From the Planck constant to the cuff.
+ The questionPut together, what makes the cuff pressure one blood-pressure monitor reports mean what it says?
Two ends of one chain
The two ends of the monitor's chain can be compared directly, because both are standard uncertainties and both can be stated as a fraction of the value they describe. At the top stands the kilogram's consensus value, from which mass calibrations have taken their traceability since 1 March 2026 (Chapter 21). The CCM, the consultative committee that advises the CIPM on mass, gives that value a standard uncertainty of 20 parts per billion. At the bottom, the engineer of Chapter 11 found the monitor's combined standard uncertainty at 150 mmHg to be 0.70 mmHg, and 0.70 divided by 150 is about 4.7 parts per thousand. The relative uncertainty at the bottom is therefore about 230,000 times as large as at the top. Chapter 19 found about 470,000 with the 10 µg, or 10 parts per billion, that the BIPM gave the prototype at the redefinition of 2019; the consensus value's uncertainty is twice that, so the 230,000 is about half.
Every link between the two ends is a comparison, and each comparison carries the uncertainty of the link above it and adds its own (Chapter 15). Uncertainties therefore "grow progressively through the chain of measurements", as Eicke and colleagues wrote in NASA's measurement assurance guidelines of 1995. They drew the consequence for the top of the chain: "as the length of the chain increases, requirements at the higher echelons become more stringent". Each added link adds its own uncertainty before the chain reaches the limit at its bottom, so a longer chain needs a smaller uncertainty at its top. The monitor's chain runs from the defining constant down to the cuff, and Chapter 15 followed its lower links at 150 mmHg, from a national pressure balance to the monitor's limit, through uncertainties and maximum permissible errors. At each link an institute, a laboratory or a maker states the figure and answers for it.
From a constant to a pressure
The chain starts with a number that has no uncertainty at all. In 2018 the General Conference on Weights and Measures, the CGPM, fixed the Planck constant, h, at exactly 6.626 070 15 ÷ 10³⁴ joule seconds, with effect from 20 May 2019 (Chapter 19). The meter and the second were already fixed through other constants, so that number defines the kilogram. A definition is not yet a mass, however, and a national metrology institute must realize the kilogram with an instrument. NIST's Kibble balance, NIST-4, measured h in 2017 with a relative standard uncertainty of 13 parts per billion, and NRC's balance in Canada reached 9.1 parts per billion. Now that h is fixed, the same measurement realizes a kilogram, so each figure is also the relative standard uncertainty of a kilogram realized on that balance (Chapter 20).
No single balance sets the world's mass scale. Several institutes realize the kilogram, their realizations are compared in key comparisons, and the CCM computes one consensus value from the results. The current consensus value carries a standard uncertainty of 20 µg (Chapter 21), and in a kilogram 20 µg is the 20 parts per billion at the top of the chain. National institutes that do not realize the kilogram themselves take the unit from the same value through the BIPM's calibrations, so their masses start from the same 20 parts per billion. India does so through its platinum-iridium copy No. 57, which CSIR-NPL keeps (Chapter 21).
On a pressure balance, masses under measured gravity press down on a piston, and the gas pressure that balances them equals their weight divided by the piston's effective area (Chapter 15). The pressure is therefore in proportion to the mass on the piston, so a fractional uncertainty in the masses becomes the same fractional uncertainty in the pressure. At 150 mmHg, 20 parts per billion is three millionths of a millimeter of mercury.
The piston's own area adds far more. NIST's gas pressure scale rests on two primary piston gauges, which cover 20 kPa to 1 MPa, as Olson and colleagues of NIST described it in 2010. Each gauge's effective area carries a relative standard uncertainty of 3.0 to 3.2 parts per million, and the other contributions, among them mass, gravity and temperature, about 2.5 parts per million. Combined by the root-sum-of-squares (Chapter 11), 3.2 and 2.5 parts per million give about 4.1 parts per million, about 200 times the kilogram's 20 parts per billion, so at this link the piston's area, not the kilogram, sets the uncertainty. The cuff's 150 mmHg is 20.0 kPa, the lowest pressure of the gauges' range, so these gauges reach down to the monitor's test pressure. There their 4.1 parts per million is 0.0006 mmHg.
