Why believe a certificate.
+ The questionWhy should anyone trust a calibration certificate from a laboratory they have never visited?
A register of laboratories
A monitor maker's reference manometer goes out for calibration and comes back with a certificate from a laboratory that the maker's staff have never visited. The maker cannot audit that laboratory itself, so it relies on an accreditation body that has audited the laboratory against a published standard. Accreditation is an independent body's assessment of a laboratory against such a standard, followed by its formal recognition that the laboratory is competent for a stated list of tasks.
In India that body is NABL, the National Accreditation Board for Testing and Calibration Laboratories, a constituent board of the Quality Council of India. NABL accredits testing and calibration laboratories against ISO/IEC 17025, the international standard for their competence. It also accredits the providers of proficiency tests, in which laboratories' results are compared, and producers of reference materials.
NABL's register holds far more laboratories than any customer could visit. In early September 2026 it held 1,305 calibration laboratories, by the count that Germany Trade & Invest, or GTAI, a German federal agency that reports on foreign markets, gave on 30 September 2026 in a market report on laboratory technology in India. The report calls the laboratories certified, but certification is the word for quality systems, and NABL accredits laboratories.
What the laboratory must show
A laboratory accredited under ISO/IEC 17025 must control its own instruments, because it calibrates its customers' instruments with them. It verifies each instrument before putting it into service, and calibrates it where its accuracy affects the validity of results or where traceability needs it, as SADCAS, which accredits laboratories in southern Africa, and PJLA, a US accreditation body, describe the standard's clause on equipment. It keeps a calibration program, labels what needs calibration, and takes doubtful equipment out of service and examines the effect. Its records for each instrument include the software and firmware version, the calibration dates and results, and the due date. ISO published the standard's third edition in November 2017, confirmed it in 2023 and sells it rather than publishing it openly. UKAS in the United Kingdom and NIST, in its checklist for certificates, also describe its requirements.
A certificate shows the numbers a laboratory reports, but not whether the laboratory reached and reported them as the standard requires. As accreditation bodies describe the standard, the laboratory traces each result to the SI through a documented unbroken chain: by calibration at a competent laboratory, by certified reference materials from a competent producer, or by realizing the SI unit directly (Chapter 15). Its certificates carry the four items that Chapter 14 listed, from the uncertainty of each result to the results before and after any adjustment, where those results are available. Where it states that an instrument conforms to a specification, it documents the decision rule it applied, and the measurement uncertainty "shall be taken into account", in UKAS's words (Chapter 13). Accreditation is the customer's evidence that a qualified body has checked this work.
The scope and its CMC
An accreditation covers only the calibrations that its scope lists, and each row of the scope gives a range, a method and a calibration and measurement capability, or CMC. The CMC is the capability "available to customers under normal conditions": the smallest uncertainty that the laboratory may state for that calibration, given as an expanded uncertainty with a coverage probability of about 95 % (Chapter 12). A national institute's CMCs are the ones that the BIPM's database publishes (Chapter 21). On a certificate the laboratory states both the coverage factor and the coverage probability, and gives the expanded uncertainty to at most two significant digits. The definition and these rules are those of ILAC, the International Laboratory Accreditation Cooperation, through which accreditation bodies cooperated until the end of 2025, in its policy on uncertainty in calibration, issued in September 2020.
A certificate's uncertainty can be larger than the CMC but never smaller: a laboratory "shall not report a smaller measurement uncertainty than the uncertainty described by the CMC", in ILAC's words. The laboratory works out its CMC from an uncertainty budget (Chapter 11) for the best existing device, ILAC's name for the best instrument of the kind that the laboratory calibrates. On a certificate it replaces that device's rows with the rows of the customer's own instrument, such as its resolution, repeatability and hysteresis, and no customer's instrument is better than the best of its kind. Reported uncertainties therefore "tend to be larger" than the CMC, as ILAC puts it, and the CMC is the best that a customer's reference manometer can receive from that laboratory.
A laboratory accredited for one quantity or range may still calibrate others, and those certificates rest on the laboratory's word alone. ILAC, in its traceability policy, accepts an accredited laboratory's calibration only where "the scope of accreditation specifically covers the appropriate calibration". The maker therefore reads the row of the scope that matches the calibration, not the logo on the certificate.
Accreditation alone does not tell the maker whether a laboratory can support the cuff-pressure check, because accredited capabilities at the cuff's pressures lie far apart. S M Gauge Company, a UK laboratory accredited by UKAS, lists dead-weight testers, pressure balances of the kind in Chapter 15, only above 2.5 MPa, so at the cuff's pressures its accreditation covers calibration of gas gauge pressure by comparison with a reference instrument. Its CMC rises with the pressure: 2.9 Pa from −15 to 15 kPa, 4.4 Pa from 15 to 25 kPa, 11 Pa from 25 to 50 kPa and 18 Pa from 50 to 100 kPa, each at a coverage probability of about 95 %. Even the largest, 18 Pa, is about 0.14 mmHg, since 1 mmHg is 133.322 Pa (Chapter 3), so every row is a small fraction of the 1 mmHg step of the monitor's display.
PSG CIRD, the industrial research and development center of PSG College of Technology in Coimbatore, offers a much coarser capability, 5.8 mmHg at k = 2. Its NABL scope, valid from 30 January 2026 to 29 January 2030, gives that CMC to its rows for sphygmomanometers from 50 to 200 mmHg and for the blood-pressure channels of patient monitors from 50 to 400 mmHg, measured "Using Vital Sign Simulator by Direct Method". PSG CIRD does not say whether those rows check the cuff pressure or the values that a simulator presents to the monitor's algorithm.
