13 — Decisions

Pass, fail, or cannot say.

+ The questionWhen a reading lands near a specification limit, how should the decision be made?

A reading on the edge

A monitor on the production test reads exactly at the edge of its permitted error, and the test's own uncertainty reaches past the edge on both sides. The reference manometer holds the air in the test vessel at 150.0 mmHg, and the monitor's display shows 153, so the measured error is +3.0 mmHg. The maker's test engineer wrote this test and must now decide whether the monitor passes. The permitted error is ±3 mmHg or 2 % of the reading, whichever is greater, at any point of the range and with the pressure rising or falling, under R 149-1, OIML's rules for automated sphygmomanometers. India's rules for the same devices, Part VII-B, whose new text has been in force since July 2025, set the same limit (Chapter 30). Since 2 % of 150 mmHg is 3 mmHg, the two branches meet at 150 mmHg, as Chapter 11 showed, so the limit there is exactly 3 mmHg.

Above 150 mmHg the documents differ, so the engineer must name the one whose limit the test applies. OIML's and India's limit grows with the reading, to 5 mmHg at 250 mmHg, since 2 % of 250 mmHg is 5 mmHg. One maker, Zhuhai Yueja, states a flat cuff-pressure limit of ±3 mmHg at any measuring point under IEC 80601-2-30, the international standard for automated sphygmomanometers, whose 2018 edition is in force as of October 2026. The maker gives that limit in the summary of its 510(k), the premarket submission that Chapter 31 describes, and FDA, the US regulator, cleared that monitor. The summary is the maker's own document, and in it the maker names the standard for the monitor's performance testing without stating the edition. At 150 mmHg all three documents give 3 mmHg, so the monitor on the edge faces the same limit whichever one the engineer names.

The measured error of +3.0 mmHg is itself uncertain, and the engineer estimates by how much from the same kinds of error that a calibration laboratory counts. Chapter 11's five rows that carry a number describe the uncertainty of that error: the reference manometer, the hysteresis, the display step, the converter step and the noise. The German calibration service, DKD, counts rows of the same kind, among them the gauge's resolution and hysteresis, in the worked example of its guideline for calibrating pressure gauges. Doubled by Chapter 12's coverage factor of 2, the combined 0.70 mmHg of the five rows becomes an expanded uncertainty of 1.4 mmHg. The engineer takes that 1.4 mmHg as the expanded uncertainty of the error that the production test measures at 150 mmHg, an assumption that leaves out the two residuals that Chapter 11 named without a number. With about 95 % coverage, the monitor's true error then lies anywhere within 1.4 mmHg of the measured 3.0, from 1.6 to 4.4 mmHg, on both sides of the limit.

33 — One measured error, three acceptance limits
PASS CONDITIONAL PASS CONDITIONAL FAIL FAIL simple acceptance: 3.0 mmHg root-difference-square (Fluke): 2.65 mmHg acceptance limit, w = U: 1.6 mmHg TOLERANCE LIMIT 3 mmHg +1.2 +2.0 +4.0 assumed measured errors the edge: true error anywhere from 1.6 to 4.4 mmHg (U = 1.4 mmHg) +3.0 0 1 2 3 4 5 MEASURED ERROR AT 150 mmHg, mmHg guard band w = U = 1.4 mmHg Sources: ILAC G8:09/2019 4.2.3; Fluke FCM 7008.1 (company's own). U from Chapter 11's five rows, k = 2 (an assumption for the production test).
UncertaintyLimits and decisionsReadingFocal detail
Plate 33 — A reading at the limit cannot be called on its own: its uncertainty straddles the limit. The rule decides, and three common rules put the acceptance limit at 3.0, 2.65 or 1.6 mmHg.

Neither the limit nor the reading decides whether that monitor passes; the engineer needs a rule that says how the uncertainty counts. Calibration laboratories face the same question, and they answer it under ISO/IEC 17025, the standard against which they are accredited (Chapter 16). That standard defines a decision rule as a "rule that describes how measurement uncertainty is accounted for when stating conformity with a specified requirement", in the words that ILAC reproduces. The engineer's rule must relate two limits that ILAC keeps apart in its guidelines on decision rules; ILAC was the international cooperation of accreditation bodies until the end of 2025 (Chapter 16). The tolerance limit bounds the values that the property itself may take, here the monitor's true error. The acceptance limit bounds the measured values that the test will accept. A decision rule states where the acceptance limit sits relative to the tolerance limit, and for the monitor at +3.0 mmHg that position decides between pass and fail.

