Devices that measure.
+ The questionHow do medical-device rules in the US, the EU and India treat a device that measures, and the equipment that tests it?
A Class I recall for readings that were too low
Some of Abbott's FreeStyle Libre 3 and 3 Plus sensors, the glucose sensors of its continuous monitoring systems, could read glucose too low, and on 13 January 2026 FDA classified their recall as Class I, its most serious type of recall. FDA gave the reason in one sentence: "Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results." Abbott had begun the recall on 24 November 2025, and FDA recorded it as one recall with a record for each of the four sensor models. In its own announcement Abbott traced the fault to "one production line among several" and stated that its internal testing had found it. By 7 January 2026 Abbott had reported 860 serious injuries and seven deaths, as FDA stated in a notice of February 2026.
The recall is a measurement case under device law. In the words of FDA's product classification, the sensor is factory calibrated: the person who wears it performs no calibration step, so the factory's calibration is the only one between the sensor's signal and the number on the screen. The faulty readings erred in one direction, low. The displayed number decides what a patient does next, so an error in that number alone can lead over time to wrong treatment decisions, such as eating excess carbohydrates or skipping or delaying insulin, as FDA's notice describes them. A maker traces a fault of this kind back through every unit it reached (Chapter 26).
FDA's records give a quantity for each of the four sensor models:
- The Libre 3 sensor: 1,060,723.
- The largest Libre 3 Plus model: 1,663,241.
- Two further models: 258,913 and 8,062.
Adding them in turn, 1,060,723 + 1,663,241 = 2,723,964; adding 258,913 gives 2,982,877; adding 8,062 gives 2,990,939, about 3.0 million sensors. The largest record holds 1,663,241 ÷ 2,990,939 of them, or 56 %. The records list distribution to 36 US states and 19 other countries. The recall's start, 24 November 2025, and FDA's classification, 13 January 2026, lie 6 + 31 + 13 = 50 days apart. In its own announcement Abbott put the US sensors from the affected line at about 3 million, about half of them already expired or used. The two counts are close in size but need not cover the same sensors. Each factory-calibrated sensor carries its line's calibration into use, so the count measures how many units one production fault reached before it was found.
The United States: a class II device and the standards FDA recognizes
In the United States the monitor is a class II device, and its maker reaches the market through a 510(k). In that premarket submission the maker shows that its device is "substantially equivalent" to one already legally marketed, and FDA clears a 510(k) rather than approving it. FDA groups devices of one type under a three-letter product code, and it lists the monitor under code DXN, "Noninvasive blood pressure measurement system", in class II. One maker whose monitor FDA cleared on 5 December 2023 states in its 510(k) summary: "The maximum error of the pressure measurement in the cuff should be ±0.4kPa(±3mmHg) at any measuring point" (Chapter 13). The figure is the maker's statement of the limit it tested against, and the maker names IEC 80601-2-30 and ISO 81060-2:2018 for its performance tests.
FDA names the standards it accepts in its list of recognized consensus standards, and a maker may support a premarket submission with a declaration that its device conforms to one of them. Such a declaration is a different document from the EU declaration of conformity of Chapter 29, which the manufacturer issues under the EU's device regulation. Since 6 July 2020 FDA has recognized the 2018 edition of IEC 80601-2-30, the particular standard for automated non-invasive sphygmomanometers. Since 23 December 2024 it has recognized ISO 81060-2, the standard for clinical validation of the estimate (Chapter 24), in its 2018 edition with amendments 1 and 2. FDA will accept declarations to its earlier recognition, of that edition with amendment 1 only, until 20 December 2026, so a maker that declares conformity after that date must cite both amendments. For the equipment that tests monitors, FDA recognizes in full ISO TS 81060-5, a specification for the repeatability and reproducibility of the simulators that test oscillometric monitors.
