Wrong by how much.
+ The questionIf no one can know the true value, what does it mean for a reading to be wrong?
Two monitors, both within their limit
Two monitors can read the same cuff pressure 6 mmHg apart, and both can be within their limit. The monitors of two makers, Microlife and A&D, read the cuff pressure to within ±3 mmHg by the makers' own figures, and OIML allows an automated monitor ±3 mmHg or 2 % of the reading, whichever is greater (Chapter 1). At a steady 120 mmHg, 2 % of the reading is only 2.4 mmHg, so the 3 mmHg branch applies. One monitor may then show 117 mmHg and another 123 mmHg, and each meets its limit, though only at its edge (Chapter 13). A tester who wanted to say which of them is wrong would need the true pressure in the cuff, and no one has the true pressure.
The air in the cuff has one definite pressure, but no one can know it exactly. In the VIM's words, a true quantity value is one "consistent with the definition of a quantity", and a note to that definition adds that, in the approach built on errors, the true value is "considered unique and, in practice, unknowable". The true value cannot be known exactly because every instrument compares the pressure with some reference, and every reference carries a doubt of its own. A tester who calls a reading wrong therefore compares it with a reference whose value is known well enough for the purpose. The tester then faces two questions: how far the reading lies from the reference, and how well anyone knows that distance. The first is the reading's error, and the second is its uncertainty.
Error is a difference from a reference
Measurement error is a difference with a sign: in the VIM's words, the "measured quantity value minus a reference quantity value". A note to the definition warns that error "should not be confused with production error or mistake", so an instrument's error is not a fault that someone made. A monitor that shows 152 mmHg while a reference manometer shows 150 mmHg has an error of +2 mmHg, and one that shows 148 mmHg has an error of −2 mmHg. In OIML's test procedure for automated monitors, the tester records exactly this difference, the monitor's reading minus the reference's, at a series of steady pressures.
Instruments in everyday use do not all read within 3 mmHg of a reference. A'Court and colleagues at the University of Oxford measured how many do in a cross-sectional survey of the sphygmomanometers, the instruments with a cuff that measure blood pressure, in use at 38 general practices in England. The team tested them between January and July 2009 and published the survey in 2011, and the National Institute for Health Research funded it. The testers compared each instrument with a reference pressure meter at 250, 200, 150, 100 and 50 mmHg. They fully tested 584 instruments, and 503 of them, or 86 %, read within 3 mmHg of the reference at every point. Another 77 instruments, 13 %, showed at least one reading 4 to 9 mmHg off, and 4 instruments, fewer than 1 %, showed one 10 mmHg or more off. Thirty-six of the 308 digital instruments failed.
Every one of those errors is a comparison with a reference meter, and no tester knows an error better than the reference that found it. The authors do not state how accurate their reference meters were, so an instrument recorded as 3 mmHg off could have been somewhat more or somewhat less off, by an amount that the reference meter's own doubt sets. The second question, how well anyone knows the error, therefore cannot be skipped.
Systematic and random error
Error has two parts, which differ in how each behaves when a measurement is repeated. In the VIM's words, a systematic error "in replicate measurements remains constant or varies in a predictable manner", while a random error "varies in an unpredictable manner". An estimate of a systematic error is called a bias in the VIM. The monitor's pressure channel produces both kinds. A single 2SMPP-03 sensor may have an offset, its output with no pressure applied, anywhere within 4.0 mV of the typical value, by Omron's figures, and whatever one sensor's offset is, it shifts every reading in the same direction by the same amount. TI's amplifier and the circuits around it add noise, less than 10 µV from peak to peak at the amplifier's input by TI's figure, and the noise moves each reading by a different amount in no predictable direction (Chapter 5).
The two parts call for different remedies. An engineer can remove a known systematic error with a correction, and in the method of the GUM, the international guide to uncertainty (Chapter 11), the correction comes before any uncertainty is stated: the GUM's working assumption is that a result "has been corrected for all recognized significant systematic effects". A monitor can remove its sensor's offset in this way. Before a measurement, its firmware reads the sensor open to the air, stores that reading and subtracts it from every later reading, as an Omron patent describes (Chapter 17). No correction can remove a random error from a single reading, but the firmware can reduce its effect by averaging many readings (Chapters 5 and 10).