Simple acceptance shares the risk

Under the simplest rule the engineer passes every monitor whose measured error does not exceed the limit, so the acceptance limit equals the tolerance limit. ILAC calls this rule simple acceptance. The JCGM, the committee behind the GUM, describes the rule as important and widely used in its guide on conformity assessment, and notes that it is also known as shared risk. The monitor at +3.0 mmHg therefore passes, and whoever relies on it later carries the risk. Accepting an item that in fact lies outside its tolerance limit is a false accept, and under simple acceptance its probability may be as high as 50 %, by ILAC's guidelines. UKAS, the UK's accreditation body, places that 50 % at values on the tolerance limit, which is exactly where this monitor's reading lies: its true error is as likely to lie above the limit as below it.

Simple acceptance is common, and it suits a test whose uncertainty is small beside the limit. Mensor's laboratory states on the certificate of Chapter 12 that it applies "simple acceptance regardless of uncertainty". Taiwan's Bureau of Standards, Metrology and Inspection set a rule of the same kind for verifying mechanical sphygmomanometers in 2006: "the indication errors shall not exceed the maximum permissible error of verification", with no mention of the uncertainty of the test. Rules of that kind rely on a separate ratio rule for the reference, like those of Chapter 12. In legal metrology, the measurement that law regulates (Chapter 30), the usual criterion is an expanded uncertainty no larger than one third of the maximum permissible error, as the JCGM writes it, which is a ratio of three to one in Chapter 12's terms. A test uncertainty that small keeps the doubtful band narrow. That band holds the measured errors within one U of the limit, and at U's coverage probability of about 95 %, only for those could the true error lie on the other side of the limit.

The monitor's test does not meet that criterion. One third of 3 mmHg is 1 mmHg, while the test's 1.4 mmHg is 47 % of the limit, since 1.4 divided by 3 is 0.47. The doubtful band therefore runs from 1.6 to 4.4 mmHg, a wide stretch of the errors that a production line can produce. Inside that band the engineer must judge each monitor by its own measured value. The risk attached to one item is its specific risk, in the JCGM's term: the probability that this accepted monitor does not conform, or that this rejected one does.

The specific risk can be large even for a reading inside the limits, as the JCGM shows with an example from motor oil. The oil's viscosity, its resistance to flow, must lie between 12.5 and 16.3 square millimeters per second, written mm²/s, and the measured value is 13.6 mm²/s. The reading lies only 1.1 mm²/s above the lower limit, about 0.6 of its standard uncertainty of 1.8 mm²/s, and by the JCGM's figures the probability that the oil conforms is only about 66 %. Although the reading lies inside the limits, the chance that the oil lies outside them is therefore about one in three. For the monitor read on the limit itself, that chance is up to one in two.

A guard band moves the limit inward

To protect the user, the engineer moves the acceptance limit inside the tolerance limit. ILAC defines the guard band as the interval between a tolerance limit and the corresponding acceptance limit, and writes its width, w, as a multiple r of the expanded uncertainty: w = rU. A common choice is r = 1, so the guard band equals U.

ILAC states that with w = U the probability of a false accept is below 2.5 %. A reading at the acceptance limit lies one U inside the tolerance limit, so its true value lies beyond that limit only past the top of its 95 % band, where half of the 5 % outside the band lies. The JCGM writes the standard uncertainty as u, so that U = 2u, and gives the same case as w = 2u, with a false-accept probability of at most 2.3 % for a reading at the acceptance limit. The JCGM's figure is half of the 4.55 % outside a 95.45 % band, so the two figures differ only through Chapter 12's two conventions for coverage.

