What the ± promises.
+ The questionWhat does a "±" at about 95 % promise, and how do you read it off a certificate?
Two pages from a pressure laboratory
An accredited calibration laboratory prints a small number beside each reading it reports, and a real certificate shows what that number promises. On 9 February 2024 Mensor's calibration laboratory in San Marcos, Texas, calibrated a CPG2400 for the Washington State Patrol and issued a two-page certificate. The CPG2400 is a digital pressure gauge with a silicon pressure sensor, made by Mensor. The gauge reads absolute pressure, measured from vacuum rather than from the air around it (Chapter 4). Its range runs from 8 to 17 pounds per square inch, written psi, about 55 to 117 kPa. That is the range of the air's own pressure, so the gauge serves as a barometric reference and not as a reference for a cuff, but the engineer reads a cuff tester's certificate in the same way.
Mensor's laboratory applied 11 pressures, rising from 8 to 17 psi and falling back, and recorded each on its own line with the reference's value, the gauge's reading, the error as a percentage of the reading and the word "Pass". At about 15.2 psi on the rising side, for example, the reference stood at 15.19995 psi and the gauge read 15.2017 psi. Mensor gives the error as 0.0116 % of the reading, inside the gauge's limit of error of 0.030 % of the reading. Beside every line stands the small number, which grows from 0.00019 at 8 psi to 0.00038 at 17 psi. The heading of that column is not legible in the copy the patrol published, so the engineer reads its figures as psi, the unit of the reference and reading columns. Mensor explains the number in one sentence: "The combined and expanded uncertainty is reported at K=2 providing an approximate 95% confidence level."
Why the factor is 2
A combined standard uncertainty is one standard deviation, a measure of spread, while anyone who uses a reading needs a band that very probably holds the true value. The engineer obtains that band, the expanded uncertainty, written U, by multiplying the combined standard uncertainty by a number called the coverage factor, k. The chance that the band holds the true value is its coverage probability. The European co-operation for Accreditation, EA, requires accredited calibration laboratories to use k = 2, which gives a coverage probability of approximately 95 %.
In Chapter 11 the engineer ended with a combined standard uncertainty of 0.70 mmHg for the monitor's reading at 150 mmHg. Twice 0.70 mmHg is 1.4 mmHg, so that reading carries an expanded uncertainty of ±1.4 mmHg. Because U is the combined standard uncertainty multiplied by k, the step also runs backward: an engineer who receives a certificate divides its U by its k to obtain the standard uncertainty for a budget like Chapter 11's.
The 95 % rests on the bell curve of Chapter 10. A band with k = 2 reaches two standard deviations on each side of the reading. A bell curve holds 95.45 % of its values within that distance of its center, and exactly 95 % would need a factor of 1.96, as UKAS, the UK's accreditation body, notes. Laboratories therefore write "approximately" or give both figures. A São Paulo laboratory writes "aproximadamente 95,45%" on its certificate for a patient simulator, and NIST, the US national metrology institute, writes "approximately 95 percent (95.45 %)" in its example certificate for weights and measures laboratories. EA and UKAS call the figure a coverage probability and the two certificates call it a confidence level, but all of them attach it to the same factor of 2.
A factor of 2 gives about 95 % only under two conditions, which EA sets out: the result follows a bell curve closely, and its combined standard uncertainty is reliable. A result built from three or more contributions of comparable size approaches a bell curve, even when the contributions themselves are flat, as EA explains. The monitor's corrected channel in Chapter 11 meets the first condition, with three flat rows of similar size.
When the factor is not 2
When a standard uncertainty rests on few readings, the second condition fails and the factor grows, because a standard deviation computed from a few readings is itself uncertain. Chapter 10's ten readings showed this: drawn from a bell curve with a standard deviation of 0.4 mmHg, they estimated it as 0.44 mmHg. EA accepts the factor of 2 as reliable only if no contribution comes from a Type A evaluation of fewer than ten readings. Otherwise the engineer computes the result's effective degrees of freedom, a count derived with EA's formula from the degrees of freedom of each contribution, and reads k from EA's table. With 4 degrees of freedom the table gives k = 2.87, with 9 it gives 2.32 and with 10 it gives 2.28. As the count grows without limit, the factor falls to 2.00.