FDA does not recognize the clause of IEC 80601-2-30 on the ingress of water and particles, because it found no rationale for the change of protection code "from IP21 to the less stringent IP20". It also asks makers to contact it before using the standard for claims on patient transport, ambulatory monitoring or stress and exercise testing. For ISO 81060-2, FDA leaves out the clauses on special and pregnant patient populations, two examples of data analysis and one clause on the reference blood pressure. A declaration to a partly recognized standard therefore covers the recognized clauses only. Neither exclusion touches the cuff-pressure limit, but a maker who reads the standard's title on FDA's list as recognition of the whole text misreads the record.
The equipment that tests a device falls under FDA's quality rule. FDA's Quality Management System Regulation, the QMSR, has been in force since 2 February 2026. The QMSR incorporates by reference ISO 13485:2016, the quality-system standard for device makers, and under the QMSR a maker must "Document a quality management system that complies with the applicable requirements of ISO 13485". The duty to control monitoring and measuring equipment is in clause 7.6 of that standard, so a maker's test stations and gauges come under the rule.
In its warning letters of 2026 (Chapter 26), FDA holds makers to that duty for their own calibration testing. On 27 May 2026 FDA wrote to Linemaster Switch Corporation, a maker of foot pedals that control medical lasers, that units "passed Linemaster's calibration testing" and were "subsequently returned because they failed customer calibration testing". FDA added that software errors in the test areas had been described as routine and "were not investigated".
The European Union: class IIa, and limits the maker states
In the European Union a home monitor is in class IIa, as two makers declare for their monitors under Rule 10 of the Medical Devices Regulation, the MDR. The MDR defines a device's class by rules. Under Rule 10, in its Annex VIII, active devices for diagnosis and monitoring are in class IIa "if they are intended to allow direct diagnosis or monitoring of vital physiological processes". They are in class IIb where they are "specifically intended for monitoring of vital physiological parameters" whose variations "could result in immediate danger to the patient". Joytech, which issues the EU declaration for the Beurer monitors of Chapter 29, names TÜV SÜD Product Service as its notified body, the independent organization designated to assess conformity, with a certificate valid to 27 April 2027. Ya Horng, which makes Garmin's Index BPM, names DNV Product Assurance, with a certificate valid to 30 August 2028.
The maker of a device with a measuring function, such as the monitor, must also state how accurate it is. Under the MDR such a device must be designed and made for sufficient accuracy, precision and stability for its intended purpose, and "The limits of accuracy shall be indicated by the manufacturer." Under the MDR, too, "The measurements made by devices with a measuring function shall be expressed in legal units", the units of the EU's units directive, 80/181/EEC (Chapter 3). Even a class I device with a measuring function needs a notified body, whose role under the MDR is limited "to the aspects relating to the conformity of the devices with the metrological requirements". A maker that otherwise declares the conformity of a class I device alone therefore calls in an outside assessor for the measurement.
The two makers declare conformity to the standards that FDA recognizes. Joytech lists IEC 80601-2-30:2018 with its European edition, EN IEC 80601-2-30:2019, and ISO 81060-2:2018 with amendment 2, also in its European edition, among 15 standards. Ya Horng lists the European edition of ISO 81060-2 and two older European standards for sphygmomanometers, EN 1060-1 and EN 1060-3, the second a predecessor of EN IEC 80601-2-30. A maker's list is not the Commission's list of harmonized standards, the EU's counterpart of FDA's list of recognized standards, and it is not known whether either European edition appears on the Commission's list for the MDR.
A maker of in vitro diagnostic devices, which test samples taken from the body, must assure traceability in the sense of Chapter 15. Under the IVDR, the parallel regulation for those devices, the metrological traceability of values assigned to calibrators and control materials "shall be assured through suitable reference measurement procedures and/or suitable reference materials of a higher metrological order". In January 2022 the Commission listed EN ISO 17511:2021, the standard for that traceability, among the IVDR's harmonized standards.
As of November 2026 the MDR as adopted governs the monitor, but both regulations may change. On 16 December 2025 the Commission proposed one regulation amending both. Among other changes, the proposal would remove the five-year limit on the validity of notified-body certificates. In September 2026 two committees of the European Parliament voted in favor, and a plenary vote is planned for early 2027, as the consultancy Emergo by UL reported on 9 October 2026.
