+ Part III · Uncertainty

How sure, in numbers.

A calibration laboratory writes a second, smaller number after a ± sign beside every pressure it reports. In 2020 a São Paulo laboratory issued a certificate for the pressure channel of a patient simulator, an instrument used to test blood-pressure monitors, and on it the line for 150 mmHg carries ±0.09 mmHg. With that number the laboratory states how sure it is of its own result. A monitor's displayed reading carries the same kind of doubt, although the monitor prints none. Each stage of the signal chain in Part II adds an error of its own to the cuff-pressure reading, and the engineer can state in numbers the doubt that those errors leave. In Chapter 10 the engineer works out how widely repeated readings scatter, and in Chapter 11 combines every source of error from the cuff to the display into one figure. In Chapter 12 the engineer reads what the ± on a certificate promises, and in Chapter 13 the maker's test engineer turns a reading near a limit into a pass, a fail or a result that the test cannot decide.

10 — Scatter

Ten readings, one answer.

+ The questionHow do you state the scatter of repeated readings as one number?

Ten numbers for one pressure

Ten readings of one steady pressure give ten slightly different numbers on the monitor's channel. The engineer holds the air in a rigid vessel at 150.0 mmHg and connects the vessel to the monitor's tube and to a reference manometer, a gauge far more accurate than the monitor (Chapter 14). The reference shows that the pressure holds steady, yet the sensor, the amplifier, the converter and the firmware return a slightly different value at each reading. In the method of the GUM, the international guide to uncertainty, repeated readings are assumed to differ because the influence quantities of Chapter 4, the conditions besides the pressure that act on a measurement, are never held completely constant.

The ten readings are assumed: a computer program generated them, because no published figure for this scatter in a commercial monitor could be found. The program drew them at random and rounded each value to 0.1 mmHg. That rounding step is an assumption too. It stands for readings that the engineer takes from the firmware before the display rounds them to whole millimeters of mercury, a rounding that would give most of them the same number and hide the scatter. In the order drawn, the ten readings are 150.9, 149.7, 150.2, 150.1, 150.3, 149.4, 149.8, 149.7, 149.6 and 149.7 mmHg.

The mean and the spread around it

From the ten readings the engineer wants two numbers: a best value for the pressure, and a measure of how widely the readings spread around it, which says how closely one reading is likely to repeat another. The best estimate of a quantity from independent readings is their arithmetic mean, or average, by the GUM's rule. The engineer adds the ten readings, which come to 1,499.4 mmHg, and divides the sum by 10 to obtain a mean of 149.94 mmHg. The mean lies 0.06 mmHg below the reference's 150.0 mmHg. The program put no systematic error into the readings, so chance alone has moved the mean.

For the spread, the engineer starts from each reading's distance from the mean, its deviation. The ten deviations run from 0.54 mmHg below the mean to 0.96 mmHg above it, and positive and negative deviations always add to zero, so their sum cannot measure the spread. To obtain the standard deviation, the engineer squares each deviation, which makes every one positive, adds the squares and divides the total by one less than the number of readings. The square root of the result, which returns it to the readings' own unit, is the standard deviation. The divisor is 9 rather than 10 because the ten deviations must add to zero, so any nine of them fix the tenth, and that count of 9 is the degrees of freedom, in the GUM's term. For the ten readings the squares add to 1.744 mmHg², and the standard deviation comes to 0.44 mmHg.

The program drew the readings from a bell curve, the normal distribution of statistics. A bell curve makes values near its center more likely than values far from it, and it falls away in the same way on both sides. Its standard deviation sets its width, so a curve with a standard deviation of 0.4 mmHg is twice as wide as one with 0.2 mmHg. The program's curve was centered on 150.0 mmHg with a standard deviation of 0.4 mmHg, and both settings are assumptions, chosen to illustrate the method. The ten readings estimate that 0.4 mmHg as 0.44 mmHg, close but not equal, because ten draws sample the curve only coarsely.

24 — Ten readings of one pressure, and the spread around them
air RIGID VESSEL 150 mmHg MONITOR REFERENCE MANOMETER 150.0 mmHg generating curve (assumed): 150.0 mmHg, 0.4 mmHg reference 150.0 mean 149.94 mmHg 149.0 149.5 150.0 150.5 151.0 PRESSURE, mmHg 0.96 mmHg above the mean: part of the scatter, not a blunder standard deviation of the mean: 0.14 mmHg (the mean of ten) one standard deviation: 0.44 mmHg (a single reading) Generated readings (assumed generating values; fixed seed), not measured; statistics worked from the figures above.
The quantity being measuredReferences and standardsUncertaintyReadingFocal detail
Plate 24 — Single readings scatter by about 0.44 mmHg, but the mean of ten is known about three times better, to 0.14 mmHg. The readings are generated from an assumed bell curve, not measured.

