+ Part I · Foundations
What a measured number is.
A bathroom scale, a clinical thermometer and a blood-pressure monitor look nothing alike, and each does the same thing: it compares. Each sets the quantity in front of it against a reference built into it, and someone once set that reference against a better one. Before any electronics or standards, this part fixes three ideas that the rest of the guide depends on. Chapter 1 shows what a measurement is, a comparison with a reference, and introduces the blood-pressure monitor that the guide follows to its end. Chapter 2 separates the error of a reading from what anyone can know about that error, its uncertainty. Chapter 3 shows why nations must agree on a unit before any two numbers can be compared.
A number is a comparison.
+ The questionWhat is an instrument doing when it shows a number?
The sensor measures air, not blood
An automatic blood-pressure monitor reports a person's blood pressure, and its sensor never touches the blood: it measures the pressure of the air in the cuff. Texas Instruments, known as TI, is a maker of chips, and it publishes a reference design for such a monitor, a worked example of its circuit, its parts and its software. In that design a pump inflates the cuff around the upper arm until the pressure reaches a target, and a valve then opens to let the air out. A tube carries the air pressure from the cuff to a small silicon sensor on the monitor's circuit board. Everything the monitor later displays, the cuff pressure and the blood pressure alike, is worked out from that one air pressure as it falls.
The sensor in TI's design is the 2SMPP-03, a gauge-pressure sensor for air only, rated from 0 to 50 kPa, as Omron's data sheet describes it. A gauge-pressure sensor measures a pressure against the pressure of the air around it (Chapter 4). A kilopascal, kPa, is a thousand pascals, the SI unit of pressure (Chapter 3). With every heartbeat the artery under the cuff pushes against it, so the falling cuff pressure carries small pulsations, one for each beat, which this guide calls oscillations. In a typical monitor the oscillations measure from less than 1 mmHg to 3 mmHg, by the figures of Freescale, another chip maker. From them the monitor works out a blood pressure that its sensor never meets (Chapter 4).
Every measurement is a comparison
The international definition of measurement rests on comparison. The International Vocabulary of Metrology, known as the VIM, defines measurement as the process of "experimentally obtaining one or more quantity values that can reasonably be attributed to a quantity". A note to the definition gives its core in one sentence: "Measurement implies comparison of quantities or counting of entities." The VIM is the shared vocabulary of measurement, and the Joint Committee for Guides in Metrology, the JCGM, publishes it. As of November 2026 the edition in force is still the third, of 2012, although in January 2026 the JCGM sent a draft of the fourth edition to a vote.
A ruler shows the comparison plainly, because the length of a part is laid beside marks that stand for known lengths. The monitor compares in several steps that no user sees, and a measurement made this way, through a chain of signals and computations rather than against a scale, is called indirect. Its sensor compares the cuff pressure with the pressure of the air around the monitor (Chapter 4). Its firmware compares the sensor's signal with numbers stored at calibration, when the monitor's readings were set against those of a reference manometer, a pressure gauge whose own readings had been compared with a better standard still.
The number on the display is the last of these comparisons, so it can be no sounder than the ones before it. The opening pages showed what happens when one of those comparisons is never made: Perkin-Elmer's mirror matched its test instrument, and the check that no one carried out, between that instrument and an independent measure of length, decided the outcome. A further note to the VIM's definition lists what every measurement presupposes: a description of the quantity suited to the intended use of the result, a measurement procedure, and "a calibrated measuring system operating according to the specified measurement procedure". The monitor needs all three. Its quantity must be described well enough to say what is measured, its procedure must say how the cuff is inflated and read, and its measuring system must have been calibrated against a reference.
Measurand, indication and result
The measurand is the "quantity intended to be measured", in the VIM's definition. The word "intended" carries weight, because an instrument can measure a quantity close to the one intended without measuring the intended quantity itself. The monitor's pressure channel senses its own measurand, the gauge pressure of the air in the cuff at each moment of the deflation. Its blood-pressure output has a different measurand: the person's systolic and diastolic pressures in the artery under the cuff, the highest and lowest pressures of each heartbeat. Specifying a measurand requires knowing the kind of quantity and the body that carries it, as a note to the definition adds. Both measurands are pressures, but one is carried by air in a cuff and the other by blood in an artery.
