Staying true after the factory.
+ The questionHow does a measuring device stay accurate for years after it leaves the factory?
A check the user can make
A blood-glucose meter can be tested at home, because its reference comes in a bottle. Control solution is a liquid of known value that tests the meter, not the patient. MicroTech Medical makes two levels of it for its GoChek meter and sells them separately from the meter. The user is told "to compare the control results with the expected control ranges printed on the strip vial", according to FDA's review of the meter. The meter, by the review's account, "automatically detects that a control solution is being used." With the bottle, anyone who owns the meter can test the meter and its strips together, at any time, without sending them back to the maker.
Two levels of control solution test the meter at two points of its range, so they can find an error that grows with the glucose value, a span error, as well as an offset, which shifts every result alike (Chapter 11). FDA asks for that check in its 2020 guidance for glucose meters sold for home use: "The use of external control solutions by the user should be promoted," and "At least two levels of control material should be specified in the labeling as available to the user." FDA also asks for labeling that gives "instructions for what to do if the control material values are not within the manufacturer's allowable range", but it sets no frequency for the test.
The blood-pressure monitor has no such bottle, because its reference is a pressure, not a liquid. To set a known pressure, a technician needs a reference manometer, a pump and a rigid vessel, and must put the monitor into the service mode that OIML and India reserve for service personnel (Chapter 14). In the manuals for two of its monitors, Microlife gives the user no calibration procedure and recommends instead a test of accuracy every 2 years or after a mechanical impact, through its own service. The monitor checks itself against a known value only through auto-zero against the open air, which tests the zero and not the span (Chapter 17). Between one service test and the next, the user has no way to tell whether the monitor's span has moved.
An instrument can also carry its span reference inside it. Sartorius builds one, an internal weight, into its Entris II analytical balance, model BCE64I-1S, and the balance adjusts itself with that weight, a feature Sartorius lists as "Internal automatic (isoCAL)". The adjustment "accounts for environmental or temporal changes that can impact the accuracy of your weighing results", in Sartorius's words, and runs when predefined time intervals or temperature values are exceeded, according to Imlab, a distributor of Sartorius balances. The internal weight is a reference like any other and needs a calibration of its own, which neither Sartorius nor Imlab describes. A monitor would need a known pressure inside it to do the same, and none of the published monitor designs in the earlier chapters carries one.
What use does to a monitor
A monitor in use is pumped up and let down thousands of times, carried, dropped and kept in rooms that are hot or cold. Before a model goes on sale, a laboratory tests samples of it, not every unit sold, against some of those stresses. OIML's recommendation for automated sphygmomanometers, R 149-1, asks that after 10,000 simulated measurement cycles the cuff-pressure reading change by no more than 3 mmHg, and India's rule sets the same figures. Microlife gives its BP B3 Basic an expected service life of 5 years or 10,000 measurements, so the durability test puts about a lifetime of cycles on a sample. R 149-1 also requires storage tests at −5 °C and at 50 °C, and requires the limit still to be met after vibration and shock.
A monitor in service must still meet the limit it met at the factory, because R 149-1 sets no wider limit for use, and its verification in service checks the same ±3 mmHg. OIML treats weighing instruments differently, and states in R 76-1 that "The maximum permissible errors in service shall be twice the maximum permissible errors on initial verification." A scale may therefore drift to twice its limit as new before it fails in law. A sphygmomanometer has no such allowance: the drift it may develop in use is limited to what remains of the ±3 mmHg after its own error and its test's uncertainty at the factory (Chapter 25).
Surveys of devices already in use show how many of them stay within that limit. A'Court and colleagues at the University of Oxford tested the blood-pressure devices of 38 general practices in England in a cross-sectional study published in 2011 (Chapter 2). They inflated each device to 280 mmHg, compared it with a reference pressure meter at 250, 200, 150, 100 and 50 mmHg, and passed it if every reading lay within 3 mmHg of the reference. The UK's National Institute for Health Research funded the work, and the authors declared no competing interests.
A second survey applied the same rule to monitors that patients use at home. In 2020 Hodgkinson and colleagues reported ACCU-RATE, a cross-sectional study of 327 home monitors owned by patients with high blood pressure. They tested each from 0 to 300 mmHg at intervals of 50 mmHg against a reference tester, an Omron PA350, and failed any monitor beyond ±3 mmHg at any interval. The National Institute for Health Research funded the study, and no monitor maker funded it. One author, Richard McManus of the University of Oxford, declared that he had received research support in the form of blood-pressure monitors from Omron, the tester's maker, and from Lloyds Pharmacies.
- A'Court and colleagues found 604 devices and fully tested 584 of them, 97 %. Of those, 503 read within 3 mmHg at all five points: 503 ÷ 584 = 0.86, or 86 %.
- The other 584 − 503 = 81 failed: 77 by 4 to 9 mmHg at one or more points, and 4 by 10 mmHg or more.
- By type, 272 of 308 digital devices passed, 88 %, against 150 of 191 aneroid devices, 79 %, 71 of 75 mercury devices, 95 %, and all 10 hybrid devices.
- In ACCU-RATE, 86 % of the 327 home monitors passed, with a 95 % confidence interval of 82 to 90 % (Chapter 24), so about 14 % failed.
