§ Glossary
Glossary.
Terms are defined as they are used in this guide.
- %GRR
- The gauge's R&R divided by the total variation of a gauge study, as a percentage; by AIAG's manual, as Minitab and BPI reproduce it, under 10 % is generally acceptable and over 30 % unacceptable (Chapter 22).
- 1/f corner
- The frequency below which 1/f noise exceeds an amplifier's flat noise floor; about 1 to 10 Hz for the best low-frequency amplifiers (Chapter 5).
- 1/f noise
- Noise whose power is inversely proportional to frequency, so that it rises at low frequencies (Chapter 5).
- 510(k)
- The premarket submission in which a US maker shows that its device is substantially equivalent to one already legally marketed; FDA clears it (Chapter 31).
- Acceptance limit
- The upper or lower bound of the measured values that a test will accept; it equals the tolerance limit under simple acceptance and lies inside it with a guard band (Chapter 13).
- Accreditation
- An independent body's assessment of a laboratory against a standard such as ISO/IEC 17025, followed by formal recognition that it is competent for a stated list of calibrations or tests; laboratories are accredited, quality systems certified (Chapter 16).
- Accuracy
- In the VIM, the closeness of a measured value to a true value; not a quantity and given no number. A maker's "accuracy ±3 mmHg" is a maximum permissible error (Chapter 2).
- Adjustment
- A set of operations that makes a measuring system give prescribed indications for given values of the quantity, such as zero, offset or span adjustment (VIM 3.11). Calibration comes before it, and the system is usually calibrated again after it (Chapter 14).
- Alias
- The false, slower frequency that a too-fast tone produces in sampled readings: mains hum at 50 Hz read 49 times a second gives an alias at 1 Hz (Chapter 7).
- Aliasing
- The distortion by which a frequency too fast for the readings appears in them as a slower one; once the readings are taken it is generally not recoverable (Chapter 7).
- Aneroid sphygmomanometer
- A mechanical dial gauge for blood pressure, used with a stethoscope; OIML R 148 covers it (Chapter 2).
- Anti-alias filter
- A low-pass filter ahead of a sampling converter that weakens frequencies above half the sampling rate before they can fold back as aliases (Chapter 7).
- Aperture jitter
- The variation, from one sample to the next, in the instant a sample-and-hold's switch opens; it matters for fast signals, not for a blood-pressure monitor's pulse band (Chapter 7).
- Appraiser variation (AV)
- In a gauge study, the reproducibility as a standard deviation: the spread between the averages of different operators on the same parts (Chapter 22).
- Arms
- A pulse oximeter's accuracy figure: the square root of the average squared difference from oxygen saturation measured in arterial blood, which counts bias and scatter together; FDA staff put 1 Arms at about one standard deviation (Chapter 24).
- As found, as left
- Trade names for the calibration results before and after an adjustment or repair; accreditation bodies describe ISO/IEC 17025 as asking for both on a certificate where they are available (Chapter 14).
- As-maintained mass unit
- The kilogram as the BIPM actually kept it in its working standards, found in 2014 to be offset by 35 µg from the international prototype (Chapter 18).
- Attribute agreement
- The test of a measuring system that reports a verdict, such as pass or fail: how often an operator repeats a verdict, how often operators agree, and how often the verdicts match the standard, the verdict known to be right for each item (Chapter 22).
- Auscultation
- Listening to the artery with a stethoscope; the reference method for the sounds by which an observer reads systolic and diastolic pressure (Chapter 7).
- Auto-zero
- In an amplifier, a technique in which the amplifier measures its own offset and subtracts it; because the noise changes slowly at low frequencies, subtracting consecutive samples cancels its low-frequency noise too (Chapter 5). In a monitor, its own zero: before measuring, at least at power-on, the firmware reads the sensor open to the air, stores that reading and subtracts it from later readings; an adjustment, not a calibration, which removes the offset but cannot see the span (Chapter 17).
- Bandwidth
- The range of frequencies over which a signal is measured; noise power grows in proportion to it (Chapter 5).
- Bell curve
- The normal distribution: values near its center are more likely than values far from it, it falls away alike on both sides, and its standard deviation sets its width (Chapter 10).
- Bias
- The value that a measurement assigns to a systematic error (Chapter 2).
- Bias (of an estimate)
- The mean of an estimate's differences from the reference, the first figure of ISO 81060-2's criterion 1 (Chapter 24); for a measuring system, the average difference between its results on a part and that part's known value (Chapter 22).
- BIPM
- The International Bureau of Weights and Measures, the laboratory and secretariat at Sèvres set up by the Metre Convention (Chapter 3).
- Bridge
- Four resistors joined in a square, powered across two opposite corners, whose output is the voltage between the other two; it is zero when the bridge is balanced (Chapter 4).
- Calibration
- In the VIM's definition (2.39), an operation in two steps: relating the values that measurement standards provide to an instrument's indications, each with its uncertainty, and then using that relation to obtain a measurement result from an indication. It is not adjustment (Chapter 14).
- Calibration and measurement capability (CMC)
- In ILAC's definition, a calibration and measurement capability available to customers under normal conditions, as a scope of accreditation describes it or as the BIPM's database publishes it; stated as an expanded uncertainty at about 95 %, the smallest uncertainty a certificate from the laboratory may state (Chapter 16).