An instrument's indication is the "quantity value provided by a measuring instrument or a measuring system", in the VIM's words. An indication and the value of the quantity being measured "are not necessarily values of quantities of the same kind", a note warns, and the monitor shows why. In TI's design the sensor's output runs from about 0 to 30 mV over 0 to 40 kPa, as TI states, so the sensor's indication is a voltage. The converter's indication is a count, a whole number. Only the display's indication is a pressure in mmHg, and each change of kind on the way rests on a relation that someone had to establish and that can be wrong.
In the VIM's definition, a measurement result is the "set of quantity values being attributed to a measurand together with any other available relevant information". A note adds that a result "is generally expressed as a single measured quantity value and a measurement uncertainty". A quantity value is a number together with its unit, so a complete result carries three things: a number, its unit and its uncertainty.
A home monitor shows its value alone, such as 150 and its unit. The VIM's note on the measurement result allows the value alone when the uncertainty is negligible for the purpose, and says that in many fields this is the common way. The monitor's maker states a limit of error in the manual. Microlife, a monitor maker, gives that limit for one of its home monitors as "pressure within ± 3 mmHg". Such a limit tells the user how far any reading may stray, but not how far a particular one did. Chapter 2 shows why a limit of error is not an uncertainty, and Part III shows how an uncertainty is worked out.
Six stages from the cuff to the display
The monitor's cuff-pressure channel, which the guide follows from here to its last chapter, passes through six stages, each named by what it does: sense, condition, digitize, compute, connect, decide. Each stage changes the signal, and each can add an error. In TI's reference design a named part does the work of each stage.
To sense, the 2SMPP-03 uses a bridge of piezoresistors, resistors formed in silicon that change their value when the silicon flexes under pressure. TI drives the bridge with a constant current of 100 µA, the drive that Omron rates it for, and pressure unbalances the bridge so that a small voltage appears at its output (Chapter 4). To condition, an instrumentation amplifier, TI's INA350, multiplies that voltage by 50. An instrumentation amplifier is built to measure a small difference of voltage such as the bridge's output (Chapter 6). A filter then passes the oscillations to a path of their own. It weakens the signal below 1.7 hertz, or cycles per second, and above 5.9 hertz, and so it blocks the steady level that the cuff pressure gives (Chapter 6).
To digitize, a 12-bit converter turns the amplified voltage into a count by comparing it with a reference voltage of 2.5 V (Chapters 7 and 8). The converter writes each count as 12 binary digits, or bits, and twelve doublings give 4,096 possible counts. It is built into TI's MSPM0L1306 microcontroller, a chip that combines a small processor with circuits such as the converter. To compute, the firmware converts counts into pressure with coefficients set at calibration, and it runs the algorithm that turns the oscillations into blood pressure (Chapter 9). TI's own software for the design "leaves the blood pressure algorithm section blank", so the maker of each monitor supplies its own.
To connect, the monitor shows the result on its display, which steps in whole millimeters of mercury in Microlife's monitors, and an instrument may also store or send its results (Chapter 9). To decide, someone acts on the number: a production test passes or fails the monitor (Chapter 13), and the user or a doctor acts on the blood pressure it reports.
The reading and the estimate
The six stages produce two outputs, and the guide keeps them apart in every chapter. The reading is the cuff pressure in mmHg, and a chain of calibrations ties it to the SI. The monitor is compared with a reference manometer and the manometer with a better standard, and the chain reaches a pressure balance, a piston loaded with masses, which sets a pressure from the weight of the piston and its masses divided by the piston's area. Through those masses the chain reaches the kilogram (Chapters 14 and 15). The estimate is the systolic and diastolic blood pressure, which the firmware's algorithm derives from the oscillations. The oscillations are always largest when the cuff pressure equals the mean arterial pressure, the average pressure in the artery over a heartbeat, as Renesas, a third chip maker, states. The rules that lead from that point to systolic and diastolic values are each maker's own (Chapter 9).