- Of the 279 ACCU-RATE monitors whose owners could estimate how long they had been in use, those in their first 4 years of service failed in 8 of 155 cases, 5.2 %, and older ones in 32 of 124, 25.8 %: 25.8 ÷ 5.2 = 5.0, five times as often.
In two surveys published nine years apart, about one device in seven failed the ±3 mmHg rule, and in the home survey a monitor in use for more than four years was five times as likely to fail as one in its first four years. Time in service therefore predicts failure, and a maker or a clinic that sets a check interval has a reason to shorten it as monitors grow older. The same home survey found a mean error, ignoring its sign, of 1.0 mmHg at 150 mmHg and 1.5 mmHg at 280 to 300 mmHg, so the errors found in monitors in use are larger at the top of the range.
Service checks and re-verification after a repair
In a service check a technician repeats the factory's comparison in the field: the monitor in its service mode, a rigid vessel in place of the cuff, and a reference manometer reading the same pressure (Chapter 14). India's rule and the makers keep that work for trained people. India's rule also requires seals against opening the case or tampering with the measuring parts. Welch Allyn's Ambulo 2400 has a second, independent safety processor that reads the pressure alongside the main processor, and Welch Allyn allows only its authorized service personnel to calibrate that processor. Mindray states that the tests in its VS-900 manual "should be performed by qualified service personnel only".
A repair is a different event from a routine check, because it can change the instrument that the last calibration described. A technician may replace the sensor, the pump, a valve or a length of tube. A new sensor brings an offset and a span of its own, which may differ from the old sensor's by as much as Omron's tolerances allow (Chapter 11), so the coefficients stored for the old sensor no longer fit. A new tube or valve can leak. In either case the certificate on file describes a monitor that no longer exists. The repaired monitor needs re-verification after a repair: a check against the same limits as when new, before it goes back into use.
OIML and India each name a repair as a reason to verify an instrument again. OIML asks in R 148-1 that a non-automated sphygmomanometer be verified "before use and during the instrument's lifetime, and after repair". For automated monitors, OIML states in R 149-1 that verification covers at least the cuff-pressure limit, the repeatability and the air leakage, the three properties that a repair is most likely to disturb. India's general rules for weights and measures require re-verification before use after any repair, dismantling or reinstallation, as the law firm S.S. Rana describes them (Chapter 30 maps the jurisdictions). For software, OIML notes in its software document that when a legally relevant parameter changes, a re-verification "might be required depending on national legislation". New firmware changes the monitor as a repair does, and it carries a new software identification (Chapter 26).
What field evidence can and cannot show
A maker learns about its monitors in use from three kinds of record: the as-found values of service checks, the complaints of users and clinics, and its own testing of returned units. Of the three, the as-found values tell the maker most, because each records a calibration made after a known time in use. A maker that collects them by model, age and serial number can see drift before a survey does. Complaints arrive without a reference, so they show where to look rather than how large an error is.
A field record can also point to a fault that a check of the cuff pressure would not find. In August 2026 Midmark began recalling four models of its IQvitals Zone vital-signs monitors across the United States, and FDA classified the recall as Class II on 18 September 2026. Class II is the middle of the three classes that FDA assigns to recalls by risk. These monitors offer more than one deflation method, the way a monitor lets air out of the cuff while its algorithm reads the oscillations. FDA gives the reason as "A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used." Midmark told its users to "use only the Step Deflation Method". FDA's record does not say that the cuff-pressure reading itself was wrong. Because the remedy changes only the deflation method, the evidence points to a fault in the method behind the estimate rather than in the reading.
A technician who checks the cuff pressure in service cannot find a fault of that kind. The technician holds the pressure steady in a rigid vessel, so the monitor never runs a deflation method at all. A check of the pressure alone leaves untested the algorithm's step from the oscillations to the systolic and diastolic values (Chapters 1 and 9). In the field, therefore, a technician tests only the reading against its limit. A test on a simulator, which produces in the cuff the oscillations of an arm, shows whether the algorithm still repeats its answers (Chapter 23), and only a clinical validation against people shows that the estimate is right (Chapter 24).
- Design the service mode with access control and seals, and write the field check with the same points, vessel and reference limit as the factory test.
- Collect every service check's as-found values by model, age and serial number, and trend them for drift before customers report it.
- Shorten the check interval for older monitors when the as-found data or field surveys show failure rising with age.
- Require re-verification after any repair that opens the air path, replaces the sensor or loads new firmware, and write new coefficients for a new sensor.
- Add a dynamic test on an oscillometric simulator to the field checks, because a static pressure check does not exercise the deflation methods that the estimate depends on.
- A glucose meter's control solution gives the user a reference to test the meter at home, while a blood-pressure monitor checks only its zero against the open air and needs a reference manometer in a service mode to check its span.
- OIML allows a monitor's reading to change by no more than 3 mmHg over 10,000 simulated cycles and sets no wider limit in service, unlike a weighing instrument, whose limit in service is twice its limit as new.
- In two surveys, 503 of 584 devices in English general practices and 86 % of 327 home monitors read within 3 mmHg, and home monitors in use for more than four years failed five times as often as those in their first four years.
- A repair can change the instrument a calibration described, so it calls for re-verification before use, and Midmark's 2026 recall shows a fault in a deflation method that a static check of the cuff pressure would not find.
Let's Talk