India: a draft class and two regimes side by side
In India the monitor is in class B, but only in a draft list. India classifies medical devices under its Medical Devices Rules, 2017, and CDSCO, its medical-device regulator, administers those rules. In a list of cardiovascular devices dated 1 April 2025, CDSCO places every sphygmomanometer in class B: the aneroid manual instrument, the automatic-inflation electronic sphygmomanometer for professional use, portable or not, the electronic sphygmomanometer for home use and the long-term recorder. A non-invasive unit that measures several cardiovascular quantities goes to class C. The list is a draft issued for comments within 30 days, as the newspaper Business Standard and the regulatory services company CliniExperts report, and no final list is known.
The class decides the licensing route. A maker of class A or B devices applies on form MD-3 and receives its manufacturing license on form MD-5 from the State Licensing Authority. Classes C and D go through forms MD-7 and MD-9 to the Central Licensing Authority, and an importer of any class applies to the central authority on forms MD-14 and MD-15. CDSCO exempts from licensing only class A devices that are both non-sterile and non-measuring. A device with a measuring function therefore needs a license even in class A, and a class B monitor made in India needs one from its State.
The same monitor also falls under Part VII-B of the Eighth Schedule (Chapter 30), which DoCA administers under the Legal Metrology Act. No published text is known to say how the two regimes apply together: none states that a monitor needs both a CDSCO license and model approval under the Act, and none states that either exempts it from the other. A maker in India therefore settles each regime on its own. For the calibration of medical devices, NABL, India's accreditation body, gives specific criteria in its document NABL 126, issued in 2018 and amended in 2021.
| United States | European Union | India | |
|---|---|---|---|
| Class of a home monitor | II, product code DXN | IIa, Annex VIII Rule 10, as makers declare | B, in a draft list of 1 April 2025 |
| Route to market | 510(k), cleared by FDA | notified body's certificate; maker's declaration of conformity | State license, forms MD-3 and MD-5; imports MD-14 and MD-15 |
| The reading's accuracy | the maker's stated limit; IEC 80601-2-30 recognized in part | limits indicated by the maker, in legal units | Part VII-B, under legal metrology (Chapter 30) |
| The estimate's accuracy | ISO 81060-2 recognized in part (Chapter 24) | standards the makers declare | Part VII-B, clause 5(2) |
| Drafts | ISO/DIS 81060-2 | the MDR and IVDR proposal | the cardiovascular class list |
Drafts on the way
Four drafts could change these rules, and none is in force as of November 2026. The vote on ISO/DIS 81060-2, the draft revision of the clinical validation standard, closed on 15 September 2026. The 2018 edition with its two amendments stays in force, and FDA does not recognize the draft. Edition 3 of IEC 80601-2-30 has reached a committee draft for vote, and on 29 May 2026 DIN, Germany's standards body, issued a German draft based on it. The 2018 edition of that standard also stays in force. CDSCO's cardiovascular list of 2025 and the Commission's proposal for the MDR and the IVDR are the other two. A maker who designs to a draft cannot be sure the final text will match it, so the records to rely on are the editions in force and the regulators' own lists, each with its date.
- Build the cuff-pressure evidence once, on the bench method of OIML R 149-2 and India's Part VII-B, with the reference manometer and its certificate recorded.
- State the device's limits of accuracy, as the MDR's Annex I, section 15.1, requires of the maker, and keep the evidence behind them in the technical file.
- Before declaring conformity to a standard in a 510(k), read FDA's recognition record for the clauses it excludes and the date on which an older recognition stops being accepted.
- Keep the estimate's clinical validation separate from the reading's calibration in every file, because each regulator treats them as two claims (Chapter 24).
- Apply ISO 13485's control of measuring equipment to the production test station itself, and write down how a failed calibration of that station leads to a look at the results it passed (Chapter 26).
- In India, obtain the CDSCO license and settle the legal-metrology model approval separately, since no published text is known to join them.