- Calibration coefficients
- Numbers measured for each unit and stored in its firmware that turn counts into the displayed unit: chiefly a zero, measured again with the sensor open to the air, and a span factor from calibration against a reference (Chapter 9).
- Calibration hierarchy
- The sequence of calibrations from a reference to the final measuring system (VIM 2.40), along which the uncertainty necessarily increases (Chapter 15).
- Calibration interval
- The time an owner allows between successive calibrations of an instrument, chosen from risk and reviewed from the results; an accredited laboratory prints none unless the customer asks (Chapter 17).
- Certified reference material
- A material whose value a competent producer certifies with its uncertainty; the links of a traceability chain in chemistry and laboratory medicine (Chapter 15).
- CGPM
- The General Conference on Weights and Measures: delegates of every Member State, normally meeting every four years, whose resolutions decide the units; its 28th meeting was in October 2026 (Chapter 3).
- Checksum
- A number computed from stored data, such as a program and its calibration coefficients, so that a changed bit gives a different number (Chapter 26).
- Chi-squared test of consistency
- A statistical test of whether a set of results scatters around its common value by more than the results' stated uncertainties allow (Chapter 19).
- CIPM
- The International Committee for Weights and Measures: eighteen members from different States, meeting every year under the CGPM's authority, advised by ten Consultative Committees (Chapter 3).
- CIPM MRA
- The mutual recognition arrangement, signed on 14 October 1999, through which national metrology institutes demonstrate the international equivalence of their measurement standards and the certificates they issue (Chapters 3, 21).
- Class E1
- The most accurate class of weight in OIML R 111; a 1 kg weight of this class may differ from its nominal value by no more than 0.5 mg (Chapters 20, 21).
- Class I recall
- FDA's most serious type of recall, a classification of the recall action separate from the device's class (Chapters 9, 31).
- Clinical validation
- A study in people that compares a device's estimates with a reference method under a protocol fixed in advance; the only proof of a blood-pressure monitor's systolic and diastolic values (Chapter 24).
- Combined standard uncertainty
- The square root of the sum of the squared contributions of every row of an uncertainty budget; 0.70 mmHg for the monitor's corrected cuff-pressure reading at 150 mmHg (Chapter 11).
- Common mode
- The voltage that both inputs of a differential amplifier share; about 1 V at the monitor's bridge outputs (Chapter 6).
- Common-mode rejection ratio (CMRR)
- How well an amplifier ignores a common-mode voltage, stated in decibels; at least 85 dB for the INA350 (Chapter 6).
- Conditional pass and conditional fail
- ILAC's outcomes for a measured result inside the guard band below the tolerance limit, or above the tolerance limit but within the guard band beyond it: results the test cannot decide (Chapter 13).
- Consensus value
- The value of the kilogram that the CCM computes from the key comparisons of the realizations and from which mass has been disseminated since February 2021; since 1 March 2026 it puts the old prototype at 1 kg − 12 µg, with a standard uncertainty of 20 µg (Chapter 21).
- Control solution
- A liquid of known value that tests a glucose meter, not the patient (Chapter 27).
- Corner frequency
- The frequency at which a filter begins to weaken a signal (Chapter 6).
- Inputs that share a cause and so move together; the simple sum of squares does not apply to them, and the GUM adds a term for each related pair (Chapter 11).
- Coverage factor
- The multiplier k that turns a combined standard uncertainty into an expanded uncertainty; accredited calibration laboratories use k = 2, about 95 % (95.45 % for a bell curve), and a larger k when a result rests on few readings (Chapter 12).
- Coverage probability
- The probability that the interval of plus or minus the expanded uncertainty contains the true value; about 95 % for k = 2. EA and UKAS call it a coverage probability, Mensor and NIST a confidence level (Chapter 12).
- Criterion 1 (ISO 81060-2)
- Over all paired readings, a mean difference from the reference of no more than 5 mmHg with a standard deviation of no more than 8 mmHg, as the AAMI/ESH/ISO collaboration states it (Chapter 24).
- Criterion 2 (ISO 81060-2)
- A limit on the standard deviation of each subject's averaged difference that falls as the mean difference grows, from 6.95 mmHg at a mean of zero to 4.79 mmHg at 5 mmHg, so that at least 85 % of errors stay within 10 mmHg (Chapter 24).
- Decibel (dB)
- A unit for ratios that counts tenfold steps: 20 dB for each tenfold step of a voltage ratio and 10 dB for each tenfold step of a power ratio (Chapter 5).
- Decision rule
- A rule that describes how measurement uncertainty is accounted for when stating conformity with a specified requirement, in ISO/IEC 17025's definition as ILAC reproduces it (Chapter 13).
- Defining constants
- The seven constants whose fixed numerical values define the SI since 20 May 2019, the Planck constant among them (Chapter 19).
- Deflation method
- The way a monitor lets air out of the cuff while its algorithm reads the oscillations (Chapter 27).
- Degree of equivalence
- The difference between an institute's result in a key comparison and the reference value, stated with the uncertainty of that difference (Chapter 21).
- Degrees of freedom
- The number of independent deviations behind a standard deviation, one less than the number of readings in the simple case; it sets the coverage factor when few readings are used (Chapters 10 and 12).