With few readings the table makes U larger. Chapter 10's ten readings meet EA's condition, so EA accepts k = 2 for their mean, although its table gives 2.32 for the 9 degrees of freedom of ten readings. Had the engineer taken k from the table instead, 2.32 times the mean's standard deviation of 0.14 mmHg would have given 0.32 mmHg, against the 0.28 mmHg that k = 2 gives.
Ramírez-Barrera and colleagues met the same effect in a real budget, in their bench calibrations of four aneroid gauges in Medellín, published in 2021 (Chapter 10). For the first gauge at 120 mmHg they report a combined standard uncertainty of 0.672 mmHg, 19.22 effective degrees of freedom and an expanded uncertainty of ±1.4 mmHg. They take k from the effective degrees of freedom, and EA directs the engineer to round 19.22 down to 19, for which its table gives 2.14. The product of 0.672 mmHg and 2.14 is 1.44 mmHg, which fits the printed ±1.4 mmHg, while k = 2 would have given 1.3 mmHg.
When one row, the budget's dominant contributor, supplies almost all of the total of squares, it also moves the factor away from 2. When that contributor is a Type A evaluation, its degrees of freedom set the factor through the same statistics as EA's table, as a worked example by UKAS shows. When it is a flat Type B limit, the result keeps the flat shape of Chapter 11 instead of approaching a bell curve, and UKAS works this case through with a digital voltmeter's limit. The monitor's corrected channel has no such row: its largest row supplies 44 % of the total of squares and two others supply most of the rest, so the factor of 2 serves it.
A certificate read line by line
The engineer reads a certificate as a series of checks, and each item on Mensor's certificate answers one of them. First the engineer confirms that the certificate describes the instrument in hand: Mensor names the issuer, its procedure and accreditation mark, the customer, and the gauge's serial number, sensor and range. Next the engineer finds the limit that the gauge was judged against, 0.030 % of the reading, because "Pass" means nothing without it. The two dates show how long the result is meant to serve: the calibration on 9 February 2024 and the recalibration due on 9 February 2025.
The as-found calibration is meant to describe the gauge as it arrived, but Mensor records that the zero was adjusted by 0.001 psi before it. Those readings therefore describe the gauge after a small adjustment, not exactly as it arrived, and the engineer keeps the adjustment and the calibration apart, as Chapter 14 does.
The word "Pass" rests on a rule that Mensor states: an asterisk would mark any point exceeding the limit of error on "simple acceptance regardless of uncertainty". "Pass" therefore compares the measured error alone with the limit and leaves the uncertainty out, a rule that Chapter 13 examines. None of the eleven lines carries an asterisk, and every error lies below 0.012 % of the reading, under half of the 0.030 % limit. The gauge passed at every point.
The engineer also follows the chain one link up from the gauge. Mensor names the reference it used, listed as DH CPD 8000, whose own calibration falls due in December 2024, and that reference is the next link (Chapter 15). Mensor also states its route of traceability: A2LA, the American Association for Laboratory Accreditation, accredits Mensor's laboratory, and A2LA had signed the arrangement in which the members of ILAC, the international cooperation of accreditation bodies, recognized one another's accreditations. On 1 January 2026 that arrangement passed into a single new one under Global ACI, which replaced ILAC and lists A2LA among the accreditation bodies it recognizes (Chapter 16).
A tester's certificate in the cuff's range
A patient simulator that tests monitors is calibrated in the cuff's own range. On 2 March 2020 LRM, the Laboratório de Referência Metrológica in São Paulo, calibrated the non-invasive pressure channel of a Rigel Uni-SiM, one such simulator, and issued a four-page certificate. LRM reports its results for non-invasive pressure in 13 lines, from 30 to 300 mmHg, with readings to 0.01 mmHg. For each it gives two pressures, the error between them in mmHg and in percent, the expanded uncertainty and k = 2.00. The uncertainty rises from 0.02 mmHg at 30 mmHg to 0.18 mmHg at 300 mmHg, and at 150 mmHg it is 0.09 mmHg. LRM states no limit and no conformity and says that its results refer only to the instrument calibrated, so the engineer receives measurements with their uncertainty and no verdict.
