI's RTD guide: current through the sensor and a reference resistor; the current cancels, the resistor must hold a precision reference (ADI ADR4550B): at most 2 ppm per °C, 30 °C → 0.009 mmHg; typically 25 ppm in 1,000 h → 0.004 mmHg TI TIDUF53 (2023); Freescale AN1571 (2005); ADI CN-0102 (2012); TI SBAA275 (2023); ADR4550 data sheet Rev. G. TI's design is not ratiometric.
References and standardsThe signal in the chainNoise and errorReadingFocal detail
Plate 14 — When the sensor and the converter take their scale from separate sources, a change in either moves the reading. When one source serves both, its change cancels, and the stable part moves elsewhere in the circuit.
Solved example: a 0.1 % change at 150 mmHg, with a separate reference and with a shared one

TI's design, in which the excitation and the reference are separate:

  • TI sets the bridge current as 0.249 V divided by the resistor's 2,490 Ω, which is 100 µA. A resistor 0.1 % low, the edge of its tolerance, therefore makes the current 0.1 % high, so the sensor's output and the reading rise by 0.1 % of 150 mmHg, 0.15 mmHg: 150 mmHg reads 150.15 mmHg until calibration sets the span (Chapter 9).
  • If the 2.5 V reference rose by 0.1 % instead, each count would stand for 0.1 % more voltage, and the same pressure would read 0.1 % low, about 149.85 mmHg.

A ratiometric design, such as Chua and Siew's meter on one 5 V supply: a 0.1 % rise in the supply raises the sensor's output and the converter's reference together, the ratio between them stays the same, and the reading stays at 150 mmHg.

The B grade of Analog Devices' ADR4550, a 5 V precision reference, shows by its maker's figures how far such a part drifts, although it is not the part in TI's design. Its temperature coefficient is at most 2 ppm, parts per million, for each degree, so a change of 30 °C moves its voltage by 30 times 2 ppm, 60 ppm, and moves 150 mmHg by 150 times 60 millionths, about 0.009 mmHg. Its long-term drift over the first 1,000 hours is typically 25 ppm, about 0.004 mmHg. Each drift is no more than a hundredth of a ±3 mmHg limit, so a precision reference keeps the scale steady with a wide margin.

The current that measures also heats

The excitation that makes a sensor's signal also puts heat into the sensor, and the warming is an error of its own, called self-heating. For this reason most excitation currents for resistance thermometers should stay below 1 mA, by TI's guide to them. The temperature rise is the power the current dissipates divided by a dissipation constant, the power that warms the element by one degree. By TI's figures that constant ranges from about 2.5 mW for each degree for a small thin-film element to 65 mW for a large wire-wound one, a milliwatt being a thousandth of a watt.

The power is the current squared times the resistance, so a current of 1 mA in a 100 Ω Pt100 dissipates 1 mA times 1 mA times 100 Ω, or 0.1 mW. That warms the small element by 0.1 divided by 2.5, about 0.04 °C. The Pt100's resistance rises by about 0.39 Ω for each degree (Chapter 4), so the small element's warming adds 0.04 times 0.39 Ω, about 0.016 Ω, and the element reads about 0.04 °C high, an error the sensor makes by being measured.

The monitor's pressure sensor heats itself in the same way. The 2SMPP-03 consumes 0.2 mW by Omron's rating, exactly the power of 100 µA flowing through its 20 kΩ bridge, 100 µA times 100 µA times 20,000 Ω, and twice the Pt100's 0.1 mW. Omron states no temperature rise for the sensor. Because the current is constant, the heating stays nearly the same at every pressure, so it shifts the sensor's temperature once instead of moving with the reading. The temperature effects of Chapter 4 then act on that shifted temperature, and a zero reading taken under the same excitation as the measurement carries the same heating.

Conditioning a bridge sensor's channel
  • Decide early whether the design is ratiometric. If the excitation and the converter's reference come from separate sources, budget the drift of both.
  • Take every amplifier error at the gain the design uses, and convert it into the display's unit before comparing it with the limit.
  • Remove fixed offset and gain errors by zeroing and calibration, and budget what changes with temperature after calibration.
  • Give the oscillation path its own filter and gain, and count the filter's loss near its corner frequency.
  • Check self-heating at the excitation the design uses, and take the zero reading under that same excitation.
+ What this chapter established
  • The converter counts in fractions of its reference voltage, so in a design whose excitation and reference are separate, such as TI's, the reference must hold its value between calibrations.
  • An instrumentation amplifier measures the bridge's millivolt difference while rejecting the voltage both outputs share, and calibration and zeroing remove its fixed gain and offset errors, but calibration does not remove their drift with temperature.
  • One sensor feeds two paths: the cuff pressure reaches the converter with little gain, and the oscillations are filtered off the cuff pressure and amplified for the algorithm toward TI's target of 70 dB or more, about 3,200 times.
  • A ratiometric design feeds the sensor and the converter's reference from one source, so a change in that source cancels, but some part, such as a reference resistor, must still stay stable.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.