A mean is surer than one reading

When the engineer averages the ten readings, high and low readings partly cancel, so the mean varies much less than a single reading, which can land anywhere under the bell curve. How much the mean varies is measured by the standard deviation of the mean, which by the GUM's rule is the standard deviation of single readings divided by the square root of their number. The square root comes from Chapter 5's rule for adding independent noises: the sum of ten readings holds ten times the pressure but only the square root of 10 times one reading's scatter. Dividing the sum by 10 therefore divides the scatter by the square root of 10, about 3.16, and the mean's standard deviation is 0.44 divided by 3.16, or 0.14 mmHg. The mean's distance of 0.06 mmHg from the reference is less than half of that figure, so chance explains it.

Because of the square root, each extra reading improves the mean less than the one before. Four times as many readings halve the standard deviation of the mean, because the square root of 4 is 2: 40 readings would bring it from 0.14 to about 0.07 mmHg, and 160 readings to about 0.035 mmHg. To divide it by ten, the engineer needs a hundred times as many readings, since the square root of 100 is 10. Repetition therefore helps most at the start, and the engineer stops adding readings once the mean's standard deviation is small beside the other sources of error that the engineer combines with it in Chapter 11.

Working out the doubt by statistics on a series of readings, as the engineer has done here, is a Type A evaluation in the GUM's term, and ten readings is the count that two common rules of practice name. Europe's accreditation bodies, working together as the European co-operation for Accreditation, EA, warn that a Type A evaluation resting on fewer than ten readings needs its reliability considered. A balance's repeatability is tested by weighing one test weight not less than ten times under the US Pharmacopeia's chapter on balances, as Mettler-Toledo and Troemner, two companies that sell weighing equipment, describe it. Neither EA nor the two companies say that ten readings are enough for every purpose: EA treats fewer than ten as a reason for caution, and the pharmacopeia's chapter, as the two companies describe it, does not allow fewer. The engineer therefore takes ten as a floor, not as a count that suits every test.

Solved example: the mean, the standard deviation and the standard deviation of the mean of the ten readings

The ten generated readings, in mmHg: 150.9, 149.7, 150.2, 150.1, 150.3, 149.4, 149.8, 149.7, 149.6 and 149.7.

  • Mean: the readings add to 1,499.4 mmHg, and 1,499.4 ÷ 10 = 149.94 mmHg.
  • Deviations from the mean: +0.96, −0.24, +0.26, +0.16, +0.36, −0.54, −0.14, −0.24, −0.34 and −0.24 mmHg, which add to zero.
  • Squares of the deviations: 0.9216, 0.0576, 0.0676, 0.0256, 0.1296, 0.2916, 0.0196, 0.0576, 0.1156 and 0.0576 mmHg², which add to 1.744 mmHg².
  • Standard deviation: 1.744 ÷ 9 = 0.1938 mmHg², and its square root is 0.44 mmHg.
  • Standard deviation of the mean: the square root of 10 is 3.162, and 0.44 ÷ 3.162 = 0.139 mmHg, about 0.14 mmHg.

The program generated the readings with a mean of 150.0 mmHg and a standard deviation of 0.4 mmHg, and the ten readings recover both to within what chance allows: the mean to 0.06 mmHg, less than half of its own standard deviation, and the scatter as 0.44 mmHg. For a channel with this assumed scatter, a single reading at 150 mmHg typically lies within about half a millimeter of mercury of the mean, half the display's step. Most repeated readings would therefore show the same number on the display, as eight of the ten do when rounded.

The standard deviation of the mean applies only to a result that is itself a mean. A monitor in use takes one reading of the cuff pressure at each moment, so the doubt that scatter adds to that reading is the standard deviation of a single reading, 0.44 mmHg, and not the 0.14 mmHg of a mean of ten. UKAS, the UK's accreditation body, applies the same rule to a weighing in a worked example: it finds a repeatability of 8.7 mg from ten readings, and because the result is the mean of three, it divides 8.7 mg by the square root of 3 and obtains 5.0 mg. Ten readings estimate the scatter, and the number of readings averaged into the result sets the divisor.