With w = U the engineer moves the monitor's acceptance limit from 3.0 to 1.6 mmHg, since 3.0 minus 1.4 is 1.6. A monitor whose measured error at 150 mmHg is 1.6 mmHg or less passes outright, because even the upper end of its uncertainty stays within the 3 mmHg limit. Near the limit the engineer need not force a pass or a fail, because ILAC allows statements with more than two outcomes in its guidelines. A measured result inside the guard band and below the tolerance limit is a "Conditional Pass". A result above the tolerance limit but below the tolerance limit plus the guard band, here 4.4 mmHg, is a "Conditional Fail", and only a result beyond 4.4 mmHg is a plain fail. A conditional result tells whoever relies on the monitor that the test cannot say on which side of the limit its true error lies.

The monitor at +3.0 mmHg falls exactly on the line between the two conditional outcomes, and ILAC's wording, "below" the tolerance limit for one and "above" it for the other, places it in neither. The engineer's rule must therefore say where a reading on a boundary goes, and ILAC expects the rule to be agreed before the measurements are taken.

A guard band set outward protects the person measured instead. In the JCGM's example of speed enforcement, the limit is 100 km/h and the threshold for a fine about 107 km/h. A driver is fined only when the measurement shows, with 99.9 % confidence, that the limit was exceeded. The inward band of a production test protects the other party, the person who later relies on the monitor.

One company's rule, with its numbers

A guard band can depend on the test uncertainty ratio of Chapter 12, as it does in the rule that Fluke Corporation publishes for stating conformity on its calibration certificates. Fluke applies no guard band when the ratio is at least four to one. Below that ratio Fluke moves the acceptance limit inward, by one of four methods that it names, two of them from ILAC's guidelines. With the first, Fluke subtracts U from the specification, the limit that the instrument must meet. With the second, a root-difference-square form, Fluke takes the square root of the specification squared minus U squared. Fluke bases its statements on an estimated false-accept risk of 2 %, the limit that ANSI/NCSL Z540.3 set (Chapter 12). Fluke issues a "conditional pass" when an error lies within the specification but the false-accept risk could exceed 2 %.

Fluke does not mention blood-pressure monitors in its policy, but the engineer can apply the policy to one. Since 3 divided by 1.4 is 2.14, the ratio of the 3 mmHg limit to the test's 1.4 mmHg is 2.1 to one, below four to one, so the policy calls for a guard band. With the first form the engineer subtracts U and reaches an acceptance limit of 1.6 mmHg, the same as ILAC's w = U. The root-difference-square form gives an acceptance limit of 2.65 mmHg: 3 squared minus 1.4 squared is 9 minus 1.96, or 7.04, its square root is 2.65, and 3 minus 2.65 leaves a guard band of only 0.35 mmHg. Under that form the monitor at +3.0 mmHg is a conditional pass in Fluke's sense, within the specification but with a false-accept risk that may exceed 2 %. The two forms differ by 2.65 minus 1.6, or 1.05 mmHg of acceptance limit, and on a production line that difference decides how many monitors are scrapped.

Solved example: four monitors at 150 mmHg under three rules

The limit at 150 mmHg is 3 mmHg, and the test's expanded uncertainty is U = 1.4 mmHg (Chapters 11 and 12). The four measured errors below are assumed, for illustration.

  • Acceptance limits: simple acceptance 3.0 mmHg; guard band w = U, 3.0 − 1.4 = 1.6 mmHg; root-difference-square, the square root of (3.0² − 1.4²) = the square root of (9 − 1.96) = the square root of 7.04 = 2.65 mmHg. A result above 3.0 + 1.4 = 4.4 mmHg fails under ILAC's statements.
  • Error +1.2 mmHg: pass under all three rules.
  • Error +2.0 mmHg: pass under simple acceptance and under the root-difference-square form; a conditional pass under w = U, because 2.0 lies between 1.6 and 3.0 mmHg.
  • Error +3.0 mmHg: pass under simple acceptance; on the boundary between conditional pass and conditional fail under w = U; a conditional pass under the root-difference-square form, as Fluke uses it.
  • Error +4.0 mmHg: fail under simple acceptance and under the root-difference-square form; a conditional fail under w = U, because 4.0 lies below 4.4 mmHg.