- Design traceability
- The documented link from each design requirement to the design output that answers it and to the verification that shows it is met; a different thing from metrological traceability (Chapters 15, 25).
- Diagnostic sensitivity
- The share of people with a condition whom a test detects; distinct from an instrument's sensitivity, its change in output for each unit of input (Chapters 4 and 24).
- Diastolic pressure
- The lowest pressure in the artery during a heartbeat; in an oscillometric monitor, an estimate derived by the algorithm from the envelope (Chapters 1, 9).
- Differential nonlinearity (DNL)
- The largest departure of any one converter step from its ideal size; a DNL worse than minus one step leaves a missing code (Chapter 8).
- Disseminate
- To pass a unit's value down to other standards by comparison, from a realization to national standards, accredited laboratories and factories (Chapter 21).
- Dither
- Noise added to a converter's input; about half a step rms randomizes the quantization error and lets averaging resolve finer than one step (Chapter 8).
- Dominant contributor
- The row of an uncertainty budget with the largest share of the total of squares: the offset, with 62 %, for the bare sensor. In the corrected channel it is the reference manometer, at 44 %, with the hysteresis close behind at 39 %, so no row supplies even half of the total (Chapters 11, 12).
- Doubtful band
- The measured values within one expanded uncertainty of a tolerance limit, the only ones whose true value, at about 95 % coverage, could lie on the other side of it (Chapter 13).
- Drift
- A slow change in an instrument's output for the same input, such as a sensor's zero moving with temperature or time (Chapter 17).
- Effective area
- The area by which a pressure balance's piston turns a force into a pressure; about 3.357 cm² for CSIR-NPL's air piston gauge, known to nearly 11 parts per million at k = 2 (Chapter 15).
- Effective degrees of freedom
- A count derived by formula from the degrees of freedom of each contribution to a budget, from which EA-4/02's table gives the coverage factor (Chapter 12).
- Effective number of bits (ENOB)
- SINAD minus 1.76 dB, divided by 6.02 dB: the bit count of a perfect converter that would add as much noise and distortion; up to 11.1 for the converter in TI's monitor design (Chapter 8).
- Envelope
- The curve traced by the sizes of the oscillations plotted against the falling cuff pressure; the firmware receives one point on it per heartbeat (Chapter 7).
- Equipment variation (EV)
- In a gauge study, the repeatability as a standard deviation: the scatter when one operator measures the same part several times (Chapter 22).
- Error budget (as a requirement)
- An allocation of one instrument's accuracy limit into shares, each with its size, its owner and the test that proves it, set before the parts are chosen; for the monitor, seven shares of ±3 mmHg at 150 mmHg (Chapter 25).
- Estimate
- In this guide, the systolic and diastolic blood pressure that the monitor's algorithm derives from the oscillations in the cuff pressure; only a clinical validation can prove it (Chapters 1, 24).
- EU declaration of conformity
- The manufacturer's statement that a device meets the requirements of the EU's device regulation (Chapters 29, 31).
- Excitation
- The current or voltage that powers a bridge or other sensor; the 2SMPP-03 is driven by a constant 100 µA (Chapter 4).
- Expanded uncertainty
- The combined standard uncertainty multiplied by a coverage factor, written U and printed after the ± on a certificate; ±1.4 mmHg for the monitor's reading at 150 mmHg (Chapter 12).
- Factory calibrated
- FDA's product-classification wording for a sensor that its user does not calibrate, so the factory's calibration is the only one between its signal and the displayed number (Chapter 31).
- False accept
- Accepting an item that in fact lies outside its tolerance limit; ILAC states that under simple acceptance its probability may be as high as 50 %, which UKAS places at values on the tolerance limit (Chapters 12, 13).
- Fault
- In OIML's weighing recommendation, the difference between an instrument's error of indication and its intrinsic error, the error it shows minus the error it had when sound (Chapter 26).
- Filter
- A circuit that passes some frequencies and weakens others; a high-pass filter weakens slow changes, a low-pass filter weakens fast ones, and a band-pass filter weakens both (Chapter 6).
- Fixed-ratio method
- The common published oscillometric algorithm: systolic and diastolic pressure are read where the envelope falls to a population-based fraction of its peak, one fraction on each side; the makers' own fractions are proprietary (Chapter 9).
- Gain
- The factor by which an amplifier multiplies its input; 50 for the in-amp in TI's design (Chapter 6).
- Gain error
- The amount by which an amplifier's actual gain differs from its nominal gain; it scales every reading in proportion (Chapter 6).
- Gauge
- In manufacturing, any measuring instrument or test system, such as a production pressure test station (Chapter 22).
- Gauge pressure
- Pressure measured against the pressure of the surrounding air; the monitor's sensor reads the cuff's pressure above the room's (Chapter 4).
- Gauge R&R (GRR)
- Repeatability and reproducibility together: the variation that a measuring system adds to what it measures, the root-sum-of-squares of the equipment and appraiser variation (Chapter 22).
- Gauge study
- The study that measures a measuring system's gauge R&R (Chapter 22).
- Global ACI MRA
- The Multilateral Recognition Arrangement of Global Accreditation Cooperation Incorporated, into which the arrangements of ILAC and IAF passed on 1 January 2026 (Chapter 16).