In a real instrument, errors that keep their sign from one test pressure to the next reveal a systematic error. Ramírez-Barrera and colleagues measured errors of this kind in a bench study published in 2021. They worked at the Instituto Tecnológico Metropolitano, ITM, in Medellín, Colombia, and calibrated four aneroid sphygmomanometers, the mechanical dial gauges used with a stethoscope, chosen at random from different brands. ITM funded the work, and no author worked for a maker. The team followed OIML R 16-1, the model rule that OIML replaced with R 148 in 2020, and compared each gauge with a digital reference manometer in two rising and two falling series. The fourth gauge read low at every point the authors report, by 0.235 to 2.710 mmHg, so it carried a systematic error.
Accuracy, trueness and precision
Accuracy, trueness and precision, three words that everyday speech uses loosely, mean three different things in measurement. Measurement accuracy is, in the VIM's words, the "closeness of agreement between a measured quantity value and a true quantity value of a measurand", and a note to the definition adds that accuracy "is not a quantity and is not given a numerical quantity value". By the same note, a monitor is called more accurate when its error is smaller.
A monitor has good trueness when the average of an unlimited number of its repeated readings lies close to a reference value. Over so many readings the random errors average out, while a systematic error stays in every one, so trueness responds to systematic error and not to random error. A monitor has good precision when its repeated readings agree closely with each other, in the VIM's words the "closeness of agreement between indications or measured quantity values obtained by replicate measurements". Engineers usually state precision as a standard deviation, the usual measure of how widely values scatter about their average (Chapter 10). A note to the definition warns that "measurement precision" is sometimes "erroneously used to mean measurement accuracy".
Precision depends on what stays the same between the repeated readings, so it is stated under one of two sets of conditions that the VIM defines. Under repeatability conditions, the same operators use the same procedure, measuring system, operating conditions and location over a short period. Under reproducibility conditions, the locations, operators and measuring systems differ. A production tester who reads one monitor several times in a minute measures its repeatability. An engineer who compares monitors from different lines, tested by different people at different sites, measures their reproducibility. The first spread shows how closely one monitor repeats itself, and the second how closely monitors agree across lines, people and sites.
Makers use the word accuracy differently. When Microlife and A&D state an accuracy of ±3 mmHg, they give a limit within which the error must lie, and that limit is a number, while the VIM's accuracy has none. Such a limit is a maximum permissible measurement error in the VIM's term, and OIML's ±3 mmHg is a limit of the same kind. The limit tells a user how wrong a monitor is allowed to be. It does not tell how wrong one particular monitor is.
High precision means only that repeated readings agree with each other. A monitor whose sensor carries a large uncorrected offset can show the same value every time it is tested at one pressure, with every reading displaced by the same amount. Its precision is then excellent and its trueness poor, and only a comparison with a reference reveals the displacement. A small spread of repeated readings is therefore no evidence of accuracy.
Resolution is not accuracy
An instrument's resolution, the smallest change it can show, sets no limit on its error. In the VIM's words, resolution is the "smallest change in a quantity being measured that causes a perceptible change in the corresponding indication", and a note adds that noise or friction can limit it. A display's resolution is the "smallest difference between displayed indications that can be meaningfully distinguished", usually its last step.
An analytical balance shows the difference between resolution and accuracy. The Sartorius CPA224S displays to 0.1 mg, repeats within 0.1 mg and departs from a straight line by less than 0.2 mg, twice its display step, by its maker's figures. Shimadzu's figures for its AUW220D balance follow the same pattern. A balance that shows a tenth of a milligram can therefore be wrong by two tenths, and Chapter 8 sets the monitor's 1 mmHg display step against its ±3 mmHg limit in the same way.