The same four monitors give three different sets of results, and only the rule differs. Under simple acceptance the maker ships the monitor at +3.0 mmHg with up to an even chance that it is outside its limit, while under w = U the engineer holds back both middle monitors for a further look. A smaller U would narrow every band at once. Halving a row quarters its square, so if the reference manometer's and the hysteresis's rows of Chapter 11 were halved, their squares of 0.213 and 0.187 mmHg² would fall to 0.053 and 0.047 mmHg², and the sum of the five squares from 0.485 to 0.185 mmHg². Its square root is 0.43 mmHg, so U, twice that, would fall from 1.4 to about 0.86 mmHg, the acceptance limit under w = U would rise to 3.0 minus 0.86, about 2.1 mmHg, and fewer good monitors would land in the doubtful band.

What each rule costs

Every decision rule trades the user's risk against the maker's yield, the share of its output that passes the test. The engineer can see that trade only by counting the whole stream of items tested. The JCGM calls the risk counted that way a global risk: the probability, judged before any item is measured, that a non-conforming item will be accepted or a conforming one rejected. In the JCGM's example, precision resistors of nominal 1500 Ω must lie between 1499.8 and 1500.2 Ω, and about 90 % of the process's output conforms. The measuring system has a standard uncertainty of 0.04 Ω, so U is 0.08 Ω, and the guard band of 0.02 Ω is a quarter of U. An inspector applying that guard band accepts 84 resistors in every 100 and rejects 16.

34 — What a rule costs, counted per hundred
JCGM 106 EXAMPLE: 100 RESISTORS nominal 1500 Ω, limits 1499.8 to 1500.2 Ω u = 0.04 Ω; guard band 0.02 Ω = a quarter of U accepted, conforming: 83 rejected, conforming: 7 global producer's risk 7 % accepted, not conforming: 1 global consumer's risk 1 % rejected, not conforming: 9 84 accepted, 16 rejected: seven good parts scrapped for each bad part shipped ANEROID GAUGES, 2008 DISSERTATION share of gauges that passed 60 % passed: measured error alone 12 % passed: uncertainty counted a fifth as many 0 % 50 % 100 % Silva Junior, PUC-Rio master's dissertation, 2008 (abstract); n not stated Sources: JCGM 106:2012, 9.5.3 (illustrative example); Silva Junior 2008, PUC-Rio (abstract).
UncertaintyLimits and decisionsFocal detail
Plate 34 — Every rule trades the user's risk against the maker's yield. In the JCGM's example seven good parts are scrapped for each bad part shipped; for aneroid gauges, counting the uncertainty cut the pass rate from 60 % to 12 %.

Of the 84 accepted resistors, 83 conform and 1 does not, a global consumer's risk of 1 in 100, or 1 %, carried by whoever receives the parts. Of the 16 rejected, 9 are truly out of tolerance and 7 conform, a global producer's risk of 7 in 100, or 7 %, carried by the maker. Seven good parts are therefore scrapped for every bad part shipped. Moving the acceptance limits inward would ship fewer bad parts and scrap more good ones, and moving them outward would do the reverse. On the monitor's production line the same count tells the engineer what each guard band costs in good monitors scrapped.

Counting the uncertainty changes the result for real sphygmomanometers as well, and by a large factor. In 2008 Silva Junior tested new and used aneroid gauges against a pressure standard for a master's dissertation in metrology at PUC-Rio, a university in Rio de Janeiro, supervised by Costa Monteiro. Of the gauges, 60 % passed when only their measured errors were counted, and 12 %, a fifth of 60 %, once the uncertainty of the measurement was counted, as the dissertation's abstract reports. The abstract gives neither the number of gauges nor the exact rule. Silva Junior proposed a lower limit that takes the uncertainty into account.

No guard band written for sphygmomanometers by OIML, a national metrology institute or an accreditation body has been found in open sources as of October 2026. Under the rules in force, a ratio rule governs the reference and simple acceptance governs the decision. The nearest is a proposal by Ramírez-Barrera and colleagues, in the 2021 bench study of aneroid gauges that Chapter 10 described. Citing the JCGM, they propose a safety margin of no more than 30 % of the tolerance interval. They put that margin at 2.7 mmHg, taking 30 % of the range of diastolic pressures associated with prehypertension, a band of slightly raised blood pressure. Applied to the 6 mmHg between the monitor's limits of −3 and +3 mmHg, the same 30 % would give 1.8 mmHg. They do not say whether the margin applies to the error alone or to the error with its uncertainty, and the margin is a proposal, not a rule in force. An engineer who adopts it must therefore still decide how to apply it.