Each national institute answers for its own link. NPL's figure of Chapter 15, about 0.003 mmHg at this pressure, is larger than NIST's because it belongs one link lower: NPL, the UK's national metrology institute, states it for calibrating a customer's pressure balance, and gives no coverage factor for it. India's primary standard is an air piston gauge kept by CSIR-NPL, which gives the gauge's effective area an expanded uncertainty of nearly 11 parts per million at k = 2 (Chapter 15). The institutes show in key comparisons that their links agree. The national institutes thus hand the laboratories below them a pressure known to parts per million.
From the laboratory to the bench
Below the national institutes, accredited laboratories calibrate reference instruments for industry, and at this link the uncertainty grows again. A laboratory of this kind compares the instrument with a reference instrument instead of generating a pressure from masses, and each comparison adds the repeatability and the resolution of the instrument being calibrated. UKAS accredits S M Gauge Company, a UK laboratory, to calibrate gas gauge pressure from 15 to 25 kPa with an expanded uncertainty of 4.4 Pa at about 95 %, by its schedule of June 2026 (Chapter 16). That figure is the laboratory's calibration and measurement capability, the smallest uncertainty it may state for such a calibration. At 150 mmHg, 4.4 Pa is 0.033 mmHg, and 4.4 Pa in 20.0 kPa is 220 parts per million. A reference manometer that this laboratory calibrates therefore reaches the maker's bench with an uncertainty no smaller than about 0.033 mmHg.
The maker owns the next link, and its budget counts the reference manometer at OIML's limit, not at the manometer's own calibration. At its calibration station the maker compares each monitor with a reference manometer (Chapter 14), and OIML allows that manometer a maximum permissible error of ±0.8 mmHg. In a 2021 bench study, Ramírez-Barrera and colleagues checked four sphygmomanometers against a reference manometer whose certificate gave an expanded uncertainty of ±0.062 mmHg, with a coverage factor of 1.99 (Chapter 15). That certificate figure is larger than the accredited laboratory's best capability above and far smaller than OIML's limit. The engineer of Chapter 11 counted the reference as 0.8 mmHg divided by the square root of 3, about 0.46 mmHg, because the limit is what the maker's procedure guarantees for every reference the station may use.
At the bench the monitor adds the largest uncertainties of the whole chain. The sensor's hysteresis and the display's 1 mmHg step join the reference's row, and together the three give most of the combined standard uncertainty of 0.70 mmHg, which becomes an expanded uncertainty of ±1.4 mmHg at k = 2 (Chapter 12). The production test then decides whether each monitor passes. With a guard band equal to the 1.4 mmHg, the maker passes a monitor only if its measured error at 150 mmHg lies within 1.6 mmHg of zero, the ±3 mmHg that OIML and India set less 1.4 mmHg (Chapter 13). Within the maker, the design team answers for the budget (Chapter 25), the test engineers for the decision (Chapters 22 and 23) and the service for keeping the monitor within its limit in use (Chapter 27).
- 150 mmHg is 150 × 133.322 = 19,998 Pa, about 20.0 kPa (Chapter 15).
- The kilogram's consensus value: 20 parts per billion of 150 mmHg is 150 × 20 ÷ 10⁹ = 0.000003 mmHg, as a standard uncertainty.
- NIST's primary piston gauges: the area's 3.2 parts per million and the other rows' 2.5 combine as √(3.2² + 2.5²) = √(10.24 + 6.25) = √16.49 = 4.1 parts per million. Of 150 mmHg that is 150 × 4.1 ÷ 10⁶ = 0.0006 mmHg, at k = 1.
- The accredited laboratory: 4.4 Pa ÷ 133.322 Pa per mmHg = 0.033 mmHg at about 95 %. Taking k = 2 for that coverage, the standard uncertainty is 0.033 ÷ 2, about 0.017 mmHg.
- The maker's reference manometer at OIML's limit: 0.8 ÷ √3 = 0.8 ÷ 1.732 = 0.46 mmHg (Chapter 11).
- The monitor: 0.70 mmHg, the square root of a total of squares of 0.485 mmHg² (Chapter 11). As a fraction, 0.70 ÷ 150 = 0.0047, which is 4.7 parts per thousand, and 0.70 ÷ 150 ÷ 0.00000002, the kilogram's 20 parts per billion, is about 230,000.