- Global risk
- The probability, over the whole stream of items before any is measured, that a non-conforming item will be accepted (consumer's risk) or a conforming one rejected (producer's risk) (Chapter 13).
- Guard band
- The interval between a tolerance limit and the corresponding acceptance limit, often equal to the expanded uncertainty; with w = U the monitor's acceptance limit at 150 mmHg moves from 3.0 to 1.6 mmHg (Chapter 13).
- Half-width
- The distance from the center of a band of possible values to either of its limits (Chapter 11).
- Harm
- In ISO 14971's words, "Injury or damage to the health of people, or damage to property or the environment" (Chapter 26).
- Harmonized standards
- The standards on the European Commission's list, the EU's counterpart of FDA's list of recognized standards (Chapter 31).
- Hazard
- In ISO 14971's words, a "Potential source of harm" (Chapter 26).
- Hazardous situation
- In ISO 14971's words, a "Circumstance in which people, property or the environment is/are exposed to one or more hazards" (Chapter 26).
- Hysteresis
- The difference between a sensor's output at one pressure reached from below and from above; up to 0.2 %FS for the 2SMPP-03 (Chapter 4).
- Image (of a tone)
- A frequency at a whole multiple of the sampling rate, plus or minus a tone's own frequency, which sampled readings cannot tell apart from the tone; a tone above half the sampling rate always has an image below half that rate, where the readings show it as an alias (Chapter 7).
- Indication
- The quantity value an instrument provides; it need not be the same kind of quantity as the measurand, as with the sensor's millivolts and the converter's counts (Chapter 1).
- Indirect measurement
- A measurement made through a chain of signals and computations rather than by setting the quantity beside a scale, as the monitor turns an air pressure into millivolts, counts and mmHg (Chapter 1).
- Influence quantity
- A quantity other than the one measured that changes an instrument's output, such as temperature for a silicon bridge (Chapter 4).
- Input-referred noise
- Noise expressed as if it entered at the input of an amplifier or front end, so that it can be compared directly with the sensor's signal; under 10 µV peak to peak for the front end of TI's design (Chapters 5, 10, 11).
- Instrumentation amplifier
- Also in-amp: an amplifier built to measure the small difference between two inputs while rejecting the voltage they share (Chapter 6).
- Integral nonlinearity (INL)
- The largest departure of a converter's actual transfer characteristic from a straight line; an accuracy figure, separate from the converter's bit count (Chapter 8).
- Interface
- The boundary between two owners of a measuring chain, where one part's output becomes another's input, such as the sensor's microvolts for each millimeter of mercury at the amplifier (Chapter 25).
- Interference
- Noise with an outside cause, such as mains hum or the noise of the monitor's motors (Chapter 5).
- Intermediate check
- A check between full calibrations that an instrument still works and keeps its calibration status, such as a monitor read at zero and at one pressure (Chapter 17).
- International prototype of the kilogram
- The platinum-iridium cylinder kept at the BIPM that defined the kilogram from 1889 to 2019; since 2019 a mass like any other, with an uncertainty (Chapters 18, 19).
- JCTLM
- The Joint Committee for Traceability in Laboratory Medicine, which the BIPM set up in 2002 with ILAC and the International Federation of Clinical Chemistry and Laboratory Medicine; its database lists the reference materials, reference measurement procedures and reference measurement services at the upper links of the traceability chains of in-vitro diagnostic tests (Chapter 15).
- Josephson effect
- The effect by which a junction between two superconductors, driven at a known microwave frequency, holds a voltage set by that frequency and the Josephson constant, twice the elementary charge divided by h; used to realize the volt (Chapter 20).
- Kappa
- A statistic of agreement between verdicts that corrects the share of agreeing verdicts for the share that chance alone would produce; AIAG suggests at least 0.75, as Minitab reports (Chapter 22).
- KCDB
- The BIPM's key comparison database, which publishes the institutes' peer-reviewed calibration and measurement capabilities with the comparisons that support them (Chapter 21).
- Key comparison
- A comparison in which several national institutes measure the same quantity, usually on one transfer standard that travels between them, under a pilot laboratory (Chapter 21).
- Key comparison reference value
- The reference value that a key comparison yields, against which each participant's result is set (Chapter 21).
- Kibble balance
- An instrument, formerly called the watt balance, that weighs a mass against the force on a current-carrying coil and then moves the coil to measure the voltage it generates, so that the mass follows from the Planck constant, the meter and the second (Chapter 20).
- Least significant bit (LSB)
- A converter's step: its full range divided by its number of codes; one count of the monitor's 12-bit converter is 0.109 mmHg at the cuff (Chapter 8).
- Least-squares adjustment
- A calculation that finds the one set of values for several constants that best fits all the measurements together, as CODATA's task group did with the data available by 1 July 2017 (Chapter 19).
- Legal metrology
- In OIML's words, the "practice and process of applying statutory and regulatory structure and enforcement to metrology": the part of measurement a state makes law and checks; its scope differs from country to country (Chapter 30).
- Limits of agreement
- The bias plus and minus about twice the standard deviation of the differences from the reference: the band that holds about 95 % of single differences when they follow a bell curve (Chapter 24).