- 150 mmHg is 20.0 kPa (Chapter 15), inside S M Gauge's row from 15 to 25 kPa, whose CMC is 4.4 Pa.
- 4.4 Pa divided by 133.322 Pa per mmHg is 0.033 mmHg, at a coverage probability of about 95 %.
- A reference manometer calibrated there carries an uncertainty of at least 0.033 mmHg. OIML's limit for the maker's reference manometer is ±0.8 mmHg, and 0.8 divided by 0.033 is about 24, so the limit is about 24 times that CMC.
- PSG CIRD's CMC is 5.8 mmHg at k = 2, and 5.8 divided by 3 is 1.93, almost twice the ±3 mmHg that OIML allows the monitor's cuff-pressure reading. The interval that CMC places around any reported error is twice 5.8, or 11.6 mmHg wide, against a limit band 6 mmHg wide.
The first scope can support the reference manometer with a wide margin. The second cannot show that a monitor meets ±3 mmHg, because even a reported error of zero would carry an interval from −5.8 to +5.8 mmHg, which extends past the ±3 mmHg limit on both sides. Under any decision rule that takes the uncertainty into account (Chapter 13), no monitor could be passed on such a certificate. Both laboratories are accredited, and only the scope shows which of them fits the cuff-pressure check.
Laboratories tested against each other
Accreditation also tests a laboratory's results, not only its procedures, through proficiency testing: a provider sends the same item, or items of known value, to several laboratories and compares what each reports. Before granting accreditation, the accreditation body asks each laboratory for a participation plan and confirms "representative and satisfactory participation" for its scope, under ILAC's policy on proficiency testing, issued in January 2024. Where a laboratory's performance is not satisfactory, the body looks for evidence of "prompt and appropriate corrective actions". A laboratory's stated capability is therefore checked against measurements of shared items, not only against its own records.
How often a laboratory takes part is its own decision, made according to risk. Participation is "considered mandatory when available, appropriate and deemed necessary", in ILAC's words, but ILAC sets no fixed frequency, and each laboratory sets a minimum for every area of its competence. Where no formal proficiency test is available for a calibration, a laboratory in India can still set its results against other laboratories' results, because NABL has a separate document for such comparisons between calibration laboratories.
Trust across borders
A certificate from Coimbatore or from the United Kingdom is believed in another country because accreditation bodies assess each other and sign an arrangement to recognize one another's accreditations. That arrangement was ILAC's, and on 1 December 2025 it had 121 signatories representing 122 economies. On 1 January 2026 Global Accreditation Cooperation Incorporated, known as Global ACI, began full operations. It replaced two bodies, ILAC and the International Accreditation Forum, known as IAF, and on that day their two arrangements passed into one, the Global ACI MRA, or Multilateral Recognition Arrangement. Existing accreditations "will continue to be recognised", as ILAC and IAF stated in their joint announcement.
During the change to Global ACI, a certificate and the rules it was issued under both stay valid. UKAS told its customers on 15 May 2026 that the new Global ACI MRA mark could not be used until a registration process was complete, and UKAS expected the old marks to end in three to five years. Documents with the old marks stay valid until they are renewed, reissued, updated or otherwise replaced, by UKAS's account, so a certificate with ILAC's mark keeps its standing in the meantime. ILAC's rules also stand: as of October 2026, ILAC still lists its policies on traceability, uncertainty and proficiency testing as current on its site. NABL states that it is a signatory to the Global ACI arrangement and to APAC, the regional cooperation for Asia and the Pacific, so the accreditations it grants in India carry into the new arrangement (Chapter 30).
The device maker's own laboratory
A device maker may calibrate its own references instead of sending them out. Under ISO 13485, the quality-system standard for device makers, the maker calibrates or verifies its measuring equipment against traceable standards (Chapter 17), but none of the open descriptions of that clause, FDA's training for auditors among them, settles whether it requires accreditation. Micro Precision, a calibration company with an interest in selling accredited calibration, writes that the clause "doesn't explicitly require" an accredited provider. ILAC wrote its traceability policy for accredited organizations, not for a maker's factory laboratory. For a maker that calibrates in-house, its own records are therefore the only evidence of the chain, the uncertainty and the competence, because no accreditation body has checked them.
- Read the scope before the certificate: the quantity, the range, the method and the CMC at the test pressure.
- Choose a laboratory whose CMC at the test pressure is a small fraction of the limit the reference must meet; at 150 mmHg, a CMC of 0.033 mmHg against the reference's ±0.8 mmHg leaves a wide margin.
- Check that the certificate carries the accreditation symbol, the coverage factor and the coverage probability, and an uncertainty no smaller than the CMC.
- Check that the scope was valid on the date of calibration, and ask for the laboratory's proficiency-testing record for that scope.
- Where the maker calibrates in-house, record the chain, the uncertainty budget and the staff's competence, because no accreditation body has checked them.
- An accreditation body assesses a laboratory against ISO/IEC 17025 on the customer's behalf, and in early September 2026 NABL's register held 1,305 calibration laboratories, by GTAI's count.
- An accreditation covers only the calibrations in its scope, each with a CMC, an expanded uncertainty at about 95 % that is the smallest a certificate from that laboratory may state.
- Two accredited scopes at the cuff's pressures differ widely: 0.033 mmHg at 150 mmHg in one UK laboratory, against 5.8 mmHg at k = 2 in an Indian one, too coarse to show that a monitor meets ±3 mmHg.
- Proficiency testing checks laboratories' results against each other, and since 1 January 2026 the Global ACI arrangement carries the mutual recognition that makes a certificate accepted in other countries.
Let's Talk