The two outputs are checked against different references. OIML, the intergovernmental body that writes model rules for measuring instruments, sets a limit for the reading of an automated monitor: within ±3 mmHg or 2 % of the reading, whichever is greater, with the pressure rising or falling, at any temperature from 10 to 40 °C. OIML's test replaces the cuff with a rigid metal vessel of 500 ml and compares the monitor with a reference manometer at steady pressures, so no oscillation reaches the sensor during the test. For the estimate, OIML sets a separate requirement, which the maker meets through a clinical investigation on people (Chapter 24). Microlife states that its home monitor's blood-pressure values correspond to the Korotkoff method, which finds blood pressure from the sounds heard through a stethoscope over the artery. The reading is defined against a manometer, and the estimate against a method that a person carries out.
Because the pressure test brings no oscillations to the sensor, it never exercises the algorithm, and a monitor can meet OIML's limit at every test pressure and still report blood pressures that are wrong. In 2021 PTB, Germany's national metrology institute, proposed a new OIML guide for testing monitors with simulators that reproduce real oscillations. Its proposal argued that "any verification of the pressure measurement alone completely ignores the second step", the algorithm's step from oscillations to blood pressure.
The algorithm works on the reading, so a trustworthy reading is necessary for a trustworthy estimate, but it is not enough. The pressure test never exercises the algorithm, the cuff on the arm or the person's arteries, and each of those can move the estimate while the reading stays within its limit. Calibration evidence for the cuff pressure therefore says nothing about the accuracy of the systolic and diastolic values. Chapter 24 shows how a clinical validation proves them.
Take a cuff pressure of 150 mmHg during the deflation, with oscillations of the size Freescale gives as typical, 1 to 3 mmHg.
- Share of the cuff pressure: 1 mmHg divided by 150 mmHg is 0.67 %, and 3 mmHg divided by 150 mmHg is 2.0 %. Renesas describes the part of the signal that is of interest as about 2 %.
- The cuff pressure in kilopascals: 1 mmHg is 133.322 Pa (Chapter 3), so 150 mmHg is 150 times 133.322 Pa, which is 19,998 Pa, or 20.0 kPa.
- At the sensor: Omron gives a typical span, the output at full scale minus the output at zero pressure, of 42.0 mV for the full 50 kPa. On a straight line, 20.0 kPa gives 42.0 mV times 20.0 divided by 50, which is 16.8 mV above the sensor's output at zero pressure.
- The oscillations at the sensor: at 133.322 Pa for each mmHg, the full 50 kPa is about 375 mmHg, so each millimeter of mercury gives 42.0 mV divided by 375, which is 0.112 mV. Oscillations of 1 to 3 mmHg are therefore 1 to 3 times 0.112 mV, or 0.11 to 0.34 mV. TI expects 100 to 300 µV at its amplifier's input, measured from the lowest to the highest point of each oscillation, which is the same size.
The reading rests on the steady 16.8 mV above the sensor's output at zero pressure, and the estimate is drawn from oscillations of a few tenths of a millivolt on top of that level. OIML's pressure test holds the pressure steady in a rigid vessel, so it checks that steady level and never produces the oscillations. A monitor's calibration therefore shows nothing about how its algorithm treats the part of the signal that carries the blood pressure.
- Name the measurand of each output in the requirements: the air pressure in the cuff for the reading, the pressure in the artery for the estimate.
- Write a separate accuracy requirement for each output, with its own evidence: a test at steady pressures against a reference manometer for the reading, and a clinical validation for the estimate.
- List the six stages with the part that does each and the relation it applies, so that every change of kind, from pressure to voltage, from voltage to count and from count to mmHg, is tied to a named part.
- Record the reference behind each comparison, up to the reference manometer's calibration certificate, so that a reviewer can trace the chain behind the reading.
- A measurement is a comparison with a reference, and the monitor's display shows the last of a chain of comparisons that runs back through its calibration to a reference manometer and beyond.
- The measurand is the quantity intended to be measured, an indication can be a quantity of another kind, such as a voltage or a count, and a complete result carries a number, its unit and its uncertainty.
- The monitor's cuff-pressure channel passes through six stages, sense, condition, digitize, compute, connect and decide, each done by a named part in TI's reference design.
- The monitor gives a reading, the cuff pressure, which calibration ties to the SI, and an estimate, the blood pressure, which an algorithm derives from oscillations of about 0.7 % to 2 % of a 150 mmHg cuff pressure and which only a clinical validation can prove.
Let's Talk