How large is 0.4 mmHg?

The electronics alone scatter far less than 0.4 mmHg. In Texas Instruments' reference design for a blood-pressure monitor, the front end of Chapter 5 adds less than 10 µV of noise from peak to peak, by TI's own claim rather than a measurement. The figure is input-referred, expressed as if the noise entered at the amplifier's input beside the sensor's signal, which is about 112 µV for each millimeter of mercury (Chapter 4). Divided by 112 µV per mmHg, 10 µV of noise is below 0.09 mmHg from peak to peak. Dividing 0.09 mmHg by 6.6, the peak-to-peak factor that the converter maker Analog Devices uses (Chapter 5), gives about 0.014 mmHg as a standard deviation. The design's converter steps in about 0.11 mmHg, the quantization step of Chapter 8, so the front end's scatter is smaller than a single count of the converter that reads it.

Laboratory calibrations of mechanical blood-pressure gauges give a measured scatter for a whole instrument. In a bench study published in 2021, Ramírez-Barrera and colleagues at the Instituto Tecnológico Metropolitano in Medellín, Colombia, calibrated four aneroid sphygmomanometers, the mechanical dial gauges of Chapter 2. The institute funded the work, and none of the authors worked for a company. Ramírez-Barrera and colleagues report their scatter as a figure for a mean, not for a single reading. For the first instrument at 120 mmHg they report a Type A figure of 0.42 mmHg, which they compute as the standard deviation of the instrument's repeated readings divided by the square root of their number, the rule for the standard deviation of a mean.

25 — How large 0.4 mmHg is, beside what sources give
×50 MCU 150 mmHg SENSOR AMPLIFIER CONVERTER FIRMWARE DISPLAY no figure given readings read here 0.01 0.1 1 10 mmHg electronics alone: under 10 µV peak to peak (TI's claim) = 0.014 mmHg as a standard deviation one converter count, 0.11 mmHg display step, 1 mmHg ±3 mmHg limit at 150 mmHg standard deviation of the generated mean, 0.14 mmHg assumed scatter of one reading, 0.4 mmHg (0.44 from the ten readings) aneroid gauge, Type A at 120 mmHg, 0.42 mmHg: standard deviation of a mean (Ramírez-Barrera et al. 2021) TI TIDUF53 (maker's claim); Ramírez-Barrera et al. 2021 (bench, 4 aneroid gauges); generated value assumed. Log scale.
The signal in the chainNoise and errorUncertaintyLimits and decisionsReadingFocal detail
Plate 25 — The assumed scatter is about 29 times the electronics' own. The aneroid gauge's 0.42 mmHg is the standard deviation of a mean, like the generated mean's 0.14 mmHg, which is about a third of it. A whole channel scattering by 0.4 mmHg would owe almost all of it to causes outside its front end and converter.

The assumed channel scatters about 29 times as much as its electronics, since 0.4 mmHg divided by 0.014 mmHg is about 29. If the whole channel scattered that much, almost all of its scatter would come from causes outside the front end and the converter, and no maker or study gives a figure for those causes in this channel. The generated mean's 0.14 mmHg is about a third of the aneroid gauge's 0.42 mmHg, and both are standard deviations of a mean. Ramírez-Barrera and colleagues do not state how many readings each mean contains, so the engineer cannot recover the aneroid gauge's single-reading scatter from their figure.

A repeatability test on the real channel would measure the channel's own scatter. In Chapter 11 the engineer combines the sources of error in the channel, and the electronics' noise is the only scatter for which a maker gives a figure. A measured scatter of the whole channel would enter there as a source of its own. At 0.44 mmHg it would be about as large as the largest source left after correction, the 0.46 mmHg that Chapter 11 derives from OIML's limit for the reference manometer. The engineer who runs the test therefore learns either that the channel repeats about as well as its electronics or that it has a source of scatter that no data sheet shows.

Repeatable is not the same as right

A small scatter shows that a channel repeats, not that it reads the true pressure. Every reading can lie the same distance from the true pressure, and the standard deviation measures distances from the mean, so an offset that every reading shares leaves the standard deviation unchanged. The aneroid gauges show the difference. The fourth of them, the gauge of Chapter 2's solved example, had a Type A figure of 0.28 mmHg at 120 mmHg, smaller than the first instrument's 0.42 mmHg. Yet at that pressure it read 2.43 mmHg below the reference, an error of −2.43 mmHg. The gauge repeated well and was still wrong by more than 2 mmHg, which is the difference between precision and accuracy that Chapter 2 named.