Uncertainty states what is known
A second term answers the second question, how well anyone knows a reading's error. Measurement uncertainty is, in the VIM's words, a "non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand, based on the information used". In the GUM's plainer words, the uncertainty of a result "reflects the lack of exact knowledge of the value of the measurand". A monitor's error at one reading is a single number with a sign, and no one ever knows it exactly. The uncertainty of that reading is a stated width of doubt around the result, without a sign, and an engineer can work it out and write it down.
Uncertainty has many sources, and the monitor supplies an example of most of those the GUM lists. A tester may define the measurand incompletely, by reading a cuff pressure without saying at which moment of the deflation. The engineer may know the conditions imperfectly, such as the sensor's temperature. The display has a finite resolution, the reference manometer used at calibration has an inexact value, the firmware's method rests on approximations, such as a straight line through a curved response, and repeated readings vary. An uncertainty expressed as a standard deviation is a standard uncertainty, in the GUM's term. The engineer finds one by statistics on repeated observations in a Type A evaluation, and from any other information, such as a data sheet or a certificate, in a Type B evaluation (Chapters 10 and 11).
Neither the monitor that shows 117 mmHg nor the one that shows 123 mmHg can be called wrong from its display alone, because each reading's error is unknown. A calibration laboratory can measure each monitor's error at a test pressure against a reference and state an uncertainty with it, and only that pair of numbers shows whether a monitor is likely to be within its limit.
Ramírez-Barrera and colleagues report three figures for their fourth aneroid gauge at 120 mmHg:
- the measured error, the gauge's reading minus the reference's: −2.430 mmHg;
- the uncertainty they state with it: ±1.1 mmHg (Chapter 12 explains what such a figure promises);
- their Type A figure, the standard uncertainty that they work out by statistics from the repeated readings: 0.28 mmHg (Chapter 10).
The measured error and its uncertainty together place the gauge's error between −2.43 minus 1.1, which is −3.53 mmHg, and −2.43 plus 1.1, which is −1.33 mmHg. OIML's current limit for such gauges, in R 148, is ±3 mmHg, and the range from −3.53 to −1.33 mmHg crosses it. The measured error lies inside the limit, while the lowest 0.53 mmHg of the range, from −3 to −3.53 mmHg, lies beyond it. Whether a test should pass such a gauge is a question of decision rules, which Chapter 13 takes up.
The error is 2.43 divided by 0.28, about 8.7 times that Type A figure. The scatter of the repeated readings is far too small to produce an error that large, so most of the error repeated in every series. It is a systematic error: reading the gauge again would not reduce it, while adjusting the gauge, or correcting its readings, would. The reference manometer contributed little, with an uncertainty of ±0.062 mmHg on its certificate. Most of the ±1.1 mmHg comes from the gauge itself: its scale steps of 2 mmHg, its scatter and its hysteresis, the difference between its readings with the pressure rising and falling.
- State every error as the instrument's reading minus the reference's, with its sign, and name the reference.
- Give the uncertainty of each measured error beside it, because an error close to a limit means nothing without it.
- Say whether repeated readings were taken under repeatability conditions or across operators, instruments and sites, since the two spreads answer different questions.
- Write a maker's "accuracy" in a requirement as what it is, a maximum permissible error, and never treat a display's resolution as a claim of accuracy.
- An error is a reading minus a reference value, with a sign, and because a true value is unknowable in practice, every error is known no better than the reference that measured it.
- Systematic error repeats and can be corrected once it is known, random error varies unpredictably and is reduced by averaging, and a gauge whose errors keep one sign at every pressure shows the first.
- Accuracy has no number in the VIM, trueness reflects systematic error and precision reflects scatter, while a maker's "±3 mmHg" is a maximum permissible error, a limit rather than a measurement.
- Uncertainty states how well an error is known, so a gauge's error measured as −2.43 mmHg with ±1.1 mmHg may lie on either side of a ±3 mmHg limit, while resolution, the smallest change an instrument shows, sets no limit on its error.
Let's Talk