A maker that tests its own monitors chooses its own rule, and the duties that ISO/IEC 17025 places on laboratories show what that choice involves. The standard's 2017 edition gives a laboratory three such duties, as ILAC and the accreditation bodies describe them: the laboratory agrees the decision rule with its customer, documents it with regard to the risk it carries, and identifies it in any statement of conformity. In UKAS's words, the rule "shall be described in the certificate or report". By the same reasoning, the engineer's rule should be fixed before testing, documented with its risk and named on each test record.

The engineer makes that choice at every pressure the test checks, not only at 150 mmHg. Under OIML's test procedure the engineer takes the pressure from 0 mmHg to the top of the scale in steps of no more than 50 mmHg, and India's rules use the same steps. At each step the engineer applies the limit that the named document sets there, with a guard band from the test's uncertainty at that step.

Test pressureOIML R 149-1 and India's Part VII-BIEC 80601-2-30, as one maker's 510(k) summary states it
50 mmHg±3 mmHg±3 mmHg
100 mmHg±3 mmHg±3 mmHg
150 mmHg±3 mmHg±3 mmHg
200 mmHg±4 mmHg±3 mmHg
250 mmHg±5 mmHg±3 mmHg
Writing the decision rule for a production pressure test
  • Write the rule into the test procedure before the first monitor is tested: the document whose limit applies, the test's expanded uncertainty at each step and the acceptance limit that follows.
  • Compute the ratio of limit to uncertainty at each step. At four to one or better, Fluke, for one, applies no guard band; below it, choose a guard band and state its form, w = U or root-difference-square, with the false-accept risk it is meant to hold.
  • State where a measured error exactly at a boundary goes, and what happens to a conditional result: a repeat test, a recalibration or rejection.
  • Count the yield each rule costs, and reduce the test's uncertainty, with a better reference or a sensor with less hysteresis, before narrowing the acceptance limit further.
  • Keep the cuff-pressure decision apart from the estimate: a monitor that passes the pressure test has shown nothing about its systolic and diastolic values (Chapter 24).
35 — The limit at each test step, by document
0 1 2 3 4 5 6 PERMITTED ERROR, mmHg 0 50 100 150 200 250 300 TEST PRESSURE, mmHg steps of no more than 50 mmHg (OIML R 149-2; India Part VII-B) 3 mmHg 3 mmHg 4 mmHg 5 mmHg all three documents: 3 mmHg branches meet: 2 % of 150 = 3 mmHg acceptance limit with w = U, 1.6 mmHg (U at 150 mmHg only) OIML R 149-1, India Part VII-B: 2 % of the reading IEC 80601-2-30, flat ±3 mmHg (K232814, maker's summary) Sources: OIML R 149-1:2020 5.1; G.S.R. 498(E) 2025; K232814 (2023). The guard band is shown only where U is computed.
Limits and decisionsFocal detail
Plate 35 — Up to 150 mmHg the documents agree on 3 mmHg; above it, OIML and India allow 2 % of the reading while one maker's description of the IEC limit stays flat. A production test names its document and applies its rule at every step.
+ What this chapter established
  • A decision rule states how measurement uncertainty is counted when conformity is declared, and without one a monitor whose measured error equals its 3 mmHg limit at 150 mmHg can be neither passed nor failed with confidence.
  • Simple acceptance sets the acceptance limit at the tolerance limit and accepts a risk of up to 50 % at the limit, which suits a test whose uncertainty is a small fraction of the limit.
  • A guard band of w = U moves the monitor's acceptance limit from 3.0 to 1.6 mmHg and holds the false-accept risk at that limit below about 2.5 %, while the root-difference-square form that Fluke names allows 2.65 mmHg.
  • Every rule trades the user's risk against the maker's yield: in the JCGM's resistor example seven good parts were rejected for each bad one accepted, and in Silva Junior's 2008 dissertation counting the uncertainty cut aneroid gauges' pass rate from 60 % to 12 %.