The engineer corrects an offset before counting any scatter, because in the GUM's method a result is assumed to have been corrected already for all recognized significant systematic effects (Chapter 2). The engineer therefore measures an error such as the fourth gauge's 2.43 mmHg by calibration and removes it by adjusting the gauge or correcting its readings (Chapter 14), and only then counts its 0.28 mmHg of scatter as part of the doubt. But the engineer does not correct an offset that chance has made, such as the generated mean's 0.06 mmHg below the reference. A correction based on it would shift every later reading by a chance amount instead of removing an error.

The engineer keeps a reading that lies far from the others unless the record shows a mistake. In the generated set, 150.9 mmHg lies 0.96 mmHg above the mean, more than twice the standard deviation. Yet the program drew it from the same bell curve as the other nine, so it belongs to the scatter. An engineer who left it out would understate the scatter, because the other nine readings have a standard deviation of only about 0.3 mmHg. A true mistake is different: blunders in recording or analyzing data can introduce a significant error of unknown size, and measures of uncertainty are not meant to account for such mistakes, as the GUM notes. The engineer therefore removes a reading written down as 159.0 instead of 150.9 mmHg because the record shows its cause, not because it lies far from the others.

The display step can hide the scatter

The monitor's display shows whole millimeters of mercury, and after rounding to that step only the readings that cross a boundary between two displayed values show any scatter. Rounded, the ten generated readings become 151, 150, 150, 150, 150, 149, 150, 150, 150 and 150 mmHg. Eight are identical, and only two, which happened to cross the boundaries at 150.5 and 149.5 mmHg, differ from them. The standard deviation of the rounded readings, 0.47 mmHg, lands close to the fine readings' 0.44 mmHg by chance, but it now depends on two readings instead of ten.

26 — What the 1 mmHg display does to the same readings
149.5 150.5 display rounding boundaries read from the firmware, 0.1 mmHg step (assumed) 149 150 151 mmHg AS DISPLAYED the only scatter left: two readings that crossed a boundary fine readings: 0.44 mmHg displayed readings: 0.47 mmHg, from two readings Generated readings, not measured; standard deviations worked from the figures above. A channel scattering by 0.2 mmHg would usually show ten readings of 150.
UncertaintyReadingFocal detail
Plate 26 — Rounding to the 1 mmHg display leaves eight identical readings, and the scatter that remains rests on two. A Type A evaluation therefore reads the channel finer than the display and counts the display step as a row of its own.

A channel centered on 150.0 mmHg that scattered by only 0.2 mmHg would seldom cross either boundary, so ten displayed readings would usually all show 150, a standard deviation of zero from a channel that still scatters. Such a display hides the scatter, because its step of 1 mmHg is coarser than the scatter it would have to show (Chapter 8). For a Type A evaluation the engineer therefore takes readings finer than the display, either from the firmware's value before rounding, as the generated set assumes, or from a service mode if it shows a finer step. The engineer then counts the display's own step as a separate source of error, so that the rounding is counted once and in its own place (Chapter 11).

A repeatability test on the bench
  • Hold one pressure against the reference manometer and take at least ten readings, the count below which EA asks for caution.
  • Read the channel finer than the display, from the firmware's value or a service mode, and count the display step as a separate source of error.
  • State which standard deviation each scatter figure uses: that of a single reading for a monitor that reports one reading, that of the mean only for a result that is itself a mean.
  • Keep the error against the reference apart from the scatter: measure the first by calibration and remove it by adjustment or correction, then count the second as a doubt.
  • Remove a reading only for a cause found in the record, never because it lies far from the others.
  • Label generated or assumed values as such in the design and development file.
+ What this chapter established
  • The mean of repeated readings is the best estimate of the quantity, and the standard deviation measures how widely single readings scatter around it: 149.94 mmHg and 0.44 mmHg for ten generated readings at 150.0 mmHg.
  • The standard deviation of the mean is the standard deviation divided by the square root of the number of readings, 0.14 mmHg for ten, and it applies only to a result that is itself a mean.
  • A small scatter does not make a reading right: an aneroid gauge with a Type A figure of 0.28 mmHg still had an error of −2.43 mmHg that repetition cannot reveal, so calibration measures such an error and adjustment or correction removes it.
  • A display that rounds to 1 mmHg can hide the scatter, so for a Type A evaluation the engineer takes readings finer than the display and counts the display step as a separate source of error.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.