- Linearity (of a measuring system)
- The way a measuring system's bias changes across its range (Chapter 22).
- Long-term stability
- The drift of a sensor's output over months of use; the 2SMPP-03's data sheet gives none (Chapter 4).
- Manometer
- A pressure gauge (Chapters 1, 14).
- Maximum permissible measurement error
- The VIM's term for a limit within which an instrument's error must lie, such as OIML's ±3 mmHg for a monitor's cuff pressure (Chapter 2).
- Measurand
- The quantity a measurement is meant to find, in the VIM's words the "quantity intended to be measured"; for the monitor's reading, the air pressure in the cuff; for its estimate, the pressure in the artery (Chapter 1).
- Measurement
- In the VIM, the process of experimentally obtaining quantity values that can reasonably be attributed to a quantity; it implies a comparison (Chapter 1).
- Measurement error
- A measured value minus a reference value, with a sign; a difference, not a mistake (Chapter 2).
- Measurement model
- The written relation between a result and everything it depends on: for the monitor's reading, the cuff pressure and the error each stage of the chain can add (Chapter 11).
- Measurement result
- The value attributed to a measurand with the other relevant information, generally a measured value and its uncertainty; a number, its unit and its uncertainty together (Chapter 1).
- Measurement uncertainty
- A non-negative parameter, stated with the result, that characterizes the spread of values that can be attributed to the measurand on the information used; the stated doubt (Chapter 2).
- Measurement-system analysis (MSA)
- The discipline, set out in AIAG's reference manual, of testing a measuring system's bias, linearity, stability, repeatability and reproducibility before trusting its numbers (Chapter 22).
- Measuring function
- In the MDR, the property of a device that measures; its maker must state the limits of accuracy and express the measurements in legal units (Chapter 31).
- MEMS
- Microelectromechanical system: a sensor whose moving part and electrical elements are made in the same piece of silicon (Chapter 4).
- Metre Convention
- The treaty signed in Paris on 20 May 1875 by seventeen states, which created the BIPM and the bodies that decide the units (Chapter 3).
- Metrological traceability
- The property of a measurement result whereby it can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty (VIM 2.41); a property of a result, not of an instrument or a laboratory (Chapter 15).
- Millimeter of mercury (mmHg)
- The unit in which medicine states blood pressure, fixed by convention at 133.322 387 415 Pa; not listed in the SI Brochure's version 4.01 (2026) and kept for blood pressure by law (Chapter 3).
- Missing code
- A converter code that never appears because its step has no width (Chapter 8).
- Model approval
- India's term for type evaluation: approval of a weight or measure's model that the Legal Metrology Act requires before it is made or imported (Chapter 30).
- Moving mode
- Also velocity mode: the Kibble balance's measurement in which the coil sweeps through the field at a known speed while the voltage it generates is measured (Chapter 20).
- National metrology institute
- A country's laboratory for its national standards, which realizes the units and passes their values on; in India, CSIR-NPL in New Delhi (Chapter 3).
- National prototype
- A platinum-iridium copy of the kilogram held by a country, such as India's No. 57 at CSIR-NPL (Chapters 18, 21).
- Noise
- Any part of a signal that the measured quantity did not cause (Chapter 5).
- Noise density
- Noise stated for each square root of a hertz of bandwidth, such as 43 nV per root hertz (Chapter 5).
- Noise floor
- The level of noise below which no change in a channel can be told apart from it (Chapter 5).
- Noise-free resolution
- A maker's measure: the full range divided by the peak-to-peak noise, taken as 6.6 times the rms noise, stated in bits; the resolution at which the output code does not flicker (Chapter 8).
- Nonlinearity
- The departure of a sensor's output from a straight line through its range; up to 1.3 %FS for the 2SMPP-03 (Chapter 4).
- Notch
- A frequency that a filter blocks almost completely; at 50 readings a second, the filter of Analog Devices' AD7190 places notches at 50 Hz and at every multiple of 50 Hz (Chapter 7).
- Notified body
- An independent organization designated under EU law to assess a device's conformity; for a class I device with a measuring function, the MDR limits its role to the metrological requirements (Chapter 31).
- Number of distinct categories (ndc)
- An estimate of how many groups of parts a measuring system can tell apart: the part variation divided by the gauge's R&R, times 1.41, cut down to a whole number; AIAG recommends five or more (Chapter 22).
- Nyquist rate
- Twice the highest frequency in the signal that reaches the converter, hum included: the lowest sampling rate the sampling theorem allows (Chapter 7).
- Official copies
- The platinum-iridium copies of the international prototype kept with it at the BIPM: two at first, brought to six by the replacement of one and the addition of four others (Chapter 18).
- Offset
- A sensor's or amplifier's output with no input applied; typically −2.5 mV for the 2SMPP-03 (Chapter 4); as an error, it shifts every result by the same amount (Chapters 2, 11, 27).
- OIML
- The intergovernmental body that writes model rules for measuring instruments, its Recommendations, among them R 148 and R 149 (2020) for sphygmomanometers (Chapter 1).
- OIML-CS
- The OIML Certification System, a system for issuing, registering and using OIML certificates and type evaluation reports for types of measuring instruments; each state decides whether to accept them (Chapter 30).