The correction fitted in December 1993
The correction of December 1993 cancelled the primary mirror's error without touching the mirror. On 2 December 1993 the Space Shuttle Endeavour lifted off with two instruments built to correct the Hubble Space Telescope (the opening story), and its crew installed them on spacewalks. The Wide Field and Planetary Camera 2, WFPC2, carried the correction on small relay mirrors inside the camera, which hold, in the words of its instrument handbook, "the same error (but with opposite sign)". COSTAR placed five pairs of corrective mirrors in front of three other instruments. The correction worked: "The operation to correct the telescope optics was a total success, and the differences are striking", ESA announced on 13 January 1994.
A correction of that kind needs the error itself, and three routes that did not depend on one another measured it. In 1991 Litvak of TRW, under contract to the Jet Propulsion Laboratory, inferred the primary mirror's conic constant, the number that fixes the shape of its curve, from star images taken in orbit. He found −1.0141 against a design value of −1.0023, a difference of 0.0118. In 1993 Meinel, Meinel and Schulte, of the Jet Propulsion Laboratory and Lockheed, measured the telescope's effective value from direct images as −1.0140 ± 0.0003. The third route went back to the factory. Lew Allen's board of investigation had measured the reflective null corrector where it stood. The field lens, about 1.3 mm out of place, "can account for the entire HST error", as George Rodney, NASA's head of safety and mission quality and a member of the board, recorded.
The camera's designers then set its prescription inside the error's stated bounds. They adopted "a prescription within the error bars for the HST primary mirror", as the Space Telescope Science Institute's handbook records, so that the camera's residual aberration is small. The correction also had to sit exactly in place: by the same handbook, a misalignment of 7 % of the pupil's diameter would bring back as much wavefront error as the uncorrected mirror had.
The correction of 1993 rested on three routes, each of which could have disagreed with the others. In 1981 Perkin-Elmer had polished the mirror against one instrument whose critical length was set once and never checked, and the two instruments that disagreed with it were discounted. The two measurements from images differed by 0.0001 and so agreed within their stated uncertainties, and the third, from the null corrector itself, accounted for the whole error.
The monitor's reading needs what the correction had. Inside the instrument, every stage from the sensor to the display adds an error that a budget bounds, as Chapter 11 bounded the reading at 150 mmHg to a standard uncertainty of 0.70 mmHg. Outside it, each reference must have been compared with a better one, link by link, from the maker's reference manometer through an accredited laboratory and a national institute's pressure balance and its masses to the Planck constant (Chapter 28). The uncertainty stated at the bench then includes the uncertainty of every link above it. That uncertainty must also be small enough for the decision the number serves: an expanded uncertainty of 1.4 mmHg, twice 0.70 mmHg, against a limit of ±3 mmHg (Chapter 12).
One further condition comes from the mirror's history: a second, independent check must be able to disagree with the first and be heard, as the two instruments of 1981 were not. The monitor's estimate of blood pressure also needs one proof more, a clinical validation against a reference method (Chapter 24), because a calibration of the cuff pressure does not test the algorithm. Hubble's primary mirror still has the shape it was polished to. In 1993 its error was known, measured, with its uncertainty stated, by independent methods that agreed, and a known error could be cancelled where the unknown one of 1981 could not.
- FDA classified the recall of Libre 3 and 3 Plus sensors as Class I on 13 January 2026, for low readings from a production fault in factory-calibrated sensors, across four records that count about 3.0 million sensors.
- In the United States the monitor is a class II device cleared through a 510(k) under product code DXN, FDA recognizes IEC 80601-2-30 and ISO 81060-2 only in part, and the QMSR, in force since 2 February 2026, now places the control of measuring equipment under ISO 13485.
- In the EU makers declare the monitor class IIa under Rule 10 and must state its limits of accuracy in legal units, while in India a draft list of 1 April 2025 puts it in class B, licensed by the State, beside a legal-metrology regime that no published text is known to join to CDSCO's.
- Hubble's correction of December 1993 worked because the mirror's error had been measured, with its uncertainty stated, by independent methods that agreed, and the monitor's reading means what it says on the same terms, while its estimate needs clinical validation as well.