Suppose the whole of the laboratory's 0.017 mmHg reached the monitor's budget as a row of its own. Its square, 0.017² = 0.0003 mmHg², would add 0.0003 ÷ 0.485, about 0.06 %, to the 0.485 mmHg², and the root would stay at 0.70 mmHg. NIST's link, squared, is 0.0006², about 0.0000004 mmHg², less than a millionth of the total. Everything above the maker's reference manometer therefore adds almost nothing to the uncertainty of the reading, and the bench sets the uncertainty that the display carries.
What the chain answers, and where it stops
What makes the reading mean what it says has two parts, because the links above the bench and the bench itself do different work. The national institutes and the accredited laboratories decide whether the cuff pressure is tied to the SI at all, and they add almost nothing to its uncertainty. The maker's bench decides how large that uncertainty is. A maker who wants a smaller uncertainty must therefore work on the sensor, the firmware and the reference used at calibration. A maker who skips a link above the bench gains no accuracy and loses the tie to the unit.
For the cuff pressure that one monitor reports to mean what it says, the three conditions set out at the start of this guide must hold, and the chain names an owner for each part. First, both chains must be known and unbroken. Laboratories whose competence others have checked must have done, dated and recorded every calibration in the chain above the monitor (Chapters 15 and 16). The design team must give the monitor's own chain, its six stages from sense to decide, a budget that fits within its limit, with every share owned and verified (Chapters 11 and 25).
Second, the uncertainty must be stated and small enough for the decision. Each laboratory must state its uncertainty, small beside the limit of the link below (Chapter 12), and the maker must count the test's uncertainty when it decides whether a monitor passes (Chapter 13). Third, tests must show that the monitor stays within its stated uncertainty between calibrations, and the maker must find any fault or wrong reference and trace each back to the results it affected (Chapters 17, 26 and 27).
A chain in which every link is checked against the next guards against the failure of the opening story. In a checked chain the next comparison can catch a wrong reference long before a telescope reaches orbit.
The chain stops at the reading. The monitor also shows systolic and diastolic values, and its algorithm derives these estimates from the oscillations that the heartbeat makes in the cuff pressure (Chapter 9). A calibration of the cuff pressure never tests the rules that turn those oscillations into the estimate. Only a clinical validation against a reference method on people can prove the estimate (Chapter 24). A monitor whose cuff pressure traces to the Planck constant can still report a blood pressure that its algorithm gets wrong, so the two claims need two kinds of proof.
- List every link from the national institute to the display, with its owner, its figure, its coverage factor, the date of its calibration and the certificate that states it.
- Convert each link's figure to the monitor's unit at its test pressure, keeping each figure's own coverage factor, so that a reviewer can see where the uncertainty grows.
- Spend effort where the uncertainty arises: the reference used at calibration, the sensor's hysteresis and the display step, not the links above the bench.
- Check each link's uncertainty against the limit of the link below, and record the ratio.
- Record changes at the top, such as the consensus value of 1 March 2026, even when they are too small to change a result, so that the chain stays complete.
- Keep the reading's chain and the estimate's validation in separate sections of the file, with separate evidence.
The platinum-iridium cylinder that defined the kilogram from 1889 to 2019 is now measured like any other mass. Until 2019 its mass was exactly 1 kg by declaration, so no one could state an uncertainty for it (Chapter 18). The BIPM still keeps the cylinder at Sèvres, and since 1 March 2026 the CCM has put its mass at 1 kg − 12 µg in the consensus value, with a standard uncertainty of 20 µg.
- At 150 mmHg the standard uncertainty grows from 20 parts per billion in the kilogram's consensus value to about 4.1 parts per million at NIST's primary piston gauges, about 0.017 mmHg at an accredited laboratory and 0.70 mmHg at the monitor, about 230,000 times the top figure in relative terms.
- The links above the maker's reference manometer would add about 0.06 % to the monitor's total of squares, so the chain above the bench ties the reading to the SI while the bench sets the size of its uncertainty.
- A cuff pressure means what it says when three conditions hold: both chains, the calibrations above the monitor and the monitor's own budget, are known and unbroken; the uncertainty is stated and small enough for the maker's decision; and tests show that the monitor stays within that uncertainty between calibrations.
- The chain ends at the reading, so the blood-pressure estimate needs a clinical validation, and the prototype itself is now 1 kg − 12 µg with a standard uncertainty of 20 µg, twice the 10 µg it was given in 2019.
Let's Talk