- Oscillometric signal
- The cuff pressure together with the small oscillations, from less than 1 mmHg to 3 mmHg, that the pulsing artery adds to it at the pace of the heart; the steady part is the reading, and the oscillations carry what the firmware turns into the estimate (Chapter 4).
- Out-of-tolerance impact assessment
- The judgment, for each result a faulty reference supported, of whether the reference's error could have made it wrong by enough to matter, and what to do about the products involved (Chapter 26).
- Oversampling
- Reading far faster than the signal needs and averaging the readings down; four times the readings buy one more bit (Chapter 8).
- Part per billion
- One part in a billion; on a kilogram, one microgram (Chapter 18).
- Part variation (PV)
- In a gauge study, the spread of the parts' true values as a standard deviation, which the measuring system exists to see (Chapter 22).
- Part VII-A
- The Part of the Eighth Schedule for non-invasive non-automated sphygmomanometers, replaced on 7 January 2026 (Chapter 30).
- Part VII-B
- The Part of the Eighth Schedule to India's Legal Metrology (General) Rules, 2011, for non-invasive automated sphygmomanometers, substituted on 28 July 2025; it sets the cuff pressure's limit at ±0.4 kPa (±3 mmHg) or ±2 % of the reading, whichever is greater (Chapter 30).
- Part-to-part variation
- The real differences between the parts or monitors in a study, as distinct from the scatter that the measuring system adds (Chapter 22).
- Pascal
- The SI unit of pressure, one newton of force on one square meter, in base units a kilogram per meter per second squared; a kilopascal is 1,000 pascals (Chapter 3).
- Peak to peak
- The distance between the highest and lowest values of a signal over a stated time; for random noise, taken as 6.6 times the rms value (Chapter 5).
- Percent contribution
- A component of a gauge study as a share of the total variance rather than of the total standard deviation; 10 % and 30 % of the standard deviation are 1 % and 9 % of the variance (Chapter 22).
- Percent of full scale (%FS)
- A sensor's error stated as a share of its whole range rather than of the reading; 1 %FS of a 50 kPa sensor is about 3.75 mmHg (Chapter 4).
- Percent within
- The share of a device's results that fall inside a fixed band around the reference, such as 95 % of a glucose meter's results within ±15 % (Chapter 24).
- Periodic verification
- One of the three campaigns, which Schwitz and colleagues date 1899 to 1911, 1947 to 1954 and 1988 to 1992, in which the BIPM compared the national prototypes with the international prototype (Chapter 18).
- Piezoresistance
- The change in a material's electrical resistance when the material is strained (Chapter 4).
- Piezoresistor
- A resistor formed in silicon whose value changes when the silicon flexes; the monitor's sensor is a bridge of them (Chapters 1, 4).
- Pilot laboratory
- The laboratory that organizes a comparison and analyzes its results; CSIR-NPL for APMP.M.P-K6 (Chapter 21).
- Planck constant
- The constant h of quantum physics, which links the energy carried by light to its frequency; since 20 May 2019 exactly 6.626 070 15 ÷ 10³⁴ joule seconds, a value that defines the kilogram (Chapter 19).
- Pounds per square inch (psi)
- A unit of pressure used on some certificates, such as Mensor's for a barometric reference from 8 to 17 psi, about 55 to 117 kPa (Chapter 12).
- Precision
- The closeness of repeated readings to one another under stated conditions, usually given as a standard deviation; not accuracy (Chapter 2).
- Predicate device
- The legally marketed device to which a 510(k) claims substantial equivalence (Chapters 24, 31).
- Pressure balance
- Also piston gauge: a piston loaded with masses in a close-fitting cylinder, which makes a known pressure as the weight of piston and masses divided by the piston's effective area; a balance because the pressure balances a weight (Chapter 15).
- Product code
- FDA's three-letter label for a type of device, which ties it to a regulation and a class; the automated monitor's is DXN (Chapter 31).
- Proficiency testing
- A check of laboratories' results against each other: a provider sends the same item, or items of known value, to several laboratories and compares what each reports (Chapter 16).
- Quantization error
- The error of up to half a step either way that every converted reading carries; modeled as a sawtooth, a flat distribution with an rms value of the step divided by the square root of 12 (Chapter 8).
- Quantum Hall effect
- The effect by which a conducting layer in a strong magnetic field at low temperature shows a resistance set by the von Klitzing constant, h divided by the square of the elementary charge; used to realize the ohm (Chapter 20).
- Random error
- The part of an error that varies unpredictably from one repeated measurement to the next, such as amplifier noise; averaging reduces its effect (Chapter 2).
- Ratiometric measurement
- A measurement in which the converter's reference is derived from the sensor's own excitation, so that a change in the excitation cancels; TI's monitor design is not ratiometric (Chapter 6).
- Re-verification after a repair
- A check of a repaired instrument against the same limits as when new, before it goes back into use, because the repair may have changed the instrument its last calibration described (Chapter 27).
- Reading
- In this guide, the monitor's displayed cuff pressure in mmHg, which a chain of calibrations ties to the SI; kept apart from the estimate (Chapter 1).
- Realize
- To make a unit in practice in a laboratory, from its definition; national laboratories compare their realizations (Chapters 3, 21).