The certificate does not give the simulator's accuracy between calibrations, and for that the engineer turns to the maker. Rigel Medical's own manual for the Uni-SiM lists an accuracy of ±0.5 %FS beneath the entry for its digital manometer, a limit for the instrument in use. The manometer's range is 0 to 410 mmHg and the simulator's built-in pump covers 0 to 350 mmHg, but the manual does not say which full scale the percentage takes. The engineer therefore cannot tell whether the limit is 0.5 % of 410 mmHg, which is 2.05 mmHg, or of 350 mmHg, which is 1.75 mmHg. The certificate's 0.09 mmHg, by contrast, describes one calibration on one day. NIST states the difference in its example certificate: "The expanded uncertainty is not to be confused with a tolerance limit for the user during application."
LRM's 0.09 mmHg describes how well the laboratory knew the simulator's pressure at 150 mmHg on 2 March 2020. That figure does not describe how far the pressure moves with use, transport or time. For a test made months after the calibration, the engineer therefore enters in the budget the maker's limit, or a drift measured over the interval (Chapter 17), and not the certificate's U alone. The engineer in Chapter 11 followed that rule by entering OIML's limit for the reference manometer rather than a certificate's figure.
The reference and the way each reading is formed decide how small a certificate's figures can be, because the reference's own uncertainty and the scatter of the readings both enter the budget (Chapter 11). The Meteorological Instrument Center of the Japan Meteorological Agency publishes a sample certificate for a barometer with its values masked. On that certificate the reference is a pressure balance, the piston loaded with masses that Chapter 15 describes. Each indication is the mean of six one-minute readings, a Type A evaluation, so averaging divides a single reading's scatter by the square root of 6, about 2.4 (Chapter 10).
One laboratory whose reference is itself a simulator claims far larger figures. In Chennai, an accredited calibration laboratory at IIT Madras uses a vital-sign simulator as its reference for the non-invasive blood-pressure channel of patient monitors, from 10 to 400 mmHg. Its scope of accreditation gives the smallest uncertainty it can claim there, its calibration and measurement capability (Chapter 16), as 9.56 % at the low end of the range and 0.93 % at the top, about 1.0 mmHg at 10 mmHg and 3.7 mmHg at 400 mmHg. Both lie far above the 0.09 mmHg that LRM printed at 150 mmHg with a standard manometer among its references.
Four to one
The engineer judges whether a certificate's U is small enough through one ratio. The test uncertainty ratio is the tolerance of the quantity tested, the limit its error must stay within, divided by the 95 % expanded uncertainty of the measurement that tests it, in ILAC's definition. In August 1988 the US Department of Defense wrote the four-to-one rule into its standard MIL-STD-45662A, whose clause on the adequacy of measurement standards reads: "the collective uncertainty of the measurement standards shall not exceed 25 percent of the acceptable tolerance for each characteristic being calibrated." An uncertainty of 25 % of the tolerance is a ratio of four to one.
The Department of Defense has since cancelled the standard, but civilian standards adopted its ratio: it passed into ANSI/NCSL Z540-1 in 1994, as Song of NIST records in a 1997 paper. The ratio is a practical substitute for controlling the risk of a wrong decision, as SIMCO, a calibration services company, describes it. A wrong decision becomes possible whenever a measured error lies within U of the limit, because the true error may then lie on the other side of it.
The successor standard replaced the ratio with the risk itself. In ANSI/NCSL Z540.3, published in 2006, NCSL International required that the probability of a false accept decision, accepting an instrument that is in fact outside its limit, "shall not exceed 2%", as Mimbs of NASA's Kennedy Space Center quotes it in a 2008 paper. The standard allowed a ratio of four to one or better when determining that probability was not practicable. NCSL International withdrew Z540.3 in October 2020 as superseded by ISO/IEC 17025:2017, the international standard for the competence of calibration and testing laboratories (Chapter 16).