- Recognized consensus standard
- A standard on FDA's list, to which a maker may declare conformity in support of a premarket submission; FDA may recognize it only in part (Chapter 31).
- Rectangular distribution
- Also flat distribution: a band in which every value between two limits is equally likely; its standard deviation is the half-width divided by the square root of 3 (Chapter 11).
- Reference
- Whatever a measurement is compared with, from the marks on a ruler to a reference manometer or a national standard (Chapters 1, 24).
- Reference manometer
- The pressure gauge whose reading serves as the reference value when a monitor or gauge is calibrated; OIML and India's rules for automated sphygmomanometers allow it a maximum permissible error of ±0.8 mmHg (Chapter 14).
- Reference method
- An independent way of finding a quantity in the same subjects, trusted because a protocol fixes how it is done, such as two observers measuring blood pressure at the same time; distinct from a reference standard, which carries a traceable value (Chapter 24).
- Relative standard uncertainty
- A standard uncertainty expressed as a fraction of the value, such as 10 parts per billion (Chapter 19).
- Repeatability (of a measuring system)
- The spread between results when one operator measures the same part several times with the same system, under repeatability conditions; a gauge study states it as the equipment variation (Chapters 2 and 22).
- Repeatability conditions
- Repeated measurements with the same procedure, operators, measuring system, conditions and location over a short period (Chapter 2).
- Reproducibility (of a measuring system)
- The spread between results when different operators measure the same part with the same system (Chapters 2 and 22).
- Reproducibility conditions
- Repeated measurements in which the locations, operators and measuring systems differ (Chapter 2).
- Resolution
- The smallest change in a measured quantity that perceptibly changes the indication; for a display, the smallest step it can show, such as 1 mmHg. It is not a limit on error (Chapters 2, 8).
- Reverse traceability
- The search backward from a reference found wrong to every result and every product that rested on it since its last good calibration (Chapter 26).
- Rms
- Root mean square: the square root of the average of a signal's squared values; for noise, its standard deviation (Chapter 5).
- Robustness (of a test method)
- In ICH's list of validation characteristics, a method's capacity to keep meeting its criteria despite the small changes that daily use brings, such as a slightly different vessel, a longer wait or a warmer room (Chapter 23).
- Root-difference-square guard band
- An acceptance limit set at the square root of the specification squared minus the expanded uncertainty squared, one of the four methods that Fluke's policy names; 2.65 mmHg for the monitor at 150 mmHg (Chapter 13).
- Root-sum-of-squares
- The rule for adding independent uncertainties: square each contribution, add the squares and take the square root of the total (Chapter 11).
- Same-arm sequential method
- In a blood-pressure validation, taking the observers' readings and the monitor's one after another on the same arm, the method the AAMI/ESH/ISO collaboration prefers (Chapter 24).
- Sample-and-hold
- A circuit that follows the input, then disconnects a small capacitor from it and holds that voltage while the converter works, so that each reading records one instant (Chapter 7).
- Sampling rate
- The number of readings a converter takes each second (Chapter 7).
- Scope of accreditation
- The list of calibrations an accreditation covers, each with its quantity, range, method and CMC; a certificate outside it does not carry the accreditation's weight (Chapter 16).
- Self-heating
- The warming of a sensor by the excitation that measures it, an error the sensor makes by being measured (Chapter 6).
- Sensitivity (of an instrument)
- The change in an instrument's output for a given change in the quantity it measures; 112 µV for each mmHg for the 2SMPP-03 (Chapter 4). Distinct from a diagnostic test's sensitivity (Chapter 24).
- Sensitivity coefficient
- In an uncertainty budget, the rate at which the result changes when one input changes, which converts each standard uncertainty into the unit of the result; distinct from an instrument's sensitivity (Chapter 11).
- Service mode
- A monitor's test mode, reserved for service personnel, in which the vent valve stays closed so that an outside pump sets the pressure and the display shows what the sensor reads (Chapter 14).
- Shot noise
- Noise from the separate, randomly arriving charges that make up a current (Chapter 5).
- SI
- The International System of Units: seven base units, the units derived from them and 24 prefixes, resting since 20 May 2019 on seven defining constants (Chapter 3).
- Sigma-delta converter
- A converter that compares its input with a 1-bit feedback loop running far faster than the signal, shapes its quantization noise out of the band of interest and filters the bit stream into slower readings of many bits; it averages over each conversion rather than sampling an instant (Chapters 7, 8).
- Signal chain
- The six stages between a physical quantity and a reported number, named by what each does: sense, condition, digitize, compute, connect, decide (Chapter 1; Part II).
- Signal-to-noise ratio (SNR)
- The size of a signal against the size of the noise that accompanies it, usually stated in decibels (Chapter 5).
- Significant fault
- In OIML's weighing recommendation, a fault greater than one verification scale interval, e; in the recommendation's 1992 edition, since superseded, OIML required the instrument to stop automatically or warn on one (Chapter 26).
- Simple acceptance
- A decision rule that accepts an item whose measured value does not exceed the tolerance limit, without allowing for uncertainty; the JCGM notes that it is also known as shared risk (Chapter 13).
- SINAD
- The ratio of a converter's rms signal to the root-sum-of-squares of everything else in its output, noise and distortion together, as Analog Devices defines it (Chapter 8).