The ratio remains in ILAC's guidelines of 2019, but the JCGM, the committee behind the GUM, warns in its guide on conformity assessment that such ratios "are sometimes ambiguously or incompletely defined". ILAC's guidelines add that a rule built on such a ratio must also specify what action to take, and one company's rule in Chapter 13 sets that action around the same 2 %.
For the monitor, the engineer sets the reference manometer against the ±3 mmHg limit. OIML allows the reference manometer a maximum permissible error of ±0.8 mmHg in its test procedure for automated sphygmomanometers. OIML states a limit rather than an uncertainty, but the engineer applies the ratio rules to it in the same way. Against ±3 mmHg, a reference within ±0.8 mmHg gives 3.75 to one, or 27 % of the tolerance, just short of the military four to one.
Germany sets a looser rule in its regulation on operating medical devices, the MPBetreibV: test equipment is accurate enough if its error limits and uncertainties do not exceed one third of those of the device tested. One third of 3 mmHg is 1 mmHg, so a reference within ±0.8 mmHg meets the German rule. The same reference fails four to one and meets one third, so the engineer records which ratio rule the test applies.
The monitor's limit at 150 mmHg is ±3 mmHg (Chapter 11).
- The simulator's certificate: U = 0.09 mmHg at k = 2, so the standard uncertainty is 0.09 ÷ 2 = 0.045 mmHg, the figure the engineer enters in a budget.
- The simulator as a reference on the day of its calibration: 3 ÷ 0.09 = 33, a ratio of 33 to one.
- OIML's limit for the reference manometer: 3 ÷ 0.8 = 3.75 to one, and 0.8 ÷ 3 = 27 % of the tolerance. Four to one would need a reference within 3 ÷ 4 = 0.75 mmHg; Germany's one-third rule allows 3 ÷ 3 = 1.0 mmHg.
- The monitor's own reading: U = 2 × 0.70 = 1.4 mmHg, so 3 ÷ 1.4 = 2.1 to one, and 1.4 ÷ 3 = 47 % of the tolerance.
The calibrated simulator meets every ratio rule by a wide margin. OIML's reference meets Germany's one third with 0.2 mmHg to spare but falls short of four to one. The monitor's own reading leaves far less margin. Its expanded uncertainty uses 47 % of the ±3 mmHg limit, so a monitor whose measured error at 150 mmHg lies between about 1.6 and 3 mmHg may or may not truly be inside its limit. The maker's test engineer therefore needs a rule for those monitors (Chapter 13).
- Find k and the stated coverage, and divide U by k before entering the figure in a budget.
- Check the unit and the basis of every figure: psi or mmHg, absolute or gauge pressure, percent of the reading or of full scale.
- Check that the calibrated points cover the range in which the instrument will be used: Mensor's barometric range does not cover a cuff.
- Note the reference listed on the certificate and the date on which its own calibration falls due.
- Read the rule by which the laboratory decides "Pass", and keep the certificate's U apart from the maker's limit for the instrument in use.
- Compute the ratio of the limit tested against to the U, and record which ratio rule the test applies.
- An expanded uncertainty is a combined standard uncertainty multiplied by a coverage factor, and accredited laboratories use k = 2, which gives about 95 %, or 95.45 % for a bell curve, so the monitor's 0.70 mmHg at 150 mmHg becomes ±1.4 mmHg.
- The factor rises above 2 when a result rests on few readings: below ten readings EA directs the engineer to its table, which gives 2.87 for 4 degrees of freedom, and in a 2021 budget for an aneroid gauge its authors printed ±1.4 mmHg where k = 2 would have given 1.3 mmHg.
- A certificate's U describes one calibration: a São Paulo laboratory's 0.09 mmHg at 150 mmHg for a patient simulator is not the simulator's ±0.5 %FS accuracy in use.
- The test uncertainty ratio, set at four to one by a US military standard of 1988, puts OIML's ±0.8 mmHg reference at 3.75 to one against the monitor's ±3 mmHg, while the monitor's own ±1.4 mmHg uses nearly half of that limit.