- Span (of a sensor)
- The sensor's output at full scale minus its output at zero; typically 42.0 mV over 50 kPa for the Omron 2SMPP-03 (Chapters 4, 11).
- Specific risk
- The probability that one particular accepted item does not conform, or that one particular rejected item does, given its measured value (Chapter 13).
- Specificity (of a test method)
- In ICH's list of validation characteristics, a method's response to the property it tests and to nothing else; for the pressure test, to the monitor's error and not to a leak, a moving pressure or a change of temperature (Chapter 23).
- Stability (of a measuring system)
- The way a measuring system's bias changes over time (Chapter 22).
- Standard deviation
- The measure of how widely single readings scatter around their mean: square each reading's deviation from the mean, add the squares, divide by one less than the number of readings and take the square root; 0.44 mmHg for the ten generated readings of Chapter 10's example (Chapter 10).
- Standard deviation of the mean
- The standard deviation of single readings divided by the square root of their number; it describes how well a mean locates the quantity, not how one reading scatters (Chapter 10).
- Standard uncertainty
- An uncertainty expressed as a standard deviation, in the GUM's definition (Chapters 2, 11).
- Systematic error
- The part of an error that stays constant or varies predictably when a measurement is repeated, such as a sensor's offset; a known one can be corrected (Chapter 2).
- Systolic pressure
- The highest pressure in the artery during a heartbeat; in an oscillometric monitor, an estimate derived by the algorithm from the envelope (Chapters 1, 9).
- Temperature coefficient
- The change of a figure, such as an offset or a reference voltage, for each degree of temperature (Chapter 6).
- Test method
- The complete procedure by which a maker decides whether a product meets a requirement: equipment, settings, steps, conditions, sample size, calculation and acceptance criteria (Chapter 23).
- Test uncertainty ratio
- The tolerance of a quantity divided by the 95 % expanded uncertainty of the measurement that tests it; the four-to-one rule dates from MIL-STD-45662A (1988) (Chapter 12).
- Test-method validation
- Evidence, fixed in advance, that a test method run as written gives results that are right and that repeat (Chapter 23).
- Thermal noise
- Also Johnson noise: the random voltage that the thermal motion of charges produces in every conductor, growing with resistance, absolute temperature and bandwidth (Chapter 5).
- Tolerance limit
- The upper or lower bound of the values that a property itself may take, such as the monitor's ±3 mmHg at 150 mmHg (Chapters 12 and 13).
- Tone
- A signal at a single frequency (Chapter 7).
- Total variation (TV)
- In a gauge study, the root-sum-of-squares of the gauge's R&R and the part variation (Chapter 22).
- Transducer
- A device that turns one physical quantity into another, such as a pressure into a voltage (Chapter 4).
- Transfer standard
- The artifact that travels between the laboratories of a comparison, such as the piston and cylinder of APMP.M.P-K6 (Chapter 21).
- True quantity value
- A value consistent with the definition of a quantity; in the VIM's traditional approach unique and, in practice, unknowable (Chapter 2).
- Trueness
- The closeness of the average of many repeated readings to a reference value; it reflects systematic error, not random error (Chapter 2).
- Type A evaluation
- Evaluation of a standard uncertainty by statistical analysis of a series of observations (Chapters 2, 10).
- Type B evaluation
- Evaluation of a standard uncertainty by any other means: a data sheet, a certificate, a standard's limit or the step of a display (Chapters 2, 11).
- Type evaluation
- An authority's tests on samples of a new design of measuring instrument before it is sold; India's law calls the step approval of the model (Chapter 30).
- Uncertainty budget
- The table of every input to a measurement with its standard uncertainty, distribution, divisor and sensitivity coefficient, from which the combined standard uncertainty is computed (Chapter 11).
- Variance
- The square of a standard deviation (Chapter 22).
- Verification (legal metrology)
- In OIML's vocabulary, a conformity assessment procedure, other than type evaluation, that ends in a verification mark or certificate on an instrument; a third sense of the word, apart from the VIM's verification and design verification (Chapter 30).
- Verification (metrological)
- Objective evidence that an item meets specified requirements, such as a maker's limit (VIM 2.44); distinct from calibration, and from design verification (Chapter 14).
- VIM
- The International Vocabulary of Metrology, the shared dictionary of measurement terms; its third edition, JCGM 200:2012, is still in force as of November 2026, and a draft of the fourth went to a vote in January 2026 (Chapter 1).
- Voltage reference
- The part that holds the voltage against which a converter counts, and so acts as the instrument's internal standard (Chapter 6).
- Weighing mode
- Also force mode: the Kibble balance's measurement in which a current through the coil balances the weight of a mass (Chapter 20).
- Working standard
- A reference used routinely to calibrate other standards; the BIPM's 1 kg working standards lost 18 to 88 µg through use between 1992 and 2014 (Chapter 18).
- Worst case
- A way of combining an error budget that adds every share at its limit, as if every error sat at its edge in the same direction at once (Chapter 25).
- XRCD (X-ray crystal density)
- The method that realizes the kilogram by counting the atoms in a silicon-28 sphere from the sphere's volume and the size of one crystal cell (Chapter 20).
Let's Talk